| Literature DB >> 35282169 |
Ramón Ruiz-Mesa1, Maria Ruiz-Santos1, Julia Blanch-Ruiz1, Ana Jiménez-Nieto1.
Abstract
Purpose: To compare the acquisition time necessary to obtain the optical biometry plus corneal power measurement using the IOLMaster 700 with central topography with that found using the standard IOLMaster 700 in combination with two corneal topographers, when acquiring biometry measurements during cataract evaluation.Entities:
Keywords: IOLMaster 700; Scheimpflug imaging; cataract; color LED; optical biometer
Year: 2022 PMID: 35282169 PMCID: PMC8906856 DOI: 10.2147/OPTH.S351845
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Mean Acquisition Time for the Different Devices
| Time (Seconds) | IOLMaster 700 with CT | IOLMaster 700 | IOLMaster 700 + Cassini | IOLMaster 700 + Pentacam HR | p value |
|---|---|---|---|---|---|
| 19.63 ± 5.02 | 20.70 ± 5.58 | 62.94 ± 15.75 | 53.27 ± 22.72 | <0.001* | |
| [11.88, 40.60] | [11.28, 40.43] | [31.28, 115.73] | [31.09, 177.94] |
Notes: p-value was obtained with the rANOVA test. *Significant differences <0.05.
Abbreviations: CT, central topography; SD, standard deviation.
Figure 1Bland-Altman plots obtained showing the mean difference versus average for the comparison between the different devices for K1 and K2. Mean (continuous line), lower and upper limits of agreement (±1.96SD, standard deviation, peripheral dotted lines) and lower and upper confidence intervals (95%) are depicted.
Figure 2Bland-Altman plots obtained showing the mean difference versus average for the comparison between the different devices for TK1, TK2, EKR1 and EKR2. Mean (continuous line), lower and upper limits of agreement (±1.96SD, standard deviation, peripheral dotted lines) and lower and upper confidence intervals (95%) are depicted.
Agreement of K1, K2, TK1, TK2, EKR1 and EKR2 (Diopters) Measurements Between the Three Devices
| Mean Difference ± SD | 95% CI | 95% LoA | p value | |
|---|---|---|---|---|
| 0.01±0.38 | [−0.06, 0.08] | [−0.73, 0.75] | 0.851 | |
| 0.05±0.30 | [−0.01, 0.11] | [−0.53, 0.64] | <0.001* | |
| 0.04±0.43 | [−0.04, 0.12] | [−0.82, 0.90] | <0.001* | |
| −0.02±0.37 | [−0.10, 0.04] | [−0.75, 0.70] | 0.337 | |
| 0.08±0.26 | [0.02, 0.13] | [−0.43, 0.60] | <0.001* | |
| 0.10±0.40 | [0.02, 0.18] | [−0.67, 0.89] | <0.001* | |
| 0.01±0.41 | [−0.07, 0.09] | [−0.80, 0.82] | 0.616 | |
| 0.30±0.43 | [0.21, 0.39] | [−0.55, 1.16] | <0.001* | |
| 0.29±0.58 | [0.17, 0.41] | [−0.85, 1.45] | <0.001* | |
| −0.02±0.39 | [−0.10, 0.04] | [−0.79, 0.74] | 0.548 | |
| 0.42±0.50 | [−0.32, 0.52] | [−0.57, 1.42] | <0.001* | |
| 0.45±0.62 | [0.32, 0.57] | [−0.77, 1.68] | <0.001* |
Notes: p-value obtained with the Tukey’s test; *Significant differences <0.05.
Abbreviations: SD, standard deviation; CI, confidence interval; LoA, limits of agreement.