A large retrospective database analysis comparing outcomes of intraoperative aberrometry with conventional preoperative planning.

Robert J Cionni1, Ramon Dimalanta2, Michael Breen2, Cody Hamilton2.   

Abstract

PURPOSE: To evaluate differences between the absolute prediction error using an intraoperative aberrometry (IA) device for intraocular lens (IOL) power determination versus the error that would have resulted if the surgeon's preoperative plan had been followed.
SETTING: Multiple centers in the United States.
DESIGN: Retrospective analysis of data collected using an IA device.
METHODS: The database information was limited according to predetermined inclusion/exclusion criteria. Primary endpoints included the difference between mean and median absolute prediction error with IA use versus preoperative calculation, and the percentage of cases were compared when the prediction error was 0.5 diopters (D) or less.
RESULTS: A total of 32 189 eyes were analyzed. The IA mean absolute prediction error was lower than the preoperative calculation, 0.30 D ± 0.26 (SD) versus 0.36 ± 0.32 D (P < .0001). The aberrometry absolute median prediction error was lower than the preoperative calculation, 0.24 D versus 0.29 D (P < .0001). There was a significantly greater percentage of eyes with an aberrometry absolute prediction error of 0.5 D or less than eyes with a preoperative absolute prediction error of 0.5 D or less (26 357 [81.9%] of 32 189 vs. 24 437 [75.9%] of 32 189, P < .0001). In addition, in those cases in which power of the IOL implanted was different than the preoperatively planned IOL power, significantly more eyes had an aberrometry absolute prediction error of 0.5 D or less (10 385 [81.3%] of 12 779 vs. 8794 [68.8%] of 12 779, P < .0001).
CONCLUSIONS: In a database of more than 30 000 eyes, calculations incorporating IA outperformed preoperative calculations. The difference was more pronounced in those cases in which the preoperatively planned IOL power was different than the power of the IOL implanted.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30104081     DOI: 10.1016/j.jcrs.2018.07.016

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Predictability of combined cataract surgery and trabeculectomy using Barrett Universal Ⅱ formula.

Authors:  Kei Iijima; Kazutaka Kamiya; Yoshihiko Iida; Masayuki Kasahara; Nobuyuki Shoji
Journal:  PLoS One       Date:  2022-06-23       Impact factor: 3.752

2.  Refractive Outcomes Following Trifocal Intraocular Lens Implantation in Post-Myopic LASIK and PRK Eyes.

Authors:  John F Blaylock; Brad J Hall
Journal:  Clin Ophthalmol       Date:  2022-07-01

3.  Predictive accuracy of an intraoperative aberrometry device for a new monofocal intraocular lens.

Authors:  Lindsay S Spekreijse; Noel J C Bauer; Frank J H M van den Biggelaar; Rob W P Simons; Claudette A Veldhuizen; Tos T J M Berendschot; Rudy M M A Nuijts
Journal:  J Cataract Refract Surg       Date:  2021-08-19       Impact factor: 3.528

4.  Clinical outcomes after aphakic versus aphakic/pseudophakic intraoperative aberrometry in cataract surgery with toric IOL implantation.

Authors:  Satish S Modi
Journal:  Int Ophthalmol       Date:  2020-08-08       Impact factor: 2.031

5.  Time Utilization and Refractive Prediction Enhancement Associated with Intraoperative Aberrometry Use During Cataract Surgery.

Authors:  Karen L Christopher; Jennifer L Patnaik; Cristos Ifantides; D Claire Miller; Richard S Davidson; Michael J Taravella; Anne Lynch; Brandie Wagner
Journal:  Clin Ophthalmol       Date:  2021-02-11

6.  Clinical Outcomes of Monofocal Toric IOLs Using Digital Tracking and Intraoperative Aberrometry.

Authors:  John F Blaylock; Brad J Hall
Journal:  Clin Ophthalmol       Date:  2021-08-26

7.  Refractive Accuracy of Barrett True-K vs Intraoperative Aberrometry for IOL Power Calculation in Post-Corneal Refractive Surgery Eyes.

Authors:  Larissa Gouvea; Kareem Sioufi; Colin E Brown; George Waring Iv; Wallace Chamon; Karolinne Maia Rocha
Journal:  Clin Ophthalmol       Date:  2021-10-27

8.  Intraoperative aberrometry compared to preoperative Barrett True-K formula for intraocular lens power selection in eyes with prior refractive surgery.

Authors:  Suzie A Gasparian; Saman Nassiri; Hyelin You; Abby Vercio; Frank S Hwang
Journal:  Sci Rep       Date:  2022-05-05       Impact factor: 4.996

9.  Comparison of Preoperative Measurements with Intraoperative Aberrometry in Predicting Need for Correction in Eyes with Low Astigmatism Undergoing Cataract Surgery.

Authors:  John A Hovanesian
Journal:  Clin Ophthalmol       Date:  2021-05-26

10.  Retrospective analysis of an intraoperative aberrometry database: a study investigating absolute prediction in eyes implanted with low cylinder power toric intraocular lenses.

Authors:  Robert J Cionni; Michael Breen; Cody Hamilton; Ravaughn Williams
Journal:  Clin Ophthalmol       Date:  2019-08-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.