Low-dose abdominopelvic computed tomography in patients with lymphoma: An image quality and radiation dose reduction study.

Sungjin Yoon1, Kwai Han Yoo2, So Hyun Park1, Hawk Kim2, Jae Hoon Lee2, Jinny Park2, Seong Ho Park3, Hwa Jung Kim4.   

Abstract

This study aimed to evaluate image quality, the detection rate of enlarged lymph nodes, and radiation dose exposure of ultralow-dose and low-dose abdominopelvic computed tomography (CT) in patients with lymphoma. Patients with lymphoma who underwent abdominopelvic CT using dual-source scanner were retrospectively recruited from a single center. CT images were obtained at 90 kVp dual-source mode reformatted in three data sets using the advanced modelled iterative reconstruction algorithm: 100% (standard-dose CT), 66.7% (low-dose CT), and 33.3% (ultralow-dose CT). Two radiologists analyzed subjective image quality and detection of abdominal enlarged lymph nodes on ultralow-dose, low-dose, and standard-dose CT blindly and independently. The results were compared with reference standards. Three readers (two radiologists and one hematologist) reviewed overall image quality and spleen size. In total, 128 consecutive CT scans (77 complete response, 44 partial response, 6 progressive disease, and 1 initial evaluation) from 86 patients (64 B-cell lymphoma, 14 T/NK-cell lymphoma, and 8 Hodgkin's lymphoma cases) were assessed. The enlarged lymph node-based detection rates for two readers were 97.0% (96/99) and 94.0% (93/99) on standard-dose CT, 97.0% (96/99) and 94.0% (93/99) on low-dose CT, and 94.0% (93/99) and 89.9% (89/99) on ultralow-dose CT. Overall image quality was 3.8 ± 0.5, 3.9 ± 0.5, and 4.1 ± 0.5 on ultralow-dose CT; 4.7 ± 0.4, 4.6 ± 0.5, and 4.8 ± 0.3 on low-dose CT; and 4.8 ± 0.4, 4.7 ± 0.4, and 4.9 ± 0.2 on standard-dose CT, according to two radiologists and one hematologist, respectively. Intraclass correlation coefficients of spleen size were 0.90 (95% confidence interval [CI], 0.87-0.93), 0.91 (95% CI, 0.88-0.93), and 0.91 (95% CI, 0.88-0.93) on ultralow-dose, low-dose, and standard-dose CT, respectively. Mean effective radiation doses of standard-dose, low-dose, and ultralow-dose CT were 5.7 ±1.8 mSv, 3.8 ± 1.2 mSv, and 1.9 ± 0.6 mSv, respectively. Our findings suggest that ultralow-dose and low-dose CT, even with radiation doses reduced by 66.7% and 33.3%, respectively, maintained adequate image quality. These imaging modalities may be employed for follow-up lymphoma evaluation in consideration of the long surveillance periods.

Entities:  

Mesh:

Year:  2022        PMID: 35951525      PMCID: PMC9371255          DOI: 10.1371/journal.pone.0272356

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.752


Introduction

Lymphomas account for 3.4% of all malignancies worldwide and consist of heterogeneous subtypes (i.e., non-Hodgkin’s lymphomas and Hodgkin’s lymphoma) [1, 2]. The majority of lymphomas involve the lymph nodes and extralymphatic organs, and occur in young adults and pediatric patients. A proportion of lymphomas are considered potentially curable diseases due to improvements in treatment protocols [3]. The Lugano classification is used for lymphoma staging and response assessment, and imaging modalities such as positron emission tomography (PET)/CT and computed tomography (CT) play key roles in response assessment and surveillance to evaluate lymph nodes and spleen [4-7]. Due to their long life expectancy, patients with lymphoma require multiple CT examinations to evaluate treatment response and surveillance after treatment. The cumulative radiation exposure of repetitive CT examinations in pediatric and young adult patients may increase baseline cancer risk [8, 9]. Therefore, dose reduction techniques for CT can be useful for patients with lymphoma in consideration of the long surveillance periods. Several studies have attempted low-dose CT to evaluate Hodgkin’s lymphoma in the thorax [10] and in patients undergoing staging or restaging of lymphoma [11, 12]. Generally, a reduction in radiation dose is related to increased image noise and decreased image quality, which may negatively impact diagnostic performance. However, in patients with lymphoma, particularly during follow-up after treatment, measurement of pre-existing lesions or improvements in lymphoma involvement constitute a major component of CT assessments. In this regard, low-dose CT with reduced image quality may be sufficient for follow-up evaluation. Recent advances in CT techniques have contributed to a reduction in radiation dose, such as the implementation of automatic adjustment of tube potential, automated tube current modulation, and iterative reconstruction (IR) [13, 14]. A new third generation of IR was recently developed based on statistical-based to model-based IR [14-16]. Advanced modelled IR (ADMIRE; Siemens Healthcare, Forchheim, Germany) [17] is a model-based IR that permits a large reduction in image noise in raw data and additional dose reduction with improved spatial resolution. Dual-energy CT with dual-source scanners can be used to compare standard and low-dose CT images by separating and combining data from each tube without the need for additional examinations [18, 19]. In this study, we evaluated radiation dose exposure and image quality of low-dose and standard-dose CT to compare the detection rates of enlarged lymph nodes of low-dose and standard-dose CT for abdominal lymphoma evaluation and post-treatment follow-up in patients with lymphoma.

