Quality control phantom for digital chest radiography.

H G Chotas1, C E Floyd, G A Johnson, C E Ravin.   

Abstract

PURPOSE: To develop and test a chest phantom for routine quality control testing of digital radiography systems.
MATERIALS AND METHODS: The phantom was constructed from sheets of copper, aluminum, and acrylic, which were cut and arranged to yield a radiographic projection resembling that of a human thorax. Regional test objects allowed quantitative assessment of optical density, contrast detail, and spatial resolution. Validation tests were performed to assess image stability in a stable imaging environment and sensitivity to changes in image quality when they occur.
RESULTS: The phantom yielded consistent pseudoclinical images when used in a routine quality control program and facilitated detection of simulated problems that were induced in imaging system performance.
CONCLUSION: The chest phantom enables quantitative, full-system testing of digital radiography system as they are used clinically for chest radiography.

Entities:  

Mesh:

Year:  1997        PMID: 8988199     DOI: 10.1148/radiology.202.1.8988199

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  4 in total

1.  Evaluation of Radiation Dose Reduction and its Effect on Image Quality for Different Flat-Panel Detectors.

Authors:  R Benedicta Pearlin; Roshan Samuel Livingstone; Anita Jasper; Shyam Kumar N Keshava; Gibikote Sridhar
Journal:  J Med Phys       Date:  2022-03-31

2.  Initial quality performance results using a phantom to simulate chest computed radiography.

Authors:  Wilbroad Muhogora; Renato Padovani; Peter Msaki
Journal:  J Med Phys       Date:  2011-01

3.  Development of a 3D-printed pelvic CT phantom combined with fresh pathological tissues of bone tumor.

Authors:  Xiaomin Li; Bing Wu; Yixuan Zou; Guozhi Zhang; Siyu Liu; Lulu Zhao; Zhengjia Zhang; Wen Wu; Chenglei Liu; Songtao Ai
Journal:  Quant Imaging Med Surg       Date:  2022-09

4.  Evaluation of a quality control phantom for digital chest radiography.

Authors:  E Mah; E Samei; D J Peck
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.