| Literature DB >> 36059823 |
Xun Chen1,2,3, Fang Chen4, Xuanqi Wang1,2,3, Yilin Xu1,2,3, Mingrui Cheng1,2,3, Tian Han1,2,3, Xiaoying Wang1,2,3, Xingtao Zhou1,2,3.
Abstract
Introduction: To evaluate the safety and anterior chamber structure of implantation of the Evolution (EVO) implantable Collamer lens (EVO-ICL) in patients with short white-to-white (WTW) corneal diameters. Materials and methods: The study population was divided into two groups: the experimental group (34 eyes of 34 patients) with WTW corneal diameters of ≤10.6 mm and the control group (59 eyes of 59 patients) with WTW corneal diameters of >10.6 mm. The outcome measures included uncorrected distance visual acuity, corrected distance visual acuity, refractive power, intraocular pressure (IOP), anterior chamber angle, depth, volume, and vault.Entities:
Keywords: EVO-ICL; anterior chamber; myopia; safety; white-to-white
Year: 2022 PMID: 36059823 PMCID: PMC9428310 DOI: 10.3389/fmed.2022.928245
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Distribution of preoperative characteristics.
| Parameter | Experimental group | Control group | P value |
| N, eyes | 34 | 59 | |
| Age, years | 28.82 ± 5.37(20∼44) | 28.17 ± 6.23(20∼45) | 0.610 |
| logMAR UDVA | 1.44 ± 0.35(1.00∼2.00) | 1.47 ± 0.25(1.00∼2.00) | 0.668 |
| logMAR CDVA | 0.06 ± 0.19(-0.08∼1.00) | 0.02 ± 0.04(-0.08∼0.15) | 0.296 |
| Refractive errors (D) | |||
| Spherical | −9.57 ± 3.28(-5.25∼-17.50) | −9.75 ± 2.85(-4.50∼-17.75) | 0.770 |
| Cylindrical | −1.05 ± 1.10(0.00∼-3.25) | −1.03 ± 0.98(0.00∼-4.50) | 0.906 |
| Spherical equivalent | −10.09 ± 3.33(-5.25∼-17.50) | −10.26 ± 2.91(-4.50∼-17.75) | 0.797 |
| WTW diameter (mm) | 10.52 ± 0.12(10.1∼10.6) | 10.87 ± 0.05(10.7∼10.9) | < 0.001 |
| Horizontal STS | 11.10 ± 0.36(10.37∼11.93) | 11.20 ± 0.35(10.26∼12.08) | 0.197 |
| Vertical STS | 11.58 ± 0.46(10.55∼12.35) | 11.74 ± 0.40(10.77∼12.62) | 0.123 |
| IOP (mm Hg) | 16.17 ± 2.71(11.5∼20.1) | 16.42 ± 2.59(10.7∼21.2) | 0.657 |
| ACA (°) | 37.42 ± 4.77(26.8∼48.7) | 37.21 ± 5.05(26.8∼49.0) | 0.842 |
| ACD(mm) | 3.00 ± 0.22(2.51∼3.49) | 3.00 ± 0.22(2.64∼3.48) | 0.970 |
| ACV (ml) | 163.15 ± 30.11(108∼242) | 168.54 ± 25.55(108∼234) | 0.361 |
| ECD (cells/mm2) | 2611.59 ± 306.37(2021∼3634) | 2647.10 ± 236.54(1934∼3218) | 0.534 |
| ICL size (mm) | 12.1 | 12.1 |
N, number of eyes; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; D, diopters; WTW, horizontal white-to-white diameter; STS, sulcus to sulcus; IOP, intraocular pressure; ACA, anterior chamber angle; ACD, anterior chamber depth; ACV, anterior chamber volume; ECD, corneal endothelial cell density; ICL, implantable collamer lens. Results are expressed as means ± standard deviation (range).
FIGURE 1The percentage of eyes that gained/lost lines of CDVA after EVO Implantable Collamer Lens implantation between experimental and control groups.
FIGURE 2The cumulative percentage of UDVA after EVO Implantable Collamer Lens implantation between experimental and control groups.
FIGURE 3Scatter plot of attempted versus achieved correction (spherical equivalent) after EVO Implantable Collamer Lens implantation between experimental and control groups. The black solid line represented achieved correction = attempted correction, the black dotted line represented achieved correction = attempted correction ± 1.00 D.
FIGURE 4Intraocular pressure (A), Anterior chamber angle (B), Anterior chamber depth (C) and Anterior chamber volume (D) between experimental and control groups before and after implantation of EVO Implantable Collamer Lens.