Pre-Operative Clinical Profile Description Segmented by Treatment Combination Cohorts
| Combination 1 | Combination 2 | Combination 3 | Combination 4 | Combination 5 | Combination 6 | Combination 7 | |
|---|---|---|---|---|---|---|---|
|
| 6.392 ± 0.327 | 6.468 ± 0.535 | 6.265 ± 0.67 | 6.601 ± 0.605 | 6.835 ± 0.63 | 6.578 ± 0.667 | 6.46 ± 0.825 |
|
| 1.358 ± 0.487 | 1.016 ± 0.589 | 1.008 ± 0.565 | 0.695 ± 0.376 | 1.029 ± 0.504 | 1.677 ± 1.227 | 1.231 ± 0.594 |
|
| 6.233 ± 2.592 | 7.182 ± 1.736 | 5.029 ± 3.263 | 7.86 ± 2.601 | 4.634 ± 2.605 | 6.497 ± 3.1 | 4.102 ± 2.137 |
|
| 58.598 ± 4.542 | 59.033 ± 3.894 | 60.774 ± 9.65 | 56.792 ± 5.601 | 54.551 ± 4.99 | 58.501 ± 7.663 | 59.591 ± 7.922 |
|
| −4.521 ± 2.662 | −9.341 ± 6.506 | −9.006 ± 7.336 | −7.688 ± 5.889 | −5.016 ± 3.333 | −4.948 ± 4.509 | −5.964 ± 3.937 |
|
| −3.682 ± 1.031 | −3.2 ± 2.683 | −3.397 ± 1.775 | −5.852 ± 2.125 | −3.321 ± 1.007 | −6.479 ± 1.866 | −3.295 ± 2.395 |
|
| 0.488 ± 0.216 | 0.497 ± 0.236 | 0.571 ± 0.19 | 0.627 ± 0.145 | 0.669 ± 0.202 | 0.408 ± 0.173 | 0.575 ± 0.246 |
Combination 1: Single 160 degrees segment. Combination 2: Two symmetrical 160 degrees segments. Combination 3: Single 210 degrees segment. Combination 4: Two symmetrical 120 degrees segments. Combination 5: Single 150 degrees segment. Combination 6: Two asymmetrical segments: 150 degrees inferior and 90 degrees superior. Combination 7: Single 210 degrees segment.
Figure 1.Main optical aberrations and corneal volume changes induced by intracorneal ring segments (ICRS) surgery implanted within 5 mm versus 6 mm of treatment diameter. Coma (a), high order aberrations (b), volume (c) and astigmatism (d) were computed considering both, anterior and posterior corneal surfaces and processed by a virtual ray tracing model.
Both Corneal Surfaces Pre and Postoperative Main Values and Geometrical Changes Caused by ISCRS Surgery
| Corneal Surface | Variable |
| Pre-Operative | Post-Operative | Change |
|
|---|---|---|---|---|---|---|
|
|
| 124 | 57.97 ± 6.94 | 55.39 ± 6.10 | −2.58 (−3.08, −2.07) | <0.001 |
|
| 124 | 6.53 ± 0.64 | 6.91 ± 0.63 | 0.38 (0.32, 0.43) | <0.001 | |
|
| 124 | −1.53 ± 1.12 | −1.96 ± 1.36 | −0.43 (−0.63, −0.22) | <0.001 | |
|
|
| 124 | 4.97 ± 0.70 | 5.124 ± 0.67 | 0.156 (0.109, 0.204) | <0.001 |
|
| 124 | −1.57 ± 1.10 | −1.64 ± 1.16 | −0.062 (−0.211, 0.087) | 0.409 |
N, sample size; BFS, best fitting sphere.
Monochromatic Optical Aberrations Grouped by Class and Visual Performance Before and After ISCRS Surgery
| N | Pre | Post | Change |
| |
|---|---|---|---|---|---|
|
| 124 | 1.32 ± 0.91 | 0.99 ± 0.52 | −0.33 (−0.453, −0.202) | <0.001 |
|
| 124 | 1.03 ± 0.67 | 0.72 ± 0.43 | −0.31 (−0.407, −0.214) | <0.001 |
|
| 124 | 0.29 ± 0.26 | 0.32 ± 0.25 | 0.024 (−0.035, 0.083) | 0.418 |
|
| 124 | −4.97 ± 2.46 | −2.92 ± 1.91 | 2.05 (1.661, 2.437) | <0.001 |
|
| 124 | 0.049 ± 0.02 | 0.065 ± 0.041 | 0.016 (0.011, 0.022) | <0.001 |
|
| 124 | 0.57 ± 0.21 | 0.69 ± 0.21 | 0.125 (0.091, 0.159) | <0.001 |
|
| 124 | −6.71 ± 5.41 | −4.28 ± 4.09 | 2.434 (1.739, 3.130) | <0.001 |
|
| 124 | −4.31 ± 2.18 | −2.05 ± 1.55 | 2.263 (1.773, 2.753) | <0.001 |
N, sample size; RMS HOAs, root mean square of the high-order aberrations (µm); coma (µm); trefoil (µm); astigmatism (D); BCVA, best-corrected visual acuity; subjective refraction: sphere and cylinder (D).
