| Literature DB >> 36068498 |
Yoshiyuki Fukugawa1, Ryo Toya2, Tomohiko Matsuyama1, Takahiro Watakabe1, Yoshinobu Shimohigashi3, Yudai Kai3, Tadashi Matsumoto1, Natsuo Oya1.
Abstract
BACKGROUND: Patients with tonsillar cancer (TC) often have dental fillings that can significantly degrade the quality of computed tomography (CT) simulator images due to metal artifacts. We evaluated whether the use of the metal artifact reduction (MAR) algorithm reduced the interobserver variation in delineating gross tumor volume (GTV) of TC.Entities:
Keywords: Gross tumor volume; Head and neck cancer; Metal artifact reduction; Oropharyngeal carcinoma; Radiotherapy; Radiotherapy planning
Mesh:
Year: 2022 PMID: 36068498 PMCID: PMC9450459 DOI: 10.1186/s12880-022-00889-0
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 2.795
GTVCONV and GTVMAR values for four observers
| Observer | GTV | Mean ± SD, cm3 | Range, cm3 | |
|---|---|---|---|---|
| A | GTVCONV | 15.7 ± 8.2 | 6.0–31.3 | 0.002 |
| GTVMAR | 12.9 ± 6.9 | 3.7–30.4 | ||
| B | GTVCONV | 16.1 ± 7.6 | 4.4–29.5 | 0.006 |
| GTVMAR | 13.3 ± 6.5 | 7.1–26.3 | ||
| C | GTVCONV | 11.4 ± 7.0 | 3.5–29.8 | 0.013 |
| GTVMAR | 9.9 ± 5.6 | 3.3–21.1 | ||
| D | GTVCONV | 12.6 ± 6.3 | 2.7–25.1 | 0.433 |
| GTVMAR | 12.3 ± 6.4 | 4.2–28.9 | ||
| All observers | GTVCONV | 14.0 ± 7.4 | 2.7–31.3 | < 0.001 |
| GTVMAR | 12.1 ± 6.4 | 3.3–30.4 |
GTV gross tumor volume, SD standard deviation, GTV gross tumor volume delineated based on conventional CT images, GTV gross tumor volume delineated based on the combination of conventional and metal artifact reduction CT images
Dice similarity coefficient of GTVCONV and GTVMAR for each pair and all four observers
| Observer | GTVCONV (Mean ± SD) | Range | GTVMAR (Mean ± SD) | Range |
|---|---|---|---|---|
| A and B | 0.71 ± 0.10 | 0.50–0.82 | 0.75 ± 0.10 | 0.49–0.84 |
| A and C | 0.76 ± 0.10 | 0.53–0.89 | 0.80 ± 0.10 | 0.63–0.87 |
| A and D | 0.75 ± 0.10 | 0.50–0.86 | 0.81 ± 0.10 | 0.48–0.89 |
| B and C | 0.71 ± 0.10 | 0.52–0.84 | 0.73 ± 0.10 | 0.51–0.87 |
| B and D | 0.75 ± 0.11 | 0.43–0.87 | 0.76 ± 0.11 | 0.47–0.87 |
| C and D | 0.76 ± 0.10 | 0.51–0.86 | 0.80 ± 0.10 | 0.58–0.91 |
| All observers | 0.74 ± 0.10 | 0.43–0.89 | 0.77 ± 0.10 | 0.47–0.91 |
GTV gross tumor volume delineated based on conventional CT images, SD standard deviation, GTV gross tumor volume delineated based on the combination of conventional and metal artifact reduction CT images
Fig. 1Examples of gross tumor volume (GTV) in patients with tonsillar cancer. Each panel includes axial (left), sagittal (center), and coronal (right) images. A GTVs were delineated by four observers using conventional computed tomography (CT) images. The mean Dice similarity coefficient (DSC) was 0.80 with a mean GTV value of 23.7 cm3. B GTVs were delineated by four observers based on conventional and metal artifact reduction fused CT images. The mean DSC improved to 0.83 with a mean GTV value of 21.6 cm3 based on the combination of conventional and metal artifact reduction CT images