Elena Minuesa-García1, José Enrique Iranzo-Cortés1, Teresa Almerich-Torres1, Carlos Bellot-Arcís1, José María Montiel-Company1, José Manuel Almerich-Silla1.
Abstract
In this study, we compare the efficacy and diagnostic concordance of the ICDAS, the radiographic criterion and the instrument known as the DIAGNOdent fluorescence laser pen on occlusal caries lesions using a histological section as the gold standard. Of 100 teeth that did not present cavitated occlusal lesions or occlusal fillings, 80 were chosen through a randomization program and examined by two previously trained and calibrated researchers. Subsequently, the teeth were sectioned with a diamond disk and observed under an optical microscope. The results were studied for caries with a limit established in enamel and caries with extension to dentin. The intra-examiner (0.821-0.933) and inter-examiner (0.817-0.924) reproducibility obtained for both ICDAS and DIAGNOdent for the diagnosis of borderline enamel caries was high. Similarly, intra-examiner (0.686-1.000) and inter-examiner (0.809-0.944) reproducibility for diagnosis of caries with dentin extension was also high for both methods. The sensitivity obtained was 0.76 (ICDAS), 0.87 (DIAGNOdent) and 0.58 (Rx), whereas the specificity obtained was 0.66 (ICDAS), 0.4 (DIAGNOdent) and 0.77 (Rx) for lesions limited to enamel. For lesions with extension to dentin, the sensitivity obtained was 0.73 (ICDAS), 0.82 (DIAGNOdent) and 0.09 (Rx), and the specificity obtained was 0.79 (ICDAS), 0.52 (DIAGNOdent) and 0.97 (Rx). Sensitivity increases in both cases by combining diagnostic methods. In conclusion, ICDAS and DIAGNOdent are better diagnostic methods than Rx for the detection of occlusal caries, and the combination of these methods helps to obtain a better diagnosis.Entities:
Keywords: DIAGNOdent; International Caries Detection and Assessment System (ICDAS); laser fluorescence; occlusal caries; radiography
Year: 2022 PMID: 35629063 PMCID: PMC9146251 DOI: 10.3390/jcm11102937
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.964
Intra- and inter-examiner Reproducibility of ICDAS and DIAGNOdent for lesions limited to enamel (D1) and dentin (D3).
| Inter-Examiner D1 | Inter-Examiner D3 | Intra-Examiner Examiner 1 D1 | Intra-Examiner Examiner 2 D1 | Intra-Examiner Examiner 1 D3 | Intra-Examiner Examiner 2 D3 | |
|---|---|---|---|---|---|---|
|
| 0.924 | 0.944 | 0.871 | 0.821 | 1.000 | 0.868 |
|
| 0.817 | 0.809 | ||||
|
| 0.862 | 0.870 | 0.933 | 0.890 | 0.923 | 0.919 |
|
| 0.825 | 0.922 |
Sensitivity, specificity and area under the ROC curve for ICDAS, DIAGNOdent, radiography and the combination of the three methods for the detection of caries lesions limited to enamel (D1) or dentin (D3).
| Sensitivity D1 | Specificity D1 | Area under the ROC Curve D1 | Sensitivity D3 | Specificity D3 | Area under the ROC Curve D3 | |
|---|---|---|---|---|---|---|
|
| 0.76 | 0.66 | 0.71 | 0.73 | 0.79 | 0.76 |
|
| 0.87 | 0.4 | 0.63 | 0.82 | 0.52 | 0.69 |
|
| 0.58 | 0.77 | 0.68 | 0.09 | 0.97 | 0.53 |
|
| 1 | 0.29 | 0.64 | 0.95 | 0.43 | 0.69 |
|
| 0.62 | 0.77 | 0.70 | 0.59 | 0.88 | 0.74 |
|
| 0.93 | 0.26 | 0.56 | 0.82 | 0.5 | 0.66 |
|
| 0.51 | 0.91 | 0.71 | 0.09 | 0.98 | 0.54 |
|
| 0.84 | 0.49 | 0.67 | 0.73 | 0.76 | 0.74 |
|
| 0.49 | 0.94 | 0.72 | 0.09 | 1 | 0.55 |
|
| 1 | 0.17 | 0.59 | 0.95 | 0.41 | 0.68 |
|
| 0.42 | 0.97 | 0.697 | 0.09 | 1 | 0.54 |
Figure 1ROC curve for the diagnostic validity of the different methods vs. histology in lesions limited to enamel (a) or dentin (b).