| Literature DB >> 34393594 |
Massimo Daves1, Andrea Piccin2,3, Vincenzo Roccaforte4, Giuseppe Lippi5.
Abstract
INTRODUCTION: The measurement of serum free light chain (FLC) represents a fundamental aspect on the assessment of patients with monoclonal gammopathies (MG). Different analytical methods for FLC have become available with the possibility to obtain different value with a substantial impact on the assessment of patients with MG. This study aimed to evaluate FLC results obtained with two different assays and how the difference value obtained can impact in the patient's assessment.Entities:
Keywords: analytical techniques and equipment; comparison; free light chain; immunoassays; monoclonal gammopathies
Mesh:
Substances:
Year: 2021 PMID: 34393594 PMCID: PMC8340501 DOI: 10.11613/BM.2021.030701
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Comparison of the Freelite and N-Latex FLC methods
| Lowest value | 1.5 | 2.2 | 0.0 | 1.1 | 1.3 | 0.01 |
| Highest value | 1870.0 | 4360.0 | 188.90 | 3105.7 | 1150.2 | 889.90 |
| Median (95% CI) | 32.9 | 23.2 | 1.10 | 31.5 | 15.0 | 1.80 |
| IQR | 16.1 - 65.0 | 14.3 - 40.4 | 0.70 - 2.20 | 15.3 - 79.3 | 7.7 - 36.9 | 1.00 - 6.20 |
| IQR – interquartile range. Cl – confidence interval. sFLC κ – serum free light chain kappa. sFLC λ – serum free light chain lambda. | ||||||
Figure 1Comparison of N-Latex with Freelite for the determination of FLC kappa using the Bland Altman plot. A positive bias indicates higher values for the determination of FLC by Freelite compared with N Latex FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
Figure 2Comparison of N-Latex with Freelite for the determination of FLC lambda using the Bland Altman plot. A negative bias indicates higher values for the determination of FLC by N Latex compared with Freelite FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
Figure 3Comparison of N-Latex with Freelite for the determination of FLC κ/λ ratio with Bland Altman plot. A positive bias indicates higher values for the determination of FLC κ/λ ratio by Freelite compared with N Latex FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
Figure 4Concordance analysis for the parameters investigated (N = 93). A: FLC κ/λ ratio (Weighted kappa 0.65 (95% CI 0.52 to 0.78). B: κ FLC (Weighted kappa 0.90 (95% CI 0.82 to 0.98). C: λ FLC (Weighted kappa 0.65 (95% CI 0.53 to 0.78). FLC – free light chain.