Simple quality assurance based on filtered back projection for geometrical/irradiation accuracy in single-isocenter multiple-target stereotactic radiotherapy.

Naoki Hayashi1, Shun Kurata2,3, Yasunori Saito2,4, Shuta Ogawa4, Keisuke Yasui2.   

Abstract

In single-isocenter multiple-target stereotactic radiotherapy (SIMT-SRT), it is difficult to evaluate both the geometrical accuracy and absorbed dose measurement when irradiating off-isocenter targets. This study aimed to develop a simple quality assurance (QA) method to evaluate off-isocenter irradiation position accuracy in SIMT-SRT and compare its feasibility with that of a commercial device. First, we created two types of inserts and metallic balls with a diameter of 5 mm to be inserted into a commercially available phantom (SIMT phantom). Second, we developed a dedicated analysis software using Python for the Winston-Lutz test (WLT). Third, an image processing software, including the filtered back-projection algorithm, was developed to analyze the images obtained using an electronic portal imaging device (EPID). Fourth, the feasibility of our method was evaluated by comparing it with the results of WLT using two commercially available phantoms: WL-QA and MultiMet-WL cubes. Notably, 92% of the results in one-dimensional deviations were within 0.26 mm (EPID pixel width). The correlation coefficients were 0.52, 0.92, and 0.96 in the left-right, superior-inferior, and anterior-posterior directions, respectively. In the WLT, a maximum two-dimensional deviation of 0.70 mm was detected in our method, while the deviation in the other method was within 0.5 mm. The advantage of our method is that it can evaluate the geometrical accuracy at any gantry angle during dynamic rotation irradiation using a filtered back-projection algorithm, even if the target is located off the isocenter. Our method can perform WLT at arbitrary positions and is suitable for the QA of dynamic rotation irradiation using an EPID.
© 2022. The Author(s), under exclusive licence to Japanese Society of Radiological Technology and Japan Society of Medical Physics.

Entities:  

Keywords:  Filtered back projection; Geometrical accuracy; Off-isocenter; Quality assurance; SIMT-SRT

Year:  2022        PMID: 36261754     DOI: 10.1007/s12194-022-00683-1

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  22 in total

1.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

2.  Evaluation of multiple factors affecting normal brain dose in single-isocenter multiple target radiosurgery.

Authors:  Yu Yuan; Evan M Thomas; Grant A Clark; James M Markert; John B Fiveash; Richard A Popple
Journal:  J Radiosurg SBRT       Date:  2018

3.  Defining the optimal planning target volume in image-guided stereotactic radiosurgery of brain metastases: results of a randomized trial.

Authors:  John P Kirkpatrick; Zhiheng Wang; John H Sampson; Frances McSherry; James E Herndon; Karen J Allen; Eileen Duffy; Jenny K Hoang; Zheng Chang; David S Yoo; Chris R Kelsey; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-10-21       Impact factor: 7.038

4.  Initial SRS for Patients With 5 to 15 Brain Metastases: Results of a Multi-Institutional Experience.

Authors:  Ryan T Hughes; Adrianna H Masters; Emory R McTyre; Michael K Farris; Caroline Chung; Brandi R Page; Lawrence R Kleinberg; Jaroslaw Hepel; Joseph N Contessa; Veronica Chiang; Jimmy Ruiz; Kounosuke Watabe; Jing Su; John B Fiveash; Steve Braunstein; Samuel Chao; Albert Attia; Diandra N Ayala-Peacock; Michael D Chan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2019-04-06       Impact factor: 7.038

5.  Restricted single isocenter for multiple targets dynamic conformal arc (RSIMT DCA) technique for brain stereotactic radiosurgery (SRS) planning.

Authors:  Jenghwa Chang; A Gabriella Wernicke; Susan C Pannullo
Journal:  J Radiosurg SBRT       Date:  2018

6.  Dosimetric Performance and Planning/Delivery Efficiency of a Dual-Layer Stacked and Staggered MLC on Treating Multiple Small Targets: A Planning Study Based on Single-Isocenter Multi-Target Stereotactic Radiosurgery (SRS) to Brain Metastases.

Authors:  Taoran Li; Peyton Irmen; Haisong Liu; Wenyin Shi; Michelle Alonso-Basanta; Wei Zou; Boon-Keng Kevin Teo; James M Metz; Lei Dong
Journal:  Front Oncol       Date:  2019-01-22       Impact factor: 6.244

7.  Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases.

Authors:  Kasirajan Tamil Selvan; Ganesan Padma; Murugesan Kesavan Revathy; N Arunai Nambi Raj; K Senthilnathan; P Ramesh Babu
Journal:  J Med Phys       Date:  2019 Apr-Jun

8.  Retrospective quality metrics review of stereotactic radiosurgery plans treating multiple targets using single-isocenter volumetric modulated arc therapy.

Authors:  Yunfeng Cui; Hao Gao; Jiahan Zhang; John P Kirkpatrick; Fang-Fang Yin
Journal:  J Appl Clin Med Phys       Date:  2020-04-02       Impact factor: 2.102

9.  Dosimetric quality and delivery efficiency of robotic radiosurgery for brain metastases: Comparison with C-arm linear accelerator based plans.

Authors:  Shuming Zhang; Ruijie Yang; Xin Wang
Journal:  J Appl Clin Med Phys       Date:  2019-10-03       Impact factor: 2.102

10.  Dosimetric effect of rotational setup errors in stereotactic radiosurgery with HyperArc for single and multiple brain metastases.

Authors:  Tomohiro Sagawa; Shingo Ohira; Yoshihiro Ueda; Yuichi Akino; Hirokazu Mizuno; Masao Matsumoto; Masayoshi Miyazaki; Masahiko Koizumi; Teruki Teshima
Journal:  J Appl Clin Med Phys       Date:  2019-09-11       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.