Literature DB >> 33769764

Surface light scattering from 1-piece hydrophobic acrylic intraocular lenses with hydroxyethyl methacrylate: contralateral observation for 7 years.

Katsuhito Kinoshita1, Kazunori Miyata, Ryohei Nejima, Masato Honbo, Yosai Mori, Keiichiro Minami.   

Abstract

PURPOSE: To quantitatively and intraindividually compare surface light scattering for 7 years after implantation of Clareon and AcrySof intraocular lenses (IOLs).
SETTING: Miyata Eye Hospital, Miyazaki, Japan.
DESIGN: Retrospective comparative case series.
METHODS: Clinical records of eyes that had contralateral implantation of SN60WF and SY60WF IOLs were reviewed. Light scattering on the anterior surface of IOLs, corrected distance visual acuity (CDVA), and mesopic and photopic contrast sensitivities were examined at 1 year, 3 years, and 7 years postoperatively, and they were intraindividually compared.
RESULTS: 34, 19, and 16 patients visited at 1 year, 3 years, and 7 years postoperatively, respectively. Surface light scattering in eyes with SY60WF IOL was significantly reduced, and the rate of increase was 2.74 computer compatible tapes per year. Although there were prominent increases in eyes with SN60WF IOL, no statistically significant difference was found in their CDVA and contrast sensitivities.
CONCLUSIONS: In eyes with Clareon IOLs, the development of surface light scattering was suppressed up to 7 years. The increase rate demonstrated that the visual acuity would be least deteriorated up to 12 years postoperatively, whereas the influence of straylight induced should be investigated.
Copyright © 2021 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.

Entities:  

Year:  2021        PMID: 33769764     DOI: 10.1097/j.jcrs.0000000000000621

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Clinical outcomes of new multifocal intraocular lenses with hydroxyethyl methacrylate and comparative results of contrast sensitivity, objective scatter, and subjective photic phenomena.

Authors:  Yong Woo Lee; Chul Young Choi; Kun Moon; Yong Jin Jeong; Sang Il An; Je Myung Lee; Jong Ho Lee; Min Cheol Seong
Journal:  BMC Ophthalmol       Date:  2022-09-21       Impact factor: 2.086

2.  One-year multicenter evaluation of a new hydrophobic acrylic intraocular lens with hydroxyethyl methacrylate in an automated preloaded delivery system.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

  2 in total