Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya.

Matilu Mwau1, Sven Schaffer2, Humphrey Kimani1, Purity Kasiano1, Francis Ogolla1, Elizabeth Ajema1, Scriven Adoyo1, Ednah Nyairo1, Norah Saleri1, Sangeetha Vijaysri Nair3.   

Abstract

INTRODUCTION: HIV-1 viral Load (VL) testing is recommended for the monitoring of antiretroviral treatment. Dried Blood Spots (DBS) are an effective sample type in resource limited settings, where safe phlebotomy and reliable shipping are hard to guarantee. In HIV high burden countries, high throughput assays can improve access to testing services. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima Assay) is a high throughput assay that runs on the CE-IVD approved Panther platform. The objectives of this study were to assess the performance characteristics of Aptima for VL monitoring using plasma and venous DBS specimens and to determine the stability of HIV-1 RNA in DBS.
MATERIALS AND METHODS: This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples while paired DBS samples and plasma were prepared from the remaining 915 samples. The agreement between the Aptima assay and the Abbott RealTime HIV-1 Assay (Abbott RT) was analysed by comparing the HIV-1 VL in both assays at the medical decision point of 1000 copies/mL. To assess stability of HIV-1 RNA in DBS, VL in DBS spotted on day 0 were compared with that from the same DBS card after 21 days of storage at room temperature.
RESULTS: Overall, 436 plasma samples had quantifiable results in both Aptima and Abbott RT. The agreement between the two assays at 1000 copies/mL was 97.48% with a Pearson's correlation coefficient (r) of 0.9589 and gave a mean bias of 0.33 log copies/mL on Bland-Altman analysis. For fresh DBS, the agreement in both assays was 94.64% at 1000 copies/mL, with an r of 0.8692 and a mean bias of 0.35 log copies/mL. The overall agreement between DBS tested in Aptima on day 0 versus day 21 was 95.71%, with a mean bias of -0.154.
CONCLUSION: The Aptima HIV-1 Quant Dx assay is an accurate test for VL monitoring of HIV-1 using DBS and plasma sample types in Kenya.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35994447      PMCID: PMC9394820          DOI: 10.1371/journal.pone.0269838

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.752


Introduction

More than 70% of the 37.7 million people living with HIV-1 in 2021 were on antiretroviral treatment (ART) [1, 2]. The World Health Organization (WHO) strongly recommends regular viral load (VL) testing to enable early detection of treatment failure, prevent emergence of drug resistance, and minimize transmission [3, 4]. Many countries especially in Africa have scaled up VL testing [5] to meet the WHO recommendations and to help end the AIDS epidemic [2]. In 2020, at least 75% of the 1.4 million people living with HIV-1 in Kenya were on ART and needed routine VL testing to monitor effectiveness of treatment [1]. Through nine reference laboratories and using both plasma and DBS, the national VL program delivered 850000 tests [6]. Most facilities offering comprehensive care for HIV in Kenya are located far from reference laboratories, and therefore DBS collection and transportation enables decentralization of specimen collection and increases access to VL monitoring. It also enables more timely detection for treatment failure which is crucial in reducing HIV transmission and emergence of drug resistance. Approximately 20% of specimens used for VL monitoring in Kenya are DBS. In Kenya, VL tests are conducted using Abbott and Roche assays running on the m2000 or COBAS Ampliprep/COBAS TaqMan platforms respectively. These platforms are medium throughput and since number of samples often exceeds the capacity to test, backlogs are quite common [7]. These backlogs could be reduced either by building the capacity of regional laboratories to test or by providing platforms with higher throughputs in the reference laboratories. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima Assay) is an in vitro nucleic acid amplification test that runs on the high throughput Panther Platform. It is intended for use in the detection of Human Immunodeficiency Virus type 1 (HIV-1) in human plasma, serum and DBS from infected individuals [8, 9]. The assay also quantifies HIV in human plasma and DBS over ranges of 30 to 10,000,000 copies/mL and 883 to 10,000,000 copies/mL, respectively. Data on the performance of the Aptima VL Assay has been published mainly from studies conducted in the United States and Europe [10-20]. Few publications on Aptima performance have come out of Africa [17, 19], although Africa has the greatest genetic diversity of HIV-1 [21, 22]. In a previous study [23], we investigated the performance of the Aptima HIV Quant Dx assay using fingerstick and venous dried blood spots prepared under field conditions. In that study, we did not compare HIV-1 VL in Aptima with those in other assays using DBS. Instead it was compared to the Aptima plasma HIV-1 VL. In fact, whereas multiple studies compared VL results of paired DBS and plasma samples at the medical decision point of 1000 copies/ mL by testing both sample types in Aptima [23-26], there is no published information comparing the performance of DBS specimens tested in Aptima and other assays at the medical decision point (MDP) of 1000 copies/mL. The purpose of this study was to compare the performance of the Aptima assay to Abbott RT assay for VL monitoring using both plasma and DBS specimens collected from patients in Kenya. The stability of HIV in DBS specimens was also assessed in this study.

Materials and methods

Ethics statement

This study was approved by the Institutional Review Board of the Kenya Medical Research Institute under protocol numbers KEMRI/SERU 2457 and 3544. It was conducted in accordance with the ethical standards of the Helsinki Declaration of 1975 as revised in 2000.

Study design

This was a cross-sectional prospective study conducted in 2017 and 2018. Firstly, the performance of Aptima assay on the Panther platform was compared with the reference Abbott-RealTime HIV-1 (Abbott RT) assay using plasma prepared from venous blood samples collected from HIV positive patients enrolled in Nairobi and Busia. The performance of DBS was evaluated from a further 915 venous blood samples collected from HIV positive patients enrolled in Nairobi. All venous blood was sent to the laboratory where it was either separated into plasma or spotted on to DBS cards. To accurately assess the stability of HIV-1 RNA in DBS, the venous DBS samples prepared for the first comparison were retested on day 21 on both Abbott RT and Aptima Assays. DBS was packaged with desiccants and stored at room temperature for the 21 days duration of the DBS stability study.

Study population

The study enrolled a cross-section of HIV positive adults receiving care and treatment in health facilities in Nairobi and Busia and who gave written informed consent. The study participants were mainly patients whose VL was being monitored because they were on antiretroviral therapy and a very few who were yet to initiate treatment. Plasma collected from both sites were tested in both assays. Venous DBS tested in this study were prepared from blood collected from a subset of study participants in Nairobi. Due to this there is a difference in number of plasma and DBS samples that were tested. A small subset of DBS samples did not have the paired plasma results due to a failed run on m2000.

Sample collection and preparation

Whole blood was collected from 2227 participants by phlebotomy, shipped to KEMRI HIV laboratories within 6 hours of collection, and processed within 12 hours of receipt. A total of 1312 samples provided only plasma. Paired plasma and DBS samples were prepared from a further 915 venous samples. To prepare plasma, the whole blood samples were centrifuged at 1,100g for 10 minutes within 24 hours of collection, and the plasma was stored at -80°C. To prepare DBS, 70μl of venous blood was spotted in each of ten spots (2 DBS cards) per patient. The DBS samples were allowed to dry overnight. The first DBS card for each patient was used for day 0 testing, while the DBS cards for the 21-day time point were packaged with desiccants and stored at room temperature. All samples were de-identified prior to Aptima testing.

Laboratory methods

All plasma and DBS samples were first tested using the Abbott platform [27]. Patient results were generated from the plasma samples tested on the Abbott platform. All DBS testing in Abbott was performed with 2 spots. Briefly, each of two spots was gently pushed out of the DBS card into a centrifuge tube, which was then filled with 1.3 mL of mDBS buffer. Centrifuge tubes were incubated at room temperature for 30 min. Thereafter, 1 mL of eluent was pipetted out of each centrifuge tube into a secondary tube for testing according to the manufacturer’s instructions [27]. To determine the performance of Aptima assay, both plasma and venous DBS were tested on Panther system [8, 9]. All DBS testing in Aptima was performed with one spot. In each case, 1 mL of DBS extraction buffer was added into a specimen aliquot tube after which one spot was carefully introduced into the tube. Specimen aliquot tubes were rocked at room temperature for 30 minutes, centrifuged for 2 min at 3000 rpm, arranged on a rack and introduced into the panther machine for processing. For plasma samples, 0.75mL was aliquoted into secondary Aptima specimen aliquot tubes, loaded onto racks and introduced into the panther machine for processing [8, 9]. The system draws 0.5mL from each tube for the assay. Fifteen tubes were loaded onto each rack, for a maximum of 6 racks. The seventh rack was loaded with four samples only while the eighth rack was loaded with a Negative Control, a Low Positive Control, a High Positive Control and a Calibrator. The racks were transported into their appropriate lanes, the bay doors closed, and processing initiated. Initial results were available in 3.5 hours, with five results received every five minutes thereafter. Results were posted as either “Not Detected” or “Invalid”, or as “copies/mL”. To assess the effect stability of HIV-1 RNA in DBS over time, results of DBS tested on day 0 were compared with DBS tested on day 21 in Aptima. For the day 0 condition, freshly prepared DBS was tested. DBS packaged with desiccants in a Ziploc bag and stored at room temperature was tested for the 21-day stability timepoint. Plasma results for Abbott and Aptima were reported without conversion. For the Abbott two spot DBS protocol, a conversion factor was applied by the software to convert results to copies/ mL of HIV-1. For Aptima DBS, a conversion factor was applied by the research team to convert results to copies/ mL of HIV-1. The linear quantitative range for Aptima using DBS was 883–10,000,000 copies/mL, while for the Abbott RT assay for plasma sample for 1 mL and 0.6mL protocol is 40–10,000,000 copies/mL, while that for the 0.5 mL protocol is 75–10,000,000 copies/mL. For this study, testing was performed with 0.6 mL of plasma sample. The linear quantitative range of Abbott Real-Time assay using the two spot DBS protocol is 550–10,000,000 copies/mL.

