Transportation and Centering Ability of Kedo-S Pediatric and Mtwo Instruments in Primary Teeth: A Cone-beam ComputedTomography Study.

Selvakumar Haridoss1, Keerthana M Rakkesh2, Kavitha Swaminathan1.   

Abstract

Background: Cleaning and debriding the canals and preserving the shape of the canal without deformation is the primary goals of pulpectomy. Transportation is a critical endodontic iatrogenic fault that could cause a catastrophe. This study evaluated the canal centering ability and canal transportation caused by Kedo-S pediatric and Mtwo instruments, using a cone-beam computed tomography (CBCT). Materials and methods: This in vitro study was performed on distal roots of 50 primary mandibular first molars. The teeth were scanned using CBCT and randomly divided into two groups. The canals were then prepared using either Kedo-S or Mtwo files (n = 25). The instrumented canals were rescanned. The scanned volumes were sectioned at 2, 4, and 6 mm from cementoenamel junction (CEJ). Canal transportation (CT) and instrument centering ability were estimated and compared in both groups.
Results: The mean values for two study groups were compared. T -test was used to determine theP value. The Levene's test was used to test the significance between two groups. The two groups showed similar results in terms of transportation and centering ability (P > 0.05).
Conclusion: Kedo-S pediatric and Mtwo instruments demonstrated similar canal centering ability and CTs. How to cite this article: Haridoss S, Rakkesh KM, Swaminathan K. Transportation and Centering Ability of Kedo-S Pediatric and Mtwo Instruments in Primary Teeth: A Cone-beam Computed Tomography Study. Int J Clin Pediatr Dent 2022;15(S-1):S30-S34.
Copyright © 2022; The Author(s).

Entities:  

Keywords:  Canal centering ability; Canal transportation; Kedo-S instrument; Pediatric rotary files; Primary teeth; Root canal preparation

Year:  2022        PMID: 35645489      PMCID: PMC9108830          DOI: 10.5005/jp-journals-10005-2127

Source DB:  PubMed          Journal:  Int J Clin Pediatr Dent        ISSN: 0974-7052


Introduction

Primary teeth with extensive caries involving pulp are indicated for a pulpectomy procedure. Traditionally pulpectomy was performed by stainless steel hand files.[1] Several NiTi rotary files were later introduced with improved shaping abilities to reduce the risk of technical errors.[2] The Mtwo system which consists of four instruments with variable tip sizes. It should be used without early coronal enlargement, suggested by the manufacturer, in single-length techniques.[3-5] To perform an effective pulpectomy in primary teeth, it is mandatory to use rotary files with customized length, taper, and tip size that is suitable for root canal morphology of primary teeth.[6] The Kedo-S rotary file (pediatric rotary system) is a file system with a length of 16 mm and the working area (cutting flutes) being 12 mm. These files have variable varying taper (4-8%).[7] Cone-beam computed tomography is widely used for evaluating improvements in the root canal anatomy and visualization of maxillofacial hard tissues regions. Cone-beam computed tomography images provide submillimeter resolution with 15 times smaller than traditional scans with high quality and quick scanning time (10-70 seconds).[8] Based on the literature search there is a lack of in vitro studies comparing Kedo-S with other NiTi rotary files. Evidence shows that Mtwo file maintains centering ability and shows better removal of debris from even in curved canals compared to other rotary systems.[9,10] Hence this study is intended to evaluate the CT and centering ability of Kedo-S and Mtwo files using CBCT.

Materials and Methods

Selection and Preparation of Samples

Fifty extracted human mandibular primary first molars with minimum of 7 mm root length were collected and stored in formalin. The study was conducted in the Department of Pediatric and Preventive Dentistry, Sri Ramachandra Faculty of Dental Sciences, Sri Ramachandra Institute of Higher Education and Research with ethical approval from the Institutional Ethical Committee (CSP/17/JUN/59/217). Occupational Safety and Health Administration guidelines and regulations were used for storage and handling of teeth. All the sample teeth had their crowns cut off horizontally at the level of the CEJ using sectioning disk. The sectioned roots were mounted in a wax block of a standardized dimension. Distal roots of the primary mandibular molar were selected as it is rounded, shorter, and taper more apically.[11] Standard access cavities were made for all teeth with a diamond bur (BR 40, MANI, INC, Japan). Preoperative radiographs were taken and canal patency was assessed using 15 root canal instrument. All the samples teeth were then divided into two groups containing 25 teeth each (Flowchart 1).
Flowchart 1

