| Literature DB >> 35548029 |
R Benedicta Pearlin1, Roshan Samuel Livingstone1, Anita Jasper1, Shyam Kumar N Keshava1, Gibikote Sridhar1.
Abstract
Purpose: To evaluate the image quality of semi-anatomical chest radiographs acquired using low radiation doses from seven different flat-panel detector (FPD) systems. Materials andEntities:
Keywords: Digital radiography; image quality; radiation dose
Year: 2022 PMID: 35548029 PMCID: PMC9084589 DOI: 10.4103/jmp.jmp_127_21
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Characteristics of the flat panel detectors assessed in this study
| Digital radiography unit | Detector name | Year of installation | Pixel pitch (µm) |
|---|---|---|---|
| Siemens fusion | Trixell | 2012 | 149 |
| Siemens aristos | Trixell | 2007 | 143 |
| GE definium 8000 | GE flashpad | 2010 | 200 |
| Philips digital diagnost | Trixell pixium 4600 | 2007 | 143 |
| Fujifilm FDR smart X FGXR | FDR D-EVO plus C43i | 2016 | 150 |
| Fujifilm FDR smart FGX | FDR D-EVO plus C35 | 2019 | 150 |
| Prognosys prorad | Canon | 2011 | 125 |
Figure 1Placement of regions of interests for quantitative assessment of image quality in the lung (1), heart (2) and diaphragm (3) using the semi anatomical chest phantom
Image quality evaluation of different flat panel detectors systems while using similar radiation dose
| kVp | DR system | mAs | DAP (µGym2) | CNR | SNR | FOM | Average number of contrast details visualized | Spatial resolution Lp/mm | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||||||||||
| Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | |||||
| 70a | Siemens fusion | 8 | 51.05 | 3.58 | 3.33 | 1.83 | 13.44 | 22.35 | 44.55 | 0.25 | 0.22 | 0.07 | 13 | 6 | 8 | 2.8 |
| Siemens aristos | 10 | 51.5 | 4.23 | 3.98 | 2.77 | 7.63 | 15.01 | 38.94 | 0.35 | 0.31 | 0.15 | 15 | 8 | 14 | 3 | |
| GE definium 8000 | 8 | 48.06 | 3.85 | 3.14 | 2.16 | 34.77 | 58.24 | 101.79 | 0.31 | 0.2 | 0.1 | 16 | 9 | 10 | 2.8 | |
| Philips digital diagnost | 8 | 48.82 | 4.32 | 4.19 | 1.45 | 5.28 | 13.38 | 24.91 | 0.38 | 0.36 | 0.04 | 15 | 7 | 7 | 3 | |
| Fujifilm FGXR | 6.4 | 49.54 | −3.07 | −2.91 | −1.08 | 23.27 | 15.39 | 4.36 | 0.19 | 0.17 | 0.02 | 14 | 8 | 11 | 3 | |
| Fujifilm FGX | 6.4 | 53.11 | −3.21 | −2.37 | −1.0 | 14.56 | 12.92 | 9.56 | 0.19 | 0.11 | 0.02 | 16 | 8 | 7 | 3 | |
| Prognosys | 5 | 51.5 | 2.9 | 3.7 | 0.4 | 12.30 | 33.14 | 34.51 | 0.16 | 0.27 | 0.003 | 14 | 8 | 5 | 2.8 | |
| 110b | Siemens fusion | 0.8 | 9.43 | 4.02 | 3.74 | 1.95 | 21.7 | 29.63 | 41.77 | 1.71 | 1.48 | 0.41 | 12 | 3 | 9 | 2.8 |
| Siemens aristos | 0.8 | 9.89 | 4.77 | 3.84 | 3.01 | 15.12 | 19.17 | 34.78 | 2.3 | 1.49 | 0.92 | 10 | 6 | 10 | 3 | |
| GE definium 8000 | 0.8 | 11.73 | 5.47 | 4.56 | 3.18 | 12.98 | 22.82 | 60.21 | 2.55 | 1.77 | 0.86 | 12 | 6 | 12 | 2.8 | |
| Philips digital diagnost | 0.6 | 8.98 | 4.67 | 3.49 | 2.61 | 5.01 | 8.93 | 13.51 | 2.43 | 1.36 | 0.76 | 10 | 5 | 11 | 3 | |
| Fujifilm FGXR | 0.32 | 6.02 | −2.59 | −1.67 | −0.43 | 10.21 | 5.32 | 1.96 | 1.11 | 0.46 | 0.03 | 11 | 5 | 8 | 3 | |