Data analysis

Only valid results were used for analyses. VL data were transformed into log10 copies/mL. Statistical analysis was performed using Stata/MP version for Mac. In the correlation analysis, only the VL data from patients who had quantitative values in both Aptima Assay and Abbott RT were included for analysis. The correlation was determined by simple linear regression with generation of Pearson’s correlation coefficient (r) as well as Bland–Altman analysis [28] to calculate the bias and the limit of agreement between assay results. The agreement between the assays for the plasma and DBS sample types were determined at the clinical decision point of 1000 copies/mL using a contingency table.

Results

Test outcomes

A total of 2227 venous samples were successfully collected. For the plasma study, 1312 plasma samples were tested on both the Abbott and Aptima Assays. On the Abbott RT Assay, 713 samples returned “not detected” results. One sample failed while 598 samples had detectable VLs; of those, 154 had <75 copies/mL while one (1) had <40 copies/mL. On the Aptima assay, 965 samples had quantifiable results while 347 returned “not detected” results. Of the quantifiable results, 451 had <30 copies/mL, while one sample had >10,000,000 copies/mL. For the DBS study, 915 samples were tested on both assays. In all, 693 DBS samples had detectable VLs on the Abbott RT assay: of those, 6 had <550 copies/mL. A further 222 samples returned “not detected” results. For the Aptima Assay, 866 samples had quantifiable VLs on DBS. Of those, 213 had 870copies/mL while 3 had 10,000,000 copies/mL. Only 49 samples had “not detected” results.

Agreement between Aptima and Abbott RT VL assay for various sample types at the clinical cut-off of 3 log copies/mL (1,000 copies/mL)

Plasma HIV-1 VL was conducted successfully for 1311 of the 1312 patients in both Aptima and Abbott RT assays and the results used for comparison. The positive agreement was 98.42% (95% CI 96.35%-99.32%), the negative agreement 97.19% (95% CI 95.97–98.05) and the overall agreement 97.48%. Fresh DBS from 915 patients were also tested and the results compared between the two assays. The positive agreement was 95.35% (95% CI 93.47%-96.70%) while the negative agreement was 92.77% (95% CI 88.86%-95.38%). The overall agreement was 94.64%. Fresh DBS was compared with three weeks old DBS for 910 patients to evaluate the stability of HIV-1 RNA in DBS; the positive agreement was 95.05% (95% CI 93.1%-96.5%) while the negative agreement was 97.34% (95%CI 94.6%-98.7%). The overall agreement was therefore 95.71% (Table 1).
Table 1

Agreement between Aptima and Abbott RT VL assay at a clinical cut-off of 3 log copies/mL (1,000 copies/mL) of HIV for the various sample types.

Result (95%CI)
Abbott RT Plasma
Aptima Plasma <3.0 >3.0 Total
<3.0 968 (97.19%)5 (1.58%)973 (74.16%)Negative Agreement97.19% (95.97%-98.05%)
>3.0 28 (2.81%)311 (98.42%)339 (25.51%)Positive Agreement98.42% (96.35%-99.32%)
Total 996 (75.91%)316 (24.09%)1,312 (100.00%) Total Agreement 97.48% (96.49%-98.20%)
Abbott RT DBS
Aptima DBS <3.0 >3.0 Total
<3.0 231 (92.77%)31 (4.58%)262 (28.63%)Negative Agreement92.77% (88.86%-95.38%)
>3.0 18 (7.23%)635 (95.35%)653 (71.37%)Positive Agreement95.35% (93.47%-96.70%)
Total 249(27.21%)666(72.79%)915 (100.00%) Total Agreement 94.64% (92.99%-95.93%)
Aptima DBS day 0
Aptima DBS day 21 <3.0 >3.0 Total
<3.0 256 (97.34%)32 (4.95%)288 (31.65%)Negative Agreement97.34% (94.61%-98.70%)
>3.0 7 (2.66%)615 (95.05%)622 (68.35%)Positive Agreement95.05% (93.10%-96.48%)
Total 263 (28.90%)647 (71.10%)910 (100.00%) Total Agreement 95.71% (94.19%-96.85%)
The discordant rates between the two assays were 2.52% (33/1312) and 5.36% (49/915 samples) for plasma and DBS respectively. Of the 33 discordants for plasma, 28 had <1000 copies/mL for Abbott and >1000 copies/mL for Aptima while 5 samples had VL >1000 copies/mL in Abbott and <1000 copies/mL in Aptima. Of the 49 patients with discordant DBS results, 31 were above MDP for Abbott RT and below MDP for Aptima with the reverse being true for the remaining 18 patients. A total of 17/31 (54.84%) patients with Aptima DBS <3.0 and Abbott DBS >3.0 log copies/mL had a plasma VL <3.0 log copies/mL (1000 copies/mL) in both Aptima and Abbott assays. For 15 of these 17 patients, DBS VL was <3.0 log copies/mL on testing on day 21 in Aptima as well. For the remaining 14/31 patients, the plasma VL was between >3.0 log copies/mL in Aptima, with 12 of them being >3.0 in Abbott assay. Of the 18 patients with discordant results with Aptima DBS VL >3.0 and Abbott DBS VL <3.0 log copies/mL on day 0, 8 (44.4%) had Aptima DBS VL >3.0 log copies/mL on day 21.The remaining 10 (55.56%) had Aptima DBS VL <3.0 log copies/mL on day 21. The Aptima plasma VL was >3.0 log copies/mL for the 8 samples and <3.0 log for all the 10 samples.

Regression analysis of the Aptima assay for VL testing

Four hundred and thirty-six (436) plasma samples had quantifiable results on both platforms and were used for correlation analysis. A significant correlation was found with a slope of 1.035 and r of 0.9589 (Fig 1).
Fig 1

HIV VL comparison for 436 plasma samples tested in Aptima and Abbott RT assays.

Bland-Altman analysis yielded a mean bias of 0.33 log copies/mL with Aptima VL results being higher than Abbott RT. The upper and lower limits of agreement were 1.018 to -0.359; 7.11% of the comparisons lay outside the limits of agreement (Fig 2).
Fig 2

Bland Altman analysis of HIV VL results for 436 plasma samples tested in Aptima and Abbott RT assays.

Comparison of fresh DBS tested on Aptima DBS and Abbott RT assays

Six hundred and thirty-six (636) fresh DBS samples had quantifiable results in both assays and were used for correlation analysis. On correlation, the slope was 0.9175 while r was 0.8692 (Fig 3).
Fig 3

HIV VL comparison in fresh DBS tested in Aptima and Abbott RT assays.

Bland-Altman analysis yielded a mean bias of 0.35 log copies/mL (95% CI 1.04 and -0.34) across the assay range; 4.72% of the comparisons lay outside the limits of agreement (Fig 4).
Fig 4

Bland Altman analysis of HIV VL results in fresh DBS tested in Aptima and Abbott RT assays.

Fresh DBS compared with 421-day old DBS on the Aptima Assay

Six hundred and three (603) samples had quantifiable DBS results on day 0 and day 21. On correlation, the slope was 0.9328 while r was 0.9339 (Fig 5).
Fig 5

HIV VL in DBS tested on day 0 and day 21 in the Aptima assay.

Bland-Altman analysis yielded a mean bias of -0.154 log copies/mL. The upper and lower limits of agreement were -0.644 to 0.336; 5.97% of the comparisons lay outside the limits of agreement (Fig 6).
Fig 6

Bland Altman analysis of HIV VL in DBS tested on day 0 and day 21 in the Aptima assay.