Diagrammatic representation of study design

Diagrammatic representation of study design

Canal Preparation

In group I, E1 Kedo-S files with X Smart Endodontic motor (Dentsply Maillefer, Switzerland) were used to prepare the canal with 300 rpm speed and torque of 2.2 N cm. Mtwo Basic Sequence NiTi rotary files (VDW, Munich, Germany) driven by an X Smart Endodontic motor (Dentsply Maillefer, Switzerland) at a speed of 300 rpm and a torque of 1.2 N cm were used for canal preparation in group II. The canals were prepared for the full length by single length technique without early coronal enlargement. Three Mtwo Basic sequence instruments (no.10 size to no. 20 size) were used in primary teeth. Canals were irrigated with 3 mL of a 5.25% NaOCl solution (27-gauge needle). Glyde (Dentsply, Maillefer) was used for lubrication during instrumentation and after instrumentation; each instrument was changed after five canals.

Specimen Scan

Teeth were scanned before and after canal preparation with CBCT (Planmeca ProMax® 3D Mid) with a field of view = 80 mm × 80 mm and voxel size of 200 mm. The exposure time of 12 seconds and slice thickness of 0.4 mm with 90 kV and 10 mA was used. Romexis® digital imaging software, version 3.5.2 (Planmeca, Helsinki, Finland) were used to assess the CBCT images. The CEJ was taken as a reference point. The canal preparation was measured at three levels. The cervical level was assessed at 2 mm below the CEJ. The middle level was assessed at 4 mm below CEJ. The apical level was assessed at 6 mm below CEJ.

Canal Transportation

Voxel measurements were used to quantify noninstrumented and instrumented canals, while M1 was used to calculate the number of voxels at the mesial wall of the noninstrumented canal from the outer surface of the mesial portion of the root. M2 was the calculation, after instrumentation, of the number of voxels from the outer root surface of the mesial part of the root to the canal wall. The calculation of the number of voxels from the outer surface of the distal root portion to the distal wall of the noninstrumented canal was D1. D2 was the calculation after instrumentation of the number of voxels from the external surface of the distal portion of the root to the distal surface of the canal (Figs. 1 and 2).
Fig. 1

A schematic diagram, showing the cross section of non-instrumented canal (blue small ovoid) and instrumented canal (red circle)

Fig. 2

Representative CBCT scanning images of Kedo-S and Mtwo rotary systems

A schematic diagram, showing the cross section of non-instrumented canal (blue small ovoid) and instrumented canal (red circle) Representative CBCT scanning images of Kedo-S and Mtwo rotary systems From the following equation, CT was assessed (CT) = (M 1 -M 2) -(D 1 -D2) CT equal to 0 (zero) meant lack of transport as regards the direction of canal transport; negative value indicates transportation to the distal trend and transportation to the mesial trend is indicated by a positive value.[12]

Canal Centering Ability

The following equation was used to determine the canal centering ability centralization ability ratio = (M 1 − M 2)/(D 1 − D 2) A result equal to 1.0 demonstrated complete centralization ability ratio. This meant that the instrument had a lesser capacity to sustain itself in the central axis of the canal while this value was closer to zero.[12]

Statistical Analysis

The mean and standard deviation for canal transportation and canal centering ability were estimated and T -test was used to calculate theP value. Levene's test was used to calculate the level of significance and was set at 0.05.

Results

In this study, canal transportation and centering ability was examined at 2, 4, and 6 mm from CEJ. The mean CT for the mentioned diameters is listed in Table 1 and Figure 3. There was no considerable difference between the two systems, based on statistical evidence, in terms of canal transportation.
Table 1

Mean and standard deviation for canal transportation

Groupn Cervical level (2 mm below CEJ)mean ± SD Middle level(4 mm below CEJ)mean ± SD Apical level(6 mm below CEJ)mean ± SD
Group I n = 25−0.008 ± 0.18−0.031 ± 0.02−0.018 ± 0.28
Group II n = 250.026 ± 0.93−0.016 ± 0.09−0.09 ± 0.94
Figs 3A and B

Distribution of the amount of displacement in the two rotary systems: (A) Kedos-S; (B) Mtwo

