Image quality evaluation of in-treatment four-dimensional cone-beam computed tomography in volumetric-modulated arc therapy for stereotactic body radiation therapy.

Yoshinobu Shimohigashi1, Yasuhiro Doi2, Yumiko Kouno2, Yohei Yotsuji2, Masato Maruyama2, Yudai Kai2, Ryo Toya3.   

Abstract

In this study, the image quality of in-treatment four-dimensional cone-beam computed tomography (In-4D-CBCT) obtained with various prescription doses (PDs) were quantitatively evaluated in volumetric-modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT) of the lungs and liver. To assess image quality, we used a dynamic thorax phantom and three-dimensional (3D) abdominal phantom; In-4D-CBCT images were acquired with various PDs (from 5 to 12 Gy). The In-4D-CBCT with various PDs were compared with the reference images (pre-4D-CBCT). The image quality was evaluated using the signal-to-noise ratio (SNR), the contrast-to-noise ratio (CNR), and the Dice similarity coefficient (DSC). The fiducial marker positions with various PDs were compared with those of the reference images. For the dynamic thorax phantom, the difference between pre- and In-4D-CBCT in terms of SNR and CNR decreased, as the PD increased from 6 to 12 Gy. The median DSC ranged from 0.7 to 0.74, and showed good similarity. For the 3D abdominal phantom, the difference between pre- and In-4D-CBCT in terms of SNR and CNR decreased as the PD increased from 5 to 6 Gy; conversely, it increased as the PD increased from 7 to 8 Gy. The fiducial marker positions were within 1.0 mm for all PDs. We concluded that the image quality of In-4D-CBCT degraded compared with the reference image; however, it was sufficiently accurate for assessing the intra-fractional tumor position in VMAT for SBRT of the lungs and liver both in terms of the target volume similarity and accuracy of the fiducial marker position.
Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Gantry rotation speed; Image quality evaluation; In-treatment four-dimensional cone-beam computed tomography; Stereotactic body radiation therapy

Mesh:

Year:  2019        PMID: 31726265     DOI: 10.1016/j.ejmp.2019.11.003

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Impact of metal artifact reduction algorithm on gross tumor volume delineation in tonsillar cancer: reducing the interobserver variation.

Authors:  Yoshiyuki Fukugawa; Ryo Toya; Tomohiko Matsuyama; Takahiro Watakabe; Yoshinobu Shimohigashi; Yudai Kai; Tadashi Matsumoto; Natsuo Oya
Journal:  BMC Med Imaging       Date:  2022-09-06       Impact factor: 2.795

2.  Hypofractionated palliative volumetric modulated arc radiotherapy with the Radiation Oncology Study Group 8502 "QUAD shot" regimen for incurable head and neck cancer.

Authors:  Ryo Toya; Tetsuo Saito; Kohsei Yamaguchi; Tomohiko Matsuyama; Takahiro Watakabe; Tadashi Matsumoto; Ryoji Yoshida; Akiyuki Hirosue; Daizo Murakami; Yorihisa Orita; Hideki Nakayama; Natsuo Oya
Journal:  Radiat Oncol       Date:  2020-05-27       Impact factor: 3.481

3.  Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study.

Authors:  Jihun Kim; Ki Chang Keum; Ho Lee; Chae-Seon Hong; Kwangwoo Park; Jin Sung Kim
Journal:  Radiat Oncol       Date:  2020-09-25       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.