| Literature DB >> 31580532 |
Shuming Zhang1, Ruijie Yang1, Xin Wang2.
Abstract
The incidence of brain metastases is increasing and various treatment modalities exist for brain metastases. The aim of this study was to investigate the dosimetric quality and delivery efficiency of robotic radiosurgery (CyberKnife) for multiple brain metastases compared with C-arm linear accelerator (linac) based plans. C-arm linac based plans included intensity-modulated radiation therapy (IMRT), volumetric modulated arc therapy (VMAT) and non-coplanar VMAT with 1, 3 and 5 non-coplanar arcs, respectively (NC1, NC3 and NC5). For 20 patients, six plans with a prescription dose of 30 Gy in three fractions were generated. The gradient index (GI), conformity index (CI), maximum dose (Dmax ) of organs at risk (OARs), normal brain tissue volume (V3 Gy -V24 Gy ), monitor units (MUs) and beam on time (BT) were evaluated. The GI of CyberKnife plans (3.60 ± 0.70) was lower than IMRT (6.21 ± 2.26, P < 0.05), VMAT (6.04 ± 1.93, P < 0.05), NC1 (5.16 ± 1.71, P < 0.05), NC3 (5.02 ± 1.59, P < 0.05) and NC5 (5.03 ± 1.72, P < 0.05). The CI of the VMAT plans (both coplanar and non-coplanar) was larger than IMRT and CK plans. The Dmax for most OARs of the CyberKnife plan was lower than the C-arm linac based plans, although some differences were not statistically significant. The normal brain tissue volume of CyberKnife plan was lower than the C-arm linac based plans, and the normal brain tissue volume of non-coplanar VMAT plans was lower than IMRT and VMAT plans at high-moderate dose level. However, the MUs and BT of CyberKnife plans was more than C-arm linac based plans. CyberKnife plan was better than C-arm linac based plans in protecting normal brain tissue and OARs for patients with multiple brain metastases. C-arm linac based plan with non-coplanar arc provided better protection of normal brain tissue than coplanar plan. However, the BT of CyberKnife plan was longer than C-arm linac based plans.Entities:
Keywords: CyberKnife; IMRT; VMAT; dosimetry; multiple brain metastases; non-coplanar
Mesh:
Year: 2019 PMID: 31580532 PMCID: PMC6839388 DOI: 10.1002/acm2.12746
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Dose distributions of CyberKnife (CK), intensity‐modulated radiation therapy (IMRT), volumetric modulated arc therapy (VMAT), non‐coplanar VMAT with one non‐coplanar arc (NC1), non‐coplanar VMAT with three non‐coplanar arc (NC3) and non‐coplanar VMAT with five non‐coplanar arc (NC5) plans in the axial plane (upper), coronal plane (center), and sagittal plane (lower) for a typical patient.
Figure 2The average normal brain tissue volume receiving specific dose for CyberKnife (CK), intensity‐modulated radiation therapy (IMRT), volumetric modulated arc therapy (VMAT), non‐coplanar VMAT with one non‐coplanar arc (NC1), non‐coplanar VMAT with three non‐coplanar arc (NC3) and non‐coplanar VMAT with five non‐coplanar arc (NC5) plans.
Evaluation parameters of CK, IMRT, VMAT, NC1, NC3 and NC5 plans ( ± SD).
| Parameters | CK | IMRT | VMAT | NC1 | NC3 | NC5 |
|
|---|---|---|---|---|---|---|---|
| GI | 3.60 ± 0.68 | 6.21 ± 2.26 | 6.04 ± 1.93 | 5.16 ± 1.71 | 5.02 ± 1.59 | 5.03 ± 1.72 | a, b, c, d, e |
| CI | 0.86 ± 0.08 | 0.81 ± 0.07 | 0.90 ± 0.04 | 0.89 ± 0.07 | 0.89 ± 0.07 | 0.91 ± 0.07 | a, b, c, d, e |
| MUs | 29 031.96 ± 6562.18 | 3319.00 ± 1278.87 | 2968.00 ± 898.66 | 2527.90 ± 376.81 | 2518.25 ± 348.13 | 2384.95 ± 231.47 | a, b, c, d, e |
| BT/min | 39.05 ± 5.11 | 5.53 ± 2.13 | 4.95 ± 1.50 | 4.22 ± 0.63 | 4.20 ± 0.58 | 4.09 ± 0.29 | a, b, c, d, e |
Statistical significance P: a = CK vs IMRT, b = CK vs VMAT, c = CK vs NC1, d = CK vs NC3, e = CK vs NC5.
Abbreviations: BT, treatment time; CI, conformity index; CK, CyberKnife; GI, gradient index; IMRT, intensity‐modulated radiation therapy; MUs, monitor units; NC1, non‐coplanar VMAT with one non‐coplanar arc; NC3, non‐coplanar VMAT with three non‐coplanar arc; NC5, non‐coplanar VMAT with five non‐coplanar arc; NS, not significant; VMAT, volumetric modulated arc therapy.
