| Literature DB >> 35297569 |
Andre Karius1,2, Christoph Bert1,2.
Abstract
The regular evaluation of imaging performance of computed tomography (CT) scanners is essential for CT quality assurance. For automation of this process, the software QAMaster was developed at the Universitätsklinikum Erlangen, which provides based on CT scans of the CatPhan® 504 (The Phantom Laboratory, Salem, USA) automated image quality analysis and documentation by evaluating CT number accuracy, spatial linearity, uniformity, contrast-noise-ratio, spatial resolution, noise, and slice thickness. Dose assessment is supported by calculations of the weighted computed tomography dose index (CTDIw ) and weighted cone beam dose index (CBDIw ). QAMaster was tested with CatPhan® 504 scans and compared to manual evaluations of these scans, whereby high consistency of the respective results was observed. The CT numbers, spatial linearity, uniformity, contrast-noise-ratio, noise, and slice thickness deviated by only (0.13 ± 0.25) HU, (0.02 ± 0.05) mm, (-0.01 ± 0.03)%, 0.8 ± 1.8, (0.131 ± 0.05) HU, and (0.004 ± 0.005) mm between both evaluations, respectively. The QAMaster results for spatial resolution did not differ significantly (p = 0.34) from the CatPhan® 504 based manual resolution assessment. Dose computations were fully consistent between QAMaster and manual calculations. Thus, QAMaster proved to be a comprehensive and functional software for performing an automated CT quality assurance routine. QAMaster will be open-source after its release.Entities:
Keywords: computed tomography; cone beam computed tomography; image quality assurance; phantom-based software framework
Mesh:
Year: 2022 PMID: 35297569 PMCID: PMC8992959 DOI: 10.1002/acm2.13588
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIGURE 1Computed tomography (CT)‐scans of the CatPhan sections CTP528 (a; used for MTF determination), CTP404 (b; contains inserts, rods and ramps for determining CT number accuracy, contrast‐noise‐ratio (CNR), spatial linearity, and reconstructed slice thickness) and CTP486 (c; used for evaluating uniformity and noise characteristics). The rods in CTP404 are labeled for distinctiveness with the numbers 1– 4. The non‐labeled insert at 6 o'clock position in (b) represents an additional, second air insert
FIGURE 2Main menu (after performing the quality assurance routine, QAR) of a computed tomography (CT) system in QAMaster, which gives a color‐coded overview of measured image parameters, baselines, and tolerances
Setting of tolerances by QAMaster considering the measured baselines. The tolerances were chosen based on the regulations given by EFOMP and the DIN 61223‐2‐66 for QA in CBCT and CT, respectively
|
|
|
|---|---|
| CT numbers | 50 HU |
| Distances between rods | 1 mm |
| Slice thickness | 0.5 mm |
| Uniformity indices | 10% |
| HWHM/HWTM of MTF, CNRs, noise | 20% |
Abbreviations: CNR, contrast‐noise‐ratio; CT, computed tomography; HWHM, half width half maximum; HWTM, half width tenth maximum; MTF, modulation transfer function; QA, quality assurance.
FIGURE 3Graphical output by QAMaster: Shown are the regions of interest (ROIs) for determining contrast‐noise‐ratio (CNR), computed tomography (CT) number accuracy (both a) and uniformity (b), the calculated point‐spread functions (c) and MTF (d) as well as the 2D (e) and 1D (f) noise‐power spectrum (NPS)
FIGURE 4Comparison of the modulation transfer functions (MTFs) calculated by QAMaster and with both human observer (HO) methods