Materials and methods

Ethics approval

Approval for this retrospective study was obtained from the institutional review board (GAIRB2021-237) of the Gil Medical Center. All CT images were acquired using standard-dose CT scan, without additional radiation exposure. Thus, the informed consent was waived because of the retrospective nature of the study.

Study participants

In total, 197 consecutive patients who underwent dual-source CT scans for hematologic assessment at the hematologic department of Gil Medical Center from December 2018 to December 2020 were enrolled. All patients diagnosed with lymphoma were included. Patients underwent CT for lymphoma evaluation, chemotherapy response, or surveillance after treatment. Among 98 consecutive patients with 140 CT scans who were eligible, 12 patients were excluded due to a protocol change (n = 7) and lack of a reference standard (n = 5). A final total of 86 consecutive patients with 128 CT examinations were included in this study (Fig 1A). The study period, from December 2018 to March 2019, partially overlapped with that of a previous dual-energy study conducted in the same institution [20]. However, the previous study only included patients from the oncologic department. Therefore, there were no overlapping patients between the two studies.
Fig 1

Flow diagram of patients.

(a) Inclusion flow chart and (b) computed tomography (CT) examinations and reconstruction methods.

Flow diagram of patients.

(a) Inclusion flow chart and (b) computed tomography (CT) examinations and reconstruction methods.

CT technique

Contrast-enhanced abdominopelvic CT examination was acquired above the dome of the diaphragm and below the symphysis pubis. All patients received intravenous injection of 1.5 mL/kg of iopamidol (Pamiray 300; Dongkook Pharm., Korea), up to a maximum dose of 120 mL. The injection was delivered using a power injector at an injection rate of 4 mL/s and fixed injection duration of 75 s. CT scans were achieved at a fixed tube potential of 90 kVp using a third-generation dual-source CT scanner (SOMATOM Force, Siemens Healthcare, Forchheim, Germany) in dual-source mode with tube detector A (reference tube current: 100 mAs) and B (reference tube current: 200 mAs), using tube current dose modulation (CARE dose 4D; Siemens Healthcare) and the ADMIRE algorithm (Fig 1B and Table 1). We used the ADMIRE algorithm at a strength level of 2 out of 5, with an axial slice thickness of 5 mm and coronal slice thickness of 3 mm.
Table 1

Reconstruction parameters.

Standard-doseLow-doseUltralow-dose
Radiation dose exposure100%66.7%33.3%
Tube detectorMix of detector A and BTube detector ATube detector B
Kilovolt (kV)909090
Automated tube voltage selectionOffOffOff
Reference tube current (mAs)300 (100%)200 (66.7%)100 (33.3%)
Automated tube current modulationOnOnOn
Thickness of axial image5 mm5 mm5 mm
Thickness of coronal image3 mm3 mm3 mm
Pitch1.151.151.15
Rotation time (sec)0.50.50.5