Discussion

To meet UNAIDS 90-90-90 treatment goals countries with high HIV burden need to perform millions of tests each year for HIV diagnosis and monitoring [1, 2]. The 10 centralized HIV testing labs in Kenya are often unable to meet this demand resulting in sample backlogs and long turnaround times to results because the equipment’s they use for testing are not high throughput. In this study we evaluated the Aptima Assay for use in Kenya not only because it is an automated high throughput assay but also because it recently received CE IVD and WHO approval for HIV VL monitoring with both plasma and DBS specimens. We observed a total agreement of 97.48% between Aptima and Abbott RT at the medical decision point of 1000 copies/mL on testing plasma specimens. In this comparison, both the positive and negative agreement were greater than 95%. Overall, the assay met the requirement of concordance at the clinical cut-off of 1000 copies/mL. However, there were 33 specimens out of 1312 (2.5%) with discordant results at the MDP of 1000 copies/mL. A majority of these (28) had HIV VL >1000 copies/mL in Aptima but < 1000 copies/mL in Abbott RT. Only 5 specimens had VL >1000 copies/mL in Abbott RT and <1000 copies/mL in Aptima of which 3 had results that were within 0.5 log of each other in the two assays. Discordant resolution was not performed for these specimens due to sample volume limitations. Linear regression showed excellent agreement in plasma VL across the assay range with an r of 0.9589, slope of 1.035 and intercept of 0.1971. The slope is similar to that reported between Aptima and Abbott RT by other investigators in Europe [11, 20] and the United States [17, 18]. Bland Altman analysis showed that the VL in Aptima was 0.33 logs higher than Abbott RT which is in line with the 0.11 to 0.30 logs difference reported between the two assays by other investigators [11, 14, 17, 18]. Most Kenyan patients are infected with HIV subtype A or C and other investigators have reported a slightly higher bias between the two assays for these subtypes [11, 17, 18]. Currently, approximately 20% of the VL monitoring in Kenya is conducted with DBS specimens most of which are tested using the Abbott RT assay. The overall agreement between Aptima and Abbott RT assays using DBS was 94.64% with positive and negative agreements of 95.35% and 92.77% respectively. Linear regression gave a slope of 0.9175 and an r of 0.8692. The 0.3 log bias seen between Aptima and Abbott for plasma specimens was also seen with DBS. Although discordant resolution was not performed by testing the DBS is an independent assay, it was possible to assess the accuracy of DBS quantification in both Aptima and Abbott RT in much more detail by comparing it to the VL result of plasma from the same patient, tested in both assays. Of the 49 discordant DBS results between Aptima and Abbott RT around the MDP of 1000 copies/mL, 31 had results below MDP in Aptima and above MDP for Abbott RT while the reverse was true for the remaining 18 samples. Among the 31 patients with Abbott DBS VL >3 log copies/mL and Aptima DBS VL <3 log copies/mL, 17 of the patients had VL <3 log copies/mL in both Aptima plasma and Abbott plasma suggesting that these were upward misclassifications by Abbott DBS. The remaining 14 of the 31 patients with Aptima DBS VL <3 log copies/mL and Abbott DBS VL>3 log copies/mL had plasma VL >3.00 log copies/mL in both Aptima plasma and Abbott plasma suggesting that these were upward misclassifications by Abbott DBS. Among the 18 discordant samples with Abbott DBS VL <3 log copies/mL, 8 of them had Aptima plasma and DBS results >3 log copies/mL. Five of the 8 also had Abbott plasma VL >3 log copies/mL. This suggests that these samples are likely to be downward misclassifications (under quantification) by Abbott DBS. Notably, among the 31 patients with Abbott DBS VL >3 log copies/mL and Aptima DBS VL < 3 log copies/mL, 17 of the patients also had VL <3 log copies/mL in both Aptima plasma and Abbott plasma suggesting that these were upward misclassifications by Abbott DBS. From these two situations, it can be concluded that significant variability may occur while testing DBS with concentrations close to the MDP of 1000 copies/mL in this study, regardless of the assay. This calls for further research, even though testing using a single replicate for DBS appears to be a contributing factor. Nonetheless, the overall misclassification rate for Aptima DBS (24/49) was comparable to that seen with Abbott DBS (25/49) at the MDP of 1000 copies/mL. Although only venous DBS were tested in this study, we previously compared the recovery of HIV-1 from venous DBS, fingerstick DBS, and plasma with the Aptima test [23] showing good agreement between plasma, venous DBS, and fingerstick DBS. One limitation of the current study is that no replicate testing or retesting of invalid specimens was performed due to volume limitation for these clinical samples. The agreement on paired DBS tested on day 0 versus after storage at room temperature for 21 days was 96.48%. There was excellent correlation (r = 0.9339) and minimal mean bias (-0.154 logs). This indicates that there was no significant degradation of HIV-1 RNA in DBS even after three weeks of storage at room temperature with desiccants. This is relevant because it often takes up to 3 weeks for DBS specimens to be delivered to the laboratory for testing. This duration of stability is in line with Hologic’s claims that HIV-1 RNA is stable in DBS stored at 30 and 40°C with relative humidity of up to 85% for at least 2 weeks [9]. Monleau et al has reported that DBS is stable for 3 months at 20°C but it is not clear whether the relative humidity for the conditions of storage was reflective of the high relative humidity seen in Nairobi, Kenya [29]. The data that we present is relevant because there is no published data currently available comparing the VL in Aptima with Abbott RT assay at the medical decision point of 1000 copies/mL for both plasma and DBS specimens. The data in this publication generated show that there is excellent agreement between the two assays at 1000 copies/mL with plasma from over 1300 Kenyan patients. The same applies for the comparison of DBS specimens tested in the two assays. This close agreement with the Abbott RT assay implies that the Aptima Assay can be used interchangeably with Abbott RT assay for HIV VL determination in plasma and DBS. The excellent performance of Aptima for both plasma and DBS sample types meet the WHO recommendations for performance for VL monitoring [30]. Apart from the good performance, we observed some operational advantages in using the Aptima assay in the laboratory workflow during the study. Aptima is run on the Panther instrument platform that is fully automated and allows random and continuous loading of test samples. It allows loading of up to 515 tests in 12 hours, with the first results being returned in about 3.5 hours. In contrast, the Abbott m2000 platform allows 186 results in 10 hours, with the first results being returned in about 5 hours. This enables high flexibility to adapt to low or high-throughput testing. The one metre square footprint of Panther also makes it a good fit for laboratories with space constraints.

