Literature DB >> 35700271

Methods for Quantifying the Metabolic Boundary Fluxes of Cell Cultures in Large Cohorts by High-Resolution Hydrophilic Liquid Chromatography Mass Spectrometry.

Ryan A Groves1, Maryam Mapar1, Raied Aburashed1, Luis F Ponce1, Stephanie L Bishop1, Thomas Rydzak1, Marija Drikic1, Dominique G Bihan1, Hallgrimur Benediktsson2,3, Fiona Clement4, Daniel B Gregson2,3,5, Ian A Lewis1.   

Abstract

Metabolomics is a mainstream approach for investigating the metabolic underpinnings of complex biological phenomena and is increasingly being applied to large-scale studies involving hundreds or thousands of samples. Although metabolomics methods are robust in smaller-scale studies, they can be challenging to apply to larger cohorts due to the inherent variability of liquid chromatography mass spectrometry (LC-MS). Much of this difficulty results from the time-dependent changes in the LC-MS system, which affects both the qualitative and quantitative performances of the instrument. Herein, we introduce an analytical strategy for addressing this problem in large-scale microbial studies. Our approach quantifies microbial boundary fluxes using two zwitterionic hydrophilic interaction liquid chromatography (ZIC-HILIC) columns that are plumbed to enable offline column equilibration. Using this strategy, we show that over 397 common metabolites can be resolved in 4.5 min per sample and that metabolites can be quantified with a median coefficient of variation of 0.127 across 1100 technical replicates. We illustrate the utility of this strategy via an analysis of 960 strains of Staphylococcus aureus isolated from bloodstream infections. These data capture the diversity of metabolic phenotypes observed in clinical isolates and provide an example of how large-scale investigations can leverage our novel analytical strategy.

Entities:  

Mesh:

Year:  2022        PMID: 35700271      PMCID: PMC9244871          DOI: 10.1021/acs.analchem.2c00078

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   8.008


Introduction

Metabolomics has emerged as a mainstream strategy for investigating cell physiology and understanding the molecular underpinnings of diseases. Liquid chromatography mass spectrometry (LC-MS) has become one of the most widely used platforms for metabolomics due to its relatively simple sample preparation requirements and its broad metabolite coverage.[1] Over the last 20 years, the depth of LC-MS methods has been expanded to capture ever broader transects of the metabolic network. These efforts have dramatically increased the scope of routine metabolomics[2] and are now allowing researchers to map the previously unexplored “dark metabolome”.[3] Although these methods are indispensable for capturing the broadest possible scope of metabolism, not all metabolomics studies require this depth of coverage. This is particularly relevant in the context of large-cohort studies, where the objective is to quantify a select set of molecules consistently across a large number of samples. The analysis of metabolic boundary fluxes is one application where it is particularly important to have LC-MS methods optimized for long-term stability, rather than the depth of metabolite coverage. Boundary fluxes refer to the rates at which metabolites are consumed or secreted by cells[4] and can be quantified by measuring changes in the composition of cell culture growth media over time. Quantifying boundary fluxes can provide a comprehensive understanding of cellular function by enabling flux balance models of cell metabolism to be optimized,[5] nutritional dependencies to be mapped,[6] and metabolic networks of model organisms to be optimized for industrial applications.[7,8] Moreover, we have recently shown that metabolic boundary fluxes can be used as a clinical microbiology assay for identifying microbial species and measuring antibiotic susceptibility profiles.[9] One critical distinction between boundary flux analysis and routine metabolomics is the complexity of the samples. Whereas intracellular metabolic phenotypes affect thousands of molecules,[10] cell media can contain 100 or fewer organic molecules.[11] The long chromatographic gradients that have been developed for intracellular metabolomics are therefore not optimized for the modest complement of molecules found in culture media and needlessly increase the analysis time, cost, and error arising from time-dependent changes in LC-MS response factors. Another key distinction between boundary flux studies and routine metabolomics is scale. Although relatively few metabolites need to be quantified in cell media, the long-term quantitative stability of these analyses is critical for minimizing batch effects across large cohorts. A variety of chromatographic,[12] ionization,[13] and other instrument factors contribute to the loss of LC-MS performance over time. One major factor is the deposition of lipids, proteins, salts, and other insoluble materials on the electrospray ionization source and ion optics.[14] These deposits cause progressive changes in the response factors that generally result in a loss of ionization efficacy. A related problem is the progressive deterioration of the chromatography column, which results from degradation of the stationary phase.[12] The time-dependent loss of chromatographic and ionization performance on LC-MS instruments is exacerbated in metabolomics studies because the complex samples can quickly foul the instrument. In contrast, the cell culture medium analyzed for boundary flux studies tends to be less complex and less chemically diverse than cell extracts, thereby enabling the analysis of larger cohorts of samples than would be practical for many metabolomics studies. However, the hydrophilic nature of most media components means that they are best resolved by hydrophilic interaction liquid chromatography (HILIC),[4,15,16] which requires long equilibration times between injections to stabilize the stationary phase.[17,18] These long delays between injections are counterproductive in large-cohort studies when the primary objective is to maximize sample throughput. Novel methods that mitigate these sorts of shortcomings are thus necessary for large-scale analyses of metabolic boundary fluxes. In summary, analyzing boundary fluxes provides a powerful approach for investigating cell physiology in large-cohort studies, but modern metabolomics methods are not optimized for this application. Whereas conventional metabolomics methods are tuned to maximize the breadth of chemical diversity and the depth of metabolite coverage, boundary flux studies track a small collection of aqueous metabolites, have less complex input samples, and emphasize quantitative stability across large cohorts. To address this discrepancy, we developed a zwitterionic HILIC (ZIC-HILIC) LC-MS method that is specifically optimized to enable the quantification of microbial boundary fluxes. Our approach uses sample injection multiplexing to minimize column re-equilibration times and a standardized LC-conditioning workflow for stabilizing the performance of the instrument. We show that this approach enables over 397 metabolites to be resolved in a 4.5 min gradient. Moreover, we show that metabolites observed in Staphylococcus aureus cultures can be quantified over 1100 injections with a median coefficient of variation of 0.127 across the observed metabolites.

