Johanna Vuille-Dit-Bille1, Nicole Weingand1, Rahel Jud Schefer2, Martina Stirn3, Katja-Nicole Adamik1, Justus M K Rathmann4, Nadja E Sigrist2.
Abstract
Rotational Thromboelastometry (ROTEM) allows for the global assessment of hemostasis in whole blood samples. Preanalytical and analytical factors may influence test results, and data about the reliability and reproducibility of lyophilized ROTEM tests are scarce. Therefore, the objective of this study was to evaluate the influence of blood collection site on ROTEM S parameters and to assess intrarater and in-between device variability. A total of thirty, healthy, staff-owned dogs were included. Blood collection and ROTEM analysis were performed by trained staff according to a standardized protocol. Extrinsically activated (tissue factor; Ex-TEM S), with the addition of cytochalasin for platelet inhibition (Fib-TEM S), and intrinsically activated (In-TEM) analyses were performed. Analysis of our data showed significant variability for various Ex-TEM S and Fib-TEM S parameters from different collection sites and intrarater and in-between device measurements. We conclude that serial monitoring with ROTEM should be performed on the same device, with blood always taken from the same collection site using a standardized blood sampling technique. While In-TEM S, apart from maximum lysis, showed very stable and reliable results, we suggest interpreting especially clotting and clot formation parameters from Ex-TEM S and Fib-TEM S tests with caution and using duplicate measurements to detect outliers and to prevent initiation of incorrect therapies.Entities:
Keywords: ROTEM; canine; coagulation; coefficient of variation; sampling site; thromboelastometry
Year: 2022 PMID: 36009691 PMCID: PMC9405132 DOI: 10.3390/ani12162101
Source DB: PubMed Journal: Animals (Basel) ISSN: 2076-2615 Impact factor: 3.231
Comparison of jugular and venous ROTEM parameters in simultaneously measured blood samples.
| Parameter | Jugular | Saphenous | Jugular-Saphenous | Intraclass Correlation Coefficient (ICC) | CoV | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Unit |
| Mean | SD | Mean | SD | Mean Difference * | ICC | Lower CI | Upper CI |
| % |
| |
| Ex-TEM | |||||||||||||
| CT | s | 36 | 53✓ | 31 | 41✓ | 21 | 21 | 0.290 | −0.014 | 0.552 | 0.030 | 27 | <0.001 |
| CFT | s | 36 | 433 | 783 | 234✓ | 123 | 219 | 0.198 | −0.117 | 0.484 | 0.108 | 20 | 0.069 |
| α | ° | 36 | 50✓ | 11 | 55✓ | 10 | 7 | 0.545 | 0.200 | 0.755 | 0.002 | 11 | <0.001 |
| MCF | mm | 36 | 42✓ | 8 | 45✓ | 7 | 5 | 0.646 | 0.287 | 0.825 | 0.001 | 8 | <0.001 |
| MCE | 36 | 74✓ | 23 | 85✓ | 24 | 13 | 0.712 | 0.359 | 0.865 | <0.001 | 13 | <0.001 | |
| G | 36 | 3711✓ | 1162 | 4242✓ | 1191 | 675 | 0.712 | 0.351 | 0.866 | 0.001 | 13 | <0.001 | |
| ML | % | 35 | 4✓ | 2 | 4✓ | 2 | 1 | 0.290 | −0.014 | 0.552 | 0.030 | 21 | <0.001 |
| In-TEM | |||||||||||||
| CT | s | 18 | 202✓ | 20 | 188✓ | 17 | 26 | 0 | −0.383 | 0.369 | 0.588 | 9 | <0.001 |
| CFT | s | 18 | 127✓ | 48 | 132✓ | 43 | 22 | 0.814 | 0.574 | 0.926 | <0.001 | 12 | <0.001 |
| α | ° | 18 | 67✓ | 7 | 66✓ | 6 | 4 | 0.789 | 0.527 | 0.915 | <0.001 | 4 | <0.001 |