Conclusions

In conclusion the performance of Aptima HIV-1 Quant Dx Assay was equivalent to that of Abbott RT for quantification of both plasma and DBS specimens from patients in Kenya. The high throughput, random access and complete automation provided by the Panther system enabled us to scale up testing up to 1000 specimens per day. This makes the Aptima Assay on Panther a good solution for clinical laboratories that need to close the gap between current testing capacity and UNAIDs goals for HIV-1 testing. It can help meet the increased demand for testing in any laboratory that performs VL testing. 22 Oct 2021
PONE-D-21-27672
Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya
PLOS ONE Dear Dr. Mwau, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Two reviewers and I have provided comments that must be addressed for further consideration of the manuscript, especially in addressing the lack of replicate testing and problems with the statistical analysis.
Please submit your revised manuscript by Dec 06 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Julie AE Nelson, PhD Academic Editor PLOS ONE Additional Editor Comments (if provided): The major issues to address are the inconsistent numbers throughout, the lack of replicate testing, and the higher rate of quantitative values from DBS than from plasma. These and other issues are detailed below: Line 17: The background section is too long—drop one or two of the sentences. Line 32: Add that venipuncture was the source of blood for the DBS in this study. Line 33: Use the correct name of the Abbott assay: Abbott RealTime HIV-1 Assay Line 37: Drop the sentence about the Stata software, since that is not needed in the abstract. Make sure everything is the same between the text and the abstract and the figures (436 vs 439; r = 0.96 vs R = 0.92—need to check both the number and the capitalization for that one; r = 0.87 vs R = 0.8692—need to check number of significant digits and capitalization; mean bias for DBS storage = -0.154 vs -0.15). Be consistent with your significant digits and with small case r throughout. Line 46: Drop the sentence about the discordance percentages since they are just obtained by subtracting the agreement percentages from 1 and provide nothing new. Line 48: Remove this last sentence of the results about the Panther platform throughput—not needed in the abstract. Line 52: this section needs more about the interpretation of the data in the results section and could mention that the assay worked well on samples in Kenya specifically. Line 74: small case m2000 Line 75: there is a (ref) that needs the citation added. Line 79: add the full name of the Aptima assay here: Hologic Aptima HIV-1 Quant Dx Assay. Line 110: Fix the name of the Abbott assay: Abbott RealTime HIV-1 Assay. Line 136: clarify what is meant by testing DBS on day 0. Is that the day after spotting with no time in the presence of desiccants? Line 144: the lower limit for the Abbott assay is 40 for both the 1mL and 0.6mL protocols, and 75 for the 0.5mL protocol, but all the linear ranges can be removed from the methods except possibly for the DBS protocols for each assay. Since all the analysis is with the 1000cp/mL cut-off, the lower limits for plasma are not needed. However, it would be useful to specify which volume protocol was used for the Abbott plasma testing since that is not stated. Line 167: The use of positive predictive values and negative predictive values in this context is incorrect since there does not appear to be a “gold standard” to show which samples should be positive and which should be negative, suggesting that one of the assays was considered the gold standard for the calculations. All these calculations should be removed from the manuscript. Line 169: There is a difference in number of plasma and DBS samples included in the data—add an explanation of the difference, whether this was because DBS hadn’t been made for some, the DBS data were invalid, or what other reason. Line 176: remove the word “therefore”. Line 178: remove the sentence about discordance rate Line 184: discordance rates not shown in Table 1 but can be obtained with the positive agreement values in Table 1, so re-write this statement. This paragraph (lines 184-191) is difficult to understand in terms of useful results. It is useful to the reader to know how many of the discordant results were due to higher detection in one assay vs the other, but the value of the rest of the “results” is less clear. What is the goal of reporting that 17 of the discordant samples (Abbott higher than Aptima in DBS) had corresponding plasma VL that were higher than 3631cp/mL? Is that VL cutoff a magic number for these assays? There is no mention of these “results” in the discussion, providing no context for why this “result” should be noted. Table 1: the two sides of the table are not related because different numbers of samples are used for each side. Therefore, remove the agreement data from the table and make a new table of these values, adding the number of samples included in each calculation. Also add the DBS day 21 calculations that are not shown in the current table. As mentioned above, remove the PPV and NPV data. Line 277: add the daily throughput of the Abbott m2000 as a comparator here to strengthen your argument that the Panther has higher throughput. Alternatively, add the throughput (tests per day) in line 75 where you describe it as medium throughput. Line 218: the discussion needs to start with discussing the results that have just been presented in the previous section, so move the first paragraph down. Reorganize the discussion to move the summary of the results all to the top paragraph(s), then start to discuss them with other published studies in subsequent paragraphs. Line 232: remove the discordance rate here since it is essentially a retelling of the initial result of agreement. Line 234: change “several” to the actual number of these specimens. This is also a good place to add discussion of the higher number of DBS with quantifiable VL than matched plasma with quantifiable VL, which is a critical question in this study. No replicate testing is mentioned, not even for invalid results. When running assays on samples that are near the lower limit of detection/quantitation, replicate testing can be critical. Therefore, this must be discussed as a limitation of the study in the discussion. Check all the references that they are cited in the manuscript. Your previous publication in PLoS One from April is highly relevant, especially when discussing the limitation that you did not compare fingerstick DBS since you previously looked at that in the April publication. Add the versions and dates used the package inserts in references 8 and 26. Figures 1-3: All the (a) figure legends must be changed to be more like the (b) legends where the “Figure 1a” is not part of the title and contain sufficient information to stand alone. For example: Figure 1a: HIV-1 viral load comparison for 439 plasma samples tested on both the Aptima HIV-1 Quantitative RNA assay and the Abbott HIV-1 RealTime assay. Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Please amend your current ethics statement to address the following concerns: a) Did participants provide their written or verbal informed consent to participate in this study? b) If consent was verbal, please explain i) why written consent was not obtained, ii) how you documented participant consent, and iii) whether the ethics committees/IRB approved this consent procedure. 3. Thank you for stating the following in the Competing Interests/Financial Disclosure * (delete as necessary) section: “Hologic, Inc provided the funds and materials for the study.” We note that you received funding from a commercial source: “Hologic, Inc.” Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc. Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your amended Competing Interests Statement within your cover letter. We will change the online submission form on your behalf. 4. Thank you for stating the following in the Competing Interests section: “Sangeetha Nair and Sven Schaffer were employees of Hologic, Inc at the time of the study. The rest of the authors, based at Kenya Medical Research Institute, declare no conflict of interest.” Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: No ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: No ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This is an easy to read manuscript providing important data relevant to increasing access to HIVVL testing in Africa. have the following comments. line 55 - Update the HIV statistics to the most recent. line 95 - I suggest "preferred reference in Kenya' rather than gold standard. line 107 - Study Design -Was there sufficient plasma to test the same samples on both assays i.e. same plasma samples tested on Aptima and Abbott? line 118 - Study Population- inclusion criteria needs to be more detailed, e.g., on ARVs or not, consecutive patients, routine clinic visits line 127 - Was plasma separated on the same day as DBS preparation? line 132 - How soon after collection were the Abbott tests performed? line 134- What is the time difference between Aptima and Abbott tests? Some detail is required on the Aptima procedure as this is the "study" assay. line 142 - Input volume for the Aptima ? line 144 - Which protocol was used for the Abbott? line 167 - Results - was prevalence taken into account for the PPV and NPV? line 184 - It maybe easier to describe discordance as upward and downward misclassification and the range of the VL obtained for the misclassified samples according to sample type. lines 281 - 226 -Discussion - This part is more suited to the introduction. line 234 - is there an explanation for this finding? line 242 - is there a reason for this observation? line 250 - is there a reason for this limitation? line 277 - is this per 8/12/24 hrs? This will give a better idea of throughput. Reviewer #2: Mwau et al. have investigated the performance of the aptima HIV1 assay compared to the Abbott real-time assay on plasma and DBS samples, at the 1000cp/mL decision point, as well as data on HIV RNA stability at room temperature. 1312 paired plasma and venous blood DBS samples were used. Total agreement between Abbott and Aptima assays at the 3 log threshold was 97.48% for plasma and 94.64% for DBS. Data on quantification agreement between assays, and over time for DBS are provided. Major points It is not clear at abstract’s reading that authors were working on DBS prepared from venous blood and not from finger-stick. The number of quantifiable samples is different in abstract 436 and results 439. The direction of the quantification bias (higher in aptima or abbott) is not clear. % of agreements are given but which % is needed to say that an assay is accurate? In the background, authors acknowledge that several papers have been published on the aptima assay performance compared to other platforms on plasma samples, but not assessing agreement at the medical decision threshold. However, authors have published such a study in Pone in 2021, with the same ethics number, though with fewer samples. In that study, authors have already evaluated FS and VB DBS samples. This work is not discussed, nor referenced. In the results total, positive and negative agreements between assays at the 1000 cp MDP are given, but there is no data on assay variability at this level of VL. This repeatability and reproducibility assessment is mandatory for the interpretation of assays’ discordances. Evaluations of the aptima assay from DBS samples have been published, but authors did not discuss these previously published results, nor the limitations of HIV RNA testing from DBS at low viral loads where HIV DNA, of cell-associated DNA may lead to over quantification of VL. 439 quantifiable plasma VL but 636 quantifiable fresh VB DBS: this discrepancy is not discussed Minor points • The daily throughput of abbott and hologic platforms is not specified. • Works of Yek, Carrera, and Sahoo appear in the reference list but are not cited in the text. These authors have published on DBS and on the aptima assay • In the methods, a flow chat of tested samples is lacking ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 10 Dec 2021 Additional Editor Comments (if provided): The major issues to address are the inconsistent numbers throughout, the lack of replicate testing, and the higher rate of quantitative values from DBS than from plasma. These and other issues are detailed below: a. Corrected the inconsistent in N from 439 to 436 for quantified plasma samples throughout the document b. Clinical studies like the one included in this paper do not have enough specimen volume to permit replicate testing in 2 different assays. Replicate testing is more commonly performed for panels than for clinical samples c. The number of quantifiable results in plasma and DBS sample sets are different because not all plasma and DBS were prepared from the same patients. Blood samples from only a subset of patients were used to prepare plasma and DBS for testing in the two assays. The study also included some DBS samples that did not have corresponding plasma samples and vice versa. Line 17: The background section is too long—drop one or two of the sentences. Deleted a couple of sentences from the background section Line 32: Add that venipuncture was the source of blood for the DBS in this study. Added the information that DBS was prepared using venous blood Line 33: Use the correct name of the Abbott assay: Abbott RealTime HIV-1 Assay Corrected the name Line 37: Drop the sentence about the Stata software, since that is not needed in the abstract. Removed the statement Make sure everything is the same between the text and the abstract and the figures (436 vs 439; r = 0.96 vs R = 0.92—need to check both the number and the capitalization for that one; r = 0.87 vs R = 0.8692—need to check number of significant digits and capitalization; mean bias for DBS storage = -0.154 vs -0.15). Be consistent with your significant digits and with small case r throughout. Changed correlation coefficient from “R” to “r” in the document Line 46: Drop the sentence about the discordance percentages since they are just obtained by subtracting the agreement percentages from 1 and provide nothing new. Removed the statement on discordant rate Line 48: Remove this last sentence of the results about the Panther platform throughput—not needed in the abstract. Removed the statement Line 52: this section needs more about the interpretation of the data in the results section and could mention that the assay worked well on samples in Kenya specifically. Added a few words to make the conclusion specific to samples in Kenya Line 74: small case m2000 Changed