Improved agreement and diagnostic accuracy of a cuffless 24-h blood pressure measurement device in clinical practice.

Thenral Socrates1, Philipp Krisai2,3, Annina S Vischer4, Andrea Meienberg4, Michael Mayr4, Thilo Burkard4,2.   

Abstract

A cuffless blood pressure (BP) device (TestBP) using pulse transit time is in clinical use, but leads to higher BP values compared to a cuff-based 24 h-BP reference device (RefBP). We evaluated the impact of a recent software update on BP results and TestBP's ability to differentiate between normo- and hypertension. 71 individuals had TestBP (Somnotouch-NIBP) and RefBP measurements simultaneously performed on either arm. TestBP results with software version V1.5 were compared to V1.4 and RefBP. Mean 24 h (± SD) BP for the RefBP, TestBP-V1.4 and TestBP-V1.5 were systolic 134.0 (± 17.3), 140.8 (± 20) and 139.1 (± 20) mmHg, and diastolic 79.3 (± 11.7), 85.8 (± 14.1) and 83.5 (± 13.0) mmHg, respectively (p-values < 0.001). TestBP-V1.5 area under the curve (95% confidence interval) versus RefBP for hypertension detection was 0.92 (0.86; 0.99), 0.94 (0.88; 0.99) and 0.77 (0.66; 0.88) for systolic and 0.92 (0.86; 0.99), 0.92 (0.85; 0.99) and 0.84 (0.74; 0.94) for diastolic 24 h, awake and asleep BP respectively. TestBP-V1.5 detected elevated systolic/diastolic mean 24 h-BP with a 95%/90% sensitivity and 65%/70% specificity. Highest Youden's Index was systolic 133 (sensitivity 95%/specificity 80%) and diastolic 87 mmHg (sensitivity 81%/specificity 98%). The update improved the agreement to RefBP. TestBP was excellent for detecting 24 h and awake hypertensive BP values but not for asleep BP values.

Entities:  

Year:  2021        PMID: 33441915      PMCID: PMC7806663          DOI: 10.1038/s41598-020-80905-x

Source DB:  PubMed          Journal:  Sci Rep        ISSN: 2045-2322            Impact factor:   4.379


  19 in total

1.  Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor.

Authors: 
Journal:  Blood Press Monit       Date:  1998-10       Impact factor: 1.444

2.  European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults.

Authors:  Eoin O'Brien; Neil Atkins; George Stergiou; Nikos Karpettas; Gianfranco Parati; Roland Asmar; Yutaka Imai; Jiguang Wang; Thomas Mengden; Andrew Shennan
Journal:  Blood Press Monit       Date:  2010-02       Impact factor: 1.444

3.  Cuffless Blood Pressure Measurement: How Did Accuracy Become an Afterthought?

Authors:  Raj Padwal
Journal:  Am J Hypertens       Date:  2019-08-14       Impact factor: 2.689

4.  Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.

Authors:  Philipp Krisai; Annina Salome Vischer; Leo Kilian; Andrea Meienberg; Michael Mayr; Thilo Burkard
Journal:  Heart       Date:  2018-09-18       Impact factor: 5.994

5.  A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research.

Authors:  Terry K Koo; Mae Y Li
Journal:  J Chiropr Med       Date:  2016-03-31

6.  Global burden of hypertension: analysis of worldwide data.

Authors:  Patricia M Kearney; Megan Whelton; Kristi Reynolds; Paul Muntner; Paul K Whelton; Jiang He
Journal:  Lancet       Date:  2005 Jan 15-21       Impact factor: 79.321

7.  The history of blood pressure measurement.

Authors:  E O'Brien; D Fitzgerald
Journal:  J Hum Hypertens       Date:  1994-02       Impact factor: 3.012

8.  Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method.

Authors:  Heiko Gesche; Detlef Grosskurth; Gert Küchler; Andreas Patzak
Journal:  Eur J Appl Physiol       Date:  2011-05-10       Impact factor: 3.078

9.  [2018 ESC/ESH Guidelines for the management of arterial hypertension. The Task Force for the management of arterial hypertension of the European Society of Cardiology (ESC) and the European Society of Hypertension (ESH)].

Authors:  Bryan Williams; Giuseppe Mancia; Wilko Spiering; Enrico Agabiti Rosei; Michel Azizi; Michel Burnier; Denis L Clement; Antonio Coca; Giovanni de Simone; Anna Dominiczak; Thomas Kahan; Felix Mahfoud; Josep Redon; Luis Ruilope; Alberto Zanchetti; Mary Kerins; Sverre E Kjeldsen; Reinhold Kreutz; Stephane Laurent; Gregory Y H Lip; Richard McManus; Krzysztof Narkiewicz; Frank Ruschitzka; Roland E Schmieder; Evgeny Shlyakhto; Costas Tsioufis; Victor Aboyans; Ileana Desormais
Journal:  G Ital Cardiol (Rome)       Date:  2018-11

10.  European Society of Hypertension practice guidelines for ambulatory blood pressure monitoring.

Authors:  Gianfranco Parati; George Stergiou; Eoin O'Brien; Roland Asmar; Lawrence Beilin; Grzegorz Bilo; Denis Clement; Alejandro de la Sierra; Peter de Leeuw; Eamon Dolan; Robert Fagard; John Graves; Geoffrey A Head; Yutaka Imai; Kazuomi Kario; Empar Lurbe; Jean-Michel Mallion; Giuseppe Mancia; Thomas Mengden; Martin Myers; Gbenga Ogedegbe; Takayoshi Ohkubo; Stefano Omboni; Paolo Palatini; Josep Redon; Luis M Ruilope; Andrew Shennan; Jan A Staessen; Gert vanMontfrans; Paolo Verdecchia; Bernard Waeber; Jiguang Wang; Alberto Zanchetti; Yuqing Zhang
Journal:  J Hypertens       Date:  2014-07       Impact factor: 4.844

View more
  3 in total

1.  A fusion approach to improve accuracy and estimate uncertainty in cuffless blood pressure monitoring.

Authors:  Cederick Landry; Sean D Peterson; Arash Arami
Journal:  Sci Rep       Date:  2022-05-13       Impact factor: 4.996

Review 2.  How Should We Measure and Deal with Office Blood Pressure in 2021?

Authors:  Annina S Vischer; Thilo Burkard
Journal:  Diagnostics (Basel)       Date:  2021-02-03

Review 3.  The Advent of Cuffless Mobile Device Blood Pressure Measurement: Remaining Challenges and Pitfalls.

Authors:  Hae-Young Lee; Thilo Burkard
Journal:  Korean Circ J       Date:  2022-03       Impact factor: 3.243

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.