| MCF | mm | 18 | 57✓ | 7 | 57✓ | 7 | 2 | 0.934 | 0.832 | 0.975 | <0.001 | 2 | <0.001 |
| MCE | 18 | 139✓ | 42 | 140✓ | 49 | 11 | 0.941 | 0.850 | 0.978 | <0.001 | 5 | 0.001 | |
| G | 18 | 6966✓ | 2114 | 7012✓ | 2430 | 539 | 0.941 | 0.850 | 0.978 | <0.001 | 5 | 0.001 | |
| ML | % | 18 | 0.2✓ | 0.4 | 0.2✓ | 0.6 | 0.2 | 0.361 | −0.130 | 0.704 | 0.070 | 71 | 0.187 |
| Fib-TEM | |||||||||||||
| CT | s | 19 | 106✓ | 266 | 120 | 269 | 21 | 0.982 | 0.954 | 0.993 | <0.001 | 18 | 0.066 |
| MCF | mm | 19 | 4✓ | 2 | 6✓ | 4 | 2 | 0.246 | −0.189 | 0.613 | 0.133 | 21 | 0.027 |
| MCE | 19 | 5✓ | 2 | 6✓ | 5 | 2 | 0.168 | −0.269 | 0.561 | 0.227 | 21 | 0.053 | |
| G | 19 | 237✓ | 87 | 313✓ | 242 | 105 | 0.167 | −0.267 | 0.558 | 0.228 | 20 | 0.054 | |
✓ within the reference interval. CI—confidence interval; CT—clotting time; CFT—clot formation time; α—alpha-angle; MCF—maximum clot firmness; MCE—maximum clot elasticity (E = 100 × MCF/(100−MCF); G—calculated measure of total clot strength (G = 5000 × MCF/(100−MCF); ML—maximum lysis; Ex-TEM—tissue factor-activated temogram; In-TEM—elagic acid-activated temogram; Fib-TEM—tissue factor-activated temogram with thrombocyte inhibition. * average absolute difference between the saphenous and jugular measurement for each group.
Change of coagulation status with samples measured simultaneously from 2 different sampling sites (jugular vs. saphenous), as duplicates and on 2 different devices.
| Change of Coagulation Status Based on Collection Site | |||
|---|---|---|---|
| TEST | Coagulation Status | V. Jugularis | V. Saphena |
| Ex-TEM | normocoagulable | 33/36 (92%) | 32/36 (89%) |
| hypocoagulable | 2/36 (5%) | 1/36 (3%) | |
| hypercoagulable | 1/36 (3%) | 3/36 (8%) | |
| In-TEM | normocoagulable | 16/18 (89%) | 15/18 (83%) |
| hypocoagulable | 2/18 (11%) | 2/18 (11%) | |
| hypercoagulable | 0/18 (0%) | 1/18 (6%) | |
| Fib-TEM | normocoagulable | 18/19 (95%) | 17/19 (89%) |
| hypocoagulable | 1/19 (5%) | 1/19 (5%) | |
| hypercoagulable | 0/19 (0%) | 1/19 (5%) | |
|
| |||
|
|
|
|
|
| Ex-TEM | normocoagulable | 16/23 (70%) | 15/23 (65%) |
| hypocoagulable | 6/23 (26%) | 6/23 (26%) | |
| hypercoagulable | 1/23 (4%) | 2/23 (9%) | |
| In-TEM | normocoagulable | 13/17 (76%) | 13/17 (76%) |
| hypocoagulable | 4/17 (24%) | 4/17 (24%) | |
| hypercoagulable | 0/17 (0%) | 0/17 (0%) | |
| Fib-TEM | normocoagulable | 18/21 (86%) | 17/21 (81%) |
| hypocoagulable | 2/21 (9%) | 2/21 (10%) | |
| hypercoagulable | 1/21 (5%) | 2/21 (10%) | |
|
| |||
|
|
|
|
|
| Ex-TEM | normocoagulable | 4/7 (57%) | 3/7 (43%) |
| hypocoagulable | 3/7 (43%) | 4/7 (57%) | |
| hypercoagulable | 0/7 (0%) | 0/7 (0%) | |
| In-TEM | normocoagulable | 4/6 (67%) | 4/6 (67%) |
| hypocoagulable | 2/6 (33%) | 2/6 (33%) | |
| hypercoagulable | 0/6 (0%) | 0/6 (0%) | |
| FibTEM- | normocoagulable | 4/4 (100%) | 3/4 (75%) |
| hypocoagulable | 0/4 (0%) | 1/4 (25%) | |
| hypercoagulable | 0/4 (0%) | 0/4 (0%) | |
Ex-TEM—tissue factor-activated temogram; In-TEM—elagic acid-activated temogram; Fib-TEM—tissue factor-activated temogram with thrombocyte inhibition.
