Performance Characteristics of a Dual-Sided Position-Sensitive Sparse-Sensor Detector for Gamma-ray Imaging.

William C J Hunter1, Donald Q DeWitt1, Robert S Miyaoka1.   

Abstract

Purpose: We characterize the performance of a dualsided position-sensitive sparse sensor (DS-PS3) array detector for positron emission tomography (PET). The DS-PS3 detector is designed as a high performance, cost effective PET detector for organ-specific imaging systems (e.g., brain, breast, etc.).
Methods: Two sparse 4-by-4 arrays of silicon photomultipliers (18.5% SiPM fill-factor) coupled through segmented light guide are used to readout a 15-by-15 array of 2-mm-pitch, 20-mm-long LSYO crystals. Uniform flood data were used for crystal identification, depth determination, and position-dependent energy resolution. Intrinsic-spatial and depth-of-interaction (DOI) resolutions were determined by stepping a collimated gamma-ray source over the front and side, respectively.
Results: We measured an average intrinsic spatial resolution of 2.14 ± 0.07 mm full width at half maximum (FWHM). DOI FWHM resolution varied from 2.2 mm for crystals over sensors to 5.3 mm for crystals between sensors. Average DOI resolution was 3.6 ± 0.8 mm FHWM. Average energy resolution for the detector module was 16.6% with a range of 11.3% to 25.8%. Conclusions: We have demonstrated use of a dual-sided sparse sensor arrays to enable low-cost high-performance decoding of three-dimensional positioning within a PET detector using an 18.5% sensor fill-factor.

Entities:  

Year:  2021        PMID: 35372738      PMCID: PMC8974312          DOI: 10.1109/trpms.2021.3087465

Source DB:  PubMed          Journal:  IEEE Trans Radiat Plasma Med Sci        ISSN: 2469-7303


  23 in total

1.  Development and Initial Results of a Brain PET Insert for Simultaneous 7-Tesla PET/MRI Using an FPGA-Only Signal Digitization Method.

Authors:  Jun Yeon Won; Haewook Park; Seungeun Lee; Jeong-Whan Son; Yina Chung; Guen Bae Ko; Kyeong Yun Kim; Junghyun Song; Seongho Seo; Yeunchul Ryu; Jun-Young Chung; Jae Sung Lee
Journal:  IEEE Trans Med Imaging       Date:  2021-06-01       Impact factor: 10.048

2.  Design and evaluation of the MAMMI dedicated breast PET.

Authors:  L Moliner; A J Gonzalez; A Soriano; F Sanchez; C Correcher; A Orero; M Carles; L F Vidal; J Barbera; L Caballero; M Seimetz; C Vazquez; J M Benlloch
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

3.  Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards.

Authors:  Kanae K Miyake; Keiichi Matsumoto; Mika Inoue; Yuji Nakamoto; Shotaro Kanao; Tae Oishi; Shigeto Kawase; Keishi Kitamura; Yoshiyuki Yamakawa; Ayako Akazawa; Tetsuya Kobayashi; Junichi Ohi; Kaori Togashi
Journal:  J Nucl Med       Date:  2014-05-08       Impact factor: 10.057

4.  National Electrical Manufacturers Association and Clinical Evaluation of a Novel Brain PET/CT Scanner.

Authors:  Kira S Grogg; Terrence Toole; Jinsong Ouyang; Xuping Zhu; Marc D Normandin; Quanzheng Li; Keith Johnson; Nathaniel M Alpert; Georges El Fakhri
Journal:  J Nucl Med       Date:  2015-12-23       Impact factor: 10.057

5.  TRIMAGE: A dedicated trimodality (PET/MR/EEG) imaging tool for schizophrenia.

Authors:  Alberto Del Guerra; Salleh Ahmad; Mihai Avram; Nicola Belcari; Arne Berneking; Laura Biagi; Maria Giuseppina Bisogni; Felix Brandl; Jorge Cabello; Niccolò Camarlinghi; Piergiorgio Cerello; Chang-Hoon Choi; Silvia Coli; Sabrina Colpo; Julien Fleury; Vito Gagliardi; Giuseppe Giraudo; Karsten Heekeren; Wolfram Kawohl; Theodora Kostou; Jean-Luc Lefaucheur; Christoph Lerche; George Loudos; Matteo Morrocchi; Julien Muller; Mona Mustafa; Irene Neuner; Panagiotis Papadimitroulas; Francesco Pennazio; Ravichandran Rajkumar; Cláudia Régio Brambilla; Julien Rivoire; Elena Rota Kops; Jürgen Scheins; Rémy Schimpf; N Jon Shah; Christian Sorg; Giancarlo Sportelli; Michela Tosetti; Riccardo Trinchero; Christine Wyss; Sibylle Ziegler
Journal:  Eur Psychiatry       Date:  2018-02-01       Impact factor: 5.361

6.  Effects of Detector Thickness on Geometric Sensitivity and Event Positioning Errors in the Rectangular PET/X Scanner.

Authors:  Lawrence R MacDonald; William C J Hunter; Paul E Kinahan; Robert S Miyaoka
Journal:  IEEE Trans Nucl Sci       Date:  2013-10       Impact factor: 1.679

Review 7.  Advances in multimodal neuroimaging: hybrid MR-PET and MR-PET-EEG at 3 T and 9.4 T.

Authors:  N Jon Shah; Ana-Maria Oros-Peusquens; Jorge Arrubla; Ke Zhang; Tracy Warbrick; Jörg Mauler; Kaveh Vahedipour; Sandro Romanzetti; Jörg Felder; Avdo Celik; Elena Rota-Kops; Hidehiro Iida; Karl-Josef Langen; Hans Herzog; Irene Neuner
Journal:  J Magn Reson       Date:  2012-12-10       Impact factor: 2.229

8.  PET characteristics of a dedicated breast PET/CT scanner prototype.

Authors:  Yibao Wu; Spencer L Bowen; Kai Yang; Nathan Packard; Lin Fu; George Burkett; Jinyi Qi; John M Boone; Simon R Cherry; Ramsey D Badawi
Journal:  Phys Med Biol       Date:  2009-06-17       Impact factor: 3.609

9.  Evaluation of [18F]-4-fluoroantipyrine as a new blood flow tracer for multiradionuclide autoradiography.

Authors:  K Sako; M Diksic; A Kato; Y L Yamamoto; W Feindel
Journal:  J Cereb Blood Flow Metab       Date:  1984-06       Impact factor: 6.200

10.  Evaluation of primary breast cancers using dedicated breast PET and whole-body PET.

Authors:  Deep K Hathi; Wen Li; Youngho Seo; Robert R Flavell; John Kornak; Benjamin L Franc; Bonnie N Joe; Laura J Esserman; Nola M Hylton; Ella F Jones
Journal:  Sci Rep       Date:  2020-12-14       Impact factor: 4.379

View more
  1 in total

Review 1.  Silicon photomultiplier signal readout and multiplexing techniques for positron emission tomography: a review.

Authors:  Haewook Park; Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-07-16
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.