John Davidson1, Xiaolin Gu2, Michael Breen2, Raiju J Babu2.
Abstract
Purpose: To evaluate absolute prediction errors following phacoemulsification with implantation of a multifocal toric intraocular lens (IOL) using intraoperative aberrometry for IOL power selection and to compare findings with the globally optimized and manufacturer's recommended lens constants and regression coefficients.Entities:
Keywords: intraocular lens implantation; intraoperative aberrometry; lens constant optimization; prediction error; refractive error; visual acuity
Year: 2022 PMID: 36193510 PMCID: PMC9526440 DOI: 10.2147/OPTH.S369797
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Baseline Characteristics for IOL Models Implanted, and Pre-Operative Biometric Measurements in the Pre-Optimization and Post-Optimization Groups
| Pre-Optimization Group (n=1027) | Post-Optimization Group (n=419) | All (N=1446) | |
|---|---|---|---|
| IOL model implanted, n (%) | |||
| SND1T3 | 508 (49.4) | 215 (51.3) | 723 (50.0) |
| SND1T4 | 304 (29.6) | 115 (27.5) | 419 (29.0) |
| SND1T5 | 191 (18.6) | 66 (15.8) | 257 (17.8) |
| SND1T6 | 24 (2.3) | 23 (5.5) | 47 (3.3) |
| Days to refraction, mean (SD) | 47.7 (41.6) | 48.2 (47.7) | 47.9 (43.4) |
| Keratometry* (D), mean (SD) | 44.11 (1.49) | 44.14 (1.56) | 44.12 (1.51) |
| White-to-white distance (mm), mean (SD) | 12.03 (0.44) | 12.06 (0.43) | 12.04 (0.44) |
| Axial length (mm), mean (SD) | 24.30 (1.39) | 24.10 (1.34) | 24.24 (1.38) |
| Axial Length (mm), n (%) | |||
| ≤ 22 mm | 27 (2.6) | 19 (4.5) | 46 (3.2) |
| > 22 - < 26 mm | 877 (85.4) | 361 (86.2) | 1238 (85.6) |
| ≥ 26 mm | 123 (12.0) | 39 (9.3) | 162 (11.2) |
Note: *Keratometry = (Ksteep + Kflat)/2.
Abbreviations: D, diopter; IOL, intraocular lens; SD, standard deviation.
Figure 1Proportions of (A) first eyes and (B) all eyes with absolute ORA SYSTEM prediction error ≤0.25 D and ≤0.50 D before and after the first global optimization of the lens constant.
Figure 2(A) Mean and (B) median absolute ORA SYSTEM prediction error before and after optimization of the lens constant in first eyes and in all eyes.
Effect of Axial Length on Absolute ORA SYSTEM Prediction Error After the First Global Lens Constant Optimization in First Eyes
| Outcome | Axial Length | Pre-Optimization | Post-Optimization | Difference |
|---|---|---|---|---|
| n/N (%) | n/N (%) | Post – Pre % (95% CI) | ||
| Proportion of eyes with absolute ORA SYSTEM prediction error ≤0.25 D | ≤22 mm | 9/23 (39.1) | 8/16 (50.0) | 10.9 (−20.7, 42.5) |
| >22 to <26 mm | 238/713 (33.4) | 175/326 (53.7) | 20.3 (13.9, 26.7) | |
| ≥26 mm | 43/96 (44.8) | 17/36 (47.2) | 2.43 (−16.7, 21.5) | |
