Lexiaozi Fan1,2, Kyungpyo Hong1, Li-Yueh Hsu3, James C Carr1, Bradley D Allen1, Daniel C Lee4, Daniel Kim1,2. 1. Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA. 2. Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA. 3. Department of Radiology and Imaging Sciences, National Institutes of Health, Bethesda, Maryland, USA. 4. Division of Cardiology, Internal Medicine, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
Abstract
PURPOSE: The purpose of this study was to determine an optimal saturation-recovery time (TS) for minimizing the underestimation of arterial input function (AIF) in quantitative cardiac perfusion MRI without multiple gadolinium injections per subject. METHODS: We scanned 18 subjects (mean age = 59 ± 14 years, 9/9 males/females) to acquire resting perfusion data and 1 additional subject (age = 38 years, male) to obtain stress-rest perfusion data using a 5-fold accelerated pulse sequence with radial k-space sampling and applied k-space weighted image contrast (KWIC) filters on the same k-space data to retrospectively reconstruct five AIF images with effective TS ranging from 10 to 21.2 ms (2.8 ms steps). Undersampled images were reconstructed using a compressed sensing framework with temporal-total-variation and temporal-principal-component as 2 orthogonal sparsifying transforms. The image processing steps included, same motion correction across five different AIF images, signal normalization by the proton-density-weighted-image, signal-to-T1 conversion using a Bloch equation, T1 -to-gadolinium-concentration conversion assuming fast water exchange, T2 * correction to the AIF, and gadolinium-concentration to myocardial blood flow (MBF) conversion based on a Fermi model. RESULTS: Among five TS values, the shortest TS (10 ms) produced significantly (P < 0.05) higher peak AIF and lower resting MBF (13.73 mM, 0.73 mL g-1 min-1 ) than 12.8 ms (11.24 mM, 0.89 mL g-1 min-1 ), 15.6 ms (9.56 mM, 1.05 mL g-1 min-1 ), 18.4 ms (8.55 mM, 1.17 mL g-1 min-1 ), and 21.2 ms (7.95 mM, 1.27 mL g-1 min-1 ). Similarly, shorter TS reduced underestimation of AIF (or overestimation of MBF) for both during stress and at rest, but this effect was canceled in myocardial-perfusion-reserve (MPR). CONCLUSION: This study demonstrates that TS of 10 ms reduces the underestimation of AIF and, hence, the overestimation of MBF compared with longer TS values (12.8-21.2 ms).
PURPOSE: The purpose of this study was to determine an optimal saturation-recovery time (TS) for minimizing the underestimation of arterial input function (AIF) in quantitative cardiac perfusion MRI without multiple gadolinium injections per subject. METHODS: We scanned 18 subjects (mean age = 59 ± 14 years, 9/9 males/females) to acquire resting perfusion data and 1 additional subject (age = 38 years, male) to obtain stress-rest perfusion data using a 5-fold accelerated pulse sequence with radial k-space sampling and applied k-space weighted image contrast (KWIC) filters on the same k-space data to retrospectively reconstruct five AIF images with effective TS ranging from 10 to 21.2 ms (2.8 ms steps). Undersampled images were reconstructed using a compressed sensing framework with temporal-total-variation and temporal-principal-component as 2 orthogonal sparsifying transforms. The image processing steps included, same motion correction across five different AIF images, signal normalization by the proton-density-weighted-image, signal-to-T1 conversion using a Bloch equation, T1 -to-gadolinium-concentration conversion assuming fast water exchange, T2 * correction to the AIF, and gadolinium-concentration to myocardial blood flow (MBF) conversion based on a Fermi model. RESULTS: Among five TS values, the shortest TS (10 ms) produced significantly (P < 0.05) higher peak AIF and lower resting MBF (13.73 mM, 0.73 mL g-1 min-1 ) than 12.8 ms (11.24 mM, 0.89 mL g-1 min-1 ), 15.6 ms (9.56 mM, 1.05 mL g-1 min-1 ), 18.4 ms (8.55 mM, 1.17 mL g-1 min-1 ), and 21.2 ms (7.95 mM, 1.27 mL g-1 min-1 ). Similarly, shorter TS reduced underestimation of AIF (or overestimation of MBF) for both during stress and at rest, but this effect was canceled in myocardial-perfusion-reserve (MPR). CONCLUSION: This study demonstrates that TS of 10 ms reduces the underestimation of AIF and, hence, the overestimation of MBF compared with longer TS values (12.8-21.2 ms).
