| Literature DB >> 35449382 |
Xi Liao1, Yuwei Liu2, Shiqi Liang1, Ka Li3.
Abstract
BACKGROUND: Hydrophilic coated catheters are recommended to reduce the side effects of intermittent catheterization (IC) in patients with bladder dysfunction. However, there is lack of Level one evidence to support the use of this intervention. SEARCHEntities:
Keywords: Adverse events; Hematuria; Hydrophilic catheters; Intermittent catheterization; Urethral stricture
Mesh:
Year: 2022 PMID: 35449382 PMCID: PMC9184422 DOI: 10.1007/s11255-022-03172-x
Source DB: PubMed Journal: Int Urol Nephrol ISSN: 0301-1623 Impact factor: 2.266
Fig. 1PRISMA flow diagram—clinical search strategy
Summary of extracted clinical data
| Study | Location | Age | Gender a | Catheter material (brand), C/Tb | Size |
|---|---|---|---|---|---|
| William DeFoor (2017) [ | America | 12.9/13.6 | 38/40 | C: unknown T: unknown/lofric (Wellspect Healthcare) | Unknown |
| De Ridder (2005) [ | Spain, Belgium | 37.5±14.6/ 36.7±14.6 | M | C:PVCc (Conveen,Coloplast) T:PUd/Speedicath (Coloplast) | ch10,12,14 |
| Pachler (1999) [ | Denmark | 71.3 | M | C:PVC (Mentor santa barbara) T:PVC/lofric (Astra pharmaceuticals) | Unknown |
| Diana D. Cardenas (2011) [ | America, Canada | 35.1±13.2/ 37.2±14.4 | 100/39 | C:PVC (Conen)T: PU/Speedicath | Unknown |
| Tariq Burki (2019) [ | Saudi Arabia | 5 | 47/54 | C:PEe T: unknown | Unknown |
| Ronald (1996) | America | 11.7±3.8/ 12.1±5.7 | M | C:PVC (Mentor) T:unknown/Lofric | 11.5 ± 2.5/11.1 ± 2.1 |
| Wyndaele (2000) [ | Belgium | 45±15 | M | C:unknown T:unknown/Urocath-Gel1 | 12–14 French |
| Luca Cindolo (2003) [ | Italy | 62.3/67.4 | 80/20 | C:PVC T:PVC/EasiCath (Coloplast, Denmark) | 12-Charr |
| Sataa Sallami (2010) [ | Tunisia | 62/60.9 | M | C:PVC T:unknown/LoFric (Astra Tech; Molndal, Sweden) | Unknown/Number 16 or 18 |
| Jonathan et al. (2003) [ | America | 39.8±12.9/ 39.6±16 | M | C:PVC T: unknown/Lofric | MOST are 14Fr, a few are 16Fr, 12Fr |
| Stensballe (2005) [ | Denmark | 24 | M | C:silica gel or PVC (incare1 advance plus, Hollister inc,USA) T:unknown/speedicath (Conveen, ColoplastA/S Denmark) | CH12 |
| Kjaergaard (1994) [ | Denmark | 68 | M | C:no T: unknown/LoFric | Unknown |
aGender: M/ F (male/ female)
bT: hydrophilic coated (HC); C: non-hydrophilic catheters (NHC)
cPVC: polyvinyl chloride
dPU: polyurethane
ePE: polyethylene
Fig. 2A Risk of bias summary for RCT (n = 9) and NRS (n = 3); B risk of bias graph for all included studies (n = 12). RCT randomized controlled trials, NRS non-randomized controlled trials
Fig. 3Meta-analysis comparing hydrophilic catheter with non-hydrophilic catheter, evaluating gross hematuria
Fig. 4Meta-analysis comparing hydrophilic catheter with non-hydrophilic catheter, evaluating subgroup analysis of gross hematuria
Fig. 5Meta-analysis comparing hydrophilic catheter with non-hydrophilic catheter, evaluating microscopic hematuria
Fig. 6Meta-analysis comparing hydrophilic catheter with non-hydrophilic catheter, evaluating urethral stricture