PURPOSE: To evaluate the effect of plan parameters on volumetric modulated arc therapy (VMAT) dosimetric accuracy, together with the possibility of scoring plan complexity. METHODS: 142 clinical VMAT plans initially optimized using a 4° control point (CP) separation were evaluated. All plans were delivered by a 6 MV Linac to a biplanar diode array for patient-specific quality assurance (QA). Local Γ index analysis (3%, 3 mm and 2%, 2 mm) enabled the comparison between delivered and calculated dose. The following parameters were considered for each plan: average leaf travel (LT), modulation complexity score applied to VMAT (MCSv), MU value, and a multiplicative combination of LT and MCSv (LTMCS). Pearson's correlation analysis was performed between Γ passing rates and each parameter. The effects of CP angular separation on VMAT dosimetric accuracy were also analyzed by focusing on plans with high LT values. Forty out of 142 plans with LT above 350 mm were further optimized using a finer angle spacing (3° or 2°) and Γ analysis was performed. The average Γ passing rates obtained at 4° and at 3°∕2° sampling were compared. A further correlation analysis between all parameters and the Γ pass-rates was performed on 142 plans, but including the newly optimized 40 plans (CP every 3° or 2°) in place of the old ones (CP every 4°). RESULTS: A moderate significant (p < 0.05) correlation between each examined parameter and Γ passing rates was observed for the original 142 plans at 4° CP discretization. A negative correlation was found for LT with Pearson's r absolute values above 0.6, suggesting that a lower dosimetric accuracy may be expected for higher LT values when a 4° CP sampling is used. A positive correlation was observed for MCSv and LTMCS with r values above 0.5. In order to score plan complexity, threshold values of LTMCS were defined. The average Γ passing rates were significantly higher for the plans created using the finer CP spacing (3°∕2°) compared to the plans optimized using the standard 4° spacing (Student t-test p < 0.05). The correlation between LT and passing rates was strongly diminished when plans with finer angular separations were considered, yielding Pearson's r absolute values below 0.45. CONCLUSIONS: At 4° CP sampling, LT, MCSv, and LTMCS were found to be significantly correlated with VMAT dosimetric accuracy, expressed as Γ pass-rates. These parameters were found to be possible candidates for scoring plan complexity using threshold values. A finer CP separation (3°∕2°) led to a significant increase in dosimetric accuracy for plans with high leaf travel values, and to a decrease in correlation between LT and Γ passing rates. These results indicated that the influence of LT on VMAT dosimetric accuracy can be controlled by reducing CP separation. CP spacing for all plans requiring large leaf motion should not exceed 3°. The reported data were integrated to optimize our clinical workflow for plan creation, optimization, selection among rival plans, and patient-specific QA of VMAT treatments.
PURPOSE: To evaluate the effect of plan parameters on volumetric modulated arc therapy (VMAT) dosimetric accuracy, together with the possibility of scoring plan complexity. METHODS: 142 clinical VMAT plans initially optimized using a 4° control point (CP) separation were evaluated. All plans were delivered by a 6 MV Linac to a biplanar diode array for patient-specific quality assurance (QA). Local Γ index analysis (3%, 3 mm and 2%, 2 mm) enabled the comparison between delivered and calculated dose. The following parameters were considered for each plan: average leaf travel (LT), modulation complexity score applied to VMAT (MCSv), MU value, and a multiplicative combination of LT and MCSv (LTMCS). Pearson's correlation analysis was performed between Γ passing rates and each parameter. The effects of CP angular separation on VMAT dosimetric accuracy were also analyzed by focusing on plans with high LT values. Forty out of 142 plans with LT above 350 mm were further optimized using a finer angle spacing (3° or 2°) and Γ analysis was performed. The average Γ passing rates obtained at 4° and at 3°∕2° sampling were compared. A further correlation analysis between all parameters and the Γ pass-rates was performed on 142 plans, but including the newly optimized 40 plans (CP every 3° or 2°) in place of the old ones (CP every 4°). RESULTS: A moderate significant (p < 0.05) correlation between each examined parameter and Γ passing rates was observed for the original 142 plans at 4° CP discretization. A negative correlation was found for LT with Pearson's r absolute values above 0.6, suggesting that a lower dosimetric accuracy may be expected for higher LT values when a 4° CP sampling is used. A positive correlation was observed for MCSv and LTMCS with r values above 0.5. In order to score plan complexity, threshold values of LTMCS were defined. The average Γ passing rates were significantly higher for the plans created using the finer CP spacing (3°∕2°) compared to the plans optimized using the standard 4° spacing (Student t-test p < 0.05). The correlation between LT and passing rates was strongly diminished when plans with finer angular separations were considered, yielding Pearson's r absolute values below 0.45. CONCLUSIONS: At 4° CP sampling, LT, MCSv, and LTMCS were found to be significantly correlated with VMAT dosimetric accuracy, expressed as Γ pass-rates. These parameters were found to be possible candidates for scoring plan complexity using threshold values. A finer CP separation (3°∕2°) led to a significant increase in dosimetric accuracy for plans with high leaf travel values, and to a decrease in correlation between LT and Γ passing rates. These results indicated that the influence of LT on VMAT dosimetric accuracy can be controlled by reducing CP separation. CP spacing for all plans requiring large leaf motion should not exceed 3°. The reported data were integrated to optimize our clinical workflow for plan creation, optimization, selection among rival plans, and patient-specific QA of VMAT treatments.
Authors: Sophie Chiavassa; Igor Bessieres; Magali Edouard; Michel Mathot; Alexandra Moignier Journal: Br J Radiol Date: 2019-07-24 Impact factor: 3.039
Authors: Mallory C Glenn; Victor Hernandez; Jordi Saez; David S Followill; Rebecca M Howell; Julianne M Pollard-Larkin; Shouhao Zhou; Stephen F Kry Journal: Phys Med Biol Date: 2018-10-17 Impact factor: 3.609
Authors: Conor K McGarry; Christina E Agnew; Mohammad Hussein; Yatman Tsang; Alan McWilliam; Alan R Hounsell; Catharine H Clark Journal: Br J Radiol Date: 2015-10-29 Impact factor: 3.039
Authors: Vincent Caillet; Ricky O'Brien; Douglas Moore; Per Poulsen; Tobias Pommer; Emma Colvill; Amit Sawant; Jeremy Booth; Paul Keall Journal: Med Phys Date: 2019-03-04 Impact factor: 4.071
Authors: Christos Moustakis; Mark K H Chan; Jinkoo Kim; Joakim Nilsson; Alanah Bergman; Tewfik J Bichay; Isabel Palazon Cano; Savino Cilla; Francesco Deodato; Raffaela Doro; Jürgen Dunst; Hans Theodor Eich; Pierre Fau; Ming Fong; Uwe Haverkamp; Simon Heinze; Guido Hildebrandt; Detlef Imhoff; Erik de Klerck; Janett Köhn; Ulrike Lambrecht; Britta Loutfi-Krauss; Fatemeh Ebrahimi; Laura Masi; Alan H Mayville; Ante Mestrovic; Maaike Milder; Alessio G Morganti; Dirk Rades; Ulla Ramm; Claus Rödel; Frank-Andre Siebert; Wilhelm den Toom; Lei Wang; Stefan Wurster; Achim Schweikard; Scott G Soltys; Samuel Ryu; Oliver Blanck Journal: Strahlenther Onkol Date: 2018-05-25 Impact factor: 3.621