Qualitative visual image analysis

Two radiologists (S.J.Y and S.H.P, with 5 and 10 years of abdominal radiologic experience radiologists, respectively) reviewed the image analysis, and one hematologist (K.H.Y) reviewed the overall image quality and spleen size only. The images of 384 CT examinations were reviewed independently by three readers in a blinded manner. These interpretations were analyzed in three reading sessions, some of which included one-third of the three CT image sets. The images were reviewed anonymously, and the order of review was randomized with a 1-month washout period between sessions. Lymphoma involvement was evaluated based on the Lugano classification [4], with modifications: lymph nodes, spleen, liver, and other sites. An enlarged lymph node was regarded as a short-axis diameter > 1 cm, and splenomegaly was defined as the longest length of the spleen > 12 cm. Suspicious masses or nodules in the adrenal glands or liver were analyzed. Enlarged lymph nodes exhibiting a fatty hilum were considered to indicate a reactive change. We measured the average diameter of lymph nodes (short-axis diameter >0.5 cm in each data set) on standard-dose, low-dose, and ultralow-dose CT. The overall image quality score for assessing CT images was subjectively measured using a 5-point (Table 2).
Table 2

Qualitative visual image analysis.

VariablesAnalysis
Organ Positive findings
LiverNodules (except definite hemangiomas or cysts)
Spleen> 12 cm in length, mass, or nodule
Adrenal glandNodules
Lymph nodesEnlarged, short diameter > 1 cm
Overall image quality 1, nondiagnostic quality, extremely severe artifacts, insufficient for diganosis
2, poor image quality, severe artifacts causing uncertainty
3, moderate image quality, moderate artifacts with mild restricted evaluation
4, good quality, slight artifacts with sufficient for diagnosis
5, excellent image quality, no artifacts

Quantitative image noise analysis

To achieve objective image quality, regions of interest (ROI) were placed in four regions including psoas muscle, subcutaneous fat in the anterior abdominal wall, right hepatic lobe parenchyma, aorta lumen in L1 vertebral body level with same location in three image sets. The standard deviations (SDs) in Hounsfield units (HU) were measured by a 1–3 cm2 ROI by a single-blinded reader (S.J.Y.), as image noise. Mean attenuation values (HU) were measured for each ROI.

Radiation dose

The volume CT dose index (CTDIvol) and dose-length product were documented in the dose page of the scanner. In the calculation of effective dose, tissue-weighting factors for the abdomen was used in millisieverts (mSv) (male, k = 0.013; female, k = 0.017) and pelvis (male, k = 0.010; female, k = 0.016) using average values (male, k = 0.012; female, k = 0.017) [21, 22].

Reference standards

Reference standards comprised PET-CT or MRI within 3 months, interval changes of the lesion compared to serial CTs with patient’s symptoms, and clinician’s judgment based on electronic medical records (EMRs). Response assessment was based on EMRs and was categorized into complete, partial, stable, and progressive disease.

Statistical analysis

Radiation dose and objective image analyses were compared among the three CT scans using analysis of variance followed by post hoc Bonferroni correction. Subjective image evaluations of the three image analyses were compared using Kruskal-Wallis test adjusted with Monte Carlo simulation. In each image set, the detection rate for enlarged lymph nodes were compared with each other using Generalized Estimating Equations (GEE) with adjustment for multiple comparisons using Bonferroni correction. Interobserver agreement of enlarged lymph node detection among two readers was analyzed using kappa, and interobserver agreement of spleen size among three readers was assessed using intraclass correlation coefficients, defined as follows: 0.01–0.20, slight; 0.21–0.40, fair; 0.41–0.60, moderate; 0.61–0.80, substantial; and 0.81–1, excellent. Statistical significance was set at P < 0.05. After post hoc analysis, P-values < 0.01 were considered statistically significant. Statistical analyses were performed using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA) for all data analyses.

Results

Patient characteristics

The clinical characteristics of the patients are summarized in Table 3. Of the 86 patients with 128 CT scans, 57 were men and 29 were women, with a mean age ± standard deviation of 58.4 ± 16.3 years. Of patients, 33 underwent two or more CT examinations during the study period (25 patients, 2 CT examinations; 8 patients, 3 CT examinations). Among patients, B-cell lymphoma was the most common disease (n = 63, 73.3%), followed by T/NK-cell lymphoma (n = 14, 16.3%) and Hodgkin’s lymphoma (n = 8, 9.3%).
Table 3

Clinical characteristics of patients.

