Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.

Philipp Krisai1,2, Annina Salome Vischer3, Leo Kilian3, Andrea Meienberg3, Michael Mayr3, Thilo Burkard1,3.   

Abstract

OBJECTIVE: Recently, a cuffless blood pressure (BP) measurement device using pulse transit time (PTT) for beat-to-beat calculation of BP values has been validated over a short time period. However, it remains unknown how values obtained with this device compare with standard ambulatory measurements over a 24-hour period. We hypothesised that BP values measured by a cuffless PTT device (TestBP) are comparable with measurements by a standard upper arm cuff-based BP device (RefBP) in clinical practice over 24 hours.
METHODS: Between May and December 2017, 71 individuals were prospectively included. Cuffless using the Somnotouch-NIBP (Somnomedics GmbH, Randersacker, Germany) and cuff-based standard 24-hour BP measurements were performed simultaneously on the left and right arm, respectively. The first RefBP measurement was used as calibration measurement for the TestBP.
RESULTS: Mean (±SD) age was 49.3 (15.1) years, and 51% were male. Mean 24-hour BP for TestBP and RefBP were 140.8 (20.0) versus 134.0 (17.3) mm Hg for systolic (p<0.0001) and 85.8 (14.1)versus79.3 (11.7) mm Hg for diastolic (p<0.0001) measurements, respectively. Mean absolute systolic and diastolic disagreements between TestBP and RefBP were 10.2 (7.2) and 8.2 (5.5) mm Hg, respectively. The number (percentage) of absolute differences between the mean 24-hour BP values of the TestBP and RefBP within 5, 10 and 15 mm Hg were 23 (32.4), 43 (60.6) and 54 (74.6) for systolic and 24 (33.8), 51 (71.8) and 65 (91.6) for diastolic measurements, respectively.
CONCLUSION: In clinical practice over 24 hours, there was a significant difference between the TestBP and RefBP with higher systolic and diastolic BP measured with the cuffless PTT device. Reasons for this difference need to be investigated. CLINICAL TRIAL REGISTRATION: NCT03054688; Results. © Author(s) (or their employer(s)) 2019. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  hypertension

Year:  2018        PMID: 30228251     DOI: 10.1136/heartjnl-2018-313592

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  9 in total

1.  Blood Pressure Cuff Inflation Briefly Increases Female Adolescents' Restlessness During Sleep on the First But Not Second Night of Ambulatory Blood Pressure Monitoring.

Authors:  H Matthew Lehrer; Gehui Zhang; Karen A Matthews; Robert T Krafty; Marissa A Evans; Briana J Taylor; Martica H Hall
Journal:  Psychosom Med       Date:  2022-06-24       Impact factor: 3.864

Review 2.  The Potential Role of Sensors, Wearables and Telehealth in the Remote Management of Diabetes-Related Foot Disease.

Authors:  Jonathan Golledge; Malindu Fernando; Peter Lazzarini; Bijan Najafi; David G Armstrong
Journal:  Sensors (Basel)       Date:  2020-08-13       Impact factor: 3.576

3.  Improved agreement and diagnostic accuracy of a cuffless 24-h blood pressure measurement device in clinical practice.

Authors:  Thenral Socrates; Philipp Krisai; Annina S Vischer; Andrea Meienberg; Michael Mayr; Thilo Burkard
Journal:  Sci Rep       Date:  2021-01-13       Impact factor: 4.379

4.  Beat-to-beat blood pressure measurement using a cuffless device does not accurately reflect invasive blood pressure.

Authors:  Mohammed A Moharram; Luke C Wilson; Michael Ja Williams; Sean Coffey
Journal:  Int J Cardiol Hypertens       Date:  2020-04-14

5.  The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring.

Authors:  Jakob Nyvad; Kent L Christensen; Niels Henrik Buus; Mark Reinhard
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-12-17       Impact factor: 3.738

6.  Cuff-less blood pressure measurement with pulse transit time: The importance of rigorous assessment.

Authors:  Noud van Helmond; Timothy B Plante
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-12-12       Impact factor: 3.738

Review 7.  How Should We Measure and Deal with Office Blood Pressure in 2021?

Authors:  Annina S Vischer; Thilo Burkard
Journal:  Diagnostics (Basel)       Date:  2021-02-03

8.  Patients with Dipper and Nondipper High-Normal Blood Pressure Were Associated with Left Ventricular Mass.

Authors:  Fan-Kai Xiao; Ping Li; Zhan-Ying Han; Li Jing; Shaohua Hua; Luo-Sha Zhao
Journal:  Int J Hypertens       Date:  2021-12-21       Impact factor: 2.420

Review 9.  The Advent of Cuffless Mobile Device Blood Pressure Measurement: Remaining Challenges and Pitfalls.

Authors:  Hae-Young Lee; Thilo Burkard
Journal:  Korean Circ J       Date:  2022-03       Impact factor: 3.243

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.