case for m in m2000 Line 75: there is a (ref) that needs the citation added. This was the scenario in 2018. It has been resolved by combining two sentences into one: “In Kenya, viral load tests are conducted using Abbott and Roche assays running on the m2000 or COBAS Ampliprep/COBAS Taqman platforms”. Line 70 Line 79: add the full name of the Aptima assay here: Hologic Aptima HIV-1 Quant Dx Assay. Added the full name of the Aptima assay Line 110: Fix the name of the Abbott assay: Abbott RealTime HIV-1 Assay. Fixed the name of the Abbott RT assay Line 136: clarify what is meant by testing DBS on day 0. Is that the day after spotting with no time in the presence of desiccants? Yes. For the day 0 condition, freshly prepared DBS was used for testing while DBS was stored in the presence of dessicants in a Ziploc bag prior to testing on day 21 Line 144: the lower limit for the Abbott assay is 40 for both the 1mL and 0.6mL protocols, and 75 for the 0.5mL protocol, but all the linear ranges can be removed from the methods except possibly for the DBS protocols for each assay. Since all the analysis is with the 1000cp/mL cut-off, the lower limits for plasma are not needed. However, it would be useful to specify which volume protocol was used for the Abbott plasma testing since that is not stated. This has been resolved. The section has been rewritten to read as follows: “The linear quantitative range for Aptima using DBS was 883-10,000,000 copies/mL, while for the Abbott RT assay for plasma sample for 1 mL and 0.6mL protocol is 40–10,000,000 copies/mL, while that for the 0.5 mL protocol is 75-10,000,000 copies/mL. For this study, testing was performed with 0.6 mL of plasma sample. The linear quantitative range of Abbott Real-Time assay using the two spot DBS protocol is 550-10,000,000 copies/mL.” The Abbott protocol used is 0.6mL (line 166-167) Line 167: The use of positive predictive values and negative predictive values in this context is incorrect since there does not appear to be a “gold standard” to show which samples should be positive and which should be negative, suggesting that one of the assays was considered the gold standard for the calculations. All these calculations should be removed from the manuscript. Removed NPV and PPV from this paragraph and Table 1 Line 169: There is a difference in number of plasma and DBS samples included in the data—add an explanation of the difference, whether this was because DBS hadn’t been made for some, the DBS data were invalid, or what other reason. DBS was not collected for all the patients from whom plasma was collected. Line 176: remove the word “therefore”. Removed “therefore” Line 178: remove the sentence about discordance rate Removed the sentence on discordant rate Line 184: discordance rates not shown in Table 1 but can be obtained with the positive agreement values in Table 1, so re-write this statement. This paragraph (lines 184-191) is difficult to understand in terms of useful results. It is useful to the reader to know how many of the discordant results were due to higher detection in one assay vs the other, but the value of the rest of the “results” is less clear. What is the goal of reporting that 17 of the discordant samples (Abbott higher than Aptima in DBS) had corresponding plasma VL that were higher than 3631cp/mL? Is that VL cutoff a magic number for these assays? There is no mention of these “results” in the discussion, providing no context for why this “result” should be noted. Removed the reference to Table 1 from the statement of discordant rate. The purpose of pointing out that 31 samples with discordant DBS results at MDP of 1000 c/mL had plasma results <3631 c/mL is to demonstrate that these samples may flip above and below a 1000 c/mL in different replicates because the HIV concentration is close to a 1000 c/mL. Also added a couple of sentences under discussion to explain the DBS discordants to link with the results in this section. Table 1: the two sides of the table are not related because different numbers of samples are used for each side. Therefore, remove the agreement data from the table and make a new table of these values, adding the number of samples included in each calculation. Also add the DBS day 21 calculations that are not shown in the current table. As mentioned above, remove the PPV and NPV data. The two sides of Table 1 are related and use the same number of samples. The right side of Table 1 gives the positive agreement, negative agreement and total agreement for the results included in the 2x2 table on the left side of Table 1. The sample numbers on both sides of the table have been highlighted to show this and the corresponding results on each side of Table 1 have been moved to the same rows DBS day 21 are already part of Table 1 Removed PPV and NPV data Line 277: add the daily throughput of the Abbott m2000 as a comparator here to strengthen your argument that the Panther has higher throughput. Alternatively, add the throughput (tests per day) in line 75 where you describe it as medium throughput. The Abbott m2000 platform allows 186 results in 10 hours, with the first results being returned in about 5 hours. This is clarified in lines 356-358 Line 218: the discussion needs to start with discussing the results that have just been presented in the previous section, so move the first paragraph down. Reorganize the discussion to move the summary of the results all to the top paragraph(s), then start to discuss them with other published studies in subsequent paragraphs. Shortened the general information in the 1st paragraph Line 232: remove the discordance rate here since it is essentially a retelling of the initial result of agreement. Removed the discordance rate Line 234: change “several” to the actual number of these specimens. This is also a good place to add discussion of the higher number of DBS with quantifiable VL than matched plasma with quantifiable VL, which is a critical question in this study. 28 of the 1312 specimens had VL >1000 c/mL in Aptima and <1000 c/mL in Abbott RT including 5 that were not detected by Abbott RT. Five plasma specimens had a VL >1000 c/mL in Abbott RT and <1000 c/mL in Aptima including one sample that was not detected. No replicate testing is mentioned, not even for invalid results. When running assays on samples that are near the lower limit of detection/quantitation, replicate testing can be critical. Therefore, this must be discussed as a limitation of the study in the discussion. Added a limitation for not performing replicate testing and retesting of invalid specimens Check all the references that they are cited in the manuscript. Your previous publication in PLoS One from April is highly relevant, especially when discussing the limitation that you did not compare fingerstick DBS since you previously looked at that in the April publication. Added a reference to Aptima DBS manuscript published in PLOS One in April 2021 titled “Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya” in the reference section (reference 31) and referenced it in the text in the discussion Add the versions and dates used the package inserts in references 8 and 26. Added version numbers for references 8 and 26 Figures 1-3: All the (a) figure legends must be changed to be more like the (b) legends where the “Figure 1a” is not part of the title and contain sufficient information to stand alone. For example: Figure 1a: HIV-1 viral load comparison for 439 plasma samples tested on both the Aptima HIV-1 Quantitative RNA assay and the Abbott HIV-1 RealTime assay. Updated Figure legends for Figure 1-3 Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf [journals.plos.org] and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf [journals.plos.org] 2. Please amend your current ethics statement to address the following concerns: a) Did participants provide their written or verbal informed consent to participate in this study? The study enrolled a cross-section of HIV positive adults receiving care in health facilities in Nairobi and Busia and who gave written informed consent. Please see Line 121 and 122. b) If consent was verbal, please explain i) why written consent was not obtained, ii) how you documented participant consent, and iii) whether the ethics committees/IRB approved this consent procedure. We took written informed consent. 3. Thank you for stating the following in the Competing Interests/Financial Disclosure * (delete as necessary) section: “Hologic, Inc provided the funds and materials for the study.” We note that you received funding from a commercial source: “Hologic, Inc.” Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc. Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authorshttp://journals.plos.org/plosone/s/competing-interests [journals.plos.org]). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your amended Competing Interests Statement within your cover letter. We will change the online submission form on your behalf. As stated above, Hologic, Inc. provided the funds and materials for the study. This does not alter our adherence to PLOS ONE policies on sharing data and materials. Sangeetha Nair and Sven Schaffer were employees of Hologic, Inc. at the time of the study. This does not alter our adherence to PLOS ONE policies on sharing data and materials. In addition, there are restrictions on the sharing of data and materials originating from the study. The rest of the authors are employees of the Kenya Medical Research Institute and declare no conflict of interest. We have put this statement in the cover letter. 4. Thank you for stating the following in the Competing Interests section: “Sangeetha Nair and Sven Schaffer were employees of Hologic, Inc at the time of the study. The rest of the authors, based at Kenya Medical Research Institute, declare no conflict of interest.” Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests [journals.plos.org]). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf. As stated above, Hologic, Inc. provided the funds and materials for the study. This does not alter our adherence to PLOS ONE policies on sharing data and materials. Sangeetha Nair and Sven Schaffer were employees of Hologic, Inc. at the time of the study. This does not alter our adherence to PLOS ONE policies on sharing data and materials. In addition, there are restrictions on the sharing of data and materials originating from the study. The rest of the authors are employees of the Kenya Medical Research Institute and declare no conflict of interest. We have put this statement in the cover letter. 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This is an easy to read manuscript providing important data relevant to increasing access to HIVVL testing in Africa. I have the following comments. line 55 - Update the HIV statistics to the most recent. Updated the statistics and relevant reference line 95 - I suggest "preferred reference in Kenya' rather than gold standard. Updated this change line 107 - Study Design -Was there sufficient plasma to test the same samples on both assays i.e. same plasma samples tested on Aptima and Abbott? Yes. line 118 - Study Population- inclusion criteria needs to be more detailed, e.g., on ARVs or not, consecutive patients, routine clinic visits. The section has been beefed up to read as follows: The study enrolled a cross-section of HIV positive adults receiving care and treatment in health facilities in Nairobi and Busia and who gave written informed consent. The study participants were mainly patients whose viral load was being monitored because they were on antiretroviral therapy and a very few who were yet to initiate treatment. Plasma collected from both sites were tested in both assays for this study Venous DBS tested in this study were prepared from blood collected from a subset of study participants in Nairobi . Due to this there is a difference in number of plasma and DBS samples that were tested. A small subset of DBS samples did not have the paired plasma results due to a failed run on m2000. line 127 - Was plasma separated on the same day as DBS preparation? Yes. The venous blood was shipped to the lab where it was spotted onto DBS cards and then centrifuged to separate the plasma on the same day line 132 - How soon after collection were the Abbott tests performed? Whole blood was collected by phlebotomy and shipped to KEMRI HIV laboratories within 6 hours of collection, and tested within 12 hours of receipt. Please see lines 132-133 line 134- What is the time difference between Aptima and Abbott tests? The Aptima assay returned the first batch of results within 3.5 hours while the Abbott assay did so within 5 hours. Please see lines 356,357 and 358 Some detail is required on the Aptima procedure as this is the "study" assay. The following text has been added from lines 146-154: To determine the performance of Aptima assay, both plasma and venous DBS were tested on Panther system according to manufacturer’s instructions [8,9,]. Briefly, after initial preparation, 0.75mL of plasma or DBS samples were aliquoted into secondary Aptima specimen aliquot tubes. The system draws 0.5mL from each tube for the assay. Fifteen tubes were loaded onto each rack, for a maximum of 6 racks. The seventh rack was loaded with four samples only while the eighth rack was loaded with a Negative Control, a Low Positive Control, a High Positive Control and a Calibrator. The racks were transported into their appropriate lanes, the bay doors closed, and processing initiated. Initial results were available in 3.5 hours, with five results received every five minutes thereafter. Results were posted as either “Not Detected” or “Invalid”, or as “copies/mL”. line 142 - Input volume for the Aptima ? 