ParameterValue
Number of patients86
Age (years), mean ± SD58.4 ± 16.3
Men: women57: 29
Height (cm)164.1 ± 8.3
Weight (kg)66.4 ± 13.6
Effective diameter (cm)26.6 ± 3.2
BMI (kg/m2)24.5 ± 3.7
 < 18.5: thin2 (2.3)
 18.5–24.9: normal47 (54.7)
 25–29.9: overweight29 (33.7)
 30–34.9: moderate obesity8 (9.3)
 35–39.9: severe obesity0
Histology
 Non-Hodgkin’s lymphoma78 (90.7)
 B-cell lymphoma63 (73.3)
 T/NK-cell lymphoma14 (16.3)
 Hodgkin’s lymphoma8 (9.3)
Patient class
 Inpatient5 (5.8)
 Outpatient81 (94.2)
CT examinations128
Disease status
 Initial1 (0.8)
 Progressive disease6 (4.7)
 Partial response44 (34.4)
 Complete response77 (60.2)
Reference standard
 PET/CT46 (35.9)
 CT81 (63.3)
 PET/CT and liver MRI1 (0.8)

Note: Data are presented as number (%), unless indicated otherwise.

Note: Data are presented as number (%), unless indicated otherwise.

Lesion detection analysis

Table 4 compares standard-dose, low-dose, and ultralow-dose CT for the detection of lesions in the aforementioned organs and enlarged lymph node analysis in patients with lymphoma. The number of enlarged lymph nodes detected by two readers was higher on standard-dose (n = 97 and 95) and low-dose CT (n = 97 and 95) than on ultralow-dose CT (n = 94 and n = 93). The enlarged lymph node-based detection rates (i.e., number of detected true enlarged lymph nodes/number of true enlarged lymph nodes) for two readers were 97.0% (96/99) and 94.0% (93/99) on standard-dose CT, 97.0% (96/99) and 94.0% (93/99) on low-dose CT, and 94.0% (93/99) and 89.9% (89/99) on ultralow-dose CT (Table 4A and Figs 2 and 3). After GEE with adjustment for multiple comparisons using Bonferroni correction, there was no statistical difference among three data sets by two readers (reader 1, Standard-dose CT vs. Ultralow-dose CT, P = 0.256; Low-dose CT vs. Ultralow-dose CT, P = 0.256; reader 2, Standard-dose CT vs. Ultralow-dose CT, P = 0.132; Low-dose CT vs. Ultralow-dose CT, P = 0.132). Pairwise comparisons revealed different numbers of enlarged lymph nodes in the common hepatic, paraaortic, aortocaval, and internal iliac areas (standard-dose CT vs. ultralow-dose CT and low-dose CT vs. ultralow-dose CT) among the three CT doses (Tables 4B and S1). The highest number of nodules in the adrenal glands was observed on standard-dose CT (n = 5, both), followed by low-dose CT (n = 3 and 4) and ultralow-dose CT (n = 3 and 2) by two readers.
Table 4

Lesion detection in organs (a) and detailed enlarged lymph node (b) analysis.

(a) Number of detected lesions in organs
Standard-dose CTLow-dose CTUltralow-dose CTReference standard
• Enlarged lymph node*99
 Lymph node based
Reader 1
 Detection979794
 False positive111
 False negative336
 Detection rate (%)97.0 (96/99)97.0 (96/99)94.0 (93/99)
Reader 2
 Detection959591
 False positive222
 False negative6610
 Detection rate (%)94.0 (93/99)94.0 (93/99)89.9 (89/99)
 Examination based26
 Reader 1262626
 Reader 2262626
• Liver lymphoma involvement0
Reader 1000
Reader 2000
• Nodule in adrenal glands5
Reader 1533
Reader 2542
• Splenomegaly
Reader 1202223
Reader 2242422
(b) Number of enlarged lymph nodes
Standard-dose CTLow-dose CTUltralow-dose CT
Left gastric area
Reader 1555
Reader 2555
Common hepatic area
Reader 1887
Reader 2998
Portocaval area
Reader 1666
Reader 2777
Retrocrural area
Reader 1222
Reader 2222
Paraaortic area
Reader 1171717
Reader 2151514
Aortocaval area
Reader 1141413
Reader 2111110
Common iliac area
Reader 1151515
Reader 2161616
Internal iliac area
Reader 1443
Reader 2554
External iliac area
Reader 1262626
Reader 2252525
Fig 2

Abdominopelvic computed tomography (CT) images of a 23-year-old woman with Hodgkin’s lymphoma, complete remission state (body mass index, 24.2 kg/m2; effective diameter, 24.0 cm).

