PURPOSE: To compare mechanical characteristics and stability of Clareon CNA0T0 intraocular lens (IOL) with 4 currently marketed monofocal IOLs and to evaluate axial displacement and simulated dioptric power shift using a range of compression diameters.
SETTING: Alcon Laboratories, Inc., Fort Worth, Texas, USA.
DESIGN: Experimental study.
METHODS: Clareon CNA0T0, AcrySof SN60WF, enVista MX60, Tecnis ZCB00, and Vivinex iSert XY1 IOLs (10 per group) were assessed using standardized methods for axial displacement, optic decentration, and optic tilt per the International Organization for Standardization ISO11979-3. Axial displacement was also measured over a range of compression diameters (9.0 to 11.0 mm).
RESULTS: At 10.0 mm compression, the mean axial displacement for the CNA0T0 IOL (0.02 mm ± 0.01 [SD]) was significantly lower than for the MX60, ZCB00, and XY1 IOLs (P < .005). At all compression diameters, the CNA0T0 and SN60WF IOLs had the lowest levels of axial displacement and corresponding simulated dioptric power shift at the corneal plane versus other IOL models. At 10.0 mm compression, the mean optic decentration was within ±0.06 mm for all models, and there were no significant differences between the CNA0T0 IOL and other IOLs. At 10.0 mm compression, the mean optic tilt was no greater than 1.2 degrees for all IOL models evaluated; however, the mean optic tilt for the CNA0T0 IOL was significantly lower than for the MX60 IOL (P < .005).
CONCLUSION: The CNA0T0 and SN60WF IOLs showed the lowest levels of axial displacement and corresponding simulated dioptric power shift over all tested compression diameters, indicating they might provide the most consistent refractive outcomes.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2019        PMID: 30686704     DOI: 10.1016/j.jcrs.2018.10.043

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  The Impact of IOL Abbe Number on Polychromatic Image Quality of Pseudophakic Eyes.

Authors:  Arthur Bradley; Renfeng Xu; Huachun Wang; Matt Jaskulski; Xin Hong; Nicolette Brink; Stephen Van Noy
Journal:  Clin Ophthalmol       Date:  2020-08-11

2.  Rotational Stability of the Clareon Monofocal Aspheric Hydrophobic Acrylic Intraocular Lens 6 Months After Implantation.

Authors:  Thomas R Walters; Robert Lehmann; Andrew Moyes; John W French; Vidhyapriya Sreenivasan; Satish S Modi
Journal:  Clin Ophthalmol       Date:  2022-02-15

3.  Geometry of Acrylic, Hydrophobic IOLs and Changes in Haptic-Capsular Bag Relationship According to Compression and Different Well Diameters: A Bench Study Using Computed Tomography.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein
Journal:  Ophthalmol Ther       Date:  2022-02-05

4.  Clinical outcomes of new multifocal intraocular lenses with hydroxyethyl methacrylate and comparative results of contrast sensitivity, objective scatter, and subjective photic phenomena.

Authors:  Yong Woo Lee; Chul Young Choi; Kun Moon; Yong Jin Jeong; Sang Il An; Je Myung Lee; Jong Ho Lee; Min Cheol Seong
Journal:  BMC Ophthalmol       Date:  2022-09-21       Impact factor: 2.086

5.  Comparison of two one-piece acrylic foldable intraocular lenses: Short-term change in axial movement after cataract surgery and its effect on refraction.

Authors:  So Goto; Naoyuki Maeda; Kazuhiko Ohnuma; Toru Noda
Journal:  PLoS One       Date:  2022-08-30       Impact factor: 3.752

6.  Refractive stability of a new single-piece hydrophobic acrylic intraocular lens and corneal wound repair after implantation using a new automated intraocular lens delivery system.

Authors:  Kazuno Negishi; Sachiko Masui; Hidemasa Torii; Yasuyo Nishi; Kazuo Tsubota
Journal:  PLoS One       Date:  2020-09-02       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.