Performance evaluation of Mindray CAL 8000(BC-6800 and SC-120) hematology analyzer and slidemaker/stainer.

Hwan Tae Lee1, Pil-Whan Park1, Yiel-Hea Seo1, Kyung-Hee Kim1, Ja Young Seo1, Ji-Hun Jeong1, Moon Jin Kim1, Jeong-Yeal Ahn1.   

Abstract

INTRODUCTION: The Mindray CAL 8000 is a cellular analysis line that consists of the BC-6800, an automated hematology analyzer, and the SC-120, an automated slidemaker/stainer. We evaluated the performances of the BC-6800 and the SC-120.
METHODS: Four hundred and eight normal and abnormal samples were analyzed. The performance of the BC-6800 and Sysmex XE-2100 were compared, and blood films by the SC-120 and manual method were compared according to the CLSI guideline H26-A2 and H20-A2.
RESULTS: Most parameters measured by the BC-6800 matched well with the XE-2100 and manual differential. The flag efficiency of the BC-6800 for blasts (95.3%) and atypical lymphocytes (92.6%) were higher while immature granulocytes (89.7%) and NRBCs (94.1%) were lower than that of the XE-2100. Additionally, the BC-6800 detected four of five samples infected with plasmodium parasites. The SC-120 showed no carry-over and expected repeatability. There was good agreement on the five-part differential including abnormal cells between blood films by the SC-120 and manually prepared blood films. The shape of the RBC was also comparable between blood films.
CONCLUSION: The CAL-8000 analysis line is beneficial for precise, fast hematology work, and even more useful in malaria endemic areas.
© 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  CAL 8000; automatic slidemaker/stainer; hematology analyzer; performance evaluation; peripheral blood films

Mesh:

Year:  2016        PMID: 27660110      PMCID: PMC6817277          DOI: 10.1002/jcla.22065

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  7 in total

1.  Initial performance evaluation of the UniCel® DxH slide maker/stainer Coulter® cellular analysis system.

Authors:  W Brown; M Keeney; B D Hedley
Journal:  Int J Lab Hematol       Date:  2013-09-13       Impact factor: 2.877

2.  Evaluation of the fully automated hematological analyzer Mindray BC 6800: comparison with Horiba ABX Pentra DX120.

Authors:  R Grillone; E Grimaldi; F Scopacasa; B Dente
Journal:  Int J Lab Hematol       Date:  2013-11-14       Impact factor: 2.877

3.  ICSH recommendations for the standardization of nomenclature and grading of peripheral blood cell morphological features.

Authors:  L Palmer; C Briggs; S McFadden; G Zini; J Burthem; G Rozenberg; M Proytcheva; S J Machin
Journal:  Int J Lab Hematol       Date:  2015-03-02       Impact factor: 2.877

4.  Preliminary evaluation of complete blood cell count on Mindray BC-6800.

Authors:  Giuseppe Lippi; Clarissa Cattabiani; Sabrina Bonomini; Mirco Bardi; Silvia Pipitone; Franco Aversa
Journal:  Clin Chem Lab Med       Date:  2013-04       Impact factor: 3.694

5.  Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions.

Authors:  David A Novis; Molly Walsh; David Wilkinson; Mary St Louis; Jonathon Ben-Ezra
Journal:  Arch Pathol Lab Med       Date:  2006-05       Impact factor: 5.534

6.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

7.  Performance of automated slidemakers and stainers in a working laboratory environment - routine operation and quality control.

Authors:  E Simson; M G Gascon-Lema; D L Brown
Journal:  Int J Lab Hematol       Date:  2009-02-11       Impact factor: 2.877

  7 in total
  2 in total

1.  Consistency Analysis of 3 Detection Systems for Measuring Serum C-Reactive Protein.

Authors:  Xueyan Chen; Chuanyihong Huang; Weichuan Zhong; Suwen Qi; Ting Huang; Zihua Yang
Journal:  Med Sci Monit       Date:  2022-05-11

2.  Performance evaluation of routine blood and C-reactive protein analysis using Mindray BC-7500 CRP auto hematology analyzer.

Authors:  Zhen Lin; Qiu Lin; Pingli Yu; Zhixin Chen; Haifeng Lin; Bin Zhu; Meihua Wang; Yingping Cao
Journal:  Ann Transl Med       Date:  2022-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.