| Literature DB >> 35365884 |
Madeline E Carr1,2,3, Kathryn E Keenan4, Robba Rai2,3,5, Michael A Boss6, Peter Metcalfe1,2, Amy Walker1,2,3,5, Lois Holloway1,2,3,5,7.
Abstract
PURPOSE: To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice-encoded imaging direction and the reference sample ADC value.Entities:
Keywords: apparent diffusion coefficient; diffusion-weighted imaging; quantitative imaging biomarker alliance; quantitative magnetic resonance imaging; reproducibility
Mesh:
Substances:
Year: 2022 PMID: 35365884 PMCID: PMC9543906 DOI: 10.1002/mp.15645
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.506
FIGURE 1Repeatability coefficient of variation (CVST), precision (CVP) and reproducibility (CVLT) derived for each vial for axial acquisitions, calculated as an average over the 12‐month study. A vial arrangement diagram has been included in the top left for positional reference for all 13 vials. PVP concentrations (conc.) are shown (by mass fraction (%)) for inner‐ and outer‐ring vials. Note that the central vial is to be positioned at isocenter and contains distilled water
Overview of tests completed to assess conformance to the Profile
|
|
|
| |
|---|---|---|---|
|
|
| DCT | = True diffusion coefficient |
| μ | = Mean of measurements | ||
|
|
| RC | = Repeatability coefficient |
|
| CV | = Coefficient of variation | |
| SD | = Standard deviation | ||
| ST | = Over 4 × short‐term measurements | ||
| LT | = Over 12 × long‐term measurements | ||
|
|
AND
|
| = Coefficient of determination |
| RSS | = Sum of squares of residuals | ||
| TSS | = Total Sum of Squares | ||
| Y | = Measured ADC (all vials/months) | ||
| β0 | = Intercept | ||
| β1 | = Slope | ||
|
|
|
| = Region of interest (isocenter vial) |
|
| = Over ADC values within the ROI | ||
|
|
| SNR | = Signal to noise ratio |
| Signal | = Average of pixel values for each ROI over the 4 x ST repetitions | ||
| Noise | = Average of pixel SD values for each ROI over the 4 x ST repetitions | ||
|
|
| Dep
| = |
|
| = ADC generated using | ||
|
| = 500, 900, or 2000 s/mm2 | ||
Note: Excluding test D, Profile testing was only required to be performed on the central water vial (at isocenter) using axial acquisitions. Further, short‐term (ST) refers to the intraday measurements acquired, while long‐term (LT) refers to the intramonth measurements acquired
Accuracy, repeatability, reproducibility, linearity, random error, SNR, and b‐value dependence (tests A →G) tolerance limits and mean value ± SD (where applicable), obtained from the 12 monthly measurements of the central water vial (as per Profile requirements)
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
| |bias (%)| | ≤ 3.60 | +0.05 ± 0.01 | +0.83 ± 0.00 | +0.29 ± 0.01 |
|
| RCST (μm2/ms) | ≤ 0.015 | 0.003 ± 0.001 | 0.005 ± 0.002 | 0.003 ± 0.001 |
|
| ≤0.5 | 0.1 ± 0.0 | 0.1 ± 0.1 | 0.1 ± 0.0 | |
|
| RCLT (μm2/ms) | ≤0.065 | 0.028 | 0.011 | 0.027 |
|
| ≤ 2.2 | 0.9 | 0.3 | 0.9 | |
|
|
| > 0.9 | 1.0 | 1.0 | 1.0 |
|
| 0.95 ≤ | 1.00 | 1.02 | 1.02 | |
|
| CVP (%) | <2 | 0.38 ± 0.10 | 0.43 ± 0.04 | 0.38 ± 0.04 |
|
| SNRa | ≥ 50 ± 5 | 332 ± 146 | 269 ± 93 | 356 ± 68 |
|
| Dep
| <2 | 0.4 ± 0.3 | 0.3 ± 0.2 | 1.4 ± 2.1 |
aCertain month's data have been excluded from the presented results due to retrospective findings of signal saturation occurring within the data sets. For SNR, this included excluding axial results acquired for months 1, 2, 6, and 9, and coronal and sagittal results for months 1 and 5, and 1 and 6, respectively. For Dep , results for month 1 were excluded in calculations for all directions
FIGURE 2Bland–Altman plots for the central water vial (reference ADC = 1.109 μm2/ms) coronal (a), sagittal (b), and axial (c) acquisitions show the difference (%bias) between measured and reference ADC values over the 12‐month study. The average %bias (and 95% confidence intervals) is displayed and includes +0.830% (+0.724 to +0.936), +0.288% (+0.042 to +0.534), and +0.053% (–0.204 to +0.309) for coronal, sagittal, and axial acquisitions, respectively
FIGURE 3Correlation plots for coronal (a), sagittal (b), and axial (c) acquisitions with measured ADC values compared to the NIST reference ADC values for each vial. Note that all axes have employed a logarithmic scale and error bars are given as the standard deviation found between all 12 months ADC values