Mindray BC-6800 haematological analyser: 3D-DIFF scattergram usefulness in infectious mononucleosis diagnosis.

Antonio La Gioia1, Maurizio Fumi2, Fabiana Fiorini3, Maria Bombara4, Nicola La Gioia1, Ylenia Pancione2, Silvia Sale2, Marcello Fiorini3, Vincenzo Rocco2.   

Abstract

INTRODUCTION: 3D-DIFF scattergram of the Mindray BC-6800 haematological analyser shows morphological abnormalities and lymphocyte cluster splitting related to the presence of reactive lymphocytes. This study aims to assess whether these cytographic changes are useful in detecting both activated and apoptotic lymphocytes, leading to an improvement in the laboratory diagnostic process of infectious mononucleosis.
METHODS: Two hundred three samples with modified shape and doubled lymphocyte cluster of DIFF scattergram (study group) were divided into two different subgroups: with and, respectively, without serological evidence of ongoing IM. Activated and apoptotic cells in peripheral blood were counted by light microscopy or gating in the instrumental dot plots. Values of apoptotic cells counted by microscopy were compared with those resulting from gating.
RESULTS: Samples with both shape change and doubled lymphocyte cluster had serological profiles according to the diagnosis of ongoing infectious mononucleosis. Blood smears review was positive for reactive lymphocytes in all 112 samples (100%). An underestimation of apoptotic cell count by light microscopy compared with the gating in the instrumental scatterplot was also observed (96 out of 112, 85.7%).
CONCLUSION: The additional lymphocyte cluster was significantly associated with activated and apoptotic lymphocytes in samples with serology suggesting ongoing infectious mononucleosis. Considering the significance of clue for infectious mononucleosis assigned to the apoptotic lymphocytes, a specific flag such as "apoptotic cells?" could be associate with the related cluster. Such a flag could be used for dedicated rules for smears review, thus increasing infectious mononucleosis detection in laboratories that do not usually practise instrumental cytograms observation.
© 2021 John Wiley & Sons Ltd.

Entities:  

Keywords:  activated lymphocytes; apoptotic lymphocytes; infectious mononucleosis; mindray BC-6800 haematological analyser

Year:  2021        PMID: 33891809     DOI: 10.1111/ijlh.13552

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  1 in total

1.  Consistency Analysis of 3 Detection Systems for Measuring Serum C-Reactive Protein.

Authors:  Xueyan Chen; Chuanyihong Huang; Weichuan Zhong; Suwen Qi; Ting Huang; Zihua Yang
Journal:  Med Sci Monit       Date:  2022-05-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.