Experimental Section

Analytical Standard Preparation

For all qualitative work, the mass spectrometry metabolite library service (MSMLS; Sigma-Aldrich) was used to prepare standards for determining chromatographic performance. Standards were received preweighed in 96-well plates with 5 μg in each well and were subsequently reconstituted to a concentration of 25 μg/mL using 50% methanol (1:1 Optima LC-MS grade water and methanol; Thermo Fisher Scientific). Following this, individual standard solutions were then combined equally into 65 pooled mixtures with either 12 compounds or 6 compounds with 6 equiv of 50% methanol so that all mixtures reached a final concentration per metabolite of 2.08 μg/mL. These mixtures were assigned such that they excluded isobaric compounds to minimize any possible cross-interference when determining retention characteristics. Retention times were then annotated manually based on the peak signal intensity maxima in extracted ion chromatograms (±10 ppm) for all observed [M + H]+ and [M – H]− ions with a signal intensity greater than 1e.[4] Pooled standard mixtures were run in both positive and negative ion modes, and all chemical data provided by the manufacturers (chemical name, elemental composition, mass characteristics, associated database entries), subpool mixture number assignments, and observed retention characteristics are recorded in Table S1. Summary information of the number of compounds detected in each ionization mode is also tabulated in Table S2 based on these data. For all quantitative work, a mixture of 85 commonly observed compounds was prepared in 50% methanol at varying concentrations (Table S3) and stored at −80 °C in 1 mL aliquots. Samples were then thawed on the day of use, and a 16-point standard curve was prepared by serial dilution (see Table S3 for concentrations). Each metabolite concentration was calibrated to ensure that the metabolites quantified were within the linear range of response factors for each target metabolite. These samples were used in the determination of lower limit of detection (LLOD) and lower limit of quantification (LLOQ) values in positive and negative ion modes, which were respectively defined as the lowest observed concentration in this standard curve with a signal ratio >3× and >10× greater than the signal observed in a 50% methanol blank injection. Compounds observed in these data that were not present in the MSMLS set are annotated in Table S1. Some experiments in this work used additional 8- or 6-point standard curves where appropriate and are annotated in Table S3.

Biological Samples

Clinical isolates of S. aureus from patients with bloodstream infections were collected from Calgary (Alberta, Canada) hospitals over a 27 month period by Alberta Precision Laboratories. 960 S. aureus isolates were identified following Clinical and Laboratory Standards Institute (CLSI) guidelines, and single colonies isolated from microbial cultures were stored as glycerol stocks. All activities were approved by the Conjoint Health Research Ethics Board (CHREB) of the University of Calgary under certificate REB17-1525. To achieve the uniform growth of selected S. aureus cultures, microbes were inoculated in 96-well plates filled with Mueller–Hinton broth originating from a single batch and were cultured overnight. Samples for all stages of growth were cultured aerobically at 37 °C in a humidified incubator with a controlled 5% CO2, 21% O2 atmosphere, without shaking. These were then subcultured and grown to an OD between 0.35 and 0.4, and culture supernatants were then fixed in methanol (1:1 v/v). Samples were centrifuged for 5 min at 4000g, and the cleared supernatants were retrieved and stored at −80 °C. On the date of LC-MS analysis, the stored supernatants were further diluted at 1:10 with 50% methanol. Genomic sequencing, as described previously,[19] was performed on these cultures to verify microbial species and to screen for cross-contamination.

