Patrizia Natali1, Roberta Bedin2, Gaetano Bernardi3, Elena Corsini3, Eleonora Cocco4, Lucia Schirru4, Ilaria Crespi5, Marta Lamonaca5, Arianna Sala6, Cinzia Nicolò7, Massimiliano Di Filippo8, Alfredo Villa9, Viviana Nociti10, Teresa De Michele11, Paola Cavalla12, Paola Caropreso13, Francesca Vitetta14, Maria Rosaria Cucinelli1, Matteo Gastaldi15, Tommaso Trenti1, Patrizia Sola14, Diana Ferraro2,14.
Abstract
The kappa index (K-Index), calculated by dividing the cerebrospinal fluid (CSF)/serum kappa free light chain (KFLC) ratio by the CSF/serum albumin ratio, is gaining increasing interest as a marker of intrathecal immunoglobulin synthesis. However, data on inter-laboratory agreement of these measures is lacking. The aim was to assess the concordance of CSF and serum KFLC measurements, and of K-index values, across different laboratories. KFLC and albumin of 15 paired CSF and serum samples were analyzed by eight participating laboratories. Four centers used Binding Site instruments and assays (B), three used Siemens instruments and assays (S), and one center used a Siemens instrument with a Binding Site assay (mixed). Absolute individual agreement was calculated using a two-way mixed effects intraclass correlation coefficient (ICC). Cohen's kappa coefficient (k) was used to measure agreement on positive (≥5.8) K-index values. There was an excellent agreement in CSF KFLC measurements across all laboratories (ICC (95% confidence interval): 0.93 (0.87-0.97)) and of serum KFLC across B and S laboratories (ICC: 0.91 (0.73-0.97)), while ICC decreased (to 0.81 (0.53-0.93)) when including the mixed laboratory in the analysis. Concordance for a positive K-Index was substantial across all laboratories (k = 0.77) and within S laboratories (k = 0.71), and very good (k = 0.89) within B laboratories, meaning that patients rarely get discordant results on K-index positivity notwithstanding the testing in different laboratories and the use of different platforms/assays.Entities:
Keywords: cerebrospinal fluid; free light chains; kappa index; multiple sclerosis
Mesh:
Substances:
Year: 2022 PMID: 35625604 PMCID: PMC9138559 DOI: 10.3390/biom12050677
Source DB: PubMed Journal: Biomolecules ISSN: 2218-273X
Figure 1Box-plots showing median (±IQR) CSF (a) and serum (b) KFLC concentrations (mg/L), and K-index (CSF KFLC/Serum KFLC)/(CSF Albumin/serum Albumin) values (c) of the 15 samples across laboratories using Binding Site instruments and assays (B), Siemens instruments and assays (S) or a Siemens instrument coupled with a Binding Site assay (mixed). Upper whiskers show the upper adjacent value (UAV) (largest observation ≤ (third quartile + 1.5 × IQR)); lower whiskers show the lower adjacent value (LAV) (smallest value ≥ (lower quartile − 1.5 × IQR)); black dots show outliers (greater than the UAV or lower than the LAV).
Intraclass Correlation Coefficient (ICC) measured for kappa free light chain (KFLC) and albumin concentrations in cerebrospinal fluid (CSF) and serum and for K-Index = (CSF KFLC/Serum KFLC)/(CSF Albumin/serum Albumin); 95%CI: 95% confidence interval.
| ICC | CSF KFLC (95%CI) | Serum KFLC (95%CI) | K-Index | CSF Albumin (95%CI) | Serum Albumin (95%CI) |
|---|---|---|---|---|---|
| Binding Site platform and assay (B) (nr = 4) | 0.96 (0.9–0.98) | 0.93 (0.63–0.98) | 0.97 (0.94–0.99) | 1.00 (0.99–1) | 0.81 (0.41–0.94) |
| Siemens platform and assay (S) (nr = 3) | 0.99 (0.97–100) | 0.93 (0.48–0.98) | 0.95 (0.89–0.98) | 0.99 (0.98–1) | 0.93 (0.8–0.98) |
| B + S | 0.93 (0.85–0.97) | 0.91 (0.73–0.97) | 0.91 (0.82–0.96) | 0.98 (0.96–0.99) | 0.83 (0.58–0.94) |
| All laboratories (nr = 8) | 0.93 (0.87–0.97) | 0.81 (0.53–0.93) | 0.65 (0.43–0.84) | 0.97 (0.96–0.99) | 0.79 (0.55–0.92) |
Figure 2Lin’s concordance correlation coefficient for CSF KFLC measurements in the least concordant (a) pair of laboratories (0.769 (95%CI: 0.666–0.873)) and in the best concordant pair (b) (0.997 (95%CI: 0.994–0.999)).
K-index measurements across laboratories using Binding Site instruments and assays (B), Siemens instruments and assays (S) or a Siemens instrument coupled with a Binding Site assay (mixed). Discordant results, considering a cut-off for a positive K-index of 5.8, in bold.
| Sample | B1 | B1 | B3 | B4 | mixed | S1 | S2 | S3 |
|---|---|---|---|---|---|---|---|---|
| 1 | 197.0 | 252.1 | 108.3 | 205.3 | 479.5 | 170.5 | 197.9 | 185.6 |
| 2 | 6.9 | 6.4 |
| 11.8 | 10.5 | 5.8 | 6.0 | 6.7 |
| 3 | 1.6 | 2.5 | 2.0 | 2.6 | 4.8 | 2.3 | 1.3 | 1.4 |
| 4 | 7.0 | 8.6 |
| 14.1 | 18.4 |
| 6.3 | 6.4 |
| 5 | 19.3 | 24.3 | 16.0 | 15.5 | 57.0 | 14.4 | 18.0 | 14.8 |
| 6 | 3.4 | 3.1 | 2.4 | 3.2 | 3.6 | 2.2 | 2.7 | 3.4 |
| 7 | 68.5 | 51.4 | 42.3 | 36.5 | 103.8 | 36.5 | 42.8 | 42.1 |
| 8 | 18.2 | 24.6 | 21.1 | 21.3 | 42.3 | 18.8 | 17.5 | 15.4 |
| 9 | 19.2 | 17.6 | 11.5 | 10.1 | 22.7 | 12.2 | 15.6 | 11.1 |
| 10 | 1.8 | 2.9 | 3.2 | 0.4 |
| 3.2 | 2.5 | 2.3 |
| 11 | 7.7 | 10.8 | 9.6 | 11.5 | 19.3 | 10.3 | 7.8 | 10.2 |
| 12 | 27.1 | 28.5 | 18.6 | 20.0 | 86.5 | 19.3 | 23.1 | 20.1 |
| 13 | 1.1 | 1.2 | 1.2 | 1.2 | 1.2 | 1.0 | 1.0 | 0.9 |
| 14 |
| 8.2 |
| 7.0 | 14.2 | 7.5 |
| 6.1 |
| 15 | 7.0 | 12.4 | 9.5 | 9.7 | 23.0 | 13.5 | 8.4 | 8.7 |