The three different types of CT images, acquired according to the radiation dose (a-b: standard-dose CT, 5.6 mSv; c-d: low-dose CT, 3.7 mSv; e-f: ultralow-dose CT, 1.8 mSv), show the absence of enlarged lymph nodes and splenomegaly. Identical reports of the aforementioned features were also obtained from two radiologists. The overall image quality score was 4 for the ultralow-dose CT images and 5 for the low-dose and standard-dose CT images, according to three readers (two radiologists and one hematologist).

Fig 3

Abdominopelvic computed tomography (CT) images of a 54-year-old man with follicular lymphoma, partial response state (body mass index, 29.4 kg/m2; effective diameter, 27.0 cm).

The three different types of CT images, acquired according to the radiation dose (a-c: standard-dose CT, 6.8 mSv; d-f: low-dose CT, 4.5 mSv; g-i: ultralow-dose CT, 2.3 mSv), show multiple enlarged paraaortic (white arrow), aortocaval (gray arrow), right common iliac, and bilateral external iliac lymph nodes (black arrows). Identical findings for the aforementioned lesions were reported by two radiologists. The overall image quality score was 4 for ultralow-dose CT images and 5 for low-dose and standard-dose CT images, according to two radiologists; the image quality score was 5 for all the three types of CT images according to one hematologist.

Abdominopelvic computed tomography (CT) images of a 23-year-old woman with Hodgkin’s lymphoma, complete remission state (body mass index, 24.2 kg/m2; effective diameter, 24.0 cm).

The three different types of CT images, acquired according to the radiation dose (a-b: standard-dose CT, 5.6 mSv; c-d: low-dose CT, 3.7 mSv; e-f: ultralow-dose CT, 1.8 mSv), show the absence of enlarged lymph nodes and splenomegaly. Identical reports of the aforementioned features were also obtained from two radiologists. The overall image quality score was 4 for the ultralow-dose CT images and 5 for the low-dose and standard-dose CT images, according to three readers (two radiologists and one hematologist).

Abdominopelvic computed tomography (CT) images of a 54-year-old man with follicular lymphoma, partial response state (body mass index, 29.4 kg/m2; effective diameter, 27.0 cm).

The three different types of CT images, acquired according to the radiation dose (a-c: standard-dose CT, 6.8 mSv; d-f: low-dose CT, 4.5 mSv; g-i: ultralow-dose CT, 2.3 mSv), show multiple enlarged paraaortic (white arrow), aortocaval (gray arrow), right common iliac, and bilateral external iliac lymph nodes (black arrows). Identical findings for the aforementioned lesions were reported by two radiologists. The overall image quality score was 4 for ultralow-dose CT images and 5 for low-dose and standard-dose CT images, according to two radiologists; the image quality score was 5 for all the three types of CT images according to one hematologist. Lesion detection in organs (a) and detailed enlarged lymph node (b) analysis. The number of patients with splenomegaly on ultralow-dose, low-dose, and standard-lose CT was determined to be 23, 22, and 23 for radiologist 1; 22, 24, and 24 for radiologist 2; and 16, 16, and 18 for the hematologist, respectively. Spleen size measured by three readers (radiologists and one hematologist was 10.7 ± 1.69, 10.6 ± 1.78, and 10.0 ± 0.28 cm on ultralow-dose CT; 10.7 ± 1.70, 10.7 ± 1.80, and 10.0 ± 0.14 on low-dose CT; and 10.7 ± 1.70, 10.7 ± 1.79, and 10.0 ± 0.28 cm on standard-dose CT, respectively). We measured the average diameter of lymph nodes (short-axis diameter >0.5 cm, a total of 170 lymph nodes in each data set) on standard-dose, low-dose, and ultralow-dose CT, with no significant differences in the mean lymph node diameter among them (0.7±0.3, 0.7±0.3, and 0.7±0.4, respectively).

Quantitative and qualitative image analysis

A comparison of objective CT image quality based on measurement of image noise in subcutaneous fat, psoas muscle, liver, and abdominal aorta is presented in Table 5. Standard-dose CT exhibited the lowest image noise (9.4–11.5 HU), followed by low-dose CT (11.0–14.3 HU) and ultralow-dose CT (12.2–15.8 HU; P-value < 0.001; all post-hoc analysis). With regard to subjective image quality assessed by all readers, ultralow-dose CT exhibited lower overall image quality (3.8–4.1; P < 0.001; Kruskal-Wallis test), while standard-dose CT (4.7–4.9) and low-dose CT (4.6–4.8) exhibited higher overall image quality. There was no significant difference in overall image quality between standard-dose and low-dose CT (P-value [adjusted with Monte Carlo simulation] = 0.167, 0.121, and 0.088 for each reader, respectively).
Table 5

Quantitative and qualitative image analysis of three CT image sets.