Liquid Chromatography Parameters

Chromatographic separation was accomplished on two Syncronis ZIC-HILIC columns (inner diameter, 2.1 mm; length, 100 mm; particle size, 1.7 μm) using a UHPLC Vanquish Integrated biocompatible system (Thermo Fisher Scientific) equipped with an auxiliary vanquish UHPLC pump to perform the analytical gradient, while the main pump performs sample deposition and offline column re-equilibration. Mobile phases consisted of solvent A (20 mM ammonium formate, pH 3.0 in Optima LC-MS grade water; Thermo Fisher Scientific) and solvent B (Optima LC-MS grade acetonitrile with 0.1% formic acid; Thermo Fisher Scientific). Solvent A was constituted by the addition of ammonium formate salt followed by pH adjustment through titration with formic acid until a stable pH of 3.0 (±0.05) was reached overnight. For each experiment, all required solvents were produced in one large batch to minimize the possibility of solvent batch effects. All gradients were run at a flow rate of 0.6 mL/min with a sample injection volume of 2 μL using this binary solvent system. The column temperature was maintained in a sealed compartment at 30 °C (±0.005 °C). During LC-MS analysis, two chromatographic gradients were employed simultaneously using the auxiliary vanquish UHPLC pump. This setup also utilized two sets of multiport switch valves (MSVs) to direct alternating sample plugs to each column and manage the sample analyte elution into the mass spectrometer (Figure S1). For the auxiliary pump, the following gradient was used with respect to solvent B: 95% from 0 to 0.5 min; 95% to 5% from 0.5 to 3.5 min; and 5% from 3.5 to 4.5 min. Concurrently, the main pump gradient used was 95% from 0 to 0.5 mins; 95% to 5% from 0.5 to 0.75 min; 5% from 0.75 to 2 min; 5% to 95% from 2 to 2.25 min; and 100% from 2.25 to 4.5 min. The lower MSV controlled sample introduction to each column and switched at the 0.4 min mark to allow sample plug deposition on the column head before switching to the auxiliary pump to begin the analytical gradient. The upper MSV switched between columns at the beginning of each run to direct the flow from the column currently eluting an analytical gradient into the mass spectrometer and the equilibrating column to the waste.

Mass Spectrometry Parameters

Electrospray ionization was undertaken at a sheath gas flow rate of 35, an auxiliary gas flow rate of 15, and a sweep flow gas rate of 2 (all rates in arbitrary units), and all gas flows provided were high-purity N2 gas. A spray voltage of −2.5 kV was used in the negative ion mode and +3.0 kV was used in the positive ion mode. The capillary temperature was maintained at 275 °C, and the auxiliary gas heater was set at 300 °C. A Thermo Fisher Q Exactive HF Hybrid Quadrupole-Orbitrap mass spectrometer was used to acquire data in negative and positive ion full-scan modes, within a range of 50–750 m/z, at a resolution power of 120,000. The automatic gain control target was 1e,[6] and the maximum injection time was 200 ms.

LC-MS System Conditioning

To assess the long-term effects of chromatographic drift and column conditioning, the performance of a previously unused pair of chromatographic columns was monitored under standardized conditions over the course of 3192 injections. To track the performance across this process, two types of samples were repeatedly injected: a 6-point standard curve (Table S3) and a pooled S. aureus sample made from 160 culture supernatants mixed at an equal ratio. These samples were collected on a repeating schedule, wherein the six analytical standards were collected followed by 100 injections of the S. aureus pooled sample. Each sample was injected onto both columns, with injections continuously alternating between them until each column had reached a total of 1500 S. aureus samples (3192 total injections across the two columns, including standards).

Boundary Flux Analysis of S. aureus Clinical Isolates

Our methods for quantifying boundary fluxes have been described in detail elsewhere.[9] Briefly, metabolites present in microbial cultures were collected prior to and after incubating samples for 4 h in the Mueller–Hinton growth medium. Microbial culture media were diluted 1:20 (v/v final dilution) in 50% HPLC-grade methanol (Optima LC-MS grade methanol Optima LC-MS grade water; Thermo Fisher Scientific) and were analyzed by LC-MS. An 8-point analytical standard curve was collected interspersed between every 80 isolates, and metabolite concentrations were calculated based on the observed response factors for each molecule. The amount of each metabolite produced/consumed over this period was then calculated.

