Effect of the calibration method of a laser fluorescence device for detecting occlusal caries in primary molars.

Mariana Minatel Braga1, Fausto Medeiros Mendes, Célia Regina Martins Delgado Martins, José Carlos Pettorossi Imparato.   

Abstract

PURPOSE: The purpose of this study was to verify the influence of individual calibration on sound tooth surfaces and the number of readings on in vitro performance of a laser fluorescence (LF) device to detect occlusal caries in primary teeth.
METHODS: After standard calibration (on the porcelain reference object), 72 clinically suspect sites on occlusal surfaces of 54 exfoliated or extracted (for orthodontic purposes) primary molars were assessed using the DIAGNOdent. First, after individual calibration (on a sound surface of each tooth), 3 readings were performed for each site. Subsequently, another 3 readings were taken, but without individual calibration on a sound surface of the tooth. After these assessments, sites were also evaluated with only one reading after individual calibration. Histological validation was performed as the gold standard. LF values, sensitivity, specificity, accuracy (number of correct diagnosis in both sound and diseased teeth) were calculated and compared using the McNemar change test. The area under receiver operating characteristics (ROC) curves was also compared.
RESULTS: Readings with standard calibration only (mean= 11.7 +/- 10.6 SD) were significantly higher statistically than assessment with 3 readings after individual calibration (10.7 +/- 10.7) and with 1 reading (10.2 +/- 8.3) after individual calibration. Nevertheless, sensitivity, specificity, accuracy, and area under ROC curve did not change significantly.
CONCLUSION: Absence of individual calibration does not affect the laser fluorescence device's performance in detecting occlusal caries in primary teeth.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17036712

Source DB:  PubMed          Journal:  Pediatr Dent        ISSN: 0164-1263            Impact factor:   1.874


  4 in total

1.  The influence of zero value subtraction on the performance of a new laser fluorescence device for approximal caries detection.

Authors:  Jonas Almeida Rodrigues; Jonas de Almeida Rodrigues; Isabel Hug; Adrian Lussi
Journal:  Lasers Med Sci       Date:  2008-02-29       Impact factor: 3.161

2.  Effect of sealants on laser fluorescence caries detection in primary teeth.

Authors:  Elina Askaroglou; Katerina Kavvadia; Panagiotis Lagouvardos; Lisa Papagiannoulis
Journal:  Lasers Med Sci       Date:  2010-01-23       Impact factor: 3.161

3.  Diagnostic Validity in Occlusal Caries Detection of ICDAS II, DIAGNOdent, Radiography and a Combination of the Three Methods: An In Vitro Study.

Authors:  Elena Minuesa-García; José Enrique Iranzo-Cortés; Teresa Almerich-Torres; Carlos Bellot-Arcís; José María Montiel-Company; José Manuel Almerich-Silla
Journal:  J Clin Med       Date:  2022-05-23       Impact factor: 4.964

4.  Fluorescence devices for the detection of dental caries.

Authors:  Richard Macey; Tanya Walsh; Philip Riley; Anne-Marie Glenny; Helen V Worthington; Patrick A Fee; Janet E Clarkson; David Ricketts
Journal:  Cochrane Database Syst Rev       Date:  2020-12-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.