Coronary artery disease (CAD) is a leading cause of sickness and death in the United States.
Noninvasive myocardial perfusion imaging modalities such as single‐photon emission computerized tomography (SPECT), (positron emission tomography) PET, and cardiovascular magnetic resonance (CMR) play an important role as a gatekeeper to cardiac catheterization.
CMR offers several advantages over SPECT and PET, including high spatial resolution, lack of ionizing radiation, and capability to provide additional information such as myocardial function, edema, and scarring/fibrosis and hemodynamics evaluation with a single examination. Although a qualitative evaluation of perfusion CMR is the clinical norm, myocardial blood flow (MBF) quantification from perfusion CMR images offers incremental value, including: improved diagnostic accuracy,
particularly in multi‐vessel obstructive disease and microvascular dysfunction
; prognostic value
,
; and improved precision for longitudinal studies.There are several technical challenges to achieve high accuracy in MBF quantification, including, but not limited to, low signal‐to‐noise ratio (SNR), nonlinear relationship between the measured signal and concentration of gadolinium ([Gd]), and T2* decay in arterial input function (AIF) signal at peak blood enhancement because of high [Gd] during first‐passage. This study revisits the accuracy in AIF, which is often underestimated because of high [Gd] (>5 mM) in the left ventricular (LV) blood pool at peak enhancement. The two main imaging approaches to compensate for this nonlinearity issue are the dual‐bolus
and dual‐imaging methods.
,
The dual‐bolus technique, wherein the AIF image is acquired with a low [Gd] and tissue function (TF) images are acquired with a high [Gd], is not practical in a busy radiology clinic because it requires contrast dilution and twice as many bolus injections and image acquisitions. Furthermore, a meticulous setup is necessary to assure that 2 different concentrations are at equal volumes and prevent backflow of solutions in the injector apparatus between saline and gadolinium syringes.
A more practical approach is to perform one set of stress‐rest imaging with the dual‐imaging sequence,
,
,
,
wherein the AIF images are acquired with a short saturation recovery time (TS) to sample the center of k‐space and TF images are acquired with a long TS.The previous studies utilizing the dual‐imaging method
,
,
,
used TS ranging from 20‐27 ms to linearize the relationship between the AIF signal and [Gd], in order to reduce the underestimation of AIF. However, none of the studies have investigated an optimal TS for minimizing the underestimation of AIF. The main reason why it is impractical to conduct such a study is that it requires multiple injections of gadolinium. In this study, we sought to identify an optimal TS for minimizing the underestimation of AIF without multiple injections of gadolinium per subject. This was accomplished by scanning patients using a radial k‐space sampling pattern with a minimum TS of 10 ms and retrospectively reconstructing five different AIF images derived from the same radial k‐space data with effective TS ranging from 10 to 21.2 ms (2.8 ms steps) using k‐space weighted image contrast (KWIC) filters.
METHODS
Study population
All patients provided consent in writing. This study was performed in accordance with protocols approved by our institutional review board and was Health Insurance Portability and Accountability Act (HIPAA) compliant. We prospectively enrolled 18 patients (mean age = 59 ± 14 years, 9/9 males/female) with left ventricular ejection fraction (LVEF) > 50% and no evidence of vascular late gadolinium enhancement to acquire resting perfusion data: 13 patients were enrolled in a study examining the chronic (medium days since hospitalization = 102 days) impact of coronavirus disease (COVID‐19) on the heart; 3 patients were enrolled in a study examining heart failure (HF) with recovered LVEF; 2 subjects were volunteers. In addition, 1 patient (age = 38 years, male) with pulmonary hypertension, dilated right ventricle, and right ventricular dysfunction was enrolled to acquire stress‐rest perfusion data, and another subject (age = 46 years, male) was scanned to collect resting perfusion data to determine the lower limit of TS that achieves a good balance between linearity and SNR. Our justification for mixing data from different patient cohorts is based on our stated goal to identify an optimal TS for minimizing the underestimation of AIF. Other clinical profiles were omitted because they were irrelevant to this technical study.