The maximum dose for OARs of CK, IMRT, VMAT, NC1, NC3 and NC5 plans ( ± SD).
| OAR | CK | IMRT | VMAT | NC1 | NC3 | NC5 |
|
|---|---|---|---|---|---|---|---|
| Brainstem | 4.22 ± 1.7 | 5.32 ± 2.37 | 5.09 ± 2.48 | 4.27 ± 1.28 | 3.66 ± 0.95 | 3.84 ± 0.92 | a |
| Lens_R | 0.30 ± 0.33 | 1.11 ± 0.83 | 1.07 ± 0.73 | 0.99 ± 0.33 | 0.97 ± 0.33 | 0.96 ± 0.30 | a, b, c, d, e |
| Lens_L | 0.19 ± 0.15 | 0.88 ± 0.83 | 1.12 ± 0.76 | 0.98 ± 0.43 | 0.96 ± 0.33 | 0.96 ± 0.30 | a, b, c, d, e |
| Eye_R | 1.72 ± 1.23 | 2.78 ± 1.68 | 2.29 ± 1.37 | 2.04 ± 0.64 | 2.03 ± 0.61 | 1.96 ± 0.60 | a, b |
| Eye_L | 1.55 ± 1.26 | 2.33 ± 1.72 | 2.25 ± 1.38 | 1.88 ± 0.94 | 1.90 ± 0.83 | 1.80 ± 0.73 | a, b |
| Optic nerve_R | 1.18 ± 1.04 | 1.77 ± 1.25 | 1.78 ± 1.16 | 1.74 ± 0.64 | 1.72 ± 0.65 | 1.78 ± 0.56 | b, c, d, e |
| Optic nerve_L | 1.09 ± 0.80 | 1.75 ± 1.65 | 1.60 ± 0.85 | 1.82 ± 0.73 | 1.73 ± 0.63 | 1.84 ± 0.52 | b, c, d, e |
| Optic chiasm | 1.96 ± 0.89 | 2.65 ± 1.71 | 2.32 ± 1.31 | 2.37 ± 0.65 | 2.25 ± 0.47 | 2.31 ± 0.47 | NS |
| Pituitary | 1.64 ± 0.73 | 1.94 ± 1.25 | 1.76 ± 0.95 | 1.71 ± 0.48 | 1.59 ± 0.39 | 1.82 ± 0.30 | NS |
Statistical significance P: a = CK vs IMRT, b = CK vs VMAT, c = CK vs NC1, d = CK vs NC3, e = CK vs NC5.
Abbreviations: CK, CyberKnife; IMRT, intensity‐modulated radiation therapy; NC1, non‐coplanar VMAT with one non‐coplanar arc; NC3, non‐coplanar VMAT with three non‐coplanar arc; NC5, non‐coplanar VMAT with five non‐coplanar arc; NS, not significant; OARs, organs at risk; VMAT, volumetric modulated arc therapy.
The normal brain tissue volume (receiving specific dose) of NC1, NC3 and NC5 plans ( ± SD).
| Normal brain tissue volume | NC1 | NC3 | NC5 |
|
|---|---|---|---|---|
| V24 Gy/cm3 | 17.69 ± 9.61 | 18.38 ± 10.26 | 18.15 ± 9.48 | b |
| V21 Gy/cm3 | 29.9 ± 16.24 | 29.72 ± 16.48 | 29.33 ± 15.35 | NS |
| V18 Gy/cm3 | 48.45 ± 28.28 | 46.43 ± 26.21 | 45.49 ± 24.65 | b |
| V15 Gy/cm3 | 81.92 ± 52.03 | 77.74 ± 46.91 | 74.5 ± 43.21 | b |
| V12 Gy/cm3 | 137.96 ± 91.07 | 134.66 ± 86.35 | 126.63 ± 78.96 | b, c |
| V9 Gy/cm3 | 240.62 ± 160.65 | 235.2 ± 156.24 | 221.65 ± 144.33 | b, c |
| V6 Gy/cm3 | 417.3 ± 252.41 | 415.78 ± 255.96 | 399.14 ± 252.78 | c |
| V3 Gy/cm3 | 728.75 ± 291.03 | 753.14 ± 318.48 | 748.91 ± 330.21 | NS |
Statistical significance P: a = NC1 vs NC3, b = NC1 vs NC5, c = NC3 vs NC5.
Abbreviations: NC1, non‐coplanar VMAT with one non‐coplanar arc; NC3, non‐coplanar VMAT with three non‐coplanar arc; NC5, non‐coplanar VMAT with five non‐coplanar arc; NS, not significant.