Peak Integration and Statistical Analysis

Extracted ion chromatograms for figures were generated in Thermo Scientific Xcalibur 4.0.27.19 software using a mass window of (±) 10 ppm. Peak assignment and quantification were performed using the open-source MAVEN software package.[20,21] Statistical analyses were performed using Python (Version 3) with seaborn and matplotlib packages and with an in-house R software package (https://zenodo.org/record/6403220#.Ykuh0mTF2_a). Retention factors (k) were calculated, as shown in eq , where tR is the observed retention time and t0 is the time required to clear the system void volume (0.42 min)Coefficient of variation (CV) values were calculated, as shown in eq , where σ is the standard deviation of LC-MS signals for a specific metabolite across the data set and μ is the mean LC-MS signal for the same metabolite across the data setMetabolite signals were calculated as z-scores, as shown in eq , where the signal for each metabolite (x) is shown relative to the mean signal for the same metabolite observed in Mueller–Hinton media samples (μ) divided by the standard deviation (σ) of the same metabolite observed across the Mueller–Hinton media samplesGraphically displaying z-scores of metabolite consumption/production is challenging because the data span several orders of magnitude and because z-scores include both positive and negative data. To display metabolomic boundary flux data, enrichment factors (EFs) were calculated for each metabolite signal. Enrichment factors were defined, as shown in eq , where c was set to 10. This transformation is adapted from the Rocke and Durbin variance stabilization algorithm[22]

Safety Considerations

Reagents used in this investigation do not pose any significant safety risks outside of those experienced regularly when working with moderate strength acids and solvents.

Results and Discussion

Multiplexed Chromatographic Approach

Metabolomics methods are not optimized for analyzing low-complexity media samples to generate quantitatively stable readouts over thousands of injections. To address this, we developed a sulfobetaine-based ZIC-HILIC chromatography method for assessing microbial boundary fluxes that separate the media components of interest that are commonly found in microbial cultures. The stationary phase we selected was previously identified as one of the better-performing chemistries for capturing a broad coverage of aqueous metabolites.[23,24] Using this stationary phase, we evaluated 40 combinations of solvent elution profiles, pH values, organic solvents (acetonitrile, methanol), and buffer salts (ammonium formate, ammonium acetate, ammonium bicarbonate) to find a method that optimizes run time while maintaining broad metabolite coverage. These refinements were made using a mixture of 85 metabolite standards that are commonly found in cell culture media (Table S3). The optimized conditions we present here (Figure S1) use a 4.5 min linear chromatographic gradient with acetonitrile, using 20 mM ammonium formate at pH 3 (see the Liquid Chromatography Parameters section for details). Figure shows 23 representative extracted ion chromatograms from our pooled standard set, showing that the metabolites resolved using this method have an average peak width of ∼2.4 s (Table S5) and are evenly distributed across the elution profile.
Figure 1

Chromatographic performance of a broad transect of metabolites resolved by our multiplexed ZIC-HILIC method observed in negative ion mode. The main panel displays stacked extracted ion chromatograms of 23 representative standards selected from the MSMLS library covering a diversity of metabolite chemistries of biological relevance, with optimal peak shapes. Observed signal intensities are normalized across the set for visual comparison without the interference of differential ionization effects. The inset displays the distribution of retention factors (k) of all 313 compounds detected in a negative ion mode from the MSMLS library. The solid line indicates a theoretical perfect spacing of compounds across the run between the void volume and the last observed standard retention time.

Chromatographic performance of a broad transect of metabolites resolved by our multiplexed ZIC-HILIC method observed in negative ion mode. The main panel displays stacked extracted ion chromatograms of 23 representative standards selected from the MSMLS library covering a diversity of metabolite chemistries of biological relevance, with optimal peak shapes. Observed signal intensities are normalized across the set for visual comparison without the interference of differential ionization effects. The inset displays the distribution of retention factors (k) of all 313 compounds detected in a negative ion mode from the MSMLS library. The solid line indicates a theoretical perfect spacing of compounds across the run between the void volume and the last observed standard retention time. To assess the qualitative performance of this method, we analyzed a commercially available MSMLS library of metabolite reference standards containing 639 compounds (Table S1). These standards have representatives from a broad transect of polar metabolites including amino acids, nucleic acids, organic acids, energy metabolites, vitamins, and compounds important to cellular redox control. From this library, 397 compounds were ionizable and chromatographically resolvable. The chromatographic separation of these diverse analytes was close to the theoretical optimum peak spacing (Figure , inset). One of the primary disadvantages of HILIC is that it frequently requires long re-equilibration times between injections.[25] To address this, we plumbed two HILIC columns into an LC-MS system using a secondary HPLC pump (Figure S1). This arrangement allowed columns to be reconditioned offline and thereby compresses the overall analytical run times without sacrificing time on column re-equilibration. Using this strategy, we were able to reduce the total sample-to-sample cycle time to 4.5 min. Another factor affecting the length of analytical analyses is that some metabolites do not ionize in both positive and negative modes. This problem can be addressed by collecting data in consecutive injections using the two modes but this strategy has the obvious disadvantage of doubling the analytical run time. Alternatively, some LC-MS platforms support polarity switching during a single run.[26] Unfortunately, the scanning rate on the instrument used in this study (Thermo Fisher Orbitrap, QE-HF) was not fast enough to adequately sample the narrow elution peaks we observed when using these polarity switching methods. However, our analysis of the MSMLS library indicated that 313 of the 391 detectable compounds (80%) could be analyzed in negative mode (Table S2). Given that the goal of our method was to maximize sample throughput rather than increase the breadth of metabolite coverage, we elected to only collect negative mode data.