MRI hardware
MRI was performed on one 1.5 T whole‐body MRI scanner (MAGNETOM Aera, Siemens Healthineers, Erlangen, Germany) equipped with a gradient system capable of achieving a maximum gradient strength of 45 mT m−1 and a slew rate of 200 T m−1 s−1. MRI signal reception was made using anterior and posterior coil arrays with 30 total coil elements.
Pulse sequence
Relevant imaging parameters for our 2D multi‐slice cardiac perfusion pulse sequence using gradient echo readout with radial k‐space sampling included: FOV = 384 mm × 384 mm, matrix size =192 × 192, spatial resolution = 2 mm × 2 mm, slice thickness = 8 mm, TE/TR = 1.5/2.8 ms, flip angle = 15°, minimum TS = 10 ms, B1‐insensitive hybrid pulse train as the saturation pulse,
38 radial spokes per frame (corresponding to an acceleration factor of 5), single‐shot readout duration per frame = 106 ms, 100 repetitions, electrocardiogram triggering every heartbeat, and 4‐7 slices per heartbeat depending on heart rate. Each patient was instructed to breathe normally during scanning. For more details on the pulse sequence, please see Naresh et al.
Each perfusion scan was performed with administration of either 0.075 (estimated glomerular filtration rate [eGFR] = 45‐60) or 0.1 mmol kg−1 (eGFR > 60) of gadobutrol (Gadavist, Bayer HealthCare, Whippany, USA) at 3 mL s−1 via a power injector.
Minimum TS determination
We elected minimum TS as 10 ms to achieve a good balance between linearity and SNR. To support our choice, we conducted a theoretical analysis and scanned a T1mes phantom and one additional subject at rest to identify the lower bound of TS that achieves a good balance between linearity and SNR. For details, please see Minimum TS experiment in Supporting Information.
Monte Carlo simulation: SNR for TF images
Intrinsic SNR is a critical ingredient for achieving high accuracy in MBF quantification. We performed a theoretical analysis to estimate the intrinsic SNR of TF images, to justify KWIC filtering to achieve TS of 113.6 ms for TF images. For additional details, please see Monte Carlo analysis in Supporting Information.
KWIC filters
In radial k‐space sampling, the center of k‐space is sampled by each radial spoke. This provides an opportunity to retrospectively choose an arbitrary TS by applying KWIC filters. For myocardial wall assessment, images were reconstructed using a KWIC filter to exclude the first 19 radial spokes and maintain the center of the last radial spoke only (i.e., effective TS = 113.6 ms as shown in Figure 1) to maximize the SNR (see the Monte Carlo analysis in Supporting Information). To determine an optimal TS for minimizing the underestimation of AIF, five different KWIC filtered images were reconstructed from the same raw k‐space data by maintaining the center of k‐space from the first to fifth radial spoke only (i.e., effective TS from 10‐21.2 ms [2.8 ms steps] as shown in Figure 1). Because the expected AIF signal is relatively low for the first to fifth radial spokes, we zeroed out the outer one third of k‐space in the frequency‐encoding direction, to reduce noise and aliasing artifacts.
