Purpose: We report the initial development of an image-based solution for robotic assistance of pelvic fracture fixation. The approach uses intraoperative radiographs, preoperative CT, and an end effector of known design to align the robot with target trajectories in CT. The method extends previous work to solve the robot-to-patient registration from a single radiographic view (without C-arm rotation) and addresses the workflow challenges associated with integrating robotic assistance in orthopaedic trauma surgery in a form that could be broadly applicable to isocentric or non-isocentric C-arms.
Methods: The proposed method uses 3D-2D known-component registration to localize a robot end effector with respect to the patient by: (1) exploiting the extended size and complex features of pelvic anatomy to register the patient; and (2) capturing multiple end effector poses using precise robotic manipulation. These transformations, along with an offline hand-eye calibration of the end effector, are used to calculate target robot poses that align the end effector with planned trajectories in the patient CT. Geometric accuracy of the registrations was independently evaluated for the patient and the robot in phantom studies.
Results: The resulting translational difference between the ground truth and patient registrations of a pelvis phantom using a single (AP) view was 1.3 mm, compared to 0.4 mm using dual (AP+Lat) views. Registration of the robot in air (i.e., no background anatomy) with five unique end effector poses achieved mean translational difference ~1.4 mm for K-wire placement in the pelvis, comparable to tracker-based margins of error (commonly ~2 mm). Conclusions: The proposed approach is feasible based on the accuracy of the patient and robot registrations and is a preliminary step in developing an image-guided robotic guidance system that more naturally fits the workflow of fluoroscopically guided orthopaedic trauma surgery. Future work will involve end-to-end development of the proposed guidance system and assessment of the system with delivery of K-wires in cadaver studies.

Entities:  

Keywords:  3D-2D registration; image-guided surgery; known-component registration; surgical robotics

Year:  2020        PMID: 36082206      PMCID: PMC9450105          DOI: 10.1117/12.2549713

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  10 in total

Review 1.  Minimally invasive mitral valve surgery: from Port Access to fully robotic-assisted surgery.

Authors:  Harry W Donias; Hratch L Karamanoukian; Giuseppe D'Ancona; Eddie L Hoover
Journal:  Angiology       Date:  2003-01       Impact factor: 3.619

2.  Robotic drill guide positioning using known-component 3D-2D image registration.

Authors:  Thomas Yi; Vignesh Ramchandran; Jeffrey H Siewerdsen; Ali Uneri
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-06

3.  Robotic assisted laparoscopic radical prostatectomy versus retropubic radical prostatectomy: a prospective assessment of postoperative pain.

Authors:  Todd M Webster; S Duke Herrell; Sam S Chang; Michael S Cookson; Roxelyn G Baumgartner; Laura W Anderson; Joseph A Smith
Journal:  J Urol       Date:  2005-09       Impact factor: 7.450

4.  Known-component 3D-2D registration for quality assurance of spine surgery pedicle screw placement.

Authors:  A Uneri; T De Silva; J W Stayman; G Kleinszig; S Vogt; A J Khanna; Z L Gokaslan; J-P Wolinsky; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-09-30       Impact factor: 3.609

5.  Intraoperative evaluation of device placement in spine surgery using known-component 3D-2D image registration.

Authors:  A Uneri; T De Silva; J Goerres; M W Jacobson; M D Ketcha; S Reaungamornrat; G Kleinszig; S Vogt; A J Khanna; G M Osgood; J-P Wolinsky; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2017-02-24       Impact factor: 3.609

6.  Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory.

Authors:  Bowen Jiang; A Karim Ahmed; Corinna C Zygourakis; Samuel Kalb; Alex M Zhu; Jakub Godzik; Camilo A Molina; Ari M Blitz; Ali Bydon; Neil Crawford; Nicholas Theodore
Journal:  Chin Neurosurg J       Date:  2018-09-03

7.  2D-fluoroscopic navigated percutaneous screw fixation of pelvic ring injuries--a case series.

Authors:  Florian Gras; Ivan Marintschev; Arne Wilharm; Kajetan Klos; Thomas Mückley; Gunther O Hofmann
Journal:  BMC Musculoskelet Disord       Date:  2010-07-07       Impact factor: 2.362

8.  3D-2D image registration for target localization in spine surgery: investigation of similarity metrics providing robustness to content mismatch.

Authors:  T De Silva; A Uneri; M D Ketcha; S Reaungamornrat; G Kleinszig; S Vogt; N Aygun; S-F Lo; J-P Wolinsky; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2016-03-18       Impact factor: 3.609

9.  Atlas-based automatic planning and 3D-2D fluoroscopic guidance in pelvic trauma surgery.

Authors:  R Han; A Uneri; T De Silva; M Ketcha; J Goerres; S Vogt; G Kleinszig; G Osgood; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2019-05-02       Impact factor: 4.174

Review 10.  Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review.

Authors:  Chandrasekharan Nair Kesavachandran; Frank Haamann; Albert Nienhaus
Journal:  Eur J Med Res       Date:  2012-10-30       Impact factor: 2.175

  10 in total
  2 in total

1.  Fluoroscopic Guidance of a Surgical Robot: Pre-clinical Evaluation in Pelvic Guidewire Placement.

Authors:  R C Vijayan; R Han; P Wu; N M Sheth; P Vagdargi; S Vogt; G Kleinszig; G M Osgood; J H Siewerdsen; A Uneri
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-25

2.  Preclinical evaluation of a prototype freehand drill video guidance system for orthopedic surgery.

Authors:  Niral Sheth; Prasad Vagdargi; Alejandro Sisniega; Ali Uneri; Gregory Osgood; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2022-08-26