Hae-Young Lee1, Thilo Burkard2,3.
Abstract
Blood pressure measurement (BPM) is an essential part of medical examination, and therefore accuracy of BPM devices is crucial. Over the past few years, there has been a rise in new BPM techniques using photoplethysmographic (PPG) signals and complex algorithms for blood pressure estimation. Especially the combination of a mobile device or a smartphone with a camera using PPG may potentially revolutionize BPM in the future. The first-ever BPM application to be approved as a medical device was one by the Korean Ministry of Food and Drug Safety in 2020, despite the lack of robust scientific evidence proving its validity. While the prospect of using these novel BPM devices is an opportunity, there are also some critical issues around calibration and utility in different patient groups that need to be resolved before they can be incorporated into daily clinical practice.Entities:
Keywords: Blood pressure measurement; Cuffless blood pressure measurement; Photoplethysmographic blood pressure measurement; Smart devices; m-Health
Year: 2022 PMID: 35257532 PMCID: PMC8907988 DOI: 10.4070/kcj.2021.0405
Source DB: PubMed Journal: Korean Circ J ISSN: 1738-5520 Impact factor: 3.243
Figure 1Measurement sequences for the validation of BPM devices. (A) Measurement sequence according to the Universal AAMI/ESH/ISO Standard for the validation16) of cuff-based BP measurement devices, started by entry measurements with the reference method and the test device, directly followed by the validation measurement set. (B) The proposal for a validation procedure for PPG Apps needs an initial calibration to the reference method. After a first calibration set, validations are set at different time points covering one calibration’s recommended time frame.
Source of illustrations: Shutterstock.com (Zern Liew; marina_ua; Victor Metelskiy).
AAMI/ESH/ISO = Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization; App = application; BP = blood pressure; BPM = blood pressure measurement; PPG = photoplethysmographic.