LC-MS Conditioning for Large Cohorts

To assess the quantitative stability of our method in large cohorts, we injected 1500 pooled S. aureus samples alternately (3000 injections in total) on two previously unused columns. A 6-point pooled analytical standard curve was collected for every 100 samples on each column (192 injections in total). Variations in the signals from 55 metabolites present in the S. aureus samples and 77 metabolites present in the standard mixture were then tracked across all 3192 injections (Figure a) as well as retention time stability (Table S7). The overall performance of our method, as judged by the run-to-run stability and progressively smaller technical errors, stabilized after ∼400 injections with incrementally smaller improvements observed thereafter (Figure a). After 378 injections, the median CV for all metabolites was 0.15 (median CV across injections 0–378 = 0.171; median CV across injections 379–1500 = 0.127). Given this behavior, we recommend 400 replicate injections of biological samples to fully stabilize the LC-MS system prior to the analysis of large cohorts using this method.
Figure 2

Quantitative stability of metabolites from S. aureus cultures over 1500 injections. Variations in signal intensities were monitored for 55 metabolites using our multiplexed ZIC-HILIC MS analytical strategy. For this study, two new Syncronis ZIC-HILIC columns were used. Performance was monitored over repeated injections with pooled S. aureus extracellular extracts (n = 3000). Blue and orange points differentiate data originating from each of the two columns. (a) To illustrate the most variable signals for each injection, the z-score for each metabolite was computed (relative to the median signal and standard deviation observed across all injections) and the upper quartile of z-scores observed across all metabolites in each injection was computed. These data were then plotted for each column independently as absolute values and overlayed. The median z-scores observed across metabolites were then fitted to a decaying exponential (dashed line). The point where the CV values crossed the 0.15 point is noted (378). The inset shows a histogram of CV values observed across all detected metabolites from 1 to 378 sample injections and from 379 to 1500 sample injections. (b) Compound-to-compound difference and column-to-column differences were observed in the instrument response factors over the conditioning period. A representative selection of these differences is illustrated (shown as z-scores for each individual metabolite). The patterns for all 55 metabolites are provided in Figure S2.

Quantitative stability of metabolites from S. aureus cultures over 1500 injections. Variations in signal intensities were monitored for 55 metabolites using our multiplexed ZIC-HILIC MS analytical strategy. For this study, two new Syncronis ZIC-HILIC columns were used. Performance was monitored over repeated injections with pooled S. aureus extracellular extracts (n = 3000). Blue and orange points differentiate data originating from each of the two columns. (a) To illustrate the most variable signals for each injection, the z-score for each metabolite was computed (relative to the median signal and standard deviation observed across all injections) and the upper quartile of z-scores observed across all metabolites in each injection was computed. These data were then plotted for each column independently as absolute values and overlayed. The median z-scores observed across metabolites were then fitted to a decaying exponential (dashed line). The point where the CV values crossed the 0.15 point is noted (378). The inset shows a histogram of CV values observed across all detected metabolites from 1 to 378 sample injections and from 379 to 1500 sample injections. (b) Compound-to-compound difference and column-to-column differences were observed in the instrument response factors over the conditioning period. A representative selection of these differences is illustrated (shown as z-scores for each individual metabolite). The patterns for all 55 metabolites are provided in Figure S2. Importantly, the conditioning pattern observed for each compound was not uniform: significant compound-to-compound differences were observed as well as some column-to-column to differences (Figures b and S2). For example, phosphoserine and other phosphate-containing compounds showed a rapid increase in response factors over the first 400 injections, whereas acetyl-threonine and several other acetylated amino acids showed a decrease in response factors over the same period. Interestingly, several amino acids (asparagine, serine, glutamine, histidine, and lysine) showed inverse patterns between the two columns: compounds eluting from one column progressively gained signal intensity, whereas the same compounds eluting from the other column progressively lost signal intensity. However, after the recommended 400-sample conditioning period, we did not observe any significant compound-to-compound changes in response factors: all signals followed a progressive loss of signal that is consistent with the slow accumulation of insoluble materials (salts, lipids, proteins, carbohydrates, etc.) on the column, source, and ion optics. Collectively, these data show that 400 injections are sufficient to stabilize CVs, minimize compound-to-compound differences in the behavior of response factors over time, and standardize the behavior of metabolites across the two chromatographic columns. After this conditioning period, we showed that response factors are stable over more than 1100 injections (median CV = 0.127; Table S6).

