Ioana Boca Bene1, Anca Ileana Ciurea1, Cristiana Augusta Ciortea2, Ștefan Cristian Vesa3, Sorin Marian Dudea1.
Abstract
The purpose of this study was to determine the observer agreement in assessing the enhancement pattern of suspicious breast lesions with contrast-enhanced ultrasound (CEUS) using high and low frequency transducers.Entities:
Keywords: breast cancer; contrast-enhanced ultrasound; convex-array transducer; linear-array transducer
Year: 2022 PMID: 35453846 PMCID: PMC9025659 DOI: 10.3390/diagnostics12040798
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1Flowchart of patient’s inclusion in the study.
Patient’s characteristics.
| Patient’s Characteristics | Number | Frequency (%) |
|---|---|---|
|
| 48.5 (28–83) | |
|
| 26.37 (11–87.5) | |
|
| ||
|
| 8 |
|
| Fibroadenoma | 6 | 8.57 |
| Radial scar | 1 | 1.43 |
| Intraductal papilloma | 1 | 1.43 |
|
| 62 |
|
| DCIS | 1 | 1.43 |
| DCIS + Borderline phyllodes | 1 | 1.43 |
| IDC NST | 53 | 75.72 |
| IDC with mucinous components | 1 | 1.43 |
| Tubular carcinoma | 1 | 1.43 |
| Papillary carcinoma | 2 | 2.85 |
| Malignant phyllodes | 1 | 1.43 |
| ILC | 2 | 2.85 |
|
| ||
| I | 8 | 13.12 |
| II | 34 | 55.73 |
| III | 19 | 31.15 |
|
| ||
| Positive | 49 | 80.32 |
| Negative | 12 | 16.68 |
|
| ||
| Positive | 29 | 47.54 |
| Negative | 32 | 52.46 |
|
| ||
| Positive | 35 | 57.38 |
| Negative | 26 | 42.62 |
|
| ||
| <20% | 26 | 42.62 |
| ≥20% | 35 | 57.38 |
|
| ||
| A | 22 | 36.07 |
| B | 27 | 44.27 |
| Her2+ | 6 | 9.83 |
| Triple-negative | 6 | 9.83 |
Agreement between the linear-array transducer and convex-array transducer regarding lesion’s enhancement pattern for the first, second and third reader.
| Lesion’s Enhancement Pattern at CEUS | First Reader | Second Reader | Third Reader | |||
|---|---|---|---|---|---|---|
| Kappa Coefficient | Kappa Coefficient | Kappa Coefficient | ||||
| Enhancement degree (hypo-, iso-, hyper-enhancement) | 0.320 | <0.001 | 0.644 | <0.001 | 0.498 | <0.001 |
| Internal homogeneity (homogeneous/inhomogeneous) | 0.301 | 0.015 | 0.711 | <0.001 | 0.485 | <0.001 |
| Presence/absence of perfusion defects | 0.300 | 0.017 | 0.711 | <0.001 | 0.485 | <0.001 |
| Uptake pattern (centripetal/centrifugal) | 0.614 | <0.001 | 1 | <0.001 | 1 | <0.001 |
| Lesion margins (circumscribed/non-circumscribed) | - | - | 0.841 | <0.001 | 1 | <0.001 |
| Presence/absence of perilesional enhancement | 0.320 | 0.002 | 0.517 | <0.001 | 0.538 | <0.001 |
Kappa < 0—no agreement; Kappa 0–0.2—slight agreement; Kappa = 0.21–0.40—fair agreement; Kappa = 0.41–0.60—moderate agreement; Kappa = 0.61–0.80—substantial agreement; Kappa = 0.81–1—almost perfect agreement.
Inter-observer agreement between the three readers related to the linear-array and convex-array transducer.
| Lesion’s Enhancement Pattern at CEUS | Linear-Array Transducer | Convex-Array Transducer | ||
|---|---|---|---|---|
| Cronbach’s Alpha Coefficient | Cronbach’s Alpha Coefficient | |||
| Enhancement degree (hypo-, iso-, hyper-enhancement) | 0.960 | <0.001 | 0.903 | <0.001 |
| Internal homogeneity (homogeneous/inhomogeneous) | 0.703 | <0.001 | 0.757 | <0.001 |
| Presence/absence of perfusion defects | 0.703 | <0.001 | 0.757 | <0.001 |
| Uptake pattern (centripetal/centrifugal) | 0.831 | <0.001 | 0.832 | <0.001 |
| Lesion margins (circumscribed/non-circumscribed) | 0.792 | <0.001 | 0.330 | <0.001 |
| Presence/absence of perilesional enhancement | 0.853 | <0.001 | 0.762 | <0.001 |
Cronbach’s Alpha = 0.01–0.60—not acceptable internal consistency; Cronbach’s Alpha = 0.61–0.70—acceptable internal consistency; Cronbach’s Alpha = 0.71–0.80—good and acceptable internal consistency; Cronbach’s Alpha = 0.81–0.90—good internal consistency; Cronbach’s Alpha = 0.91–1—excellent internal consistency.