Standard-dose CTLow-dose CTUltralow-dose CTP-value
Quantitative analysis (Hounsfield unit, HU)
Noise
 Subcutaneous fat9.4 ± 3.111.0 ± 3.112.2 ± 3.6< 0.001
 Psoas muscle10.5 ±1.813.2 ± 2.614.5 ± 2.9< 0.001
 Liver10.3 ± 2.012.7 ± 2.814.7 ± 2.7< 0.001
 Abdominal aorta11.5 ± 2.914.3 ± 3.015.8 ± 3.3< 0.001
Attenuation
 Subcutaneous fat-113.8 ± 9.0-113.8 ± 9.1-112.1 ± 9.4< 0.001
 Psoas muscle63.1 ± 7.663.5 ± 7.263.2 ± 7.2< 0.001
 Liver122.0 ± 18.7123.2 ± 18.9120.6 ± 19.4< 0.001
 Abdominal aorta188.9 ± 31.1189.3 ± 31.3190.4 ± 38.9< 0.001
Overall image quality*
Reader 14.8 ± 0.44.7 ± 0.43.8 ± 0.5< 0.001
Reader 24.7 ± 0.44.6 ± 0.53.9 ± 0.5< 0.001
Reader 34.9 ± 0.24.8 ± 0.34.1 ± 0.5< 0.001

*Readers 1 and 2 were both radiologists, and reader 3 was a hematologist.

*Readers 1 and 2 were both radiologists, and reader 3 was a hematologist.

Inter-observer agreement

S1 Table presents the inter-reader agreement of enlarged lymph node detection. Inter-reader agreement for enlarged lymph node detection by the two readers was excellent (κ = 0.83–1). Enlarged aortocaval lymph nodes on ultralow-dose CT exhibited the lowest inter-reader agreement score (κ = 0.83) in both common iliac areas. Intraclass correlation coefficient of spleen size was 0.90 (95% confidence interval [CI]), 0.87–0.93), 0.91 (95% CI, 0.88–0.94), and 0.91 (95% CI, 0.88–0.93) on ultralow-dose, low-dose, and standard-dose CT, respectively.

Radiation dose parameters

Table 6 summarizes the dose parameters of the three CT image sets. Mean CTDIvol of standard-dose, low-dose, and ultralow-dose CT was 5.6 ± 1.5 mGy, 3.7 ± 1.0 mGy, and 1.9 ± 0.5 mGy, respectively. Mean effective radiation dose of standard-dose, low-dose, and ultralow-dose CT was 5.7 ± 1.8 mSv, 3.8 ± 1.2 mSv, and 1.9 ± 0.6 mSv, respectively.
Table 6

Dose parameters of three CT image sets.

Standard-dose CTLow-dose CTUltralow-dose CTP-value
CTDIvol (mGy)5.6 ± 1.5 (2.5–11.0)3.7 ± 1.0 (1.7–7.4)1.9 ± 0.5 (0.8–3.6)< 0.001
Dose-length product (mGy-cm)335.7 ± 108.4 (137.9–710.1)223.9 ± 72.3 (92.0–473.6)72.5 ± 20.1 (45.9–236.5)< 0.001
Effective dose (mSv)5.7 ± 1.8 (2.3–12.1)3.8 ± 1.2 (1.5–8.1)1.9 ± 0.6 (0.8–4.0)< 0.001

