Literature DB >> 33625980

Development and Initial Results of a Brain PET Insert for Simultaneous 7-Tesla PET/MRI Using an FPGA-Only Signal Digitization Method.

Jun Yeon Won, Haewook Park, Seungeun Lee, Jeong-Whan Son, Yina Chung, Guen Bae Ko, Kyeong Yun Kim, Junghyun Song, Seongho Seo, Yeunchul Ryu, Jun-Young Chung, Jae Sung Lee.   

Abstract

In study, we developed a positron emission tomography (PET) insert for simultaneous brain imaging within 7-Tesla (7T) magnetic resonance (MR) imaging scanners. The PET insert has 18 sectors, and each sector is assembled with two-layer depth-of-interaction (DOI)-capable high-resolution block detectors. The PET scanner features a 16.7-cm-long axial field-of-view (FOV) to provide entire human brain images without bed movement. The PET scanner early digitizes a large number of block detector signals at a front-end data acquisition (DAQ) board using a novel field-programmable gate array (FPGA)-only signal digitization method. All the digitized PET data from the front-end DAQ boards are transferred using gigabit transceivers via non-magnetic high-definition multimedia interface (HDMI) cables. A back-end DAQ system provides a common clock and synchronization signal for FPGAs over the HDMI cables. An active cooling system using copper heat pipes is applied for thermal regulation. All the 2.17-mm-pitch crystals with two-layer DOI information were clearly identified in the block detectors, exhibiting a system-level energy resolution of 12.6%. The PET scanner yielded clear hot-rod and Hoffman brain phantom images and demonstrated 3D PET imaging capability without bed movement. We also performed a pilot simultaneous PET/MR imaging study of a brain phantom. The PET scanner achieved a spatial resolution of 2.5 mm at the center FOV (NU 4) and a sensitivity of 18.9 kcps/MBq (NU 2) and 6.19% (NU 4) in accordance with the National Electrical Manufacturers Association (NEMA) standards.

Entities:  

Mesh:

Year:  2021        PMID: 33625980     DOI: 10.1109/TMI.2021.3062066

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  5 in total

1.  A time-based single transmission-line readout with position multiplexing.

Authors:  Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-01-17

2.  Performance Characteristics of a Dual-Sided Position-Sensitive Sparse-Sensor Detector for Gamma-ray Imaging.

Authors:  William C J Hunter; Donald Q DeWitt; Robert S Miyaoka
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-06-07

3.  Investigation of a Model-based Time-over-threshold Technique for Phoswich Crystal Discrimination.

Authors:  Haithem Bouziri; Catherine M Pepin; Konin Koua; Maher Benhouria; Caroline Paulin; Jinsong Ouyang; Marc Normandin; Jean-François Pratte; Georges El Fakhri; Roger Lecomte; Réjean Fontaine
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-05-04

4.  Monte Carlo Characterization of the Trimage Brain PET System.

Authors:  Luigi Masturzo; Pietro Carra; Paola Anna Erba; Matteo Morrocchi; Alessandro Pilleri; Giancarlo Sportelli; Nicola Belcari
Journal:  J Imaging       Date:  2022-01-23

Review 5.  Silicon photomultiplier signal readout and multiplexing techniques for positron emission tomography: a review.

Authors:  Haewook Park; Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-07-16
  5 in total