Shahram Sayyadi1, Mohammad Mahdi Omidian2, Ali Pourmojarab2, Abouzar Khodayi2, Mojtaba Baroutkoub2, Sohrab Salimi1, Alireza Manafi Rasi2.
Abstract
Background: The present study aims to evaluate the diagnostic exactitude of the intraoperative Chertsey test in tibiofibular syndesmotic injuries in patients with malleolar fractures, in comparison with a computed tomography (CT) scan. Materials andEntities:
Keywords: Ankle; chertsey test; computed tomography scan; malleolar fractures; syndesmosis
Year: 2022 PMID: 35386542 PMCID: PMC8977619 DOI: 10.4103/abr.abr_258_21
Source DB: PubMed Journal: Adv Biomed Res ISSN: 2277-9175
Figure 1Intraoperative Chertsey test demonstrating the contrast medium spread into the tibiofibular syndesmosis that was removed after screw fixation
Description and comparison of basic characteristics of patients based on the Chertsey test results
| Variable | The chertsey test results frequency (%) |
| |
|---|---|---|---|
|
| |||
| Negative | Positive | ||
| Age (mean±SD) | 33.57±10.98 | 39.13±11.84 | 0.128 |
| Gender | |||
| Male | 11 (52.4) | 10 (47.6) | 0.366 |
| Female | 12 (66.7) | 6 (33.3) | |
| Morphology | |||
| Concave | 13 (56.5) | 10 (43.5) | 0.709 |
| Shallow | 10 (62.5) | 6 (37.5) | |
| Type of fracture | |||
| BM | 4 (44.4) | 5 (55.6) | <0.001 |
| LM | 15 (78.9) | 4 (21.1) | |
| MM | 4 (100) | 0 | |
| TM | 0 | 7 (100) | |
BM: Bi-malleolar, LM: Lateral malleolus, MM: Medial malleolus, TM: Tri-malleolar, SD: Standard deviation
Correlation of the parameters of preoperative computed tomography scan in patients with positive and negative Chertsey test results and the control group
| Variable | Positive test | Control (positive) |
| Negative test | Control (negative) |
| ||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||||
| Median | Interquartile range | Median | Interquartile range | Median | Interquartile range | Median | Interquartile range | |||
| ATFD | 4.25 | 4.03-4.30 | 2.25 | 2.13-2.58 | <0.001 | 2.40 | 2.20-2.60 | 2.40 | 2.20-2.60 | 0.037 |
| MTFD | 5.05 | 5.05-5.10 | 2.70 | 2.60-2.88 | <0.001 | 2.80 | 2.70-2.90 | 2.80 | 2.70-2.90 | 0.846 |
| PTFD | 5.25 | 5.10-5.58 | 3.20 | 3-3.30 | <0.001 | 3.30 | 3.10-3.40 | 3.30 | 3.10-3.40 | 0.419 |
| MAX TFD | 6.40 | 6.30-6.50 | 4 | 4-4.10 | <0.001 | 4.20 | 4.10-4.30 | 4.20 | 4.10-4.30 | 0.005 |
| Volume | 2.50 | 2.40-2.58 | 1.50 | 1.40-1.50 | <0.001 | 1.60 | 1.40-1.70 | 1.60 | 1.40-1.70 | 0.224 |
TFD: Tibiofibular distances, ATFD: Anterior TFD, MTFD: Middle TFD, PTFD: Posterior TFD, MAX: Maximal
Figure 2Comparison of anterior tibiofibular distances (a), Middle tibiofibular distances (b), Posterior tibiofibular distances (c), Maximal tibiofibular distances (d), and volume (e) parameters of patients based on the Chertsey test results
Figure 3Operating characteristic curve of the different tibiofibular distance positions and volume obtained from the preoperative computed tomography scan
Cut-off points based on Yuden criterion for computed tomography scan parameters and preoperative volume and diagnostic values corresponding to cut-off points
| Variable | Cut-off point | 95% CI | |||
|---|---|---|---|---|---|
|
| |||||
| Sensitivity | Specificity | PPV | NPV | ||
| Anterior | <2.7 | 100 (79.4-100) | 100 (85.2-100) | 100 (79.4-100) | 100 (85.2-100) |
| Middle | <3.1 | 100 (79.4-100) | 100 (85.2-100) | 100 (79.4-100) | 100 (85.2-100) |
| Posterior | <3.6 | 100 (79.4-100) | 100 (85.2-100) | 100 (79.4-100) | 100 (85.2-100) |
| MAX | <4.5 | 100 (79.4-100) | 100 (85.2-100) | 100 (79.4-100) | 100 (85.2-100) |
| Volume | <2 | 100 (79.4-100) | 100 (85.2-100) | 100 (79.4-100) | 100 (85.2-100) |
CI: Confidence interval, PPV: Positive predictive value, NPV: Negative predictive value, MAX: Maximal