FIGURE 1
(A) Raw k‐space data; KWIC filters used for TF (B) and AIF (C) image reconstruction. For TF, the effective TS after the particular KWIC was 113.6 (= 10 + 2.8 × 37) ms. For AIF, five different KWIC filters were applied to generate five different images with effective TS ranging from 10 to 21.2 ms (2.8 ms steps). (D) k‐Space trajectories without and with KWIC filters. The darkest blue line represents the first radial spoke. The color bar indicates the radial spoke number following the saturation pulse (i.e., k‐space ordering) or effective TS. KWIC, k‐space‐weighted image contrast; TF, tissue function; AIF, arterial‐input‐function; TS, saturation‐recovery time
(A) Raw k‐space data; KWIC filters used for TF (B) and AIF (C) image reconstruction. For TF, the effective TS after the particular KWIC was 113.6 (= 10 + 2.8 × 37) ms. For AIF, five different KWIC filters were applied to generate five different images with effective TS ranging from 10 to 21.2 ms (2.8 ms steps). (D) k‐Space trajectories without and with KWIC filters. The darkest blue line represents the first radial spoke. The color bar indicates the radial spoke number following the saturation pulse (i.e., k‐space ordering) or effective TS. KWIC, k‐space‐weighted image contrast; TF, tissue function; AIF, arterial‐input‐function; TS, saturation‐recovery time
Image reconstruction
We applied KWIC filters as a pre‐processing step before compressed sensing
reconstruction of AIF and TF images from the same raw k‐space data. To accelerate image reconstruction, coil compression was used to produce 8 virtual coils,
and gpuNUFFT
was used throughout the reconstruction pipeline. The multi‐coil k‐space datasets were reconstructed using compressed sensing with temporal total variation and temporal PCA (TPCA) as the 2 orthogonal sparsifying transforms and nonlinear conjugate gradient with backtracking line search as the optimization algorithm with 30 iterations. Normalized regularization weight for temporal total variation was set as 0.005 (0.003 for TF images) of the maximum signal of the reconstructed image in each iteration, and normalized regularization weight for TPCA was set as one quarter of temporal total variation. We established the optimal regularization weights by sweeping over a range from 0.001 to 0.01 (0.001 steps) to identify an optimal regularization weight that achieves a good balance between suppression of aliasing artifacts and temporal blurring of myocardial wall motion. We determined this optimal regularization weight based on visual inspection of 6 training datasets. For more details on the image reconstruction pipeline, please see reference.
MBF quantification
Image processing was conducted by one investigator (L.F.). Five pixel‐by‐pixel MBF maps (same TF image, five different AIF images) were quantified for each slice using the following steps as shown in Figure 2. First, the motion correction was conducted on the TF images using a previously described method.
The same deformation field from the TF image was then applied to all 5 AIF images to ensure consistency. Second, T1‐weighted images were normalized by the corresponding PD‐weighted images to correct for the unknown equilibrium of magnetization and the surface‐coil effects.
Third, the normalized signal was converted to [Gd] using the Bloch equation described by Mendes et al.,
using 5.3 and 5.2 mM−1 s−1 as γ1 of gadobutrol in the blood and myocardium, respectively,
and assuming fast water exchange.
The AIF signal‐time curves were derived by manually drawing a region of interest inside the left ventricular cavity. For consistency, the same region of interest was used for all five reconstructed AIF images. Fourth, T2* decay in the AIF was corrected using the previously described theoretical approach.
Fifth, pixel‐wise MBF maps were quantified using a Fermi function
as the transfer function to deconvolve the myocardial perfusion [Gd]‐time curves with the AIF. To speed up data processing, we cropped the field of view to exclude extra cardiac tissues. For more details on the quantification of MBF, please see Naresh et al.
