BACKGROUND: We investigate the performance of new hydrophobic diffractive multifocal intraocular lenses (IOL) with hydroxyethyl methacrylate (HEMA) and compare their optical quality, contrast sensitivity, and subjective photic phenomena.Entities:
Keywords: Contrast sensitivity; Hydroxyethyl methacrylate; Multifocal intraocular lens
Mesh:
Substances:
Year: 2022 PMID: 36131264 PMCID: PMC9494844 DOI: 10.1186/s12886-022-02600-x
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.086
Fig. 1Photic phenomena images using the glare and halo simulator were provided to the patients for subjective photic phenomena survey
Comparison of preoperative ocular parameters and cataract stages between the Clareon Panoptix and Acrysof Panoptix groups
| Clareon Panoptix Group | Acrysof Panoptix Group | ||
|---|---|---|---|
| Eyes (n) | 71 | 64 | |
| Age (years) | 58.06 ± 5.03 [50, 65] | 57.67 ± 4.54 [41, 65] | 0.64 |
| Female (n) | 49 (69.0%) | 44 (68.8%) | 0.97 |
| LOCS III grade | |||
| Cortical | 2.63 ± 1.29 [0, 5] | 2.81 ± 1.42 [0, 6] | 0.45 |
| Nuclear | 3.55 ± 0.94 [0, 5] | 3.41 ± 1.11 [0, 5] | 0.42 |
| Posterior subcapsular | 0.44 ± 0.97 [0, 4] | 0.50 ± 0.98 [0, 3] | 0.71 |
| LogMAR CDVA | 0.34 ± 0.22 [0.15, 1.00] | 0.35 ± 0.27 [0.15, 1.00] | 0.80 |
| Pupil size (mm) | 5.37 ± 0.74 [3.2, 6.8] | 5.22 ± 0.82 [3.1, 6.9] | 0.25 |
| Axial length (mm) | 24.42 ± 1.51 [21.82, 27.17] | 24.13 ± 1.34 [22.38, 29.07] | 0.24 |
| ACD (mm) | 3.29 ± 0.37 [2.50, 4.16] | 3.25 ± 0.31 [2.49, 4.39] | 0.51 |
| Average K (D) | 43.75 ± 1.57 [41.22, 46.62] | 43.86 ± 1.90 [41.51, 46.78] | 0.70 |
| K astigmatism (D) | 0.81 ± 0.53 [0.00, 3.02] | 0.83 ± 0.59 [0.00, 2.82] | 0.86 |
| IOL power (D) | 18.51 ± 4.52 [7.5, 26.5] | 19.48 ± 3.15 [9.5, 24.5] | 0.16 |
| IOL cylindrical power (D) | 0.92 ± 0.83 [0.00, 3.75] | 0.96 ± 0.91 [0.00, 3.75] | 0.74 |
Values are presented as mean ± standard deviation. [minimum, maximum]
LOCS lens opacities classification system, LogMAR logarithm of minimal angle of resolution, CDVA corrected distance visual acuity, ACD anterior chamber depth, K keratometry, IOL intraocular lens
Fig. 2Photopic uncorrected (refractive error within 0.5D) LogMAR monocular and binocular defocus curves of multifocal intraocular lenses with Clareon Panoptix and Acrysof Panoptix. There were no statistically significant differences between the two groups in all diopters. D diopters; LogMAR logarithm of the minimal angle of resolution
Comparison of visual acuity, optical quality analysis and contrast sensitivity between the Clareon Panoptix and Acrysof Panoptix groups
| Clareon Panoptix ( | Acrysof Panoptix ( | ||
|---|---|---|---|
| LogMAR UDVA | 0.029 ± 0.050 | 0.027 ± 0.065 | 0.75 |
| SE (D) | −0.026 ± 0.273 | −0.027 ± 0.246 | 0.98 |
| Refractive cylinder (D) | 0.368 ± 0.107 | 0.353 ± 0.122 | 0.68 |
| OQAS | |||
| OSI | 1.594 ± 0.868 | 1.856 ± 1.138 | 0.13 |
| MTF cut-off value (CPD) | 30.723 ± 11.191 | 29.193 ± 10.719 | 0.42 |
| SR | 0.147 ± 0.060 | 0.138 ± 0.055 | 0.38 |
| Contrast sensitivity | |||
| AULCSF | 1.449 ± 0.168 | 1.358 ± 0.233 | 0.01 |
| 6.3° | 0.019 ± 0.010 | 0.026 ± 0.026 | 0.04 |
| 4.0° | 0.020 ± 0.009 | 0.030 ± 0.037 | 0.03 |
| 2.5° | 0.026 ± 0.013 | 0.037 ± 0.035 | 0.02 |
| 1.6° | 0.045 ± 0.020 | 0.059 ± 0.045 | 0.02 |
| 1.0° | 0.078 ± 0.035 | 0.105 ± 0.079 | 0.01 |
| 0.64° | 0.156 ± 0.074 | 0.201 ± 0.126 | 0.01 |
Values are presented as mean ± standard deviation
LogMAR logarithm of minimal angle of resolution, UDVA uncorrected distance visual acuity, SE spherical equivalent, D diopters, OQAS optical quality analysis system, OSI objective scattering index, MTF modulation transfer function, CPD cycle per degree, SR Strehl ratio, AULCSF area under the log contrast sensitivity function
Fig. 3Subjective photic phenomena survey from the Clareon Panoptix and Acrysof Panoptix groups