Literature DB >> 35310443

Accuracy of the Horizontal Calibrator in Correcting Leg Length and Restoring Femoral Offset in Total Hip Arthroplasty.

Xing Chen1, Shuxing Xing1, Zhiyong Zhu2, Huisheng Wang2, Zhongshen Yu2, Xizhuang Bai2, Xi Li2.   

Abstract

Background: Limb length discrepancy (LLD) is one of the most common postoperative complications and can cause serious consequences. Poor recovery of femoral offset (OD) will result in weakness of the patient's external rotator muscles and affect the patient's postoperative function. The study is aimed to present a simple approach that compensates for the shortcomings of previous measuring devices and combines the advantages of different measuring devices to provide more accurate limb length and femoral offset restoration in total hip arthroplasty (THA).
Methods: This study was a prospective controlled trial involving 89 patients with THA. Group I (n = 44) was used for intraoperative measurement of THA with our self-designed horizontal calibrator. Group II (n = 45) was measured by a traditional freehand technique. The main outcome indicators were measured on the Neusoft PACS, including LLD, femoral offset deviation, and operative time. IBM SPSS 23.0 was used for data analysis.
Results: The independent sample t-test was performed for all the data. The operative time, preoperative radiographic LLD, and OD of Group I and Group II had no statistical significance. Postoperative LLD of Group I and Group II were 2.5 ± 2.1 mm (range -5.7 to 8.3 mm) and 6.2 ± 4.3 mm (range -18.0 to 15.2 mm), and the independent sample t-test data of both (P < 0.001; 95%CI = -5.1, -2.2) showed statistical significance. In Group I, there were 38 THAs with LLD <5 mm, accounting for 86% and there were 44 THAs with LLD <10 mm, accounting for 100%. In Group II, there were 20 THAs with LLD <5 mm, accounting for 44%. There were 36 THAs with LLD <10 mm, covering for 80%. There was no significant difference in postoperative femoral offset and OD.
Conclusion: The horizontal calibrator can provide more accurate limb length and femoral offset recovery in THA. It is a simple surgical technique that does not add additionally surgical costs and does not significantly increase operative time, providing a new solution for surgeons to resolve postoperative LLD and restore femoral offset.
Copyright © 2022 Chen, Xing, Zhu, Wang, Yu, Bai and Li.

Entities:  

Keywords:  calibrator; hip arthroplasty; intraoperative; leg length discrepancy; offset

Year:  2022        PMID: 35310443      PMCID: PMC8927056          DOI: 10.3389/fsurg.2022.845364

Source DB:  PubMed          Journal:  Front Surg        ISSN: 2296-875X


  40 in total

1.  L-shaped caliper for limb length measurement during total hip arthroplasty.

Authors:  K Shiramizu; M Naito; T Shitama; Y Nakamura; H Shitama
Journal:  J Bone Joint Surg Br       Date:  2004-09

2.  A simple intraoperative technique to avoid limb length discrepancy in total hip arthroplasty.

Authors:  Aravind Desai; Bilal Barkatali; Asterios Dramis; Timothy N Board
Journal:  Surgeon       Date:  2010-02-11       Impact factor: 2.392

3.  Limb-length measurement in total hip arthroplasty using a calipers dual pin retractor.

Authors:  Iori Takigami; Mansho Itokazu; Yoshiki Itoh; Kazu Matsumoto; Takatoshi Yamamoto; Katsuji Shimizu
Journal:  Bull NYU Hosp Jt Dis       Date:  2008

4.  Leg length discrepancy after total hip arthroplasty: a review of literature.

Authors:  Aravind S Desai; Asterios Dramis; Tim N Board
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

5.  Use of intraoperative X-rays to optimize component position and leg length during total hip arthroplasty.

Authors:  Kace A Ezzet; Julie C McCauley
Journal:  J Arthroplasty       Date:  2013-09-26       Impact factor: 4.757

6.  Digital templating facilitates accurate leg length correction in total hip arthroplasty.

Authors:  Shivan Sabah Jassim; Christopher Ingham; Matthew Keeling; James Anthony Wimhurst
Journal:  Acta Orthop Belg       Date:  2012-06       Impact factor: 0.500

7.  A simple method of obtaining equal leg length in total hip arthroplasty.

Authors:  H M McGee; J H Scott
Journal:  Clin Orthop Relat Res       Date:  1985-04       Impact factor: 4.176

8.  A method of intraoperative limb length measurement in total hip arthroplasty.

Authors:  S T Woolson; W H Harris
Journal:  Clin Orthop Relat Res       Date:  1985-04       Impact factor: 4.176

9.  An intraoperative device to restore femoral offset in total hip arthroplasty.

Authors:  Enxing Xue; Zhen Su; Chengwang Chen; Paul Kim Chiu Wong; Hong Wen; Yu Zhang
Journal:  J Orthop Surg Res       Date:  2014-07-19       Impact factor: 2.359

10.  A Vertical Measurement System to Predict the Change in Leg Length in Total Hip Arthroplasty.

Authors:  Garth Grobler; Marc Nortje; Brendan Dower; Dave Chivers
Journal:  Arthroplast Today       Date:  2020-06-02
View more
  1 in total

1.  Effects of fascia iliaca compartment block as an adjunctive management to parecoxib for pain control after total hip arthroplasty.

Authors:  Xiao-Yan Li; Liang Zhang; Yu-Mei Ding; Cai-Xia Wang; Yi Qiu
Journal:  Medicine (Baltimore)       Date:  2022-07-29       Impact factor: 1.817

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.