Jorge A Calvo-Sanz1, Francisco Poyales2, Ying Zhou2, Alfonso Arias-Puente1, Nuria Garzón3.
Abstract
Purpose: To evaluate the agreement between the biometric measurements used to calculate the size of the implantable collamer lenses (ICL) with different technologies: swept-source optical coherence tomography, spectral domain optical coherence tomography, and Scheimpflug tomography.Entities:
Keywords: Anterior Chamber Depth; ICL; phakic lens; white to white
Mesh:
Year: 2022 PMID: 35502031 PMCID: PMC9333023 DOI: 10.4103/ijo.IJO_2217_21
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 2.969
Statistical demographic data
| Mean (mm) | SD (mm) | Range (mm) | |
|---|---|---|---|
| Pentacam WTW | 11.84 | 0.31 | 11.00-12.50 |
| IOLMaster 700 WTW | 12.05 | 0.30 | 11.20-12.60 |
| Cirrus WTW | 11.88 | 0.65 | 10.08-13.08 |
| Visante ATA | 12.10 | 0.32 | 11.35-12.65 |
| Pentacam ACD | 3.25 | 0.23 | 2.84-3.74 |
| IOLMaster 700 ACD | 3.19 | 0.21 | 2.83-3.59 |
| Cirrus ACD | 3.29 | 0.43 | 2.42-4.01 |
| Visante ACD | 3.28 | 0.22 | 2.90-3.74 |
ACD=anterior chamber depth (from endothelium to anterior lens surface), ATA=angle to angle (horizontal), WTW=white to white (horizontal)
Mean differences between horizontal and anterior chamber depth measurements for each pair of variables
| Variables in pairs | Mean (mm) | SD (mm) | Range (mm) |
|
|---|---|---|---|---|
| WTW Pentacam-WTW IOLMaster 700 | -0.21 | 0.09 | -0.10--0.50 | <0.001 |
| WTW Pentacam-WTW Cirrus | -0.10 | 0.46 | -0.88-1.12 | 0.156 |
| WTW Pentacam-ATA Visante | -0.26 | 0.26 | -0.85-0.18 | <0.001 |
| WTW Cirrus-WTW IOLMaster 700 | -0.11 | 0.49 | -1.62-0.60 | 0.146 |
| ATA Visante-WTW IOLMaster 700 | 0.05 | 0.27 | -0.45-0.65 | 0.227 |
| ACD Pentacam-ACD IOLMaster 700 | 0.06 | 0.06 | -0.11-0.15 | <0.001 |
| ACD Pentacam-ACD Cirrus | -0.04 | 0.09 | -0.43-0.23 | 0.003 |
| ACD Pentacam-ACD Visante | -0.03 | 0.06 | -0.23-0.11 | 0.003 |
| ACD Cirrus-ACD IOLMaster 700 | 0.09 | 0.05 | -0.14-0.23 | <0.001 |
| ACD Visante-ACD IOLMaster 700 | 0.09 | 0.05 | -0.04-0.21 | <0.001 |
ACD=anterior chamber depth (from endothelium to anterior lens surface), ATA=angle to angle (horizontal), WTW=white to white (horizontal)
Figure 1Linear correlation graphs of the horizontal diameter measurements from different devices in pairs
Linear correlation and significance for different devices in pairs for the horizontal diameter measurements and anterior chamber depth
| Variables in pairs | Correlation |
|
|
|---|---|---|---|
| WTW Pentacam-WTW IOLMaster 700 | 0.950 | 0.902 | <0.001 |
| WTW Pentacam-WTW Cirrus | 0.703 | 0.494 | <0.001 |
| WTW Pentacam-ATA Visante | 0.673 | 0.452 | <0.001 |
| WTW IOLMaster-WTW Cirrus | 0.654 | 0.426 | <0.001 |
| WTW IOLMaster-ATA Visante | 0.621 | 0.385 | <0.001 |
| ACD Pentacam-ACD IOLMaster 700 | 0.965 | 0.931 | <0.001 |
| ACD Pentacam-ACD Cirrus | 0.982 | 0.964 | <0.001 |
| ACD Pentacam-ACD Visante | 0.958 | 0.917 | <0.001 |
| ACD IOLMaster-ACD Cirrus | 0.992 | 0.984 | <0.001 |
| ACD IOLMaster-ACD Visante | 0.975 | 0.952 | 0.023 |
ACD=anterior chamber depth (from endothelium to anterior lens surface), ATA=angle to angle (horizontal), WTW=white to white (horizontal)
Figure 2Linear correlation graphs of the anterior chamber depth, from corneal endothelium to anterior capsule of the lens, from different devices in pairs
ICC between pairs of WTW, ATA, and ACD measurements
| ICC | 95% CI (mm) |
| |
|---|---|---|---|
| WTW Pentacam-WTW IOLMaster 700 | 0.973 | 0.949-0.986 | <0.001 |
| WTW Pentacam-WTW Cirrus | 0.788 | 0.605-0.886 | <0.001 |
| WTW Pentacam-ATA Visante | 0.803 | 0.627-0.896 | <0.001 |
| WTW IOLMaster 700-ATA Visante | 0.764 | 0.652-0.873 | <0.001 |
| WTW IOLMaster 700-WTW Cirrus | 0.754 | 0.542-0.868 | <0.001 |
| ACD Pentacam-ACD IOLMaster 700 | 0.980 | 0.963-0.989 | <0.001 |
| ACD Pentacam-ACD Cirrus | 0.990 | 0.981-0.995 | <0.001 |
| ACD Pentacam-ACD Visante | 0.979 | 0.960-0.989 | <0.001 |
| ACD IOLMaster 700-ACD Visante | 0.986 | 0.975-0.993 | <0.001 |
| ACD IOLMaster 700-ACD Cirrus | 0.996 | 0.993-0.998 | <0.001 |
ACD=anterior chamber depth (from endothelium to anterior lens surface), ATA=angle to angle (horizontal), CI=confidence interval, ICC=intraclass coefficient, WTW=white to white (horizontal)
Figure 3Bland–Altman representation for the horizontal corneal diameter between means of measurements and differences between devices
Figure 4Bland–Altman representation for the anterior chamber depth between means of measurements and differences between devices