FIGURE 2
A schematic overview of the image reconstruction and processing pipeline for radial perfusion data obtained during free breathing. The sequence of steps includes: KWIC filtering of the radial k‐space data, compressed sensing reconstruction of undersampled and filtered images, motion correction, signal normalization, [Gd] calculation (T2* correction to the AIF), Fermi‐model deconvolution for pixel‐by‐pixel MBF quantification. MOCO, motion correction; KWIC, k‐space‐weighted image contrast; [Gd], gadolinium concentration; AIF, arterial‐input‐function; MBF, myocardial blood flow
A schematic overview of the image reconstruction and processing pipeline for radial perfusion data obtained during free breathing. The sequence of steps includes: KWIC filtering of the radial k‐space data, compressed sensing reconstruction of undersampled and filtered images, motion correction, signal normalization, [Gd] calculation (T2* correction to the AIF), Fermi‐model deconvolution for pixel‐by‐pixel MBF quantification. MOCO, motion correction; KWIC, k‐space‐weighted image contrast; [Gd], gadolinium concentration; AIF, arterial‐input‐function; MBF, myocardial blood flow
Statistical analysis
The statistical analyses were conducted by one investigator (L.F.). A Lilliefors test was performed to test the null hypothesis that both peak AIF and mean resting MBF is normally distributed at the 5% significance level. We compared both the peak AIF and mean resting MBF values calculated from the AIF images reconstructed with five different effective TS using repeated measures ANOVA with Bonferroni correction as the post hoc test for rest perfusion data. A P < 0.05 was considered statistically significant.
RESULTS
According to the Lilliefors test, both the peak AIF and resting MBF were normally distributed (statistic: [0.05, 0.2]; P > 0.05) for the five different effective TS groups.Figure 3A shows theoretical plots of the normalized signal as a function of ground truth [Gd] for five different TS values (10 ms, 12.8 ms, 15.6 ms, 18.4 ms, and 21.2 ms). As shown, shorter TS achieves a higher degree of linearization between the normalized signal and [Gd] at the expense of lower normalized signal (i.e., SNR). Figure 3B shows actual plots of the AIF signal‐time curves and the corresponding AIF [Gd]‐time curves without and with T2* correction from one representative patient. Although shorter TS produced lower signal (i.e., SNR), it resulted in higher AIF because of better linearity compared with longer TS values. T2* correction does not impact the overall comparison across TS. Figure 3C shows the corresponding resting MBF maps; MBF increased with effective TS because of underestimation of AIF.
FIGURE 3
(A) Theoretical plots describing nonlinear relationships between the normalized signal and [Gd] for five different TS values. (B) Example plots of the AIF signal‐time curves and AIF [Gd]‐time curves without and with T2* correction from one representative patient. (C) The corresponding resting MBF maps from five different AIFs with mean and SD values as shown. [Gd], gadolinium concentration; TS, saturation‐recovery time; AIF, arterial‐input‐function; MBF, myocardial blood flow
(A) Theoretical plots describing nonlinear relationships between the normalized signal and [Gd] for five different TS values. (B) Example plots of the AIF signal‐time curves and AIF [Gd]‐time curves without and with T2* correction from one representative patient. (C) The corresponding resting MBF maps from five different AIFs with mean and SD values as shown. [Gd], gadolinium concentration; TS, saturation‐recovery time; AIF, arterial‐input‐function; MBF, myocardial blood flowFour subjects were excluded from AIF and resting MBF quantifications because of poor image quality or motion correction. As summarized in Table 1 (n = 14), compared to TS = 10 ms, higher TS resulted in significantly decreased peak AIF as well as significantly increased MBF (both P < 0.05). Specifically, the peak AIF was 13.73 ± 2.25, 11.24 ± 1.77, 9.56 ± 1.45, 8.55 ± 1.32, and 7.95 ± 1.24 mM for TS 10, 12.8, 15.6, 18.4, and 21.2 ms, respectively; the corresponding resting MBF was 0.73 ± 0.12, 0.89 ± 0.15, 1.05 ± 0.17, 1.17 ± 0.19, and 1.27 ± 0.21 mL g−1 min−1 for TS 10, 12.8, 15.6, 18.4, and 21.2 ms, respectively.
TABLE 1
A statistical summary of peak AIF and resting MBF values from 14 subjects
TS
10 ms
12.8 ms
15.6 ms
18.4 ms
21.2 ms
AIF (mM)
13.73 ± 2.25
11.24 ± 1.77*
9.56 ± 1.45*
8.55 ± 1.32*
7.95 ± 1.24*
MBF (mL g−1 min−1)
0.73 ± 0.12
0.89 ± 0.15*
1.05 ± 0.17*
1.17 ± 0.19*
1.27 ± 0.21*
Note: Reported values represent mean ± SD. Compared with TS = 10 ms as the control, both AIF and MBF values for different TS values were significantly different.
