Literature DB >> 36201009

Low tibial baseplate migration 1 year after unrestricted kinematically aligned total knee arthroplasty using a medial conforming implant design.

Abigail E Niesen1, Anna L Garverick1, Stephen M Howell1, Maury L Hull2,3,4.   

Abstract

PURPOSE: Varus alignment of the tibial baseplate and limb > 3° might adversely affect baseplate fixation after total knee arthroplasty (TKA), especially for unrestricted kinematically aligned (KA) TKA which aligns a majority of baseplates in varus. The purposes of this study were to determine whether baseplate migration at 1 year (1) was significantly less than a stability limit of 0.5 mm, (2) increased over time, and (3) was related to varus alignment of the baseplate and limb after unrestricted KA TKA.
METHODS: Thirty-five patients underwent unrestricted KA TKA using a fixed-bearing, cemented, medial conforming tibial insert with posterior cruciate ligament retention. Using model-based radiostereometric analysis, maximum total point motion (MTPM) (i.e., largest displacement on the baseplate) was computed at 6 weeks, 3 months, 6 months, and 1 year postoperatively relative to the day of surgery. Baseplate and limb alignment were measured postoperatively on long-leg CT scanograms.
RESULTS: At 1 year, mean MTPM of 0.35 mm was significantly less than the 0.5 mm stability limit (p = 0.0002). Mean MTPM did not increase from 6 weeks to 1 year (p = 0.3047). Notably, 89% (31/35) of tibial baseplates and 46% (16/35) of limbs were > 3° varus. Baseplate and limb alignment had no relationship to MTPM at 1 year (|r|≤ 0.173, p ≥ 0.3276).
CONCLUSION: Low and non-progressive tibial baseplate migration 1 year after unrestricted KA TKA with a medial conforming design should allay any concern that unrestricted KA TKA increases risk of baseplate loosening due to varus alignment of the baseplate and limb. LEVEL OF EVIDENCE: Level II, therapeutic prospective cohort study.
© 2022. The Author(s) under exclusive licence to European Society of Sports Traumatology, Knee Surgery, Arthroscopy (ESSKA).

Entities:  

Keywords:  Aseptic loosening; Baseplate stability, stable baseplate; Maximum total point motion; Migration stability limit; Model-based radiostereometric analysis; Nonprogressive migration; TKA alignment technique; TKA outcomes; Varus baseplate alignment

Year:  2022        PMID: 36201009     DOI: 10.1007/s00167-022-07171-4

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.114


  37 in total

1.  Posterior rim loading of a low-conforming tibial insert in unrestricted kinematic alignment is caused by rotational alignment of an asymmetric baseplate designed for mechanical alignment.

Authors:  Maury L Hull; Stephanie Nicolet-Petersen; Augustine Saiz; Connor Delman; Stephen M Howell
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2022-05-31       Impact factor: 4.342

Review 2.  The movement of the normal tibio-femoral joint.

Authors:  M A R Freeman; V Pinskerova
Journal:  J Biomech       Date:  2005-02       Impact factor: 2.712

3.  Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables.

Authors:  J W Bartlett; C Frost
Journal:  Ultrasound Obstet Gynecol       Date:  2008-04       Impact factor: 7.299

4.  Functional joint line obliquity after kinematic total knee arthroplasty.

Authors:  Jonathan Hutt; Vincent Massé; Martin Lavigne; Pascal-André Vendittoli
Journal:  Int Orthop       Date:  2015-03-21       Impact factor: 3.075

5.  Hot topic: alignment in total knee arthroplasty-systematic versus more individualised alignment strategies.

Authors:  Michael T Hirschmann; Jon Karlsson; Roland Becker
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-04-21       Impact factor: 4.342

6.  Does a kinematically aligned total knee arthroplasty restore function without failure regardless of alignment category?

Authors:  Stephen M Howell; Stacey J Howell; Kyle T Kuznik; Joe Cohen; Maury L Hull
Journal:  Clin Orthop Relat Res       Date:  2012-09-21       Impact factor: 4.176

7.  Implant Survival and Function Ten Years After Kinematically Aligned Total Knee Arthroplasty.

Authors:  Stephen M Howell; Trevor J Shelton; Maury L Hull
Journal:  J Arthroplasty       Date:  2018-07-31       Impact factor: 4.757

8.  Kinematic alignment in total knee arthroplasty better reproduces normal gait than mechanical alignment.

Authors:  William Blakeney; Julien Clément; François Desmeules; Nicola Hagemeister; Charles Rivière; Pascal-André Vendittoli
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-10-01       Impact factor: 4.342

9.  Kinematically aligned TKA can align knee joint line to horizontal.

Authors:  Hyung-Min Ji; Jun Han; Dong San Jin; Hyunseok Seo; Ye-Yeon Won
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-01-25       Impact factor: 4.342

View more