| Literature DB >> 35444163 |
Van-Khoa Pham1, Tran-Lan-Khue Pham2.
Abstract
The aim of the present study was to evaluate the agreements between the root canal length estimations using cone-beam computed tomography (CBCT) at different slice thicknesses, dedicated software, or an electronic apex locator (EAL) and the actual lengths (AL). In total, 111 extracted human molars with 302 root canals were chosen. Teeth were scanned using a CBCT device at a voxel size of 0.075 mm. Root canal lengths were estimated using CBCT software at different slice thicknesses (0.6, 1.2, and 2.4 mm) and dedicated software for proposed or operator lengths. The endodontic access cavities were created, and root canal lengths were estimated with an EAL for electronic length (EL) and a ruler for AL. Data were tested using paired t-tests and Bland-Altman plots to detect the differences between the methods in length estimation at a significance of 0.05. The accuracy in the range of ± 0.5 mm was 100% for the EAL. There was an agreement between the EL and CBCT at a slice thickness of 1.2 mm (p = 0.349). CBCT at the smallest slice thickness estimation was not the best modality in agreement with the AL. The EAL was an accurate and reliable method for root canal length measurement.Entities:
Mesh:
Year: 2022 PMID: 35444163 PMCID: PMC9021240 DOI: 10.1038/s41598-022-10534-z
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1CBCT measurements of root canal length at slice thickness of 1.2 mm.
The incidence (%) of differences between the four methods and the AL measurements.
| Groups | Shorter than AL N (%) | Equal to AL N (%) | Longer than AL N (%) | ± 0.5 mm (%) | ||
|---|---|---|---|---|---|---|
| > 0.5 mm | ≤ 0.5 mm | ≤ 0.5 mm | > 0.5 mm | |||
| 3D-PL–AL | 8 (2.6) | 46 (15.2) | 23 (7.6) | 214 (70.9) | 11 (3.6) | 93.8a |
| 3D-OL–AL | 4 (1.3) | 164 (54.3) | 27 (8.9) | 105 (34.8) | 2 (0.7) | 98.0b |
| CL0.6–AL | 48 (15.9) | 47 (15.5) | 0 (0.0) | 73 (24.2) | 134 (44.4) | |
| CL1.2–AL | 69 (22.8) | 50 (16.6) | 1 (0.3) | 70 (23.2) | 112 (37.1) | |
| CL2.4–AL | 42 (13.9) | 34 (11.3) | 1 (0.3) | 81 (26.8) | 144 (47.7) | |
| EL–AL | 0 (0.0) | 218 (72.2) | 54 (17.9) | 30 (9.9) | 0 (0.0) | 100.0d |
a,b,c1,2,3,dDifferent superscript letters indicate that there were significant differences at a level of 0.05 ().
3D-PL: Proposed length; 3D-OL: Operator length; EL: Electronic length; AL: Actual length.
CL0.6: CBCT length at slice thickness of 0.6 mm.
CL1.2: CBCT length at slice thickness of 1.2 mm.
CL2.4: CBCT length at slice thickness of 2.4 mm.
Mean biases, confidence intervals, p values in two statistical tests, fixed or proportional biases between experimental methods and AL measurements.
| Groups | Paired t-test | Linear regression | Fixed bias | Proportional bias | ||
|---|---|---|---|---|---|---|
| Mean bias | 95% CI | |||||
| 3D-PL–AL | -0.109 | -0.138 to -0.080 | < 0.001* | 0.239 | Yes | No |
| 3D-OL–AL | 0.026 | 0.008 to 0.045 | 0.006* | 0.890 | Yes | No |
| CL0.6–AL | 0.346 | 0.218 to 0.473 | < 0.001* | < 0.001* | Yes | Yes |
| CL1.2–AL | 0.131 | -0.038 to 0.301 | 0.129 | < 0.001* | No | Yes |
| CL2.4–AL | 0.468 | 0.343 to 0.593 | < 0.001* | < 0.001* | Yes | Yes |
| EL–AL | 0.050 | 0.042 to 0.058 | < 0.001* | 0.628 | Yes | No |
*Differences at significant level of 0.05.
3D-PL: Proposed length; 3D-OL: Operator length; EL: Electronic length; AL: Actual length.
CL0.6: CBCT length at slice thickness of 0.6 mm.
CL1.2: CBCT length at slice thickness of 1.2 mm.
CL2.4: CBCT length at slice thickness of 2.4 mm.
Mean biases, confidence intervals, p values in two statistical tests, fixed or proportional biases among methods’ estimations.
| Groups | Paired t-test | Linear regression | Fixed bias | Proportional bias | ||
|---|---|---|---|---|---|---|
| Mean bias | 95% CI | |||||
| 3D-OL–3D-PL | -0.136 | -0.158 to -0.113 | < 0.001* | 0.013 | Yes | No |
| CL0.6–CL1.2 | -0.214 | -0.379 to -0.050 | 0.011* | 0.483 | Yes | No |
| CL0.6–CL2.4 | 0.122 | 0.039 to 0.205 | 0.004* | 0.453 | Yes | No |
| CL1.2–CL2.4 | 0.337 | 0.174 to 0.499 | 0.001* | 0.764 | Yes | No |
| EL–3D-OL | -0.024 | -0.044 to -0.0030 | 0.025* | 0.535 | Yes | No |
| EL–3D-PL | -0.159 | -0.190 to -0.129 | < 0.001* | 0.025 | Yes | Yes |
| EL–CL0.6 | 0.296 | 0.167 to 0.424 | < 0.001* | 0.342 | Yes | No |
| EL–CL1.2 | 0.081 | -0.089 to 0.251 | 0.349 | 0.951 | No | No |
| EL–CL2.4 | 0.418 | 0.292 to 0.543 | < 0.001* | 0.647 | Yes | No |
*Differences at significant level of 0.05.
3D-PL: Proposed length; 3D-OL: Operator length; EL: Electronic length; AL: Actual length.
CL0.6: CBCT length at slice thickness of 0.6 mm.
CL1.2: CBCT length at slice thickness of 1.2 mm.
CL2.4: CBCT length at slice thickness of 2.4 mm.