| Literature DB >> 35979002 |
Cheng Cheng1, Banglong Xu1, Jianlong Sheng1, Zheng Huang1, Fei He1, Feng Gao1, Xiaochen Wang1.
Abstract
Objectives: To investigate the procedural efficiency, efficacy, and safety of high-power, short-term radiofrequency ablation delivered by the SmartTouch Surround Flow (STSF) catheter for paroxysmal atrial fibrillation (AF).Entities:
Year: 2022 PMID: 35979002 PMCID: PMC9377865 DOI: 10.1155/2022/6009275
Source DB: PubMed Journal: Evid Based Complement Alternat Med ISSN: 1741-427X Impact factor: 2.650
Figure 1PVI in paroxysmal AF. (a) LPLD (40 W/31 s, AI 500) ablation delivered by an ST catheter. (b) HPSD (50 W/19 s, AI 490) ablation delivered by an STSF catheter.
RF power, AI thresholds, and RF time per lesions in different left atrial sites.
| Group | Anterior | Ridge | Roof | Inferior | Posterior | |
|---|---|---|---|---|---|---|
| RF power | Control | 35 W | 35 W | 35 W | 35 W | 30 W |
| Study | 50 W | 50 W | 50 W | 50 W | 45 W | |
|
| ||||||
| AI | Control | 450–480 | 480–500 | 400–450 | 400–450 | 380–420 |
| Study | 450–480 | 480–500 | 400–450 | 400–450 | 380–420 | |
|
| ||||||
| RF time per lesions | Control | 30–40 s | 35–40 s | 25–35 s | 25–35 s | 20–25 s |
| Study | 15–20 s | 15–20 s | 10–20 s | 10–20 s | 10–15 s | |
RF, radiofrequency; AI, ablation index.
Baseline demographic and patient characteristics.
| Characteristics | Control group ( | Study group ( |
|
|---|---|---|---|
| Sex (male/female) | 20/16 | 27/9 | 0.083 |
| Age (years) | 63.11 ± 9.96 | 58.92 ± 11.92 | 0.110 |
| BMI (kg/m2) | 25.38 ± 3.24 | 24.11 ± 2.98 | 0.089 |
| Hypertension ( | 18 (50.0) | 20 (55.6) | 0.637 |
| Diabetes ( | 7 (19.4) | 6 (16.7) | 0.759 |
| Coronary heart disease ( | 11 (30.6) | 6 (16.7) | 0.165 |
| Heart failure ( | 3 (8.3) | 1 (2.8) | 0.303 |
| Stroke/TIA ( | 4 (11.1) | 2 (5.6) | 0.394 |
| Mean AF duration (months) | 28.31 ± 31.46 | 35.50 ± 36.38 | 0.716 |
| CHA2DS2-VASC | 2.31 ± 1.95 | 1.69 ± 1.49 | 0.191 |
| HAS-BLED | 0.64 ± 0.59 | 0.44 ± 0.61 | 0.130 |
| eGFR (ml/min/1.73 m2) | 89.48 ± 20.71 | 91.66 ± 26.94 | 0.165 |
| LA diameter (mm) | 38.08 ± 6.40 | 38.19 ± 6.46 | 0.942 |
| LV diameter (mm) | 45.31 ± 5.06 | 47.31 ± 4.62 | 0.084 |
| LV EF (%) | 61.53 ± 4.07 | 63.04 ± 4.40 | 0.136 |
Data are expressed as mean ± SD or n (%). BMI, body mass index; TIA, transient ischemic attack; AF, atrial fibrillation; eGFR, estimated glomerular filtration rate; LV, left ventricular; EF, ejection fraction.
Procedural characteristics and freedom from AF recurrences following a blanking period of three months.
| Characteristics | Control group ( | Study group ( |
|
|---|---|---|---|
| Left PVI time (min) | 41.3 ± 6.2 | 32.1 ± 7.0 | <0.001 |
| Right PVI time (min) | 33.1 ± 4.5 | 29.1 ± 7.0 | 0.007 |
| Procedural time (min) | 145.4 ± 34.3 | 111.8 ± 36.1 | <0.001 |
| Fluoroscopy time (min) | 4.5 (4.0–5.0) | 4.0 (3.25–5.0) | 0.129 |
| First pass isolation | 61/72 (84.7) | 69/72 (95.8) | 0.023 |
| Irrigation fluid volume (mL) | 966 ± 227 | 767 ± 171 | <0.001 |
| Steam pop ( | 2 (5.6) | 3 (8.3) | 0.643 |
| Freedom from AF recurrences ( | 31/33 (93.9) | 27/31 (87.1) | 0.348 |
Data are expressed as the mean ± SD or n (%) (fluoroscopy time was expressed in quartiles). PVI, pulmonary vein isolation; AF, atrial fibrillation.