P < 0.05.
A statistical summary of peak AIF and resting MBF values from 14 subjectsNote: Reported values represent mean ± SD. Compared with TS = 10 ms as the control, both AIF and MBF values for different TS values were significantly different.P < 0.05.Figure 4A shows actual plots of the AIF [Gd]‐time curves with T2* correction and the corresponding MBF maps during stress for different TS values; Figure 4B shows the corresponding data at rest; Figure 4C shows the myocardial perfusion reserve (MPR) values. Consistent with the remaining resting perfusion data, shorter TS reduced underestimation of AIF (or overestimation of MBF) for both stress and rest data. However, MPR did not change significantly with TS, because the impact of underestimation of AIF on MBF is canceled in MPR.
FIGURE 4
(A) Actual plots of the AIF [Gd]‐time curves with T2* correction and the corresponding MBF maps during stress for different TS values; (B) the corresponding data at rest; (C) the MPR values. AIF, arterial‐input‐function; [Gd], gadolinium concentration; MBF, myocardial blood flow; TS, saturation‐recovery time; MPR, myocardial perfusion reserve
(A) Actual plots of the AIF [Gd]‐time curves with T2* correction and the corresponding MBF maps during stress for different TS values; (B) the corresponding data at rest; (C) the MPR values. AIF, arterial‐input‐function; [Gd], gadolinium concentration; MBF, myocardial blood flow; TS, saturation‐recovery time; MPR, myocardial perfusion reserve
DISCUSSION
This study highlights the importance of reducing TS as short as 10 ms for reducing the underestimation of AIF and the overestimation of MBF. Although previous dual‐imaging studies
,
,
,
used a short TS ranging from 20‐27 ms, this work represents the first study comparing multiple TS values (10‐21.2 ms) to determine an optimal TS (10 ms) for further minimizing the underestimation of AIF.This study has several interesting points. First, we acquired resting perfusion data using radial k‐space sampling with a minimum TS of 10 ms and retrospectively reconstructing five different AIF images derived from the same radial k‐space data with TS ranging from 10 to 21.2 ms (2.8 ms steps) using KWIC filters. This experimental design enabled us to identify an optimal TS for AIF without the complexity of multiple injections of gadolinium as well as image acquisitions. Second, to reduce variables, we applied the same deformation fields to all five AIF images to correct for respiratory motion, the same region of interest to obtain the AIF signal‐time curves, and the same TF image with matching cardiac contours. We found the peak AIF and resting MBF values were significantly different between the five TS values. Given the lack of ground truth for AIF and MBF, we infer literature values. The resting MBF (0.73 mL g−1 min−1) calculated from AIF with TS of 10 ms agrees better with the resting MBF reported by two cardiac PET studies
: (1) mean resting MBF = 0.71 mL g−1 min−1 in 363 healthy subjects using 13N‐ammonia and (2) mean resting MBF = 0.74 mL g−1 min−1 in 382 healthy subjects using 82Rb. We note that this comparison across different cohorts and modalities should be interpreted with caution, because there are numerous unaccounted variables. Third, compared with TS of 21.2 ms, which is similar to TS of 23.8 ms proposed by the widely used quantitative cardiac perfusion sequence,
TS of 10 ms increased AIF by 73% and decreased resting MBF by 43%. Our finding suggests the pulse sequence proposed by Kellman et al
