Measurement of magnetostimulation thresholds in the porcine heart.

Valerie Klein1,2, Jaume Coll-Font2,3,4, Livia Vendramini2, Donald Straney2, Mathias Davids1,2,3, Natalie G Ferris2,5,6, Lothar R Schad1, David E Sosnovik2,3,4,5, Christopher T Nguyen2,3,4,5, Lawrence L Wald2,3,5,6, Bastien Guérin2,3.   

Abstract

PURPOSE: Powerful MRI gradient systems can surpass the International Electrotechnical Commission (IEC) 60601-2-33 limit for cardiac stimulation (CS), which was determined by simple electromagnetic simulations and electrode stimulation experiments. Only a few canine studies measured magnetically induced CS thresholds in vivo and extrapolating them to human safety limits can be challenging.
METHODS: We measured cardiac magnetostimulation thresholds in 10 healthy, anesthetized pigs using capacitors discharged into a flat spiral coil to produce damped sinusoidal waveforms with effective stimulus duration ts,eff  = 0.45 ms. Electrocardiography (ECG), blood pressure, and peripheral oximetry signals were recorded to determine threshold coil currents yielding cardiac capture. Dixon and CINE MR volumes from each animal were segmented to generate porcine-specific electromagnetic models to calculate dB/dt and E-field values in the porcine heart at threshold. For comparison, we also simulated maximum dB/dt and E-field values created by three MRI gradient systems in the heart of a human body model.
RESULTS: The average dB/dt threshold estimated in the porcine heart was 1.66 ± 0.23 kT/s, which is 11-fold greater than the IEC dB/dt limit at ts,eff  = 0.45 ms, and 31-fold greater than the maximum value created by the investigated MRI gradients in the human heart. The average E-field threshold estimated in the porcine heart was 92.9 ± 13.5 V/m, which is 6-fold greater than the IEC E-field limit at ts,eff  = 0.45 ms and 37-fold greater than the maximum gradient-induced E-field in the human heart.
CONCLUSION: This first measurement of cardiac magnetostimulation thresholds in pigs indicates that the IEC cardiac safety limit is conservative for the investigated stimulus duration (ts,eff  = 0.45 ms).
© 2022 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  MRI gradient field switching; cardiac stimulation; electromagnetic exposure safety; magnetostimulation thresholds

Mesh:

Year:  2022        PMID: 35906903      PMCID: PMC9420805          DOI: 10.1002/mrm.29382

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  27 in total

1.  Roles of electric field and fiber structure in cardiac electric stimulation.

Authors:  S B Knisley; N Trayanova; F Aguel
Journal:  Biophys J       Date:  1999-09       Impact factor: 4.033

Review 2.  Principles of nerve and heart excitation by time-varying magnetic fields.

Authors:  J P Reilly
Journal:  Ann N Y Acad Sci       Date:  1992-03-31       Impact factor: 5.691

3.  Magnetostimulation in MRI.

Authors:  W Irnich; F Schmitt
Journal:  Magn Reson Med       Date:  1995-05       Impact factor: 4.668

4.  Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations.

Authors:  Mathias Davids; Bastien Guérin; Axel Vom Endt; Lothar R Schad; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2018-08-09       Impact factor: 4.668

5.  Limits to neural stimulation in echo-planar imaging.

Authors:  P Mansfield; P R Harvey
Journal:  Magn Reson Med       Date:  1993-06       Impact factor: 4.668

6.  Factors determining vulnerability to ventricular fibrillation induced by 60-CPS alternating current.

Authors:  T Sugimoto; S F Schaal; A G Wallace
Journal:  Circ Res       Date:  1967-11       Impact factor: 17.367

7.  Sensitivity analysis of neurodynamic and electromagnetic simulation parameters for robust prediction of peripheral nerve stimulation.

Authors:  Valerie Klein; Mathias Davids; Lawrence L Wald; Lothar R Schad; Bastien Guérin
Journal:  Phys Med Biol       Date:  2018-12-19       Impact factor: 3.609

8.  Optimization of MRI Gradient Coils With Explicit Peripheral Nerve Stimulation Constraints.

Authors:  Mathias Davids; Bastien Guerin; Valerie Klein; Lawrence L Wald
Journal:  IEEE Trans Med Imaging       Date:  2020-12-29       Impact factor: 10.048

9.  Swine as an in vivo model for electrophysiologic evaluation of cardiac pacing parameters.

Authors:  T A Bowman; H C Hughes
Journal:  Pacing Clin Electrophysiol       Date:  1984-03       Impact factor: 1.976

10.  Investigating cardiac stimulation limits of MRI gradient coils using electromagnetic and electrophysiological simulations in human and canine body models.

Authors:  Valerie Klein; Mathias Davids; Lothar R Schad; Lawrence L Wald; Bastien Guérin
Journal:  Magn Reson Med       Date:  2020-08-19       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.