0.75 mL of sample is used per test as indicated in line 146-145. The system draws 0.5mL from each tube for the assay line 144 - Which protocol was used for the Abbott? We used the 0.6mL protocol for the Abbott test, as indicated in line 143. line 167 - Results - was prevalence taken into account for the PPV and NPV? All mention of PPV and NPV has been removed from the manuscript line 184 - It may be easier to describe discordance as upward and downward misclassification and the range of the VL obtained for the misclassified samples according to sample type. Added this to the discussion for DBS results lines 281 - 226 -Discussion - This part is more suited to the introduction. Reduced the 1st paragraph of the discussion because a lot of this covered under introduction line 234 - is there an explanation for this finding? Yes. Additional sample was not available for discordant resolution line 242 - is there a reason for this observation? HIV-1 has 4 groups (M, N, O and P) with group M having at least 9 subtypes (A, B, C,D, F, G, H, J, K) and numerous recombinants with significant differences in their nucleic acid sequences. Due to this all viral load assays have slight differences in their ability to quantify each HIV subtype although they are optimized to quantify within 0.5 log of expected concentration. Similar differences have been demonstrated between Abbott RT and Roche Cobas Ampliprep Cobas Taqman assays for HIV (see reference 17 Manak et al) line 250 - is there a reason for this limitation? The limitation regarding testing venous DBS instead of finger stick DBS has been removed from this publication. Instead, we have included a reference that shows equivalency of venous and fingerstick DBS results from the same patient on testing in Aptima line 277 - is this per 8/12/24 hrs? This will give a better idea of throughput. Up to 515 samples can be loaded within 12 hours for testing on Panther. Please see line 356. Reviewer #2: Mwau et al. have investigated the performance of the aptima HIV1 assay compared to the Abbott real-time assay on plasma and DBS samples, at the 1000cp/mL decision point, as well as data on HIV RNA stability at room temperature. 1312 paired plasma and venous blood DBS samples were used. Total agreement between Abbott and Aptima assays at the 3 log threshold was 97.48% for plasma and 94.64% for DBS. Data on quantification agreement between assays, and over time for DBS are provided. Major points 1. It is not clear at abstract’s reading that the authors were working on DBS prepared from venous blood and not from finger-stick. The number of quantifiable samples is different in abstract 436 and results 439. Added a statement in the methods that venous DBS was used for this study. Corrected the number of quantified samples to 436 2. The direction of the quantification bias (higher in Aptima or Abbott) is not clear. % of agreements are given but which % is needed to say that an assay is accurate? Added more discussion on the discordants. The discussion clarifies the number of discordant samples at the MDP of 1000 c/mL for each sample type 3. In the background, authors acknowledge that several papers have been published on the aptima assay performance compared to other platforms on plasma samples, but not assessing agreement at the medical decision threshold. However, authors have published such a study in Pone in 2021, with the same ethics number, though with fewer samples. In that study, authors have already evaluated FS and VB DBS samples. This work is not discussed, nor referenced. Added a reference to the publication in PLOS One in 2021 comparing finger stick and venous DBS in Aptima 4. In the results total, positive and negative agreements between assays at the 1000 cp MDP are given, but there is no data on assay variability at this level of VL. This repeatability and reproducibility assessment is mandatory for the interpretation of assays’ discordances. The repeatability and reproducibility results of the two assays for plasma and DBS is presented in the package inserts that are referenced The reproducibility of Aptima and Abbott assays at 1000 c/mL is published in the package inserts of these assays which is referenced in this publication 5. Evaluations of the aptima assay from DBS samples have been published, but authors did not discuss these previously published results, nor the limitations of HIV RNA testing from DBS at low viral loads where HIV DNA, of cell-associated DNA may lead to over quantification of VL. We had added a reference to our recent publication Aptima HIV Quant Dx assay uses a technology that amplifies DNA at 10,000 to 100,000 fold lower efficiency than RNA. Due to this, no overquantification of VL has been observed in DBS samples. 6. 439 quantifiable plasma VL but 636 quantifiable fresh VB DBS: this discrepancy is not discussed The number of quantifiable results in plasma and DBS sample sets are different because plasma and DBS were not prepared from all the same patients. Blood samples from only a subset of patients were used to prepare plasma and DBS for testing in the two assays. The study also included some DBS samples that did not have corresponding plasma samples and vice versa. Statements to clarify this has been added to the methods section Minor points 1. The daily throughput of Abbott and Hologic platforms is not specified. The Aptima assay allows loading of up to 515 tests in 12 hours, with the first results being returned in about 3.5 hours. In contrast, the Abbott m2000 platform allows 186 results in 10 hours, with the first results being returned in about 5 hours. Please see lines 356-358. 2. Works of Yek, Carrera, and Sahoo appear in the reference list but are not cited in the text. These authors have published on DBS and on the aptima assay References to these publications have been added in the “Background section” 3. In the methods, a flow chat of tested samples is lacking A flow chart was not added because only the blood from a subset of the 1312 patients in the plasma method comparison study were used to prepare DBS. The DBS study also had some patients who did not have plasma results. Statements to clarify this has been added to the methods section [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: No 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy [plosone.org] requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: No 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes Submitted filename: Rebuttal letter Aptima Dec 2021 MM.docx Click here for additional data file. 16 Mar 2022
PONE-D-21-27672R1
Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya
PLOS ONE Dear Dr. Mwau, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. There are still many errors in the manuscript pointed out by Reviewer #1 that need to be addressed. Reviewer #2 was unavailable , so I have added comments from my review of the manuscript. Once the comments of these two reviews have been addressed, the manuscript will be much more clear to the reader. Please submit your revised manuscript by Apr 30 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Julie AE Nelson, PhD Academic Editor PLOS ONE Journal Requirements: Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. Additional Editor Comments: Editor's review: Line 30: remove the word consenting—this is not needed in the abstract. I agree that line 33 is confusing—it could imply that DBS were made from the venous blood of some of the patients and from fingerstick from others, or it could imply that DBS were not made from the blood of all patients included, but the earlier sentence says that DBS were collected from all 1312. Suggested word changes starting at line 33: “…these patients. Agreement between the Aptima…”. I would suggest dropping the word total throughout because it does not add further understanding of what was analyzed. However, it was used in the previous publication, so keeping it would be consistent. The authors should mention their previous study in the introduction (not just as a citation), that the current study is an extension of that one with more data, and especially to note that fingerstick and venous DBS were compared in the previous study so they didn’t need to do that here. In the discussion (lines 326-327), they should acknowledge their previous study is their own and that the current study extends from the previous one. The authors need to add the specific number of samples used in this study in the methods section. This includes the number people with both plasma and DBS, the number with only plasma, and the number with only DBS. Line 283: change wording to “HIV viral load >1000 copies/mL in Aptima but <1000 copies/mL in Abbott RT” Lines 305-324: This paragraph should mostly move to the results section because it describes a new comparison of the data. In the Discussion, there should be a new paragraph about this analysis to put the results into context. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Review of responses to reviewers’ comments he major issues to address are the inconsistent numbers throughout, the lack of replicate testing, and the higher rate of quantitative values from DBS than from plasma. These and other issues are detailed below: a. Corrected the inconsistent in N from 439 to 436 for quantified plasma samples throughout the document Figure 1 legend still states 439 samples c. The number of quantifiable results in plasma and DBS sample sets are different because not all plasma and DBS were prepared from the same patients. Blood samples from only a subset of patients were used to prepare plasma and DBS for testing in the two assays. The study also included some DBS samples that did not have corresponding plasma samples and vice versa. Line 32 of the revised manuscripts states that “Both plasma and DBS samples were collected from each of 1312 adults who participated.” but line 35 DBS was prepared using venous blood from a subset of these patient.” which is conflicting. Make sure everything is the same between the text and the abstract and the figures (436 vs 439; r = 0.96 vs R = 0.92—need to check both the number and the capitalization for that one; r = 0.87 vs R = 0.8692 need to check number of significant digits and capitalization; mean bias for DBS storage = -0.154 vs - 0.15). Be consistent with your significant digits and with small case r throughout. Changed correlation coefficient from “R” to “r” in the document Still capitalized in line 231 of the revised manuscript, R of 0.92 is different to the R= 0.9589 on the figure Line 184: discordance rates not shown in Table 1 but can be obtained with the positive agreement values in Table 1, so re-write this statement. This paragraph (lines 184-191) is difficult to understand in terms of useful results. It is useful to the reader to know how many of the discordant results were due to higher detection in one assay vs the other, but the value of the rest of the “results” is less clear. What is the goal of reporting that 17 of the discordant samples (Abbott higher than Aptima in DBS) had corresponding plasma VL that were higher than 3631cp/mL? Is that VL cutoff a magic number for these assays? There is no mention of these “results” in the discussion, providing no context for why this “result” should be noted. Removed the reference to Table 1 from the statement of discordant rate. The purpose of pointing out that 31 samples with discordant DBS results at MDP of 1000 c/mL had plasma results <3631 c/mL is to demonstrate that these samples may flip above and below a 1000 c/mL in different replicates because the HIV concentration is close to a 1000 c/mL. Also added a couple of sentences under discussion to explain the DBS discordants to link with the results in this section. There is still no indication of why 3631 cp/mL (3,56 log10) specifically was chosen as a cut –off. Using a range of 2,5 – 3,5 log10 incorporating the 0,5 log10 accepted variation in VL assays, would be relevant to justify the flip argument suggested. Major points 1. It is not clear at abstract’s reading that the authors were working on DBS prepared from venous blood and not from finger-stick. The number of quantifiable samples is different in abstract 436 and results 439. Added a statement in the methods that venous DBS was used for this study. Corrected the number of quantified samples to 436 Figure 1 still states 439 samples. 6. 439 quantifiable plasma VL but 636 quantifiable fresh VB DBS: this discrepancy is not discussed The number of quantifiable results in plasma and DBS sample sets are different because plasma and DBS were not prepared from all the same patients. Blood samples from only a subset of patients were used to prepare plasma and DBS for testing in the two assays. The study also included some DBS samples that did not have corresponding plasma samples and vice versa. Statements to clarify this has been added to the methods section. Does this explanation provided by the authors indicate that the discordant DBS samples were only found in the subgroup that had both plasma and DBS samples types? There are 197 more quantifiable DBS results than plasma which is unusual in a comparison of these sample types particularly since the DBS samples were made from the 1312 enrolled patients and has a higher limit of quantification than plasma. Is this related to the run failure? It is important to clarify the reason for the difference. Minor points 3. In the methods, a flow chat of tested samples is lacking A flow chart was not added because only the blood from a subset of the 1312 patients in the plasma method comparison study were used to prepare DBS. The DBS study also had some patients who did not have plasma results. Statements to clarify this has been added to the methods section It would still have been easier to understand the sample numbers of the enrolled samples had been described more clearly in the text or in the flow chart suggested above; according the total number of samples enrolled, number of plasma only, both plasma and DBS, and DBS only. The sample numbers are raised several times in the review because it not defined clearly enough in the methods and results. Second review Line 24 What criteria were used to define high throughput? Line 112 Stating that some DBS samples did not have paired plasma samples under the study design heading suggests that was intentional. Is this correct? Lines 223 – 225 What happened to the other 10 upward misclassified DBS samples? Line 321 – Does this include the Aptima plasma? it is not clear. Line 284 – No discussion about possible under quantification but the Realtime assay. A general review of the grammar and formatting is required including spacing, punctuation and duplication of a few words. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.
