Imaging performance of a digital storage phosphor system.

W Hillen1, U Schiebel, T Zaengel.   

Abstract

The imaging performance of a storage phosphor system (SPS) for digital projection radiography is studied in which the x-ray image is temporarily stored in a light-stimulable phosphor plate which is subsequently read out by a scanning laser beam. The imaging performance of this system has been analyzed for two types of imaging plates. The spatial resolution is described by the modulation transfer function and the signal-to-noise (S/N) ratios of the recorded image data are measured in terms of noise-equivalent quanta and detective quantum efficiency. Their dependence on detector entrance dose and spatial frequency is discussed. A detailed analysis of the different sources of image noise is given to outline the intrinsic features and limits of the system. Finally, the S/N behavior of the SPS is compared with published data of screen-film systems.

Entities:  

Mesh:

Year:  1987        PMID: 3683303     DOI: 10.1118/1.596127

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; I Levis; M Sayag; R Kleehammer; R Heidsieck; C J D'Orsi
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

2.  Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography.

Authors:  Patrik Sund; Magnus Båth; Susanne Kheddache; Lars Gunnar Månsson
Journal:  Eur Radiol       Date:  2003-10-16       Impact factor: 5.315

Review 3.  Chest imaging within the radiology department by means of photostimulable phosphor computed radiography: a review.

Authors:  H Blume; R G Jost
Journal:  J Digit Imaging       Date:  1992-05       Impact factor: 4.056

4.  Image deconvolution in digital autoradiography: a preliminary study.

Authors:  Mutian Zhang; Qing Chen; Xiao-Feng Li; Joseph O'Donoghue; Shutian Ruan; Pat Zanzonico; C Clifton Ling; John L Humm
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

5.  Gain and offset calibration reduces variation in exposure-dependent SNR among systems with identical digital flat-panel detectors.

Authors:  Charles E Willis; Yevgeniy Y Vinogradskiy; Brad K Lofton; R Allen White
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

6.  Support line and tube visibility in chest examinations using computed radiography.

Authors:  W Huda; C J Belden; L A Webb; C K Palmer
Journal:  J Digit Imaging       Date:  1997-08       Impact factor: 4.056

7.  Image optimization in a computed-radiography/photostimulable-phosphor system.

Authors:  R H Sherrier; H G Chotas; G A Johnson; C Chiles; C E Ravin
Journal:  J Digit Imaging       Date:  1989-11       Impact factor: 4.056

8.  Image quality control for digital mammographic systems: initial experience and outlook.

Authors:  H Roehrig; T Yu; E Krupinski
Journal:  J Digit Imaging       Date:  1995-05       Impact factor: 4.056

Review 9.  [Digital radiography of the chest in pediatric patients].

Authors:  S Puig
Journal:  Radiologe       Date:  2003-12       Impact factor: 0.635

10.  Development of a storage phosphor imaging system for proton pencil beam spot profile determination.

Authors:  Jufri Setianegara; Thomas R Mazur; Yao Hao; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2021-08-10       Impact factor: 4.506

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.