| All | 290/832 (34.9) | 200/378 (52.9) | 18.1 (12.1, 24.0) | |
| Proportion of eyes with absolute ORA SYSTEM prediction error ≤0.50 D | ≤22 mm | 17/23 (73.9) | 12/16 (75.0) | 1.09 (−26.7,28.9) |
| >22 to <26 mm | 451/713 (63.3) | 273/326 (83.7) | 20.5 (15.1, 25.8) | |
| ≥26 mm | 80/96 (83.3) | 32/36 (88.9) | 5.56 (−7.1, 18.2) | |
| All | 548/832 (65.9) | 317/378 (83.9) | 18.0 (13.1. 22.9) | |
| Mean (SD) | Mean (SD) | Post – Pre (95% CI) | ||
| Mean absolute ORA SYSTEM prediction error (D) | ≤22 mm | 0.35 (0.24) | 0.32 (0.29) | −0.04 (−0.21, 0.14) |
| >22 to <26 mm | 0.45 (0.33) | 0.29 (0.25) | −0.16 (−0.20, −0.12) | |
| ≥26 mm | 0.32 (0.26) | 0.29 (0.21) | −0.03 (−0.13, 0.06) | |
| All | 0.44 (0.32) | 0.29 (0.25) | −0.14 (−0.17, −0.11) | |
| Median | Median | Post – Pre | ||
| Median absolute ORA SYSTEM prediction error (D) | ≤22 mm | 0.38 | 0.22 | −0.16 |
| >22 to <26 mm | 0.38 | 0.23 | −0.15 | |
| ≥26 mm | 0.30 | 0.27 | −0.03 | |
| All | 0.36 | 0.23 | −0.13 |
Abbreviations: CI, confidence interval; D, diopter; ORA, Optiwave Refractive Analysis.
Effect of Axial Length on Absolute ORA SYSTEM Prediction Error After the First Global Lens Constant Optimization in All Eyes
| Outcome | Axial Length | Pre-Optimization | Post-Optimization | Difference in Percentage |
|---|---|---|---|---|
| n/N (%) | n/N (%) | Post – Pre % (95% CI) | ||
| Proportion of eyes with absolute ORA SYSTEM prediction error ≤0.25 D | ≤22 mm | 11/27 (40.7) | 10/19 (52.6) | 11.9 (−17.2, 41.0) |
| >22 to <26 mm | 291/877 (33.2) | 192/361 (53.2) | 20.0 (14.0, 26.0) | |
| ≥26 mm | 57/123 (46.3) | 18/39 (46.2) | −0.2 (−18.1, 17.8) | |
| All | 359/1027 (35.0) | 220/419 (52.5) | 17.6 (12.0. 23.2) | |
| Proportion of eyes with absolute ORA SYSTEM prediction error ≤0.50 D | ≤22 mm | 20/27 (74.1) | 14/19 (73.7) | −0.4 (−26.2, 25.4) |
| >22 to <26 mm | 555/877 (63.3) | 300/361 (83.1) | 19.8 (14.8, 24.8) | |
| ≥26 mm | 105/123 (85.4) | 34/39 (87.2) | 1.8 (−10.4, 14.0) | |
| All | 680/1027 (66.2) | 348/419 (83.1) | 16.8 (12.2. 21.5) | |
| Mean (SD) | Mean (SD) | Post - Pre (95% CI) | ||
| Mean absolute ORA SYSTEM prediction error (D) | ≤22 mm | 0.37 (0.27) | 0.32 (0.28) | −0.05 (−0.21, 0.12) |
| >22 to <26 mm | 0.45 (0.33) | 0.30 (0.25) | −0.16 (−0.19, −0.12) | |
| ≥26 mm | 0.31 (0.25) | 0.29 (0.20) | −0.01 (−0.10, 0.07) | |
| All | 0.43 (0.32) | 0.30 (0.25) | −0.14 (−0.17, −0.11) | |
| Median | Median | Post - Pre | ||
| Median absolute ORA SYSTEM prediction error (D) | ≤22 mm | 0.38 | 0.21 | −0.17 |
| >22 to <26 mm | 0.38 | 0.23 | −0.15 | |
| ≥26 mm | 0.28 | 0.28 | 0 | |
| All | 0.36 | 0.24 | −0.12 |
Abbreviations: CI, confidence interval; D, diopter; ORA, Optiwave Refractive Analysis.