Discussion

This study compared the radiation dose exposure, the detection rate of enlarged lymph nodes, and image quality of standard-dose, low-dose, and ultralow-dose abdominopelvic CT using the ADMIRE algorithm in patients with lymphoma. The enlarged lymph node-based detection rate was 94.0−97.0% on standard-dose and low-dose CT, and 89.9−94.0% on ultralow-dose CT according to two readers, with excellent inter-reader agreement. Ultralow-dose and low-dose CT effectively reduced radiation dose by 66.7% and 33.3%, respectively, while maintaining adequate image quality. We observed the same rate of detection of enlarged lymph nodes (i.e., > 1 cm in diameter) by two readers between low-dose and standard-dose CT. Although the rate of detection of enlarged lymph nodes was slightly higher on standard-dose CT than on ultralow-dose CT, there was no statistically significant difference between standard-dose and ultralow-dose CT. We conjecture that the detection of enlarged abdominal lymph nodes can be achieved in a relatively simple and clear manner by radiologists. The development of CT techniques and image reconstruction algorithms, including ADMIRE, may facilitate higher detection rates, even on ultralow-dose CT. Several studies have reported non-inferior or comparable diagnostic performance for a specific diagnosis (e.g., urinary stones or acute appendicitis) of a relatively simple disease or organ between low-dose abdominal CT and standard-dose CT [23-25]. However, the diagnostic performance of low-dose CT for small or inconspicuous abdominal structures has been unsatisfactory, even with the use of model-based IR [26, 27]. Due to high image noise in the abdomen, low-dose CT has limitations in the evaluation of small lesions in neoplastic conditions (e.g., liver metastasis and pancreatic cancer) and inflammatory diseases [18, 27, 28]. Small lesions with high image noise may obscure lesion detection due to similar attenuation as the background [27, 29]. Compared to standard-dose CT, ultralow-dose CT exhibited a similar rate of detection of enlarged lymph nodes in this study, while reducing the radiation dose by 33.3%. The high rate of detection of enlarged lymph nodes on ultralow-dose CT could be underpinned by the abdominal lymph nodes being less affected by image noise due to definite contrast differences between retroperitoneal or peritoneal fat and surrounding lymph nodes with a clear margin (i.e., background and lesion sharpness). These findings suggest that ultralow-dose CT may be employed instead of standard-dose CT during follow-up of patients with lymphoma. Notably, the interobserver agreement for abdominal lymph nodes was excellent in our study. This could be due to the sparsity of other anatomical structures surrounding the abdominal lymph nodes; as such, the diagnosis of enlarged lymph nodes exhibited small differences between the two readers despite the presence of image noise. With regard to lymph node evaluation using low-dose CT, Paolini et al. reported that there was no significant difference in the delineation of thoracic lymph nodes between contrast-enhanced low-dose and standard-dose CT [30]. Mueller‑Lisse et al. also reported that contrast-enhanced low-dose CT with approximately 1 mSv revealed equivalent delineation of thoracic lymph nodes compared to standard-dose CT [31]. In line with prior studies on thoracic lymph nodes, our study revealed similar results regarding abdominal lymph node evaluation using low-dose and ultralow-dose CT in patients with lymphoma. Hérin et al. reported that reduced-dose CT with model-based IR could reduce the amount of radiation delivered to patients with lymphoma while maintaining image quality comparable to that of standard-dose CT with filtered-back projection [12]. Herein, we compared three CT image sets using ADMIRE (i.e., without using a filtered-back projection comparison). We implemented ADMIRE in actual practice and focused on the number of radiation doses to be reduced in patients with lymphoma while maintaining image quality. Quantitative image noise increased with a decrease in radiation dose. However, we observed that subjective visual image quality was comparable between standard-dose and low-dose CT, and there were no significant differences in the results of lymph node evaluation between these image sets. Although low-dose CT exhibited slightly increased objective image noise and comparable subjective image quality compared to standard-dose CT, this may not affect the diagnosis of lymph node detection. Nevertheless, ultralow-dose CT resulted in degradation of both qualitative visual image quality and quantitative image noise analysis. A blind image analysis was performed in this study for the independent evaluation of lymph node enlargement in each image reading session. The use of low-dose CT as an imaging modality for follow-up evaluation of lymphoma permits the use of initial standard-dose CT as a reference and comparative examinations in actual practice. We predict better diagnostic performance of enlarged lymph node detection, even using ultralow-dose CT, compared to the current results. In consideration of the long-term follow-up period, low-dose and/or ultralow-dose CT offers a safe and accurate alternative imaging modality to replace current standard-dose CT. Our study has a few limitations. First, we only assessed changes in tube load (mAs) and a single image reconstruction using ADMIRE, and we did not consider filtered-back projection. Thus, the study protocol differs from clinical protocols, as a fixed kVp was used for both detection tubes without automatic tube voltage adjustment (Care kV). As various image acquisition parameters can affect image quality in CT, our results may have limited generalizability. Second, as images reconstructed with ADMIRE have different appearances, it is difficult to achieve a true blinded analysis of subjective criteria. Third, because most of the patients who underwent dual energy CT examination at our hospital were outpatients, most of the patients with lymphoma included in the study were outpatients, and we predominantly focused on treatment response. Therefore, our results may be applicable to surveillance of patients after lymphoma treatment but not for patients with initial lymphoma evaluation, relapse, or suspected progressive disease. Finally, we analyzed CT images with a slice thickness of 5 mm in the axial direction and 3 mm in the coronal direction. Although these parameters can reflect readings in actual practice, it may also lead to partial volume effect when measuring lymph nodes since thin slices (e.g., 2 mm) were not analyzed. In conclusion, our findings suggest that ultralow-dose and low-dose CT effectively reduce radiation dose by 66.7% and 33.3%, respectively, while maintaining acceptable image quality and can be used as an imaging modality for follow-up evaluation of lymphoma, in consideration of the long-term follow-up period.