Mean and standard deviation for canal transportation Distribution of the amount of displacement in the two rotary systems: (A) Kedos-S; (B) Mtwo The frequency of the direction of transportation is shown in Table 2. In general, transportation to the distal area in both systems was higher than transportation to the mesial area, although it was not a statistically significant difference.
Table 2

Frequency of direction of transportation

Distance from CEJ System Transportation toward distal (%) No transportation (%) Transportation toward Mesial (%) P value
Cervical2 mmKedo-S12 (48)9 (36)4(16)0.489
Mtwo7(28)14(56)4(16)
Middle4 mmKedo -S12(48)9(36)4(16)0.746
Mtwo11(44)12(48)2(8)
Apical4 mmKedo-S11(44)10(40)4(16)0.882
Mtwo11(44)13(52)1(4)

*P value = Significant P value, †P value < 0.05 = Statistically significant.

Frequency of direction of transportation *P value = Significant P value, †P value < 0.05 = Statistically significant. The Kedo-S and Mtwo rotary files maintain the canal centering ability better in the middle level when compared to the cervical and apical levels. Both the files were not statistically significant (P < 0.05) in maintaining the canal centering ability (Table 3).
Table 3

Canal centering ability of files at cervical, middle, and apical level of primary molars root canals

Levels Mean ± SD Kedo-S Mean ± SD Mtwo Mean difference Mtwo File vs. Kedo -S file P value 95% Confidence interval of the difference
Lower Upper
Cervical 2 mm below CEJ0.82 ± 2.120.26 ± 3.710.5620.562−1.202.32
Middle 4 mm below CEJ0.72 ± 1.840.93 ± 1.09−0.2070.632−1.070.66
Apical 6 mm below CEJ1.3 ± 2.162.45 ± 5.58−1.1860.338−3.681.30

*P value = Significant P value, †P value < 0.05 = Statistically significant.

Canal centering ability of files at cervical, middle, and apical level of primary molars root canals *P value = Significant P value, †P value < 0.05 = Statistically significant.

Discussion

The basic principles for biomechanical preparation in endodontic treatment are complete removal of vital tissue, necrotic remnants, debris, and infected dentin thereby render tissue repair and regeneration.[13] The root canal geometry is different in primary teeth and hence it is important to assess the canal preparation using different instruments.[14] The strength of the nickel-titanium (Ni-Ti) rotary systems was to uniformly, smoothly prepare the curved canals and maintain the shape with less instrument time and canal tapering than hand instruments.[15] In the present study, teeth with a least amount 7 mm of root length were selected to simulate a clinical conditions. As it offers accurate three-dimensional (3D) observation, CBCT imaging has been used in measuring dentin thickness removal, canal curvature, transportation, and canal cantering ratio.[16] Therefore, the objective of this research was to CBCT evaluation of the transportation and centering ability with rotary files of Kedo-S and Mtwo. In our study, the Mtwo rotary system shows no transportation at 2, 4, and 6, mm. This finding was similar to the results of previous studies that evaluated the preparation of curved canals by using Mtwo files and other NiTi rotary files. It was reported that the Mtwo files conserved canal curvatures better than the K3, race,[3,4] and Protaper instruments.[17] Owing to the design of Mtwo files, fewer preparation errors have been reported.[18] However, no significant difference was noted in this respect between the two systems. Both systems in all sections mostly recorded < 0.1 mm of canal transportation, which is within the clinically acceptable range given by Peters.[19] In this study, both the rotary file systems maintain the canal centering ability better in the middle level are in agreement with Selvakumar et al. who observed K3 rotary file (2 and 4% taper) maintain the centering ability better than stainless steel.[20]On the contrary, Gambil et al. have also found no significant difference between NiTi and K- flex instruments.[12] The risk factors for canal transportation and centering ability are complex radicular anatomy, the lack of direct access, instrument design, incorrect sequences of the usage of different instruments, speed of instrument rotation, the use of inadequate irrigation, and lubricant solutions.[21] In these risk factors, the instrument design and internal canal morphology are intrinsic factors that are independent of the operator's expertise and skill. Among these two factors, the instrument design can be modified. In our study, the Kedo-S file shows more distal displacement even though it is not statistically significant by modifying the shape of the instrument might reduce the transportation of the file. The impediment of this investigation was smarter to compare the outcomes with a conventional stainless steel file. Notwithstanding, transportation has not conceded to the highest quality level.[22] Since the lower transportation of NiTi files compared to that of stainless steel hand instruments is already established,[23] we focused on recently presented newly introduced pediatric rotary with existing NiTi instruments as they were. Another limitation was that regardless of our endeavor to standardize the samples using the exclusion/inclusion criteria, extracted teeth cannot be completely standardized in terms of canal shapes and sizes.[24,25] The quality purposes of our investigation are canal preparation was performed in natural teeth consequently, its outcomes could be better generalizable to the clinical practice. The procedures were performed by one operator (high reproducibility of results), utilizing software calculations (high precision), and utilization of CBCT.

