Masahide Saito1, Takafumi Komiyama1, Kan Marino1, Shinichi Aoki1, Mitsuhiko Oguri1, Takashi Yamada1, Naoki Sano1, Hidekazu Suzuki1, Koji Ueda1, Hiroshi Onishi1.
Abstract
Purpose: We aimed to investigate the effects of different multi-leaf collimator (MLC) speed constraints in volumetric modulated radiotherapy (VMAT) on the robustness of treatment plans for central lung cancer patients. Method and Materials: Twenty patients with central lung tumor who underwent stereotactic body radiotherapy (SBRT) with the VMAT technique at our hospital were included in this retrospective study. The reference plans were created with 3 different MLC speed constraints (Plan A: 0.1 cm/deg., Plan B: 0.3 cm/deg., and Plan C: 0.5 cm/deg.) with a 50-Gy/8Fr, planning target volume (PTV) D95% prescription. In each of these plans, setup errors from 1 to 5 mm were intentionally added in the direction of the central organ at 1-mm intervals (300 plans [20 cases × 3 MLC speeds × 5 error plans] were created in total). Each plan was then calculated by the same beam conditions as each reference plan. The actual average MLC speed and dose difference between the reference plan and the error-added plan were then calculated and compared among the 3 MLC speeds.Entities:
Keywords: IMRT; lung cancer; radiation dosimetry; radiation therapy; x-ray
Mesh:
Year: 2022 PMID: 35950289 PMCID: PMC9379802 DOI: 10.1177/15330338221119752
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
The Initial Values of the Optimization Conditions in the Treatment Planning System.
| ROI | Function description | Weight |
|---|---|---|
| PTV | Min DVH 5000 cGy to 95% volume | 10 |
| Max Dose 5750 cGy | 10 | |
| External | Dose fall-off [H]5000 cGy [L]2500 cGy, Low dose distance 0.75 cm | 1 |
| Canal | Max dose 2500 cGy | 0.01 |
| Others | As low as possible (trachea, heart, esophagus, aorta, pulmonary artery) | 0.01 |
Figure 1.Outline of this study. The study protocol consisted of 3 steps: (1) creation of the reference plan, (2) creation of error plans, and (3) comparison of each plan.
Comparison of each parameter in the 3 original plans. The results of the comparison are shown for 0.1 cm/deg., 0.3 cm/deg., and 0.5 cm/deg. as A, B and C, respectively.
| A = 0.1 cm/deg. | B = 0.3 cm/deg. | C = 0.5 cm/deg. | A versus B | B versus C | A versus C | ||
|---|---|---|---|---|---|---|---|
| Actual MLC speed ( | 0.25 ± 0.04 | 0.34 ± 0.07 | 0.39 ± 0.12 | <.0001* | <.0001* | <.0001* | |
| PTV ( | Dmean | 55.29 ± 0.29 | 54.79 ± 1.01 | 54.76 ± 0.97 | .0050* | .8394 | .0010* |
| D98% | 48.39 ± 0.50 | 48.38 ± 0.56 | 48.42 ± 0.52 | .6574 | .4116 | .7754 | |
| D50% | 55.41 ± 0.97 | 55.00 ± 1.12 | 55.00 ± 1.06 | .0094* | .7419 | .0172* | |
| D2% | 60.92 ± 2.05 | 59.30 ± 1.88 | 59.11 ± 1.87 | <.0001* | .0310* | <.0001* | |
| Esophagus ( | D1cc | 19.80 ± 8.98 | 19.76 ± 9.28 | 19.45 ± 8.99 | .8501 | .0137* | .4238 |
| D5cc | 8.52 ± 6.21 | 8.55 ± 6.26 | 8.46 ± 6.21 | .9849 | .1641 | .3101 | |
| Heart ( | D10cc | 19.73 ± 5.30 | 19.36 ± 5.77 | 19.18 ± 5.64 | .5625 | .1250 | .4375 |
| D15cc | 17.68 ± 4.38 | 17.37 ± 4.93 | 17.18 ± 4.77 | .6875 | .6250 | .4375 | |
| Aorta ( | D1cc | 36.20 ± 15.22 | 35.59 ± 14.76 | 35.59 ± 14.87 | .0503 | .9541 | .0425* |
| D10cc | 21.08 ± 8.20 | 22.94 ± 13.09 | 20.42 ± 7.79 | .1099 | .1641 | .0142* | |
| Pulmonary artery ( | D1cc | 31.30 ± 15.54 | 30.76 ± 14.68 | 30.39 ± 14.71 | .4316 | .0977 | .0645 |
| D10cc | 13.61 ± 7.57 | 12.77 ± 7.23 | 13.94 ± 7.87 | .0137* | .8438 | .1602 | |
| Trachea ( | D1cc | 17.06 ± 6.95 | 16.93 ± 6.88 | 16.93 ± 6.91 | .2813 | 1.0000 | .3125 |
| D10cc | 7.28 ± 5.85 | 7.81 ± 6.45 | 7.83 ± 6.45 | .1563 | 1.0000 | .1563 | |
Figure 2.A comparison of the dose difference between the reference plan and error plan for planning target volume (PTV). The horizontal axis shows the magnitude of the error, and the vertical axis shows the dose difference from the reference plan: (A) PTV Dmean, (B) PTV D50%, (C) PTV D98%, and (D) PTV D2%. Blue, orange, and gray represent 0.1, 0.3, and 0.5 cm/deg., respectively.
Figure 3.A comparison of the dose difference between the reference plan and error plan for normal organs. The horizontal axis shows the magnitude of the error, and the vertical axis shows the dose difference from the reference plan: (A) aorta D1cc, (B) aorta D10cc, (C) esophagus D1cc, (D) esophagus D5cc, (E) pulmonary artery D1cc, (F) pulmonary artery D10cc, (G) trachea D1cc, (H) trachea D10cc, (I) heart D10cc, and (J) heart D15cc. Blue, orange, and gray represent 0.1, 0.3, and 0.5 cm/deg., respectively.