Sensitivity of Multiplexed ZIC-HILIC MS

To assess the quantitative performance of multiplexed ZIC-HILIC MS, we prepared a 16-point standard curve (n = 6) using a mixture of 85 compounds. We then determined the LLOD and LLOQ for each compound (Figures and S3 and Table S4). The median LLOD across these compounds was 24.2 nM in negative mode and 61.0 nM in positive mode ionization. Median LLOQs were 39.1 and 98.6 nM in negative mode and positive mode, respectively. Given that the primary goal of this method is to enable the quantification of abundant media compounds, which are generally present at micromolar to low millimolar concentrations, these sensitivities are more than adequate for quantifying microbial boundary fluxes.
Figure 3

Sensitivity limits for our multiplexed ZIC-HILIC MS method. Median LLOD and LLOQ levels for 77 metabolites were computed from a 16-point pooled standard curve (n = 6) in negative ion mode. LLODs and LLOQs were defined conservatively as the lowest concentration metabolite standard whose signal was empirically observed to be >3 times and >10 times the noise threshold, respectively. Similar data were observed in positive mode ionization (Figure S3); all LLOD/LLOQs are provided in Table S4.

Sensitivity limits for our multiplexed ZIC-HILIC MS method. Median LLOD and LLOQ levels for 77 metabolites were computed from a 16-point pooled standard curve (n = 6) in negative ion mode. LLODs and LLOQs were defined conservatively as the lowest concentration metabolite standard whose signal was empirically observed to be >3 times and >10 times the noise threshold, respectively. Similar data were observed in positive mode ionization (Figure S3); all LLOD/LLOQs are provided in Table S4.

Analyzing Metabolic Boundary Fluxes by Multiplexed ZIC-HILIC

S. aureus is a common cause of bloodstream infections,[27,28] and changes in its metabolism have been linked to virulence (e.g., via the arginine catabolic mobile element).[29] Thus, the metabolic profiling of S. aureus could potentially be harnessed as a tool for clinical diagnostics.[9] Currently, most S. aureus metabolomics studies have been conducted using laboratory strains[30] and relatively small collections of clinical isolates.[31] At present, it is unclear if the metabolic phenotypes described in these studies represent the full spectrum of metabolic diversity found in populations of S. aureus. To better understand the metabolic diversity of S. aureus populations and to illustrate the utility of multiplexed ZIC-HILIC MS, we analyzed the metabolic boundary fluxes of 960 clinical isolates of S. aureus collected from bloodstream infections (27 months of infections in the Calgary area). Clinical isolates were grown in 12 batches of 80 samples, and each batch was analyzed alongside an 8-point calibration curve of standards. Overall, we observed consistent metabolic phenotypes across the cohort (Figure ), as judged by signal intensities from the mixtures of standards (median CV = 0.125; Table S8).
Figure 4

Metabolic phenotypes observed via multiplexed ZIC-HILIC. Metabolite signals observed across a cohort of 960 S. aureus culture extracts originating from clinical isolates. Arginine, glucose, and succinate were chosen as biologically significant metabolites that are relevant to microbial phenotypes. The variability in arginine shown here is clinically relevant given that the arginine catabolic mobile element (ACME) is linked to virulence.[29,32]

Metabolic phenotypes observed via multiplexed ZIC-HILIC. Metabolite signals observed across a cohort of 960 S. aureus culture extracts originating from clinical isolates. Arginine, glucose, and succinate were chosen as biologically significant metabolites that are relevant to microbial phenotypes. The variability in arginine shown here is clinically relevant given that the arginine catabolic mobile element (ACME) is linked to virulence.[29,32] Our boundary flux analysis showed that S. aureus cultures preferentially consume glucose and nucleic acid-related compounds (uridine, xanthine, adenine, and adenosine) and secrete a broad range of metabolic products (mevalonate, citrulline, succinate, N-acetyl-methionine, thymine, orotate, and urocanate; Figure ). These phenotypes were surprisingly homogeneous across the cohort, with only a few molecules (e.g., citrulline) showing evidence for strain-to-strain variability. Collectively, this phenotype is consistent with a metabolic model wherein S. aureus relies on glycolysis for carbon and energy needs and uses nucleotide salvage pathways when the appropriate precursors are present in the environment. The highly stable nature of the phenotypes we observed is somewhat surprising given the prevalence of auxotrophy that has been reported previously in the literature.[33,34] However, our use of a rich microbial growth medium may provide sufficient nutritional diversity, allowing each strain to satisfy its basal biomass and energetic needs with minimal disruption in the overall flow of molecules into and out of the cells.
Figure 5

Metabolomic boundary flux of S. aureus clinical isolates. LC-MS analysis of 77 metabolite levels observed in in vitro cultures of 960 S. aureus isolates from a 27 month collection period of bloodstream infections in Calgary. Isolates were grown under standardized conditions in Mueller–Hinton broth. Orange/blue points indicate metabolite values observed in individual isolates that were run on each of the two analytical columns. Dashed lines denote the two standard deviation boundaries of metabolite signals observed in the media control. Enrichment factors are a log-transformed z-score (unitless) calculated according to eq in the Experimental Section. This transformation enables the data, which span many orders of magnitude in the z-score space, to be visualized. Raw data are provided in Table S9 prior to transformation.