| Fujifilm FGX | 0.32 | 7.82 | −4.53 | −4.48 | −2.42 | 16.83 | 15.78 | 9.16 | 2.62 | 2.56 | 0.75 | 11 | 6 | 10 | 3 | |
| Prognosys | 0.5 | 11.02 | 4.4 | 3.7 | 0.6 | 9.97 | 16.48 | 16.72 | 1.793 | 1.21 | 0.034 | 8 | 5 | 8 | 2.8 | |
aFor 70 kVp, measurements were performed without grid, bFor 110 kVp, measurements were performed with grid. DR: Digital radiography, DAP: Dose-area-product, CNR: Contrast to noise ratio, SNR: Signal to noise ratio, FOM: Figure of merit
Image quality evaluation of different flat panel detectors systems while using low radiation dose
| kVp | DR system | mAs | DAP (µGym2) | CNR | SNR | FOM | Average number of contrast details visualized | Spatial resolution Lp/mm | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||||||||||
| Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | Lung | Heart | Diaphragm | |||||
| 70a | Siemens fusion | 6.3 | 40.02 | 3.59 | 3.17 | 1.58 | 15.88 | 26.22 | 46.29 | 0.32 | 0.25 | 0.06 | 14 | 5 | 9 | 2.8 |
| Siemens aristos | 3 | 16.71 | 4.29 | 3.66 | 1.1 | 19.43 | 33.72 | 55.26 | 1.1 | 0.8 | 0.07 | 12 | 6 | 7 | 3 | |
| GE definium 8000 | 6.3 | 37.95 | 2.98 | 2.87 | 2.24 | 38.39 | 64.3 | 122.14 | 0.23 | 0.22 | 0.13 | 16 | 7 | 9 | 2.8 | |
| Philips digital diagnost | 3.1 | 18.45 | 4.7 | 4.16 | 0.87 | 6.45 | 12.42 | 17.42 | 1.2 | 0.94 | 0.04 | 11 | 4 | 4 | 3 | |
| Fujifilm FGXR | 2.56 | 19.46 | −2.78 | −2.32 | −0.51 | 24.49 | 13.31 | 2.19 | 0.4 | 0.28 | 0.01 | 13 | 6 | 10 | 3 | |
| Fujifilm FGX | 2.56 | 21.51 | −3.95 | −2.74 | −0.8 | 17.86 | 14.55 | 7.33 | 0.73 | 0.35 | 0.03 | 13 | 6 | 5 | 3 | |
| Prognosys | 4 | 41.14 | 3.3 | 2.9 | 0.3 | 11.73 | 28.3 | 23.26 | 0.27 | 0.22 | 0.003 | 4 | 4 | 3 | 2.8 | |
| 110b | Siemens fusion | 0.8 | 9.43 | 4.02 | 3.74 | 1.95 | 21.7 | 29.63 | 41.77 | 1.71 | 1.48 | 0.41 | 12 | 3 | 9 | 2.8 |
| Siemens aristos | 0.5 | 6.45 | 4.63 | 4.38 | 2.4 | 21.68 | 30.76 | 43.53 | 3.32 | 2.98 | 0.89 | 9 | 4 | 9 | 3 | |
| GE definium 8000 | 0.8 | 11.73 | 5.47 | 4.56 | 3.18 | 12.98 | 22.82 | 60.21 | 2.55 | 1.77 | 0.86 | 12 | 6 | 12 | 2.8 | |
| Philips digital diagnost | 0.5 | 7.57 | 4.63 | 4.19 | 2.29 | 4.88 | 8.34 | 11.6 | 2.83 | 2.32 | 0.69 | 11 | 5 | 10 | 3 | |
| Fujifilm FGXR | 0.32 | 6.02 | −2.59 | −1.67 | −0.43 | 10.21 | 5.32 | 1.96 | 1.11 | 0.46 | 0.03 | 11 | 5 | 8 | 3 | |
| Fujifilm FGX | 0.32 | 7.82 | −4.53 | −4.48 | −2.42 | 16.83 | 15.78 | 9.16 | 2.62 | 2.56 | 0.75 | 11 | 6 | 10 | 3 | |
| Prognosys | 0.5 | 11.02 | 4.4 | 3.7 | 0.6 | 9.97 | 16.48 | 16.72 | 1.793 | 1.21 | 0.034 | 8 | 5 | 8 | 2.8 | |
aFor 70 kVp, measurements were performed without grid, bFor 110 kVp, measurements were performed with grid. DR: Digital radiography, DAP: Dose-area-product, CNR: Contrast to noise ratio, SNR: Signal to noise ratio, FOM: Figure of merit
Figure 2Plots of measured values of figure of merit against tube potentials (kVp) for different flat panel detectors used (a) lung, (b) heart and (c) diaphragm
Figure 3Plots of measured values of noise against tube potentials (kVp) for different flat panel detectors used (a) lung, (b) heart and (c) diaphragm