Comparison of paired simultaneous ROTEM measurements analyzed by the same operator on the same device/Investigation of intra-operator/intrarater variation.
| Parameter | Measurement 1 | Measurement 2 | Measurement 1–Measurement 2 | Intraclass Correlation Coefficient (ICC) | CoV | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Unit |
| Mean | SD | Mean | SD | Mean Difference * | ICC | Lower CI | Upper CI |
| % |
| |
| Ex-TEM | |||||||||||||
| CT | s | 23 | 79✓ | 56 | 50✓ | 23 | 39 | 0.098 | −0.229 | 0.445 | 0.290 | 34 | 0.001 |
| CFT | s | 23 | 558 | 713 | 616 | 759 | 429 | 0.225 | −0.214 | 0.582 | 0.153 | 32 | 0.018 |
| α | ° | 23 | 45✓ | 12 | 43✓ | 14 | 7 | 0.674 | 0.378 | 0.847 | <0.001 | 13 | <0.001 |
| MCF | mm | 23 | 37✓ | 10 | 37✓ | 11 | 6 | 0.711 | 0.427 | 0.867 | <0.001 | 13 | <0.001 |
| MCE | 23 | 62✓ | 27 | 63✓ | 31 | 15 | 0.801 | 0.586 | 0.911 | <0.001 | 19 | <0.001 | |
| G | 23 | 3124✓ | 1328 | 3164✓ | 1552 | 733 | 0.801 | 0.585 | 0.911 | <0.001 | 19 | <0.001 | |
| ML | % | 22 | 3✓ | 2 | 3✓ | 2 | 1 | 0.761 | 0.513 | 0.893 | <0.001 | 40 | <0.001 |
| In-TEM | |||||||||||||
| CT | s | 18 | 189✓ | 35 | 179✓ | 20 | 14 | 0.779 | 0.433 | 0.916 | <0.001 | 5 | <0.001 |
| CFT | s | 18 | 139✓ | 56 | 129✓ | 40 | 23 | 0.787 | 0.525 | 0.914 | <0.001 | 12 | <0.001 |
| α | ° | 18 | 66✓ | 8 | 67✓ | 6 | 4 | 0.793 | 0.538 | 0.917 | <0.001 | 4 | <0.001 |
| MCF | mm | 17 | 55✓ | 6 | 56✓ | 6 | 1 | 0.970 | 0.917 | 0.989 | <0.001 | 1 | 0.001 |
| MCE | 17 | 129✓ | 35 | 133✓ | 34 | 6 | 0.973 | 0.927 | 0.990 | <0.001 | 3 | <0.001 | |
| G | 17 | 6437✓ | 1768 | 6662✓ | 1676 | 302 | 0.973 | 0.928 | 0.990 | <0.001 | 3 | <0.001 | |
| ML | % | 16 | 0.1✓ | 0.2 | 0.1✓ | 0.5 | 0.2 | 0 | −0.566 | 0.453 | 0.581 | 141 | 0.188 |
| Fib-TEM | |||||||||||||
| CT | s | 21 | 107✓ | 258 | 212 | 573 | 111.95 | 0.720 | 0.436 | 0.875 | <0.001 | 21 | 0.128 |
| MCF | mm | 21 | 5✓ | 2 | 5✓ | 2 | 1 | 0.870 | 0.706 | 0.945 | <0.001 | 12 | 0.008 |
| MCE | 21 | 5✓ | 2 | 5✓ | 3 | 1 | 0.861 | 0.690 | 0.941 | <0.001 | 12 | 0.006 | |
| G | 21 | 237✓ | 113 | 246✓ | 137 | 37 | 0.870 | 0.709 | 0.945 | <0.001 | 13 | 0.004 | |
✓ within the reference interval. CI—confidence interval; CT—clotting time; CFT—clot formation time; αalpha-angle; MCF—maximum clot firmness; MCE—maximum clot elasticity (E = 100 × MCF/(100−MCF); G—calculated measure of total clot strength (G = 5000 × MCF/(100−MCF); ML—maximum lysis; Ex-TEM—tissue factor-activated temogram; In-TEM—elagic acid-activated temogram; Fib-TEM—tissue factor-activated temogram with thrombocyte inhibition. * average absolute difference between the first and second measurement for each group.