Figure 2.Pre- and postoperative coma (a), high order aberrations (b), astigmatism (c), and Visual Strehl (d) comparison for each combination of intracorneal rings segments (ICRS). Combination 1: single 160 degrees, combination 2: two symmetrical 160 degrees, combination 3: single 210, combination 4: two symmetrical 120 degrees, combination 5: single 150 degrees, combination 6: 150 degrees superior and 90 degrees inferior, combination 7: single 210 degrees.
Geometrical and Aberrometric Changes After ICRS Surgery Induced by Each Combination of ICRS
| Combination 1 | Combination 2 | Combination 3 | Combination 4 | Combination 5 | Combination 6 | Combination 7 | |
|---|---|---|---|---|---|---|---|
|
| 0.45 ± 0.286** | 0.638 ± 0.256* | 0.539 ± 0.309* | 0.165 ± 0.133* | 0.293 ± 0.232* | 0.333 ± 0.536 | 0.383 ± 0.186* |
|
| 0.494 ± 0.338** | 0.571 ± 0.429** | 0.514 ± 0.298* | −0.351 ± 0.293* | 0.273 ± 0.298** | 0.214 ± 0.575 | 0.143 ± 0.452 |
|
| 0.511 ± 0.508 | 0.446 ± 0.454 | 0.558 ± 0.529** | 0.334 ± 0.309* | 0.199 ± 0.437 | 0.191 ± 0.714 | 0.320 ± 0.560 |
|
| −3.38 ± 2.355** | −4.78 ± 2.142* | −3.92 ± 3.269* | −1.91 ± 1.441* | −1.39 ± 2.708 | −2.32 ± 2.864 | −1.25 ± 3.291 |
|
| 0.325 ± 0.774 | −0.67 ± 1.455 | −0.046 ± 1.226 | −0.75 ± 0.682* | −0.344 ± 0.784 | −0.58 ± 1.047 | −0.99 ± 1.793 |
|
| 1.036 ± 1.722 | −0.78 ± 2.057 | −0.093 ± 1.075 | −0.99 ± 1.170** | −0.301 ± 0.972 | −0.73 ± 1.101 | −1.17 ± 1.941 |
|
| −0.154 ± 1.416 | −1.16 ± 1.434 | −0.396 ± 1.977 | −0.379 ± 1.012 | −0.170 ± 2.003 | −0.96 ± 2.344 | −0.85 ± 2.676 |
|
| −0.52 ± 0.316** | −0.184 ± 0.367 | −0.66 ± 0.780** | 0.085 ± 0.333 | −0.318 ± 0.791 | −0.184 ± 0.367 | −0.441 ± 0.936 |
|
| −0.53 ± 0.356** | −0.201 ± 0.298 | −0.415 ± 0.579 | −0.050 ± 0.331 | −0.323 ± 0.544 | −0.201 ± 0.298 | −0.336 ± 0.657 |
|
| 2.136 ± 1.611** | 2.658 ± 2.449 | 1.091 ± 1.998 | 3.788 ± 1.628* | 1.635 ± 1.607* | 2.658 ± 2.449 | 0.518 ± 2.100 |
|
| 0.026 ± 0.022 | 0.008 ± 0.016 | 0.021 ± 0.032 | 0.021 ± 0.018* | 0.021 ± 0.039 | 0.008 ± 0.016 | −0.009 ± 0.035 |
|
| 0.271 ± 2.913 | 6.341 ± 4.141** | 3.262 ± 4.342 | 2.551 ± 4.328 | 1.130 ± 2.368 | 6.341 ± 4.141** | 2.295 ± 2.745 |
|
| 1.409 ± 0.976** | 0.700 ± 2.280 | 0.868 ± 1.814 | 4.216 ± 2.419* | 1.810 ± 1.396* | 4.375 ± 2.018* | 0.636 ± 2.782 |
|
| 2.685 ± 1.409** | 4.231 ± 2.493** | 3.025 ± 2.177* | 6.334 ± 2.447* | 3.103 ± 1.812* | 6.156 ± 2.420* | 2.749 ± 2.293** |
*p < 0.001, **p < 0.05.
R (mm)/asph, radius of curvature and asphericity of the anterior cornea (conic fitting); R (mm)/asph ICRS meridian, radius of curvature and asphericity of the anterior cornea in the meridian of the segment; R (mm)/asph ICRS perpendicular meridian, radius of curvature and asphericity of the anterior cornea in the perpendicular of the meridian of the segment; wavefront astigmatism (D); spherical equivalent (D); refractive astigmatism (D); SIA (D), surgically induced astigmatism.
Geometric and Visual Differences Between Different Optical Zones for the 160–150 Segments and 210 Segments
| 6–150 | 5–160 |
| 6–210 | 5–210 |
| |
|---|---|---|---|---|---|---|
|
| −1.39 ± 2.71 | −3.38 ± 2.35 | 0.035 | −1.25 ± 3.29 | −3.92 ± 3.27 | 0.021 |
|
| 0.29 ± 0.23 | 0.45 ± 0.28 | 0.077 | 0.38 ± 0.18 | 0.54 ± 0.31 | 0.096 |
|
| −0.32 ± 0.54 | −0.53 ± 0.35 | 0.227 | −0.33 ± 0.65 | −0.41 ± 0.58 | 0.703 |
|
| −1.63 ± 1.61 | −2.13 ± 1.61 | 0.377 | 0.51 ± 2.10 | 1.09 ± 1.99 | 0.408 |
|
| 0.021 ± 0.04 | 0.02 ± 0.02 | 0.636 | −0.009 ± 0.035 | 0.021 ± 0.032 | 0.009 |