Metabolomic boundary flux of S. aureus clinical isolates. LC-MS analysis of 77 metabolite levels observed in in vitro cultures of 960 S. aureus isolates from a 27 month collection period of bloodstream infections in Calgary. Isolates were grown under standardized conditions in Mueller–Hinton broth. Orange/blue points indicate metabolite values observed in individual isolates that were run on each of the two analytical columns. Dashed lines denote the two standard deviation boundaries of metabolite signals observed in the media control. Enrichment factors are a log-transformed z-score (unitless) calculated according to eq in the Experimental Section. This transformation enables the data, which span many orders of magnitude in the z-score space, to be visualized. Raw data are provided in Table S9 prior to transformation. Interestingly, three of the metabolites that do show evidence for metabolic diversity across the clinical cohort are arginine and two of its catabolites, ornithine and citrulline. Changes in the boundary fluxes of these compounds are consistent with the presence/absence of the arginine catabolic mobile element (ACME),[32] a virulence-associated trait that is selected in tandem with the methicillin resistance encoding SCCmec operon in the USA300 clone[29] that has been reported in the Calgary population.[35] Detecting these phenotypes in clinical cohorts highlights one potential application of the analytical strategy introduced here. The method we introduce here can be applied to a range of applications where the primary goal is to understand the rates at which a discrete set of microbial metabolites are excreted to or consumed from the culture medium. These applications include (1) constructing/refining flux balance models,[4,5] (2) screening large libraries of isolates for auxotrophy or other metabolic variabilities,[33,34] (3) as a tool to accelerate bioengineering projects, particularly those that are seeking to use saturation mutagenesis and other high-throughput methods to maximize metabolic outputs in microbial systems,[7,8] and (4) as a strategy for rapidly screening inhibitor libraries to identify new antimicrobial lead compounds. Moreover, we have recently shown that microbial boundary fluxes can be used as a diagnostic readout for differentiating microbial pathogens and measuring their antibiotic susceptibility profiles.[9] In summary, the method we present here has broad applicability to any large-cohort study of in vitro culture systems where the primary objective is to quantify the rates at which metabolites are consumed or secreted to the culture medium.

Limitations in the Multiplexed ZIC-HILIC MS Strategy

Although the methods introduced here enable the quantification of boundary fluxes in large cohorts, they have some inherent limitations that should be considered in the context of each study. One important consideration is that this method relies on a custom hardware configuration involving two HPLC pumps and two switch ports, a setup that is not standard on most LC-MS instruments. In addition, we show that conditioning the LC-MS using 400 injections of biological samples is needed to stabilize the performance of the method. This conditioning period would be disruptive if conducted on the main LC-MS platform; however, conditioning can be completed offline on a separate HPLC. Beyond these hardware considerations, the most significant limitation of the method is its relatively limited sensitivity. Modern metabolomics methods are pushing ever further into secondary metabolism with higher sensitivity methods. The method introduced here is not suited to these in-depth applications and will not cover the chemical diversity, the low abundance molecules, and the isomer resolution that is possible with longer methods. One specific shortcoming we have identified with our multiplexed ZIC-HILIC MS strategy is that pyruvate, lactate, and short-chain fatty acids are not detected—all of which are important for the comprehensive mapping of metabolic fluxes. Despite these limitations, we have shown that the method can be applied to large microbial cohorts and introduces minimal quantitative variability across thousands of samples.

Conclusions

Metabolomics has emerged as a mainstream strategy for understanding the molecular underpinnings of life. Modern LC-MS metabolomics methods can capture over a thousand molecules and a broad transect of central carbon metabolism. Despite this, there are relatively few methods that have been optimized for large-cohort, boundary flux studies involving thousands of samples. The multiplexed ZIC-HILIC MS methods introduced here address this gap and enable consistent quantification of microbial metabolic boundary fluxes. We show how this strategy can be used to map metabolites flowing into and out of cells and stably quantify these phenotypes across thousands of clinical samples. This study provides a template for other large-cohort metabolomics studies and provides a comprehensive set of reference data for implementing this method in other laboratories.
  33 in total

Review 1.  Metabolic engineering of microorganisms for biofuels production: from bugs to synthetic biology to fuels.

