Accuracy of user-adjusted axial length measurements with optical biometry in eyes having combined phacovitrectomy for macular-off rhegmatogenous retinal detachment.

Rubina Rahman1, Simon Kolb2, Chun Xiao Bong2, John Stephenson2.   

Abstract

PURPOSE: To evaluate the accuracy of user-adjusted axial length (AL) measured by optical biometry for intraocular lens (IOL) calculations in eyes having combined phacovitrectomy for macula-off rhegmatogenous retinal detachment (RRD).
SETTING: Ophthalmology Department, Calderdale Royal Hospital, Halifax, United Kingdom.
DESIGN: Retrospective case series.
METHODS: Consecutive eyes having phacovitrectomy for macula-off RRD were evaluated. The ALs were measured using optical biometry with user adjustment to identify a posterior peak corresponding to the eye's AL and ultrasound (US). These were compared with each other and with the postoperative optical biometry and analyzed for accuracy as an indication of the eye's AL.
RESULTS: The study comprised 22 eyes of 22 patients. There was no statistically significant difference between the mean AL measurements derived from user-adjusted optical biometry and US AL (P = .964). The user-adjusted optical biometry was not statistically significantly different from the postoperative optical biometry (P = .242). Compared with the postoperative optical biometry, the IOL power was within ±0.5 diopter in 12 (92%) of 13 cases (95% confidence interval [CI], 77.8 to 100.0) for user-adjusted optical biometry and in 10 (77%) of 13 cases (95% CI, 54.0 to 99.8) for US measurements.
CONCLUSIONS: User-adjusted optical biometry could be used as an alternative method for measuring AL in macula-off RRD with combined phacovitrectomy. However, optical biometry would require assessment of agreement with US AL in cases in which a posterior peak is not easily identifiable. User-adjusted optical biometry might outperform US AL when calculating IOL power; however, a larger study should be performed to confirm this. FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27492099     DOI: 10.1016/j.jcrs.2016.04.030

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Changes in axial length after vitrectomy for rhegmatogenous retinal detachment combined with choroidal detachment.

Authors:  Ji-Peng Li; Jun Xu; Meng Zhao
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

2.  Accuracy of segmented measurement of axial length in ultra-high myopia filled with silicone oil using immersion B-scan ultrasonography.

Authors:  Qing-Hua Yang; Hong-Tao Zhang; Xiao-Qi Li; Bing Chen; Zhao-Hui Li; Yi-Fei Huang; Xin Jin; Ying Zhang; Li-Qiang Wang
Journal:  Int J Ophthalmol       Date:  2022-05-18       Impact factor: 1.645

3.  Evaluation of Axial Length Changes after Combined Phacovitrectomy for Macula-off Rhegmatogenous Retinal Detachment.

Authors:  Sadik Altan Ozal; Ruveyde Garip; Ece Ozal; Ayca Kupeli
Journal:  Beyoglu Eye J       Date:  2019-12-24

4.  Long-Term Reproducibility of Axial Length after Combined Phacovitrectomy in Macula-sparing Rhegmatogenous Retinal Detachment.

Authors:  Tae-Seen Kang; Hye-Jin Park; Young-Joon Jo; Jung-Yeul Kim
Journal:  Sci Rep       Date:  2018-10-26       Impact factor: 4.379

5.  Changes in ocular biometric measurements after vitrectomy with silicone oil tamponade for rhegmatogenous retinal detachment repair.

Authors:  Rui Liu; Qingchen Li
Journal:  BMC Ophthalmol       Date:  2020-09-02       Impact factor: 2.209

6.  Differences in Axial Length and IOL Power Based on Alternative A-Scan or Fellow-Eye Biometry in Macula-Off Rhegmatogenous Retinal Detachment Eyes.

Authors:  Rui Liu; Hongrong Li; Qingchen Li
Journal:  Ophthalmol Ther       Date:  2021-12-08