6 May 2022 Rebuttal notes Line 30: remove the word consenting—this is not needed in the abstract. The word “consenting” has been removed from line 30. I agree that line 33 is confusing—it could imply that DBS were made from the venous blood of some of the patients and from fingerstick from others, or it could imply that DBS were not made from the blood of all patients included, but the earlier sentence says that DBS were collected from all 1312. We have rewritten the methods section (lines 27-34) to read as follows: This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples while paired DBS samples and plasma were prepared from the remaining 915 samples. The agreement between the Aptima assay and the Abbott RealTime HIV-1 Assay (Abbott RT) was analysed by comparing the HIV-1 viral load in both assays at the medical decision point of 1000 copies/mL. To assess stability of HIV-1 RNA in DBS, viral load in DBS spotted on day 0 were compared with that from the same DBS card after 21 days of storage at room temperature. Suggested word changes starting at line 33: “…these patients. Agreement between the Aptima…”. I would suggest dropping the word total throughout because it does not add further understanding of what was analyzed. However, it was used in the previous publication, so keeping it would be consistent. We have deleted several “total” words throughout the manuscript. We have left a few of the words where the sense would be lost if we deleted them. The authors should mention their previous study in the introduction (not just as a citation), that the current study is an extension of that one with more data, and especially to note that fingerstick and venous DBS were compared in the previous study so they didn’t need to do that here. In the discussion (lines 326-327), they should acknowledge their previous study is their own and that the current study extends from the previous one. We have rewritten the paragraph that starts from line 85 to capture these details as follows: In a previous study, we investigated the performance of the Aptima HIV Quant Dx assay using fingerstick and venous dried blood spots prepared under field conditions, although we did not compare HIV-1 VL in Aptima with those in other assays using DBS. In fact, whereas multiple studies compared viral load results of paired DBS and plasma samples at the medical decision point of 1000 copies/ mL by testing both sample types in Aptima [21-23, 31], there is no published information comparing the performance of DBS specimens tested in Aptima and other assays at the medical decision point (MDP) of 1000 copies/mL. The authors need to add the specific number of samples used in this study in the methods section. This includes the number people with both plasma and DBS, the number with only plasma, and the number with only DBS. We have rewritten the methods section to make sure this is very clear, as follows: Whole blood was collected from 2227 participants by phlebotomy, shipped to KEMRI HIV laboratories within 6 hours of collection, and processed within 12 hours of receipt. A total of 1312 samples provided only plasma. Paired DBS and plasma samples were prepared from a further 915 venous samples. To prepare plasma, the whole blood samples were centrifuged at 1,100g for 10 minutes within 24 hours of collection, and the plasma was stored at -800C. To prepare DBS, 70µl of venous blood was spotted in each of ten spots (2 DBS cards) per patient. The DBS samples were allowed to dry overnight. The first DBS card for each patient was used for day 0 testing, while the DBS cards for the 21-day time point were packaged with desiccants and stored at room temperature. All samples were de-identified prior to Aptima testing. Line 283: change wording to “HIV viral load >1000 copies/mL in Aptima but <1000 copies/mL in Abbott RT” The word “but” has been added, apologies for the oversight in this sentence. Lines 305-324: This paragraph should mostly move to the results section because it describes a new comparison of the data. In the Discussion, there should be a new paragraph about this analysis to put the results into context. The results had already been presented in the results section, and for this reason we have carefully revised the paragraph to lighten it while still retaining its meaning, from lines 317-329 Reviewer #1: Review of responses to reviewers’ comments The major issues to address are the inconsistent numbers throughout, the lack of replicate testing, and the higher rate of quantitative values from DBS than from plasma. These and other issues are detailed below: a. Corrected the inconsistent in N from 439 to 436 for quantified plasma samples throughout the document: n is now corrected to 436 where appropriate Figure 1 legend still states 439 samples. This has been corrected. Line 234 now reads: Figure 1a: HIV viral load comparison for 436 plasma samples tested in Aptima and Abbott RT assays c. The number of quantifiable results in plasma and DBS sample sets are different because not all plasma and DBS were prepared from the same patients. Blood samples from only a subset of patients were used to prepare plasma and DBS for testing in the two assays. The study also included some DBS samples that did not have corresponding plasma samples and vice versa. In the methods section, we have clarified the exact number of samples of each type used in the study. Please see lines 127-136 Line 32 of the revised manuscripts states that “Both plasma and DBS samples were collected from each of 1312 adults who participated.” but line 35 DBS was prepared using venous blood from a subset of these patient.” which is conflicting. This is our error. We have rewritten lines 27 to 31 to read as follows: This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples. Paired DBS and plasma samples were prepared from the remaining 915 samples. Make sure everything is the same between the text and the abstract and the figures (436 vs 439; r = 0.96 vs R = 0.92—need to check both the number and the capitalization for that one; r = 0.87 vs R = 0.8692 need to check number of significant digits and capitalization; mean bias for DBS storage = -0.154 vs - 0.15). Be consistent with your significant digits and with small case r throughout. Changed correlation coefficient from “R” to “r” in the document This has now been done throughout the document. Still capitalized in line 231 of the revised manuscript, R of 0.92 is different to the R= 0.9589 on the figure. We have corrected this error in the manuscript, and resolved capitalisation around r. Line 184: discordance rates not shown in Table 1 but can be obtained with the positive agreement values in Table 1, so re-write this statement. This paragraph (lines 184-191) is difficult to understand in terms of useful results. It is useful to the reader to know how many of the discordant results were due to higher detection in one assay vs the other, but the value of the rest of the “results” is less clear. What is the goal of reporting that 17 of the discordant samples (Abbott higher than Aptima in DBS) had corresponding plasma VL that were higher than 3631cp/mL? Is that VL cutoff a magic number for these assays? There is no mention of these “results” in the discussion, providing no context for why this “result” should be noted. We have removed the reference to Table 1 from the statement of discordant rates. The purpose of pointing out that 31 samples with discordant DBS results at MDP of 1000 c/mL had plasma results <3631 c/mL is to demonstrate that these samples may flip above and below a 1000 c/mL in different replicates because the HIV concentration is close to a 1000 c/mL. We have also added a couple of sentences under the discussion to explain the DBS discordants to link with the results in this section. There is still no indication of why 3631 cp/mL (3,56 log10) specifically was chosen as a cut –off. Using a range of 2,5 – 3,5 log10 incorporating the 0,5 log10 accepted variation in VL assays, would be relevant to justify the flip argument suggested. Your point is well noted. We agree that there is nothing special about the 3631cp/ml. We have rewritten the sentence (lines 326-329) as follows: The remaining 14 of the 31 patients with Aptima DBS VL <3 log c/mL and Abbott DBS VL>3 log c/mL had plasma VL >3.00 log copies/mL in both Aptima plasma and Abbott plasma suggesting that these were upward misclassifications by Abbott DBS. Major points 1. It is not clear at abstract’s reading that the authors were working on DBS prepared from venous blood and not from finger-stick. The number of quantifiable samples is different in abstract 436 and results 439. We have added a statement in the methods that venous DBS was used for this study. Please see line 23. We have corrected the number of quantified samples to 436, and indicted that venous DBS was used, throughout the document. Figure 1 still states 439 samples. We have corrected this to 436 samples in Fig. 1. 6. 439 quantifiable plasma VL but 636 quantifiable fresh VB DBS: this discrepancy is not discussed. The 436 quantifiable plasma samples were prepared from venous blood taken from 1312 patients. The 636 quantifiable DBS samples were prepared from a further group of (915) participants, who also provided plasma. For that reason the yields are not comparable. Does this explanation provided by the authors indicate that the discordant DBS samples were only found in the subgroup that had both plasma and DBS samples types? As clarified, there were 915 paired DBS and plasma samples. The 49 discordant DBS samples are a subset of the 915 samples. There are 197 more quantifiable DBS results than plasma which is unusual in a comparison of these sample types particularly since the DBS samples were made from the 1312 enrolled patients and has a higher limit of quantification than plasma. Is this related to the run failure? It is important to clarify the reason for the difference. These results arise from different sets of participants. So that comparison does not arise. Please see lines 127-136. Whole blood was collected from 2227 participants by phlebotomy, shipped to KEMRI HIV laboratories within 6 hours of collection, and processed within 12 hours of receipt. A total of 1312 samples provided only plasma. Paired DBS and plasma samples were prepared from a further 915 venous samples. The 49 discordant DBS samples were part of the 915 paired DBS and plasma samples tested. Minor points 3. In the methods, a flow chat of tested samples is lacking. A total of 1312 samples provided only plasma. Paired DBS and plasma samples were prepared from a further 915 venous samples. Since there was no further dichotomy, a flow chart was not necessary. It would still have been easier to understand the sample numbers of the enrolled samples had been described more clearly in the text or in the flow chart suggested above; according the total number of samples enrolled, number of plasma only, both plasma and DBS, and DBS only. The sample numbers are raised several times in the review because it not defined clearly enough in the methods and results. We have extensively rewritten this section to make it as clear as possible. Please see lines 179-185. Second review Line 24 What criteria were used to define high throughput? There are no universally agreed criteria for defining high throughput. In our setting, we consider low throughput to mean <96 tests in 24 hours, medium throughput to mean 97-480 tests in 24 hours and high throughput to be >481 tests in 24 hours. Line 112 Stating that some DBS samples did not have paired plasma samples under the study design heading suggests that was intentional. Is this correct? This was our mistake. (In actual fact, this error occurred out of confusion from: 258 of our DBS samples had accompanying fingerstick samples while the rest did not, but that data is reported elsewhere already). As stated in the manuscript, a total of 1312 samples provided only plasma. Both plasma and paired DBS samples were prepared from a further 915 venous samples. Lines 223 – 225 What happened to the other 10 upward misclassified DBS samples? We have revised this section so that the discussion does not mislead, from lines 335-346. Line 321 – Does this include the Aptima plasma? it is not clear. The upward misclassification was seen only in Aptima DBS, not Aptima or Abbott plasma or Abbott DBS. Line 284 – No discussion about possible under quantification by the Realtime assay. The discussion around this finding has now been provided, lines 331-342 A general review of the grammar and formatting is required including spacing, punctuation and duplication of a few words. We have used the grammar tool to resolve spacing challenges and correct grammatical errors. Submitted filename: Rebuttal notes Aptima Adult Paper May 6 2022.docx Click here for additional data file. 30 May 2022 Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya PONE-D-21-27672R2 Dear Dr. Mwau, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Julie AE Nelson, PhD Academic Editor PLOS ONE Additional Editor Comments (optional): Changes to make before publication: Line 51: Choose one of these options: add “(VL)” after “viral load” here and change to VL for all subsequent uses, OR change all uses of “VL” to “viral load”. Do not use them interchangeable throughout. Line 66: add “, respectively” after “platforms” Line 76: add comma before “respectively” Line 130: remove “; of those” Line 132: fix “-800C” to “-80°C” Line 139: add a short description and citation for the DBS testing on the Abbott system since it is not described in the package insert. A description should include how much of a spot used and how much volume of which buffer was used for elution. It would also be nice for the reader to include a similar description for the Aptima assay at line 142 so readers don’t need to go to the package insert. There is a description of how many spots were used on lines 155-156—move this information up to line 139-143. Line 159: Specify whether the conversion factor done by the tech or by the software on the Aptima. What about the conversion factor for Abbott? Line 190: the 2.94 copies/mL and 7 copies/mL should say log copies/mL instead. However, please convert these numbers to copies/mL for the Aptima to be equivalent to the Abbott description for better readability. Line 230: You have said that 1312 plasma samples were compared in the assays and that 915 DBS samples were compared in the assays. In this sentence, however, you now indicate that the plasma samples were tested that correspond to the DBS samples. Is there a reason that you did not include these 915 plasma samples results with the 1312 results? Also, if both plasma and DBS were made for 915 patients, were both plasma and DBS tested on Abbott for delivery of patient results? It is ok to not include the plasma results with the 1312, but it would be helpful to the reader to explain why they were not reported that way. Line 296: Remove the phrase about the 0.33 log higher quantification, stopping the sentence after the first “Abbott RT” in this line. The later discussion of the 0.33 log is more clear and this one is just confusing. Line 324: This sentence got jumbled. Remove the “>3 log c/mL in both Aptima plasma and Abbott plasma”. Here I also notice that you are interchangeable using log c/mL and log copies/mL. Use just one way throughout the manuscript. Line 344: rewrite as follows: “Although only venous DBS were tested in this study, we previously compared the recovery of HIV-1 from venous DBS, fingerstick DBS, and plasma with the Aptima test [31], showing good agreement between plasma, venous DBS, and fingerstick DBS.” Line 346: change “this” to “the current” Reviewers' comments: 8 Jun 2022 PONE-D-21-27672R2 Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya Dear Dr. Mwau: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Julie AE Nelson Academic Editor PLOS ONE
  21 in total

