Literature DB >> 33675592

Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device.

Anna Vybornova1, Erietta Polychronopoulou2, Arlène Wurzner-Ghajarzadeh2, Sibylle Fallet1, Josep Sola1, Gregoire Wuerzner2.   

Abstract

OBJECTIVE: The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month.
METHODS: In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject.
RESULTS: Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP.
CONCLUSION: After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population.
Copyright © 2021 The Author(s). Published by Wolters Kluwer Health, Inc.

Entities:  

Mesh:

Year:  2021        PMID: 33675592     DOI: 10.1097/MBP.0000000000000531

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  7 in total

Review 1.  Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals.

Authors:  Ramakrishna Mukkamala; Mohammad Yavarimanesh; Keerthana Natarajan; Jin-Oh Hahn; Konstantinos G Kyriakoulis; Alberto P Avolio; George S Stergiou
Journal:  Hypertension       Date:  2021-09-13       Impact factor: 10.190

2.  Wearable Photoplethysmography for Cardiovascular Monitoring.

Authors:  Peter H Charlton; Panicos A Kyriaco; Jonathan Mant; Vaidotas Marozas; Phil Chowienczyk; Jordi Alastruey
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2022-03-11       Impact factor: 10.961

3.  Validation of the optical Aktiia bracelet in different body positions for the persistent monitoring of blood pressure.

Authors:  Josep Sola; Anna Vybornova; Sibylle Fallet; Erietta Polychronopoulou; Arlene Wurzner-Ghajarzadeh; Gregoire Wuerzner
Journal:  Sci Rep       Date:  2021-10-19       Impact factor: 4.379

Review 4.  The Advent of Cuffless Mobile Device Blood Pressure Measurement: Remaining Challenges and Pitfalls.

Authors:  Hae-Young Lee; Thilo Burkard
Journal:  Korean Circ J       Date:  2022-03       Impact factor: 3.243

5.  Smartwatch-Based Blood Pressure Measurement Demonstrates Insufficient Accuracy.

Authors:  Maarten Falter; Martijn Scherrenberg; Karen Driesen; Zoë Pieters; Toshiki Kaihara; Linqi Xu; Enrico Gianluca Caiani; Paolo Castiglioni; Andrea Faini; Gianfranco Parati; Paul Dendale
Journal:  Front Cardiovasc Med       Date:  2022-07-11

Review 6.  Advances in Non-Invasive Blood Pressure Monitoring.

Authors:  Xina Quan; Junjun Liu; Thomas Roxlo; Siddharth Siddharth; Weyland Leong; Arthur Muir; So-Min Cheong; Anoop Rao
Journal:  Sensors (Basel)       Date:  2021-06-22       Impact factor: 3.576

Review 7.  Wearable blood pressure measurement devices and new approaches in hypertension management: the digital era.

Authors:  D Konstantinidis; P Iliakis; F Tatakis; K Thomopoulos; K Dimitriadis; D Tousoulis; K Tsioufis
Journal:  J Hum Hypertens       Date:  2022-03-23       Impact factor: 3.012

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.