Literature DB >> 35913540

Risk factors for hardware-related complications after extra-articular distal humerus fracture fixation using an anatomical locking plate.

Hyoung-Seok Jung1, Yoo-Sun Won2, Yang-Seon Choi2, Jae-Sung Lee3.   

Abstract

PURPOSE: We aimed to identify the incidence and risk factors of hardware-related complications in patients treated with anatomical locking plate fixation for extra-articular distal humerus fractures.
METHODS: From 2013 to 2020, patients with extra-articular distal humerus fractures who underwent open reduction and internal fixation with an extra-articular distal humerus locking plate (EADHP) were retrospectively reviewed and categorized according to the presence/absence of hardware-related complications. Hardware-related complications were defined as the occurrence of skin prominence on the plate and discomfort in activities of daily living. Patient demographics, the lateral condylar angle, lateral body length, shaft-condylar angle of the humerus, and plate length were analyzed.
RESULTS: Of the 29 patients, 10 (34%) did not develop hardware-related complications (group A), whereas 19 (66%) did (group B). Patient demographics did not differ between the groups. However, the number of patients who underwent hardware removal was significantly greater in group B (16/19) than in group A (4/10; p = 0.032). Radiologic assessment revealed no significant difference in the lateral condylar or shaft-condylar angle. However, the lateral body length was greater in group A than in group B (44.5 ± 4.8 vs. 39.5 ± 3.7, p = 0.007). The plate length significantly differed between the groups. Twelve of 19 (63%) patients in group B received short-hole plates (six holes), while nine of ten (90%) patients in group A received long-hole plates (eight holes). In the multivariable analysis, the lateral body length of the distal humerus (p = 0.047, odds ratio = 0.734, 95% confidence interval: 0.542-0.996) and plate length (p = 0.036, odds ratio = 0.076, 95% confidence interval: 0.542-0.996) were associated with hardware-related complications.
CONCLUSIONS: Most patients developed hardware-related complications, particularly with short plates, mainly because of the narrow lateral body length of the distal humerus. Surgeons should be careful to secure EADHP in the appropriate position, especially when short plates are used in patients with narrow lateral body length.
© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany.

Entities:  

Keywords:  Complication; Fracture; Humerus; Locking plate

Year:  2022        PMID: 35913540     DOI: 10.1007/s00068-022-02064-0

Source DB:  PubMed          Journal:  Eur J Trauma Emerg Surg        ISSN: 1863-9933            Impact factor:   2.374


  21 in total

1.  Mismatch of AO anatomically shaped distal humeral plate with humeral shaft forward flexion angulation in adult Chinese population.

Authors:  Zhiyong Zhou; Zhifeng Tang; Xueqin Zhao; Wentao Chen; Xu Chen; Mingzhang Mu; Shudong Zhang; Bin Ning; Tanghong Jia
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-03-22

2.  Outcomes of distal humerus diaphyseal injuries fixed with a single-column anatomic plate.

Authors:  John T Capo; Monika P Debkowska; Frank Liporace; Bryan G Beutel; Eitan Melamed
Journal:  Int Orthop       Date:  2013-12-15       Impact factor: 3.075

3.  Surgical fixation of extra-articular distal humerus fractures with a posterolateral plate through a triceps-reflecting technique.

Authors:  John A Scolaro; Pramod Voleti; Amun Makani; Surena Namdari; Amer Mirza; Samir Mehta
Journal:  J Shoulder Elbow Surg       Date:  2013-12-14       Impact factor: 3.019

4.  Minimally Invasive Osteosynthesis with a Bridge Plate Versus a Functional Brace for Humeral Shaft Fractures: A Randomized Controlled Trial.

Authors:  Fabio Teruo Matsunaga; Marcel Jun Sugawara Tamaoki; Marcelo Hide Matsumoto; Nicola Archetti Netto; Flavio Faloppa; Joao Carlos Belloti
Journal:  J Bone Joint Surg Am       Date:  2017-04-05       Impact factor: 5.284

5.  Lateral J-plate fixation in comminuted intercondylar fracture of the humerus.

Authors:  Kyu Hyun Yang; Hui Wan Park; Sung Jin Park; Sung Hoon Jung
Journal:  Arch Orthop Trauma Surg       Date:  2003-04-25       Impact factor: 3.067

6.  Fixation of extra-articular distal humerus fractures using one locking plate versus two reconstruction plates: a laboratory study.

Authors:  Nirmal C Tejwani; Anjali Murthy; Jason Park; Toni M McLaurin; Kenneth A Egol; Fred J Kummer
Journal:  J Trauma       Date:  2009-03

7.  Plate selection for fixation of extra-articular distal humerus fractures: a biomechanical comparison of three different implants.

Authors:  John A Scolaro; Jason E Hsu; David J Svach; Samir Mehta
Journal:  Injury       Date:  2014-09-10       Impact factor: 2.586

8.  A paradigm shift in the surgical reconstruction of extra-articular distal humeral fractures: single-column plating.

Authors:  Gregory M Meloy; Matthew A Mormino; Peter A Siska; Ivan S Tarkin
Journal:  Injury       Date:  2013-08-08       Impact factor: 2.586

9.  Outcome of anatomic locking plate in extraarticular distal humeral shaft fractures.

Authors:  Deepak Jain; Gurpreet S Goyal; Rajnish Garg; Pankaj Mahindra; Mohammad Yamin; Harpal S Selhi
Journal:  Indian J Orthop       Date:  2017 Jan-Feb       Impact factor: 1.251

10.  Anatomic fit of precontoured extra-articular distal humeral locking plates: a cadaveric study.

Authors:  Joon-Ryul Lim; Tae-Hwan Yoon; Hwan-Mo Lee; Yong-Min Chun
Journal:  Clin Shoulder Elb       Date:  2021-05-27
View more