1.  Comparison of the Hologic Aptima HIV-1 Quant Dx Assay to the Roche COBAS Ampliprep/COBAS TaqMan HIV-1 Test v2.0 for the quantification of HIV-1 RNA in plasma samples.

Authors:  Kristian Schønning; Kim Johansen; Bodil Landt; Thomas Benfield; Henrik Westh
Journal:  J Clin Virol       Date:  2017-05-08       Impact factor: 3.168

2.  Evaluation of the RealTime HIV-1, Xpert HIV-1, and Aptima HIV-1 Quant Dx Assays in Comparison to the NucliSens EasyQ HIV-1 v2.0 Assay for Quantification of HIV-1 Viral Load.

Authors:  Orna Mor; Yael Gozlan; Marina Wax; Fernando Mileguir; Avia Rakovsky; Bina Noy; Ella Mendelson; Itzchak Levy
Journal:  J Clin Microbiol       Date:  2015-08-19       Impact factor: 5.948

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Evaluation of the Aptima HIV-1 Quant Dx Assay for HIV-1 RNA Quantitation in Different Biological Specimen Types.

Authors:  Christina Yek; Marta Massanella; Tashi Peling; Kristen Lednovich; Sangeetha V Nair; Andrew Worlock; Milenka Vargas; Sara Gianella; Ronald J Ellis; Matthew C Strain; Michael P Busch; C Thomas Nugent; Douglas D Richman
Journal:  J Clin Microbiol       Date:  2017-06-07       Impact factor: 5.948

5.  Evaluation of the Aptima HIV-1 Quant Dx Assay Using Plasma and Dried Blood Spots.

Authors:  Malaya K Sahoo; Vici Varghese; Elizabeth White; Meg Winslow; David A Katzenstein; Robert W Shafer; Benjamin A Pinsky
Journal:  J Clin Microbiol       Date:  2016-08-17       Impact factor: 5.948

6.  Aptima HIV-1 Quant Dx--A fully automated assay for both diagnosis and quantification of HIV-1.

Authors:  Sangeetha Vijaysri Nair; Hee Cheol Kim; Jacqueline Fortunko; Tracy Foote; Tashi Peling; Cuong Tran; Charles Thomas Nugent; Sunghae Joo; Youna Kang; Bana Wilkins; Kristen Lednovich; Andrew Worlock
Journal:  J Clin Virol       Date:  2016-02-04       Impact factor: 3.168

7.  Evaluation of Performance Characteristics of the Aptima HIV-1 Quant Dx Assay for Detection and Quantitation of Human Immunodeficiency Virus Type 1 in Plasma and Cervicovaginal Lavage Samples.

Authors:  Soya S Sam; Jaclynn R Kurpewski; Susan Cu-Uvin; Angela M Caliendo
Journal:  J Clin Microbiol       Date:  2016-02-03       Impact factor: 5.948

Review 8.  Predicting treatment failure in adults and children on antiretroviral therapy: a systematic review of the performance characteristics of the 2010 WHO immunologic and clinical criteria for virologic failure.

Authors:  George W Rutherford; Andrew Anglemyer; Philippa J Easterbrook; Tara Horvath; Marco Vitoria; Martina Penazzato; Meg C Doherty
Journal:  AIDS       Date:  2014-03       Impact factor: 4.177

9.  Multicenter Evaluation of Two Next-Generation HIV-1 Quantitation Assays, Aptima Quant Dx and Cobas 6800, in Comparison to the RealTime HIV-1 Reference Assay.

Authors:  Frank Wiesmann; Robert Ehret; Gudrun Naeth; Martin Däumer; Jörg Fuhrmann; Rolf Kaiser; Christian Noah; Martin Obermeier; Gunnar Schalasta; Carsten Tiemann; Eva Wolf; Heribert Knechten; Patrick Braun
Journal:  J Clin Microbiol       Date:  2018-09-25       Impact factor: 5.948

10.  Evaluation of the Aptima(®) HIV-1 Quant Dx assay for HIV-1 RNA viral load detection and quantitation in plasma of HIV-1-infected individuals: A comparison with Abbott RealTime HIV-1 assay.

Authors:  Alessandra Amendola; Maria Pisciotta; Loredana Aleo; Valeria Ferraioli; Claudio Angeletti; Maria Rosaria Capobianchi
Journal:  J Med Virol       Date:  2016-02-19       Impact factor: 2.327

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.