Authors:  Sung Kuk Lee; Howard Chou; Timothy S Ham; Taek Soon Lee; Jay D Keasling
Journal:  Curr Opin Biotechnol       Date:  2008-11-10       Impact factor: 9.740

2.  Reviewing the metabolome coverage provided by LC-MS: Focus on sample preparation and chromatography-A tutorial.

Authors:  Marta Roca; Maria Isabel Alcoriza; Juan Carlos Garcia-Cañaveras; Agustín Lahoz
Journal:  Anal Chim Acta       Date:  2020-12-23       Impact factor: 6.558

Review 3.  Metabolite Measurement: Pitfalls to Avoid and Practices to Follow.

Authors:  Wenyun Lu; Xiaoyang Su; Matthias S Klein; Ian A Lewis; Oliver Fiehn; Joshua D Rabinowitz
Journal:  Annu Rev Biochem       Date:  2017-06-20       Impact factor: 23.643

4.  Staphylococcus aureus physiological growth limitations: insights from flux calculations built on proteomics and external metabolite data.

Authors:  Chunguang Liang; Manuel Liebeke; Roland Schwarz; Daniela Zühlke; Stephan Fuchs; Leonhard Menschner; Susanne Engelmann; Christiane Wolz; Sarah Jaglitz; Jörg Bernhardt; Michael Hecker; Michael Lalk; Thomas Dandekar
Journal:  Proteomics       Date:  2011-04-07       Impact factor: 3.984

5.  A positive/negative ion-switching, targeted mass spectrometry-based metabolomics platform for bodily fluids, cells, and fresh and fixed tissue.

Authors:  Min Yuan; Susanne B Breitkopf; Xuemei Yang; John M Asara
Journal:  Nat Protoc       Date:  2012-04-12       Impact factor: 13.491

Review 6.  Staphylococcus aureus infections: epidemiology, pathophysiology, clinical manifestations, and management.

Authors:  Steven Y C Tong; Joshua S Davis; Emily Eichenberger; Thomas L Holland; Vance G Fowler
Journal:  Clin Microbiol Rev       Date:  2015-07       Impact factor: 26.132

7.  The arginine catabolic mobile element and staphylococcal chromosomal cassette mec linkage: convergence of virulence and resistance in the USA300 clone of methicillin-resistant Staphylococcus aureus.

Authors:  Binh An Diep; Gregory G Stone; Li Basuino; Christopher J Graber; Alita Miller; Shelley-Ann des Etages; Alison Jones; Amy M Palazzolo-Ballance; Françoise Perdreau-Remington; George F Sensabaugh; Frank R DeLeo; Henry F Chambers
Journal:  J Infect Dis       Date:  2008-06-01       Impact factor: 5.226

8.  Staphylococcus aureus small colony variants show common metabolic features in central metabolism irrespective of the underlying auxotrophism.

Authors:  André Kriegeskorte; Stephanie Grubmüller; Claudia Huber; Barbara C Kahl; Christof von Eiff; Richard A Proctor; Georg Peters; Wolfgang Eisenreich; Karsten Becker
Journal:  Front Cell Infect Microbiol       Date:  2014-10-21       Impact factor: 5.293

9.  Quantitative Fluxomics of Circulating Metabolites.

Authors:  Sheng Hui; Alexis J Cowan; Xianfeng Zeng; Lifeng Yang; Tara TeSlaa; Xiaoxuan Li; Caroline Bartman; Zhaoyue Zhang; Cholsoon Jang; Lin Wang; Wenyun Lu; Jennifer Rojas; Joseph Baur; Joshua D Rabinowitz
Journal:  Cell Metab       Date:  2020-08-12       Impact factor: 31.373

Review 10.  Biofuel production: an odyssey from metabolic engineering to fermentation scale-up.

Authors:  Whitney Hollinshead; Lian He; Yinjie J Tang
Journal:  Front Microbiol       Date:  2014-07-09       Impact factor: 5.640

View more
  3 in total

Review 1.  Moving beyond descriptive studies: harnessing metabolomics to elucidate the molecular mechanisms underpinning host-microbiome phenotypes.

Authors:  Stephanie L Bishop; Marija Drikic; Soren Wacker; Yuan Yao Chen; Anita L Kozyrskyj; Ian A Lewis
Journal:  Mucosal Immunol       Date:  2022-08-15       Impact factor: 8.701

2.  Biomarker enrichment medium: A defined medium for metabolomic analysis of microbial pathogens.

Authors:  Maryam Mapar; Thomas Rydzak; Ryan A Groves; Ian A Lewis
Journal:  Front Microbiol       Date:  2022-07-22       Impact factor: 6.064

3.  Microbial containment device: A platform for comprehensive analysis of microbial metabolism without sample preparation.

Authors:  Mehdi Mohammadi; Stephanie L Bishop; Raied Aburashed; Saad Luqman; Ryan A Groves; Dominique G Bihan; Thomas Rydzak; Ian A Lewis
Journal:  Front Microbiol       Date:  2022-09-13       Impact factor: 6.064

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.