may overestimate MBF because of the underestimation of AIF. This problem is fixable by redesigning the pulse sequence to use TS of 10 ms for AIF scanning. Fourth, our finding suggests that the impact of the underestimation of AIF may be negligible for calculation of MPR, because the underestimation/overestimation effect on AIF/MBF is canceled in MPR. However, more data from greater number of patients are needed to validate our initial finding.This study has several limitations that warrant further discussion. First, this study applied KWIC filtering to retrospectively select an effective TS. KWIC filters zero out the central parts of k‐space, but they retain signal from other parts of k‐space. Our signal model does not account for these residual signals, which may result in a bias in [Gd] calculation. The potential bias should be consistent across the five TS values because we applied similarly shaped KWIC filters, suggesting that the observed trend across TS should be valid. An alternative approach is to perform multiple TS acquisitions using a Cartesian k‐space sampling pulse sequence. However, such an approach is impractical because it necessitates multiple injections of gadolinium, requires compensation of residual gadolinium from previous injections, compensates for variations in physiology (e.g., heart rate, breathing pattern) and manual data analysis (segmentation), and increased safety concerns associated with gadolinium. Second, this study did not evaluate whether the observed trend is maintained at 3 T. Although there is no theoretical basis to expect a different trend, a future 3 T study is warranted for verification. Third, this study lacks ground truths for AIF and MBF to truly validate our approach. This is a common problem for all quantitative cardiac first‐pass perfusion studies in patients, unless one is able to conduct a PET‐MR study on the same day, which has its own challenges (limited available of equipment and radiotracer). Given these challenges, we elected to compare our results to 2 previous PET studies. Fourth, this study only included 1 patient scan during stress testing. More data from a greater number of patients during stress testing are needed to validate our findings.In summary, we conclude that an AIF scan with TS of 10 ms minimizes the underestimation of AIF and, hence, minimizes the overestimation of MBF compared with TS from 20 to 27 ms.
CONFLICT OF INTEREST
None of the authors have relationships with industry related to this study.Figure S1 (A) T1 maps of T1mes phantom calculated acquired using three different TS values; (B) the corresponding difference maps, where TS = 278 ms is the reference.Figure S2 (A) Five different KWIC filtered CS reconstructed images by maintaining the center from the first to the fifth radial spoke only (i.e., effective TS from 5 to 16.2 ms [2.8 ms steps]); (B) corresponding PD normalized images; (C) theoretical plots describing the nonlinear relationships between the normalized signal and [Gd] for five different TS values; (D) plots of the AIF signal‐time curves from one patient acquired with minimal TS = 5 ms.Click here for additional data file.
Authors: Timothy F Christian; Dan W Rettmann; Anthony H Aletras; Steve L Liao; Joni L Taylor; Robert S Balaban; Andrew E Arai Journal: Radiology Date: 2004-07-29 Impact factor: 11.105
Authors: Neil Chatterjee; Brandon C Benefield; Kathleen R Harris; Jacob U Fluckiger; Timothy Carroll; Daniel C Lee Journal: Magn Reson Med Date: 2016-09-08 Impact factor: 4.668
Authors: Amit R Patel; Patrick F Antkowiak; Kiran R Nandalur; Amy M West; Michael Salerno; Vishal Arora; John Christopher; Frederick H Epstein; Christopher M Kramer Journal: J Am Coll Cardiol Date: 2010-08-10 Impact factor: 24.094
Authors: Nivedita K Naresh; Hassan Haji-Valizadeh; Pascale J Aouad; Matthew J Barrett; Kelvin Chow; Ann B Ragin; Jeremy D Collins; James C Carr; Daniel C Lee; Daniel Kim Journal: Magn Reson Med Date: 2018-11-12 Impact factor: 4.668
Authors: Manesh R Patel; Eric D Peterson; David Dai; J Matthew Brennan; Rita F Redberg; H Vernon Anderson; Ralph G Brindis; Pamela S Douglas Journal: N Engl J Med Date: 2010-03-11 Impact factor: 91.245
Authors: Federico E Mordini; Tariq Haddad; Li-Yueh Hsu; Peter Kellman; Tracy B Lowrey; Anthony H Aletras; W Patricia Bandettini; Andrew E Arai Journal: JACC Cardiovasc Imaging Date: 2014-01
Authors: Lexiaozi Fan; Kyungpyo Hong; Li-Yueh Hsu; James C Carr; Bradley D Allen; Daniel C Lee; Daniel Kim Journal: Magn Reson Med Date: 2022-04-04 Impact factor: 3.737