Comparison of paired simultaneous ROTEM measurements analyzed by the same operator on two different devices/investigation of in-between device variation.
| Parameter | Device 1 | Device 2 | Device 1–Device 2 | Intraclass Correlation Coefficient (ICC) | CoV | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Unit | Δ Mean | SD | Δ Mean | SD | Mean Difference * | ICC | Lower CI | Upper CI |
|
|
| |||
| Ex-TEM | ||||||||||||||
| CT | s | 7 | 88 | 38 | 7 | 68✓ | 26 | 34 | 0.132 | −0.535 | 0.754 | 0.362 | 33 | 0.031 |
| CFT | s | 7 | 648 | 297 | 7 | 944 | 890 | 656 | 0 | −0.772 | 0.711 | 0.511 | 46 | 0.048 |
| α | ° | 7 | 40 | 8 | 7 | 38 | 11 | 7 | 0.565 | −0.299 | 0.911 | 0.081 | 15 | 0.026 |
| MCF | mm | 7 | 32✓ | 6 | 7 | 32✓ | 8 | 5 | 0.710 | −0.059 | 0.945 | 0.032 | 12 | <0.001 |
| MCE | 7 | 48✓ | 14 | 7 | 50✓ | 20 | 10 | 0.782 | 0.139 | 0.959 | 0.014 | 16 | 0.001 | |
| G | 7 | 2415✓ | 713 | 7 | 2498✓ | 990 | 518 | 0.784 | 0.149 | 0.960 | 0.013 | 16 | <0.001 | |
| ML | % | 7 | 3 | 1 | 7 | 3 | 2 | 1 | 0.676 | 0.030 | 0.934 | 0.021 | 45 | <0.001 |
| In-TEM | ||||||||||||||
| CT | s | 6 | 192✓ | 28 | 6 | 187✓ | 32 | 7 | 0.956 | 0.744 | 0.994 | <0.001 | 2 | 0.057 |
| CFT | sec | 6 | 143✓ | 33 | 6 | 138✓ | 26 | 14 | 0.877 | 0.412 | 0.982 | 0.003 | 8 | 0.001 |
| α | ° | 6 | 64✓ | 5 | 6 | 66✓ | 4 | 3 | 0.789 | 0.170 | 0.967 | 0.011 | 3 | 0.002 |
| MCF | mm | 5 | 56✓ | 7 | 6 | 53✓ | 4 | 3 | 0.698 | 0.011 | 0.949 | 0.024 | 3 | 0.154 |
| MCE | 5 | 136✓ | 51 | 6 | 117✓ | 17 | 20 | 0.472 | −0.300 | 0.900 | 0.110 | 8 | 0.255 | |
| G | 5 | 6826✓ | 2581 | 6 | 5831✓ | 837 | 1020 | 0.469 | −0.302 | 0.899 | 0.111 | 9 | 0.249 | |
| ML | % | 4 | 0✓ | 0 | 6 | 0.1✓ | 0.2 | 0.1 | 0.000 | −0.755 | 0.755 | 0.500 | 141 | 0.363 |
| Fib-TEM | ||||||||||||||
| CT | s | 4 | 40✓ | 8 | 4 | 226 | 361 | 187 | 0.040 | −0.819 | 0.884 | 0.471 | 45 | 0.366 |
| MCF | mm | 4 | 4✓ | 1 | 4 | 3✓ | 1 | 1 | 0.500 | −0.606 | 0.958 | 0.170 | 23 | 0.016 |
| MCE | 4 | 4✓ | 1 | 4 | 3✓ | 1 | 1 | 0.572 | −0.522 | 0.965 | 0.130 | 24 | 0.016 | |
| G | 4 | 188✓ | 51 | 4 | 160✓ | 63 | 39 | 0.761 | −0.069 | 0.982 | 0.034 | 19 | 0.016 | |
✓ within the reference interval. CI—confidence interval; CT—clotting time; CFT—clot formation time; α—alpha-angle; MCF—maximum clot firmness; MCE—maximum clot elasticity (E = 100 × MCF/(100−MCF); G—calculated measure of total clot strength (G = 5000 × MCF/(100−MCF); ML—maximum lysis; Ex-TEM—tissue factor-activated temogram; In-TEM—elagic acid-activated temogram; Fib-TEM—tissue factor-activated temogram with thrombocyte inhibition; Δ Mean—average absolute difference between the respective first and second measurement for each device. * average absolute difference between device 1 and 2.