| Literature DB >> 35265468 |
Mohamadhosein Asemanrafat1, Ali Chaparian1, Mehrzad Lotfi2, Alireza Rasekhi2.
Abstract
Background: The objective of this study was to investigate the influence of iterative reconstruction (IR) algorithm on radiation dose and image quality of computed tomography (CT) scans of patients with malignant pancreatic lesions by designing a new protocol.Entities:
Keywords: Computed tomography; image quality; iterative reconstruction; pancreas cancer; radiation dose
Year: 2021 PMID: 35265468 PMCID: PMC8804595 DOI: 10.4103/jmss.JMSS_81_20
Source DB: PubMed Journal: J Med Signals Sens ISSN: 2228-7477
Computed tomography scan parameters for the two parenchymal phase stages
| Parameters | Usual dose protocol | Reduced dose protocol |
|---|---|---|
| Tube voltage (kVp) | 120 | 120 |
| Tube current range (mA) | 80-500 | 80-500 |
| Rotation time (s) | 0.5 | 0.5 |
| Noise index | 8 | 12 |
| Field of view (mm) | 300-400 | 300-400 |
| Helical pitch (standard) | 0.813 | 0.813 |
| Detector configuration | 0.5×80 | 0.5×80 |
| Beam collimation (mm) | 40 | 40 |
| Slice thickness (mm) | 3 | 3 |
| Slice interval (mm) | 3 | 3 |
| Reconstruction algorithm | FBP | FBP and AIDR 3D |
FBP – Filtered Back Projection; AIDR 3D – Adaptive iterative dose reduction three-dimensional
Figure 1Axial contrast-enhanced computed tomography images of a 62-year-old male in the pancreatic parenchymal phase showing region of interests on the pancreas, portal vein, abdominal aorta, and liver. (a) Usual dose with Filtered Back Projection. (b) Reduced dose with Filtered Back Projection. (c) Reduced dose with adaptive iterative dose reduction three dimensional
Figure 2Axial contrast-enhanced computed tomography images of a 62-year-old male in the pancreatic parenchymal phase showing adenocarcinoma in the head of pancreas (star) and the encasement of superior mesenteric vein (long arrow) and celiac trunk (short arrow) by the lesion. (a) Usual dose with Filtered Back Projection. (b) Reduced dose with Filtered Back Projection. (c) Reduced dose with adaptive iterative dose reduction three dimensional
Quantitative assessment of image quality of the pancreas computed tomography for Protocols A, B, and C
| Parameters | Protocols* |
| ||||
|---|---|---|---|---|---|---|
|
|
| |||||
| A | B | C | A versus B | A versus C | B versus C | |
| Pancreatic parenchyma | ||||||
| Noise | 19.58±4.34 | 25.13±5.12 | 13.40±3.17 | <0.001 | <0.001 | <0.001 |
| SNR | 5.4±2.13 | 3.97±1.49 | 7.5±3.12 | <0.001 | <0.001 | <0.001 |
| CNR | 3.55±2 | 2.46±1.37 | 4.68±2.79 | 0.005 | 0.04 | <0.001 |
| Portal vein | ||||||
| SNR | 10.79±2.65 | 7.79±1.94 | 14.49±2.89 | <0.001 | <0.001 | <0.001 |
| CNR | 6.52±1.78 | 4.65±1.36 | 8.51±2.08 | <0.001 | <0.001 | <0.001 |
| Abdominal aorta | ||||||
| SNR | 10.44±3.14 | 6.98±1.65 | 12.77±2.38 | <0.001 | <0.001 | <0.001 |
| CNR | 6.16±2.73 | 3.84±1.11 | 6.87±1.67 | <0.001 | 0.045 | <0.001 |
| Liver | ||||||
| SNR | 5.28±1.56 | 4.46±1.23 | 8.36±2.09 | 0.01 | <0.001 | <0.001 |
| CNR | 1.17±0.64 | 1.40±0.57 | 2.58±1.01 | 0.045 | <0.001 | <0.001 |
| Paraspinal muscle | ||||||
| Noise | 14.98±3.23 | 20.60±4.19 | 10.80±1.95 | <0.001 | <0.001 | <0.001 |
*Protocol A: The images which obtained at usual dose and reconstructed with the FBP, Protocol B: The images which obtained at reduced dose and reconstructed with FBP, Protocol C: The images which obtained at reduced dose and reconstructed with AIDR 3D. SNR – Signal-to-noise ratio; CNR – Contrast-to-noise ratio; FBP – Filtered back projection; AIDR 3D – Adaptive iterative dose reduction three dimensional
Qualitative assessment of image quality of the pancreas computed tomography for Protocols A, B, and C from the two radiologists
| Protocol | Radiologist 1 | Radiologist 2 | Average scores of two radiologists | ||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
| |||||||
| A | B | C | A | B | C | A | B | C | |
| Image quality score | 4.35±0.48 | 3.83±0.75 | 4.15±0.58 | 4.38±0.54 | 3.55±0.50 | 4.20±0.52 | 4.36±0.50 | 3.69±0.64 | 4.18±0.54 |
|
| |||||||||
|
|
|
|
|
|
|
|
|
| |
|
| |||||||||
|
| <0.001 | 0.03 | <0.09 | <0.001 | <0.001 | 0.09 | <0.001 | <0.001 | 0.13 |
Radiation dose of the pancreas computed tomography for Protocols A, B, and C
| Protocol A (usual dose with FBP) | Protocol B or C (reduced dose) | Dose reduction (%) | |
|---|---|---|---|
| CTDIvol (mGy) | 8.17±1.84 (5.20-13.40) | 4.23±1.10 (3.00-9.30) | 48.07 |
| DLP (mGy.cm) | 155.34±57.29 (72.80-374.00) | 80.12±35.11 (42.00-259.00) | 48.31 |
| Effective dose (mSv) | 2.33±0.86 (1.09-5.61) | 1.20±0.53 (0.63-3.89) | 48.42 |
FBP – Filtered Back Projection; CT – Computed tomography; CTDIvol – CT dose index; DLP – Dose-length product