The number of enlarged lymph nodes and each interobserver agreement.

(DOCX) Click here for additional data file.

STROBE checklist.

(DOC) Click here for additional data file.

Statistical analysis of subjective image quality comparisons.

(DOCX) Click here for additional data file.
  31 in total

1.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

2.  Comparison of standard-dose and half-dose dual-source abdominopelvic CT scans for evaluation of acute abdominal pain.

Authors:  Ki Hyun Lee; Young Sup Shim; So Hyun Park; Seong Ho Park; Seung Joon Choi; Seong Yong Pak; Hyunhee Cheong
Journal:  Acta Radiol       Date:  2018-10-30       Impact factor: 1.990

3.  NCCN Guidelines Insights: Non-Hodgkin's Lymphomas, Version 3.2016.

Authors:  Steven M Horwitz; Andrew D Zelenetz; Leo I Gordon; William G Wierda; Jeremy S Abramson; Ranjana H Advani; C Babis Andreadis; Nancy Bartlett; John C Byrd; Luis E Fayad; Richard I Fisher; Martha J Glenn; Thomas M Habermann; Nancy Lee Harris; Francisco Hernandez-Ilizaliturri; Richard T Hoppe; Mark S Kaminski; Christopher R Kelsey; Youn H Kim; Susan Krivacic; Ann S LaCasce; Matthew Lunning; Auayporn Nademanee; Oliver Press; Rachel Rabinovitch; Nishitha Reddy; Erin Reid; Kenneth Roberts; Ayman A Saad; Lubomir Sokol; Lode J Swinnen; Julie M Vose; Joachim Yahalom; Nadeem Zafar; Mary Dwyer; Hema Sundar; Pierluigi Porcu
Journal:  J Natl Compr Canc Netw       Date:  2016-09       Impact factor: 11.908

4.  Diagnostic Accuracy of Low and Ultra-Low Dose CT for Identification of Urinary Tract Stones: A Systematic Review.

Authors:  Flora Rodger; Giles Roditi; Omar M Aboumarzouk
Journal:  Urol Int       Date:  2018-04-12       Impact factor: 2.089

5.  Recommendations for initial evaluation, staging, and response assessment of Hodgkin and non-Hodgkin lymphoma: the Lugano classification.

Authors:  Bruce D Cheson; Richard I Fisher; Sally F Barrington; Franco Cavalli; Lawrence H Schwartz; Emanuele Zucca; T Andrew Lister
Journal:  J Clin Oncol       Date:  2014-09-20       Impact factor: 44.544

6.  Thoracic lymph node delineation at dose-reduced (1 mSv) dose-modulated contrast enhanced MDCT: a retrospective pilot study.

Authors:  Marco Paolini; Kathrin Wirth; Amanda Tufman; Maximilian Reiser; Rudolf M Huber; Ullrich G Mueller-Lisse
Journal:  Radiol Med       Date:  2016-05-06       Impact factor: 3.469

Review 7.  Modelling the physics in the iterative reconstruction for transmission computed tomography.

Authors:  Johan Nuyts; Bruno De Man; Jeffrey A Fessler; Wojciech Zbijewski; Freek J Beekman
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

8.  Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults.

Authors:  Aaron Sodickson; Pieter F Baeyens; Katherine P Andriole; Luciano M Prevedello; Richard D Nawfel; Richard Hanson; Ramin Khorasani
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

Review 9.  CT radiation dose optimization and estimation: an update for radiologists.

Authors:  Hyun Woo Goo
Journal:  Korean J Radiol       Date:  2011-12-23       Impact factor: 3.500

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.