Conclusion

Within the restrictions of this study, no difference was noted in canal transportation and centering ability of the rotary files used in this study. Thus, both systems can be used with minimum endangerment of procedural errors in root canal preparation. In terms of canal transportation and centering ability, Kedo-S file would be safer in primary teeth. However, further investigations are needed to evaluate the performance of Kedo-S pediatric files in the uneven walls of primary teeth with larger sample size.

Clinical Significance

The Kedo-S pediatric rotary file showed clinically appropriate transportation and centering ability that can be recognized as major innovations in pediatric endodontics.
  22 in total

1.  Use of nickel-titanium rotary files for root canal preparation in primary teeth.

Authors:  E S Barr; D J Kleier; N V Barr
Journal:  Pediatr Dent       Date:  1999 Nov-Dec       Impact factor: 1.874

2.  Comparative study on the shaping ability and cleaning efficiency of rotary Mtwo instruments. Part 1. Shaping ability in simulated curved canals.

Authors:  E Schäfer; M Erler; T Dammaschke
Journal:  Int Endod J       Date:  2006-03       Impact factor: 5.264

3.  Root canal preparation with the NiTi systems K3, Mtwo and ProTaper.

Authors:  David Sonntag; Mareike Ott; Kathrin Kook; Vitus Stachniss
Journal:  Aust Endod J       Date:  2007-08       Impact factor: 1.659

4.  A comparison of apical transportation and length control between EndoSequence and Guidance rotary instruments.

Authors:  Bekir Karabucak; Adam Joseph Gatan; Chinchai Hsiao; Mian Khalid Iqbal
Journal:  J Endod       Date:  2010-01       Impact factor: 4.171

Review 5.  Cone beam computed tomography updated technology for endodontic diagnosis.

Authors:  Randolph Todd
Journal:  Dent Clin North Am       Date:  2014-07

6.  Evaluation of canal transportation and centering ability of K 3 (0.02%) and K 3 (0.04%) with hand K files in primary teeth using spiral computed tomography.

Authors:  Haridoss Selvakumar; Vasanthakumari Anandhan; Eapen Thomas; Kavitha Swaminathan; Rajendran Vijayakumar
Journal:  J Indian Soc Pedod Prev Dent       Date:  2014 Oct-Dec

7.  Mechanical Preparation Showed Superior Shaping Ability than Manual Technique in Primary Molars - A Micro-Computed Tomography Study.

Authors:  Lídia Regina da Costa Hidalgo; Léa Assed Bezerra da Silva; Graziela Bianchi Leoni; Jardel Francisco Mazzi-Chaves; Emily Eduardo da Silva Carvalho; Alberto Consolaro; Manoel Damião Sousa-Neto
Journal:  Braz Dent J       Date:  2017 Jul-Aug

8.  Radiographic observations on root resorption in the primary dentition.

Authors:  M Daito; S Kawahara; M Kato; K Okamoto; G Imai; T Hieda
Journal:  J Osaka Dent Univ       Date:  1991-04

9.  SEM evaluation of canal wall dentine following use of Mtwo and ProTaper NiTi rotary instruments.

Authors:  F Foschi; C Nucci; L Montebugnoli; S Marchionni; L Breschi; V A Malagnino; C Prati
Journal:  Int Endod J       Date:  2004-12       Impact factor: 5.264

10.  Cone-beam Computed Tomographic Assessment of Canal Centering Ability and Transportation after Preparation with Twisted File and Bio RaCe Instrumentation.

Authors:  Kiamars Honardar; Hadi Assadian; Shahriar Shahab; Zahra Jafari; Ali Kazemi; Kiumars Nazarimoghaddam; Mohammad Javad Kharrazifard; Hossein Labbaf
Journal:  J Dent (Tehran)       Date:  2014-07-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.